- Phone: +90 312 266 33 34 (9 a.m. to 5 p.m., UTC +3)
- E-mail: info@dynalabs.com.tr
Warranty
Our products are warranted against defective materials and workmanship for one year. Defects arising from user errors are not covered by the warranty.
Copyright
All copyrights of this manual belonging to Dynalabs products are reserved. It cannot be reproduced without written consent.
Disclaimer
Dynalabs Ltd. provides this publication “as is” without warranty of any kind,
express or implied, including but not limited to, the implied warranties of
merchantability or fitness for a particular purpose. This document is subject
to change without notice, and should not be construed as a commitment or
representation by Dynalabs Ltd.
This publication may contain inaccuracies or typographical errors. Dynalabs
Ltd. will periodically update the material for inclusion in new editions.
Changes and improvements to the product described in this manual may be made
at any time.
Introduction
Dynalabs Inertial Measurement Units are based on triaxial accelerometers and triaxial gyroscopes that are integrated in a single housing. The inertial measurement unit (IMU) is based on proven microelectromechanical systems (MEMS) accelerometers and gyroscopes for detecting the smallest linear accelerations and angular rates. Dynalabs IMUs enable separate analog voltage outputs for all 6 degrees of freedom (DOF). IMUs enable power supply voltage from 6 to 20 VDC. Dynalabs IMUs feature a lightweight, reliable aluminum housing with protection class IP68 and have cable with configurable length and connectors.
DYN-I-8000 sensors offer the following options;
- Custom Cable Length
- Custom Housing Material
- Custom Connector
- Base plate
General Information
Unpacking and Inspection
Dynalabs products provide adequate protection for undamaged products to be transported. Document the damages that occur indirectly during the transport and contact the customer representative.
System Components
The DYN-I-8000 has the following components:
- MEMS Sensor
- Calibration Certificate
- Product Manual
Specifications
Table 1 Gyroscopes Specifications Datasheet
Full scale angular velocity| __
(°/s)
| __
±75
| __
±150
| __
±300
| __
±900
---|---|---|---|---|---
__
Frequency range
| __
(Hz)
| __
0-150
| __
0-150
| __
0-150
| __
0-150
Non-linearity (full scale)| __
(%)
| __
0.06
| __
0.06
| __
0.06
| __
0.06
__
Noise (in band)
| __
(°/s /√Hz)
| __
0.0075
| __
0.0075
| __
0.0075
| __
0.0075
Scale factor (nominal)| __
(V/°/s)
| __
0.012
| __
0.006
| __
0.003
| __
0.001
Scale factor var. over. temp.| __
(%)
| __
0.5
| __
0.5
| __
0.5
| __
0.5
Bias variation with temp.| __
(°/s)
| __
± 1
| __
± 2
| __
± 3
| __
± 4
Table 2: Accecelerometers Specifications Datasheet
Full scale acceceleration| __
(g)
| __
±2
| __
±5
| __
±10
| __
±30
| __
±50
| __
±100
| __
±200
---|---|---|---|---|---|---|---|---
Frequency range (±5%)| (Hz)| __
150
| __
150
| __
300
| __
600
| __
750
| __
750
| __
1000
Non-linearity (full scale)| (%)| __
0.1
| __
0.1
| __
0.1
| __
0.1
| __
0.1
| __
0.1
| __
0.1
__
Noise (in band)
| (μg/√ Hz)| __
9
| __
21
| __
40
| __
100
| __
180
| __
340
| __
680
Scale factor (nominal)| (mV/ g)| __
1,350
| __
540
| __
270
| __
90
| __
54
| __
27
| __
13.5
Scale factor temp. coeff.| (ppm/° C)| __
120
| __
120
| __
120
| __
120
| __
120
| __
120
| __
120
Bias temperature coeff.| (mg/° C)| __
± 0.2
| __
± 0.5
| __
± 1
| __
± 3
| __
± 5
| __
± 10
| __
± 20
Environmental
Table 3: Environmental Specifications datasheet
Protection Level | IP 68 |
---|---|
Operating Voltage | 6 V – 20 V |
Operating Temperature | -40 °C to +100 °C |
Operating Current Consumption mA | <50 mA |
Isolation | Case isolated |
Physical
Table 4: Physical Specifications datasheet
Sensing Element | MEMS Capacitive – Vibrating ring |
---|---|
Housing Material | Aluminum or Steel |
Connector (Optional) | D-Sub 9 or 15 pin, Lemo, Binder |
Mounting | Adhesive or screw mount |
Base plate (Optional) | Aluminum or Steel |
__
Weight (without cable)
| 27.5g (aluminum)
69.5 g (steel)
Outline Drawing
The dimensional properties of DYN-I-8000 sensors are given below
Operation and Installation
General
The general sensor connector configuration is given below;
Cable Code/Pin Configuration
- Red: V + Power supply voltage +6 to +20 VDC
- Black: Ground Power GND
- X-Axis: Yellow: Signal(+) Positive, analog output voltage signal for differential mode.
- Purple: Signal(-) Negative, analog output voltage signal for differential mode.
- Y-Axis : Blue: Signal(+) Positive, analog output voltage signal for differential mode.
- Green: Signal(-) Negative, analog output voltage signal for differential mode.
- Z -Axis: White: Signal(+) Positive, analog output voltage signal for differential mode.
- Orange: Signal(-) Negative, analog output voltage signal for differential mode
Cable Code/Pin Configuration:
- Red: V + Power supply voltage +5 V to +20 VDC.
- Black: Ground Power GND
- RX: Brown: Gyro Signal X-Axis: Analog output voltage signal for single-ended mode.
- RY : Grey: Gyro Signal Y-Axis: Analog output voltage signal for single-ended mode.
- RZ: Pink: Gyro Signal Z-Axis: Analog output voltage signal for single-ended mode.
- White- Brown: NC Not connected.
WARNING
- Never connect the power supply and/or the power ground to yellow, purple, blue, green, white, orange, brown, grey, and/or pink cables.
- Never connect the power supply to the power ground. Always use a clean power source and check the voltage range.
Sensor Static Calibration Verification
- Using gravity for the accelerometer of the 8000 series IMU, voltage values are measured in the + and – gravity directions, providing a value of ±1 g. The measurement should be made as follows;
- When the acceleration value of 8000 series IMU’s accelerometers is entered into the data acquisition system, the sensor shows +1 g with the effect of gravity, which is in the direction of the axis to be calibrated.
- When the sensor is positioned in the opposite direction to the axis to be calibrated, the arrow shows -1g as shown below under the effect of gravity.
- Using gravity, the voltage values that provide 1 g in the + and – directions are measured and compared with the catalog value. The calibration value should be close to the catalog value with 10% tolerance. Sensor catalog sensitivity values are given in Table 2.
Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the manufacturer. The product(s) are developed, produced and tested according to following EC- directives:
- 2014/35/EU – Low Voltage Directive (LVD)
- 2006/42/EU – Machinery Safety Directive
- 2015/863/EU – RoHS Directive
Applied standards:
- EN 61010-1:2010
- EN ISO 12100:2010
- MIL-STD-810-H-2019 (Test Methods: 501.7 – High Temperature, 502.7 – Low Temperature, 514.8 – Vibration, 516.8 – Shock)
DYNALABS MÜHENDİSLİK SANAYİ TİCARET LİMİTED ŞİRKETİ declares that above
mentioned products meet all the requirements of the above mentioned standards
and regulations.
Murat Aykan, Technical Manager Ankara, 15.07.2021