GIMAURS-10T Urinalysis Reagent Strips Instructions
- June 7, 2024
- GIMA
Table of Contents
GIMA URS-10T Urinalysis Reagent Strips Instructions
SUMMARY AND INTENDED USE
The content of the instruction include usage, reaction principle, and
notification.
URS series Reagent strips are intended for qualitative and semi-quantitative
urinalysis and for in vitro diagnostics use.
The strips are for professional use only.
The strips may be read visually or instrumentally. Please read the instruction
carefully before use.
There are the test items of every products type.
Products Type | Test Item |
---|---|
URS 13 | Urobilinogen, Bilirubin, Ketone (acetoacetic acid), Blood, Protein, |
Nitrite, Leukocytes, Glucose, Specific Gravity, pH, Ascorbic Acid,
Microalbumin, Creatinine
URS 12, URS 11,
URS 10, URS 9,
URS 8, URS 7, URS 6,
URS 5, URS 4, URS 3,
URS 2, URS 1| The items of product type from URS 2 to URS 12 can be conbined
randomly from URS 13 test items mentioned above.US1 item can be any item from
URS 13.
SPECIMEN COLLECTION AND PREPARATION
Collect fresh urine in a clean dry container. Use uncentrifuged urine and mix
the sample before testing. The sample should not be more than 2 hours old at
the time of testing. Always handle specimens under sanitary conditions.
Note: Water should not be used as negative control. Preservatives will
not prevent the deterioration of ketones, bilirubin or urobilinogen. Bacterial
growth from contaminating organisms may affect glucose, pH, nitrite and blood
test results.
VISUAL READING TECHNIQUE
-
Immerse all reagent areas in specimen and remove strip immediately.
-
Run edge of strip against the rim of the container to remove excess urine.
-
Hold strip horizontally and compare test areas closely with colour chart on bottle label. Record the results.
For a semi-quantitative result read the reagent areas at the time specified on the colour chart. The pH and Protein areas may also be read immediately or at any time up to 60 seconds after dipping. For a qualitative result read the reagent areas between 1 and 2 minutes. If a positive result is obtained, repeat the test, reading each reagent at the time specified on the colour chart. Colour changes after 2 minutes are of no diagnostic value.
INSTRUMENT READING TECHNIQUE
Follow the directions given in the appropriate instrument-operating manual
STORAGE AND HANDLING PROCEDURES
Store only in original bottle. Do not use after expiry date. Every strip can
be used only once. Do not remove desiccant(s). Do not remove strip from the
bottle until immediately before it is to be used for testing. Replace cap
immediately and tightly after removing reagent strip.
Store at temperatures between 2℃-30℃.Don’t store in refrigerator. Keep away
from direct sunlight.
Do not touch test areas of reagent strips. PROTECTION AGAINST AMBIENT
MOISTURE, LIGHT AND HEAT IS ESSENTIAL TO GUARD AGAINST ALTERED REAGENT
REACTIVITY. Deterioration may result in discoloration or darkening of the
reagent areas. If this is evident or if test results are questionable or
inconsistent with expected results, confirm that strips are within the expiry
dates and compare with control urine. Please deal with the waste strips
according to “Treatment Regulations of Lab Biohazard Materials”. Once the
canister has been opened,the remaining strips are stable for up to 3 months in
humidity conditions lower than 65%RH.
LIMITATIONS OF PROCEDURES
As with all laboratory tests, definitive diagnostic or therapeutic decisions should not be made or based on any single result or method.
TEST PRINCIPLES
GLUCOSE: One enzyme, glucose oxidize catalyzes the formation of gluconic
acid and hydrogen peroxide from the oxidation of glucose. Hydrogen peroxide
releases neo-ecotypes oxide [O] under the function of peroxidase, [O] oxidates
potassium iodides chromogen, so that it makes the color change.
BILIRUBIN: This test is based on coupling the direct bilirubin with
diazotized dichloroaniline in a strongly acid medium, which produce the
diazotizing colors.
KETONE: This test is based on that acetoacetate acid react with sodium
nitroprusside in an alkaline medium, producing violet color.
SPECIFIC GRAVITY: Electrolyte (M+X-) in the form of salt in urine reacts
with poly (methyl vinyl ether/maleic omhydride (-COOH ), which is weakly acid
ionic exchanger. And then hydrogenous ion is replaced from it and reacts with
pH indicator in order to make the color change.
BLOOD: This test is based on the peroxidase activity of hemoglobin and
myoglobin, which makes peroxide release neo-ecotypic oxide[O]. Indicator is
oxidized by [O] and shows color change subsequently.
PH: This test is based on a double indicator principle.
PROTEIN: The test is based on the protein-error-of-indicators principle.
Anion on the specific pH indicator is absorbed by cation on protein molecule,
which make the indicator ionize and present color change at critical point of
color.
NITRITE: This test d epends u pon th e ni trite dia zotizewith aromatic
amino sulphanilamide to form a diazonium compound. This diazonium compound in
turn couples with 1,2,3,4-tetrahydro benzo(h)quinolin-3-phenol to produce a
pink color.
LEUCOCYTES: Granulocytic leukocytes in urine contain esterases that
catalyze the hydrolysis of the privatized pyrrole amino acid ester to liberate
3-hydroxy-5-pheny pyrrole. This pyrrole then reacts with a diazonium salt to
form a purple color.
ASCORBIC ACID: Ascorbic acid, with 1,2-dihydroxy alkenes, under alkaline
condition, deoxidizes blue 2,6- dichloroindophenolate form into colorless
N-(P-phenol)- 2,6-dichloro–P- amine phenol,.
MICROALBUMIN: Based on the protein deviation method, utilizing the
sulfonephthalein dyestuff only specific to Microalbumin.
CREATININE: Creatinine can act with 3, 5-di nitro benzoic acid in strong
alkalinity, generating colored compound.
Note:
GLUCOSE: The test is specific for glucose, no substance excreted in urine
other than glucose is known to give a positive result. In dilute urine
containing less than 0.28mmol/L ascorbic acid, as little as 2.2 mmol/L glucose
may produce a colour change that might be interpreted as positive. Ascorbic
acid concentrations of 2.8mmol/L or greater and/or high acetoacetic
concentrations (1.0mmol/L) may influence test. Small amounts of glucose may
normally be excreted by the kidney.
These amounts are usually below the sensitivity of this test.
BILIRUBIN: Normally no bilirubin is detectable in urine by even the most
sensitive methods. Even trace amounts of bilirubin are sufficiently abnormal
to require further investigation. Medicines that color the urine red or that
are themselves red in an acid medium, e.g., phenazopyridine may influence the
test. Large ascorbic acid concentration may cause false negatives.
KETONE: The test reacts with acetoacetic acid in urine. It does not react
with acetone or β-hydroxbutyric acid.
Normal urine specimens usually yield negative results with this reagent. False
positive results may occur with highly pigmented urine specimens or those
containing large amounts or levodopa metabolites.
SPECIFIC GRAVITY: The specific gravity permits determination of urine
specific gravity between 1.000 and 1.030.
In general, it correlates within 0.005 with values obtained with the
refractive index method. For increased accuracy, 0.005 may be added to
readings from urines with pH equal to or greater than 6.5. Strips read
instrumentally are automatically adjusted for pH by the instrument. The SG
test is not affected by certain nonionic urine constituents such as glucose or
by the presence of radiopaque dye. Highly buffered alkaline urines may cause
low readings relative to other methods. Elevated specific gravity readings may
be obtained in the presence of moderate quantities (1-7.5g/L) of protein.
BLOOD: The significance of the ‘Trace’ reaction may vary among patients,
and clinical judgment is required for assessment in an individual case.
Development of green spots (intact erythrocytes) or green colour (free
haemoglobin/ myoglobin) on the reagent area within 60 seconds indicates the
need for further investigation.
Blood is often found in the urine of menstruating females. Haemoglobin
concentration of 150-620μg/L is approximately equivalent to 5- 15/μL intact
red blood cells per microlitre.
This test is highly sensitive to haemolobin and thus complements the
microscopic examination. The sensitivity of this test may be reduced in urines
with high specific gravity. This test is equally sensitive to myoglobin as to
haemoglobin. Certain oxidizing contaminants, such as hypochlorite, may produce
false positive results.
Microbial peroxidase associated with urinary tract infection may cause a false
positive reaction. Levels of 2.8mmol/L ascorbic acid normally found in urine
do not interfere with this test.
PH: The pH test area measures pH values generally to within 1 unit in the
range of 5.0-8.5 visually and 5.0-9.0 instrumentally.
PROTEIN and MICROALBUMIN:
The Protein reagent area can detect albumin in urine and has low sensitivity
to mucoprotein, generaly up to 0.6g/L concentration.
The Microalbumin reagent area is to detect the microalbumin. Beyond 0.15 g/L
indicate albuminuria clinically. The Microalbumin reagent can detect
specificly microalbumin, and 9 times more sensitive than other protein.
Visible blood urine (≥0.05g/L) can be false negative action.
UROBILINOGEN: This test area will detect urobilinogen in concentrations as
low as 3μmol/L (approximately 0.2 Ehrlich unit/dL) in urine. The normal range
with this test is 3-16μmol/L. a result of 33 μmol/L represents the transition
from normal to abnormal, and the patient and/or urine specimen should be
evaluated further. The absence of urobilinogen cannot be determined with this
test.
NITRITE: This test depends upon the conversion of nitrate (derived from the diet) to nitrite by the action of principally Gram-negative bacteria in the urine. The test is specific for nitrite and will not react with any other substance normally excreted in urine. Pink spots or pink edges should not be interpreted as a positive result Any degree of uniform pink colour development should be interpreted as a positive nitrite test suggesting the presence of 105 or more organisms per mL, but colourdevelopment is not proportional to the number of bacteria present. A negative result does not prove that there is no significant bacteriuria. Negative results may occur.
- when urinary tract infections are caused by organisms which do not contain reductase to convert nitrate to nitrite;
- when urine has not been retained in the bladder long enough( four hours or more) for reduction of nitrate to occur,
- or when dietary nitrate is absent. Sensitivity of the nitrite test is reduced for urines with high specific gravity. It may resist 2.8mmol/L Ascorbic Acid.
LEUKOCYTES: Test area react with esterase in leucocytes (granulocytic leukocytes). Normal urine specimens generally yield negative result; positive results (+ or greater) are clinically significant. Individually observed ‘Trace’ results may be of questionable clinical significance; however ‘Trace’ results observed repeatedly may be clinically significant. ‘Positive’ results may occasionally be found with random specimens from females due to contamination of the specimen by vaginal discharge. Elevated glucose concentrations (160mmol/L) or high specific gravity may cause decreased test results
ASCORBIC ACID: The test area can detect the ascorbic acid in urine. Through the ascorbic acid detection, we will know the level of ascorbic acid in the body and the effect degree that the ascorbic acid bring to the test on glucose, bilirubin, blood and nitrite. It will reduce the sensitivity when the oxidant (such as potassium permanganate, hypochlorite ) in the urine.
CREATININE: Adult normal urine creatinine is 0.6-2.0g/24 hour (the Creatinine reagent area results is about 50-200mg/dL) . Result of random urine sample differ largely, from 10mg/dl to 300mg/dl. Concentrated urine and morning urine have high concentration (possibly over 200mg/dl). Because of diuresis, excessive drinking water, or other urine dilution, resulting in testing anlyte concentration decrease (result can be less than 50mg/dl).
SPECIFIC PERFORMANCE CHARACTERISTICS:
Specific performance characteristics are based on clinical and analytical studies. In clinical specimens, the sensitivity depends upon several factors; the variability of colour perception, the presence or absence of inhibitory factors typically found in urine, specific gravity, pH, and the lighting conditions when the product is read visually. Each colour block or instrumental display value represents a range of values. Because of specimen and reading variability, specimens with analyte concentrations that fall between nominal levels may give results at either level. Results at levels greater than the second positive level for the Protein, Glucose, Ketone, and Urobilinogen tests will usually be within one level of the true concentration. Exact agreement between visual results and instrumental results might not be found because of the inherent differences between the perception of the human eye and the optical system of the instruments.
Sensitivity and test range of urinalysis strips
Rem
| Sensitivity| Instrumental test range|
Visual test
range
---|---|---|---
Glucose (mmol/L)
| 2.8-5.5| Neg-55|
Neg-110
Protein (g/L)
| 0.15-0.3| Neg -3.0|
Neg — 20.0
Microalbumin (g/L)
| 0.08-0.15|
0-0.15
Ketone (acetoacetic acid) (mmol/L)
| 0.5-1.0| Neg-7.8|
Neg-16
Blood (Ery/uL)
| 5-15|
Neg. 200
Bilirubin(pmol /L)
| 3.3-8.6|
Neg.-100
Nitrite (pmol /L)
| 13-22|
Neg. or Pos.
Leukocytes (cells/pL)
| 5-15|
Neg. – 500
Urobilinogen (pmol /L)
| 3.3-16|
3.3-131
Ascorbic acid (mmoUL)
|
0.3-0.6
|
0-5.0
Creatinine(mg/dL)
| 25-75|
10-300
pH
| —|
5.0-8.5
Specific Gravity
| —| 1.005-1.030 I|
1.000-1.030
REACTIVE INGREDIENTS
(based on dry weight at time of impregnation)
PROTEIN: 0.1% m/m tetrabromphenol blue; 97.4% w/w buffer; 2.5% w/w
nonreactive ingredients
BLOOD: 26.0% w/w diisopropylbenzene dihydro peroxide; 1.5% w/w
tetramethyl-benzidine; 35.3% w/w buffer; 37.2%nonreactive in gredients.
GLUCOSE: 1.7% w/w glucose oxidase (microbial.123U); 0.2 % w/w
peroxidase(horseradish. 203 IU); 0.1% w/w potassium iodide; 71.8% w/w buffer;
26.2% w/w nonreactive ingredients.
KETONE: 5.7% w/w sodium nitroprusside; 29.9% w/w nonreactive
ingredients;64.4% w/w buffer;
LEUCOCYTES: 4.3% w/w pyrrole amino acid ester; 0.4% w/w diazonium salt;
92.6% w/w buffer; 2.7% w/w nonreactive ingredients.
NITRITE: 1.3% w/w p-arsanilicacid; 0.9% N-(1-Naphthol) ethylenediamine;
89.6%w/w buffer; 8.2% w/w nonreactive ingredients.
SPECIFIC GRAVITY: 4.8% w/w bromthymol blue; 90.2% w/w poly(methyl vinyl
ether co maleic anhydride); 5.0% w/w sodium hydroxide. PH: 3.3% w/w bromcresol
green; 55.0% w/w bromthymol blue; 41.7%w/w nonreactive ingredients.
BILIRUBIN: 0.6% w/w 2.4-dichlorbenzene amine diazonium salt; 57.3% w/w
buffer; 42.1% w/w nonreactive ingredients.
UROBILINOGEN: 0.2% w/w p-di ethylamino benzaldehyde 98.0% w/w buffer;
1.8% w/w nonreactive ingredients.
ASCORBIC ACID: 0.8% w/w 2,6-dichloroindophenolate hydrate; 40.7% w/w
buffer; 58.5% w/w nonreactive ingredients.
MICROALBUMIN: 2.2% w/w sulfonephthalein dyestuff; 96.0% w/w buffer; 1.8
w/w nonreactive ingredients.
CREATININE: 4.8% w/w 3, 5-2 nitro benzoic acid; 85.2% w/w buffer; 10% w/w
nonreactive ingredients.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
|
Product code
---|---
|
Consult instructions for use
|
Expiration date
|
Manufacture
|
In vitro diagnostic medical device compliant with Directive 98/79 / EC
|
Date of manufacture
|
Lot number
|
Temperature limit
|
In vitro diagnostic medical device
| Disposable device, do not re-use
|
Authorized representative in the European community
URS-10T (24076)
Zhejiang Orient Gene Biotech Co., Ltd 3787#, East Yangguang Avenue, Dipu Street Anji 313300, Huzhou, Zhejiang, China Made in China
Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537, Hamburg, Germany
Importado por:
Gima S.p.A. Via Marconi, 1 – 20060 Gessate (MI) Italy
gima@gimaitaly.com –
export@gimaitaly.com
www.gimaitaly.com
PROFESSIONAL MEDICALE PRODUCTS
http://www.gimaitaly.com/