hygiena BAX System Q7 Software User Guide
- June 3, 2024
- Hygiena
Table of Contents
hygiena BAX System Q7 Software User Guide
BAX System Q7 software release v4.21 contains several user interface features, new target options, and error logging improvements.
We strongly recommended that you always install the most recent version as soon as possible for optimal performance. Please read all instructions thoroughly and ensure all required prerequisites are confirmed prior to prior to beginning installation of the software update.
If you currently have a WIN7 computer system, please contact diagnostics.support@hygiena.com or your local Hygiena® representative for upgrade support. This release is NOT applicable for WIN7 systems.
Software Updates and Features for Release v4.21.3.9366 for WIN10 ONLY
Install Package
- Single executable install with reduced size (2 GB)
- Reorganization of software versioning:
- Package: 4.21.3.9366
- BAX System: 4.21.3.9316
- Bridge: 4.21.3.9359
- SDS: 1.3.1035.1
- Analysis: 4.21.3.8313
SDS Equipment Driver Improvements:
- Auto-recovery of instrument communication disconnection with notifications
User Interface Enhancements:
- Option to export results as a .csv file
- About Dialog updates
- Removed ability to initialize a run without instrument connection
- Feature to preserve calibration file during installation
Application Insights:
- Improvements with more detailed logs for troubleshooting
Certified Algorithms
- Additional targets:
- STEC Screener eae ONLY
- STEC Screener stx ONLY
- Real-Time E. coli O157:H7 S-Target ONLY
- Display messages and confidence levels for:
- Real-Time E. coli O157:H7 Exact
- Real-Time Salmonella
Software Upgrade Instructions:
1. Required Prerequisites:
Regional Settings are in English
Read/Write Permissions for the below file locations for all users
- C:\Program Files\MATLAB
- C:\Program Files (x86)\BAX
- C:\ProgramData\BAX Q7 System
- All Power and Sleep Settings are set to Never
2. Calibration storage and removal of current software:
Back up the system’s calibration file
- Select Operations > BAX Maintenance > Export Calibration from the menu bar
- Navigate to a new folder in which to store the backup file and click Save
- Name the calibration file (example: Q7Cal_date.bqc)
Uninstall the previous version of the BAX Q7 application and ANY VERSION of MATLAB (REQUIRED)
- Navigate to Windows Settings and select Apps
Click on BAX System Q7, select Uninstall, and agree (click Yes) on next
window to initiate removal
Click on ANY version of the MATLAB application, select Uninstall, and click
Agree on next window (this may take several minutes)
Restart Computer BEFORE continuing to step 3
Software Installation Instructions for step 3 and 4
3. Download 4.21.3.9366 software by clicking this link or the button from the BAX Q7 Page
- Transfer download to a memory/USB stick if computer doesn’t have internet access
A windows protect screen may pop-up. Select More Info, then select Run Anyway.
- On the next pop-up, select Yes to allow install to begin
4. MATLAB updates will install first. Follow prompts below and select the
default responses:
Install error: VCRUNTIME140.dll was not found. (shown below)
- The install will end after you click on OK. To resolve this issue, please download and install the patch vc_redist.x64.exe from https://www.microsoft.com/en-us/download/details.aspx?id=52685.
After installing this patch, go back and retry installing the updated BAX System Q7 software.
5. Following MATLAB installation, the BAX Q7 application install will
initiate.
Read the prompts carefully. Choose the default responses.
- Welcome Window – click NexT.
Calibration File Warning – ENSURE YOU HAVE BACKED UP THE CALIBRATION FILE
- Check box and click Next
License Agreement
- Check Accept Terms box and click Next.
Custom Setup
- Click Next
Password Protection
- BAX default is disabled password protection. Choose an option for your lab, then click Next
Ready to Install Program
- Click Install
Installing BAX System Q7 Software
- During installation, an Error Log Reporting window will pop-up. We recommend NOT checking the box so the system can collect software and/or hardware errors in the Log Files to assist Diagnostics Support in troubleshooting. However, please review and decide what is appropriate for your lab.
- Select Next to complete install
InstallShield Wizard Complete
-
Select Finish
-
Click on the desktop icon to initiate program. New open screen will have software version 4.21.3.9366
-
Enter Instrument Name and Serial Number
- This will appear on reports and should be utilized
- This will appear on reports and should be utilized
6 CHECK system’s calibration (NEW FEATURE)
From the menu bar, select Diagnostics > Calibration to open the Q7 Calibration Diagnostics window
- Click the Check Calibration button to determine the status:
- If calibration is available, no further actions are needed
- If calibration is NOT optimal or NOT found:
Import the backed-up calibration file by selecting Operations > BAX Maintenance > Import Calibration from the menu bar and selecting the backed-up calibration file from step 2
7. Ready to perform BAX System Analysis
Define proper targets that are used in your lab by selecting View > Select
Targets > then in window, highlight targets being used, and then use arrow
buttons to move targets from Define in System to Available for Selection
window
References
- Home | Hygiena
- Download Microsoft Visual C++ 2015 Redistributable Update 3 RC from Official Microsoft Download Center
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>