GIMA V6k5c Electric Hospital Bed Instruction Manual
- June 2, 2024
- GIMA
Table of Contents
V6k5c Electric Hospital Bed
Product Information
Specifications
Manufacturer: Jiangsu Saikang Medical Equipment
Co.,Ltd
Origin: Made in China
Model: V6k5c (Gima 44744)
Drainage Hook: 2pcs
Product Usage Instructions
1. General Information
Before using the product, read the user manual thoroughly and
ensure all parts are complete as per the specifications.
2. Warnings
Duty of the Users: Users must comply with all
usage and safety precautions outlined in the manual.
Electrical Cabling and Connections: Connect the
bed only to a mains electricity supply with an earth wire.
3. Installation Guide
3.1 Product Installation
Follow the step-by-step guide provided in the manual for proper
installation of the bed, including attaching side rails, casters,
IV pole, bed platform, and bed board.
4. Maintenance
4.1 Spare Parts Replacement and Maintenance
Refer to the manual for instructions on motor removal and
replacement, troubleshooting, bed maintenance, mattress
maintenance, casters inspection, side rails maintenance, and list
of wearing and spare parts.
4.2 Bed Cleaning Precautions
Follow recommended cleaning precautions to ensure hygiene and
longevity of the product.
5. Technical Specifications
The product comes with aluminum alloy side rails, covered
casters, IV pole, and specific bed board dimensions as per the
specifications provided.
FAQ
Question: What should I do if I encounter unexpected situations
while using the product?
Answer: In case of unexpected situations, refer
to the troubleshooting section in the manual or contact the
after-sale service center for assistance.
LETTO OSPEDALIERO ELETTRICO ELECTRIC HOSPITAL BED LIT MÉDICALISÉ ÉLECTRIQUE CAMA ELÉCTRICA DE HOSPITAL
MANUALE D’USO USER MANUAL MODE D’EMPLOI MANUAL DE USO
· È necessario segnalare qualsiasi incidente grave verificatosi in relazione
al dispositivo medico da noi fornito al fabbricante e all’autorità competente
dello Stato membro in cui si ha sede.
· All serious accidents concerning the medical device supplied by us must be
reported to the manufacturer and competent authority of the member state where
your registered office is located.
· Il est nécessaire de signaler tout accident grave survenu et lié au
dispositif médical que nous avons livré au fabricant et à l’autorité
compétente de l’état membre où on a le siège social.
· Es necesario informar al fabricante y a la autoridad competente del Estado
miembro en el que se encuentra la sede sobre cualquier incidente grave que
haya ocurrido en relación con el producto sanitario que le hemos suministrado.
V6k5c (Gima 44744)
Jiangsu Saikang Medical Equipment Co.,Ltd No. 35 Lehong Road, Modern
Agriculture Demonstration Park, Zhangjiagang City, Jiangsu Province, China
Made in China
SUNGO Europe B.V. Fascinatio Boulevard 522, Unit 1.7, 2909VA Capelleaan den
IJssel, The Netherlands
Gima S.p.A. Via Marconi, 1 – 20060 Gessate (MI) Italy gima@gimaitaly.com –
export@gimaitaly.com www.gimaitaly.com
106kPa
85%
+40°C
%
70kPa
0%
-10°C
M44744-M-Rev.0-12.23
2
SUMMARY
USER MANUAL ……………………………………………………………………….. 2
GENERAL INFORMATION…………………………………………………………. 2
WARNINGS…………………………………………………………………………….. 2
TECHNICAL SPECIFICATION ……………………………………………………… 2
1.PRODUCT INSTALLATION GUIDE …………………………………………… 2 1.1 List of standard
accessories……………………………………………………2 1.2 Caster
installation……………………………………………………………….. 3 1.3 Bed board
installation………………………………………………………….. 3 1.4 Side rails installation
……………………………………………………………………3 1.5 Head foot board installation
…………………………………………………. 4
1.6 IV pole installation………………………………………………………………. 4
2.BED OPERATION………………………………………………………………….. 4 2.1 Casters
usage……………………………………………………………………… 4 2.2 Side rail usage
……………………………………………………………………. 5 2.3 Drainage hooks usage ………………………………………………………….
5 2.4 IV pole usage ……………………………………………………………………… 5 2.5 Hand controller
usage………………………………………………………….. 5
3.SPARE PARTS REPLACEMENT AND MAINTENANCE …………………. 5 3.1 Motor removal and
replacement …………………………………………… 5
4. TROUBLESHOOTING…………………………………………………………….. 5
5.BED MAINTENANCE …………………………………………………………….. 6
6.BED CLEANING PRECAUTIONS………………………………………………. 6
7.MATTRESS MAINTENANCE …………………………………………………… 6
8.CASTERS INSPECTION AND MAINTENANCE ……………………………. 6
9.SIDE RAILS MAINTENANCE …………………………………………………… 6
10.LIST OF WEARING AND SPARE PARTS…………………………………… 6
11.PACKAGING, TRANSPORTATION AND STORAGE CONDITIONS … 7 Symbol explanation:
…………………………………………………………………. 7
12.AFTER-SALES SERVICE ………………………………………………………… 7
ENGLISH
USER MANUAL
GENERAL INFORMATION Essential Performance: This bed is designed for
positioning and transporting patients, as an aid to diagnosing, monitoring,
treating and alleviating illnesses or compensating for injuries or
disabilities. The bed itself is not life sustaining or life supporting. The
bed has no medical indication. User Environment: hospitals and medical clinics
in application environments 2 or 3. The installation room must be equipped
with electrical installation in accordance with current standards. Personnel
for use of the product: specialized operators and medical personnel.
Supervision and responsibility: the bed should be used under the supervision
of a doctor. Limits of use: the bed can be used only as described in this
manual Essential Performance: This bed is designed for positioning and
transporting m Note for Users: Please carefully read this manual and strictly
follow the steps before installing and using. We require all users to have the
knowledge of user’s manual and all the cautions. Before carrying out any
operation on the device, qualified operators and technicians should read
carefully the instruction contained in this publication. This manual is
applicable to V6k5c electric hospital bed. It contains the installation
method, operating instructions and maintenance check points. The graphics in
this manual are for users’ reference only. In actual use, please refer to the
real object. This user manual includes the description, instruction,
maintenance and technical parameters of the product.
Any questions, please contact the after-sale service center or distributor.
Anyone using or operating this product should read and comply with it. Except
for some unexpected situations, users are required to be familiar withthe
usage and precautions
ENGLISH
WARNINGS
- DUTY OF THE USERS
– Before using a hospital bed, you, as the user, must check each time that the hospital bed is fully functional and in perfect working order and must observe the instructions in the instruction manual, particularly in the safety information, during operation and maintenance. This is the only way to prevent operating errors and ensure correct handling in order to prevent from injuries and damage from occurring. – Pay special attention here to the safe routing of all loose connector cables, tubing etc… if any. – Ensure that no obstacles such as bedside cabinets, supply rails or chairs could impede adjustments to the bed – Consult the manufacturer of the equipment in question, or the distributor, if there are any uncertainties. - ELECTRICAL CABLING AND CONNECTIONS Connect the bed only to a mains electricity supply with an earth wire
TECHNICAL SPECIFICATION
Model:V6K5C External Size (LxWxH): 2085995(420-730)mm Mattress Platform:
1925x900mm Safe Working Load: 200kg Back-rest Adjustment: 0-75°(±5°) Knee-rest
Adjustment: 0-46°(±5°) Technical Configuration: PP Platform with Integrated
Mattress Retainer, removable for sufficient sterilization Linear Motor: 3pcs
Hand controller: 1pc 5″ Covered Castors: 4pcs Foldable Side Rail: 1set PP Bed
Ends plug-in type, with safe lock: 1set Bumper wheels diameter 65mm: 4pcs IV
Pole 4 hooks: 1set IV Pole Prevision: 4pcs
Drainage Hook: 2pcs
1. PRODUCT INSTALLATION GUIDE 1.1 LIST OF STANDARD ACCESSORIES
After opening the packing case, please check whether the following parts are
complete. Table a:
N° Name
Picture
Specification
Qty
Unit
1
Head foot board packing box
Check table a.1 below
1
box
2 Side rail (left) 3 Side rail (right)
Aluminum alloy side rails
1
pc
Aluminum alloy side rails
1
pc
4 Caster
5″ covered caster
4
pc
5 IV pole
16-19/offwhite
1
pc
6 Bed platform
7 Bed board packing box
8 Screw pack
Table a.1
N° Name 1 Head board
2 Foot board
/ /
Picture
2048*1047
1
pc
check table a.2
1
box
below
check table a.3
1
bag
below
Specification
Qty
Unit
990*460
1
pc
990*460
1
pc
ENGLISH
3
Table a.2
N° Name
Picture
Specification
Qty
Unit
1.3 BED BOARD INSTALLATION
1 V21 bed board 1
733*903/PP/ white
1
pc
2 V21 bed board 2
3 V21 bed board 3
4 V21 bed board 4
Table a.3
N° Name 1 Hexagon head bolt 2 Flat washer 3 Spring washer 4 Jam nut 5 30-50
drainage hook 6 Asta 154
Picture
333*903/PP/ white
1
pc
373*903/PP/ white
1
473*903/PP/ white
1
pc Put the bed frame on legs, install the bed board; (the picture does not
exactly match the real object, for installation guide reference only) align
the holes on the sides of the bed board with the corresponding positioning
posts.
Board hole 1
pc
Positioning post
Specification
Qty
Unit
8
8
pc
8
16
pc
M8
12
pc
30-50 (plastic)
8
pc
Positioning post Board hole 2
Board hole 1 Board hole 2
21.7*154
2
pc
21.7*184
2
pc
Positioning post Board hole 3
Positioning post
Board hole 3
7 Pole 184
12
2
pc
8 Flat washer
M12
4
pc
9 Jam nut
M8*16
4
pc
10 Hexagon head bolt
M8*16
8
pc
1.2 CASTER INSTALLATION
Lay down a protective pad, and place the bed frame upside down; prepare the screws shown in the picture below, using 14″inch wrench
first, according to the pic(1), install the casters and fix bed leg, then
using 17″inch wrench
according to the pic(2), fix bed legs (for easy
observation the bed board is hidden here).
Positioning post Board hole 4
Positioning post Board hole 4
Positioning post
Foro bordo 4
Positioning post
Paletto di posizionamento
1.4 SIDE RAILS INSTALLATION
Put the bed upright, install side rails; (the picture does not exactly match the real
object, for installation guide reference only) prepare the screws shown in the picture
below, use 14″ inch wrench.
.
(1)
Spring
(2)
Flat washer washer 8 8
Bolt M8*16
Bolt M8*16
Spring washer 8
Flat washer 8
4
ENGLISH
IV pole prevision 1
IV pole prevision 3
Bullone M8*50 Shim 8
Controdado M8
1.5 HEAD FOOT BOARD INSTALLATION (the picture does not exactly match the real
object, for installation guide reference only). Head board poles: 21.7×154,
foot board poles: 21.7×184, the installation method is as shown in the figure
below:
IV pole prevision 2
IV pole prevision 4
Secondary hook part Primary hook part
2. BED OPERATION
13 predisposizione
2
Pole
Shim 12 Jam nut M12
Unlocked state
Locked state
1.6 IV POLE INSTALLATION
Primary part installation: place the primary plastic hook part on top of the
16mm Ø stainless steel pole and press it down.
Secondary part installation: place the secondary plastic hook part on top of
primary, align the secondary hook part’s snap with the primary’s socket and
press it down, after you hear a “click” sound, check if the clip is tight, if
so, the installation is completed, if not, continue to press down. Place IV
pole vertically in one of the IV
pole previsions.
1- IV pole 2- Head board
3- Side rail 4- Bed platform
5- Foot board
6- Bed frame
7- Drainage hook
8- Caster
length
1925±10
Bed platform dimension
width
900±10
height
420±10
Overall dimension Tilt range
length width height back rest knee rest
2085±10
mm
995±10
(420-730)±10
0-75±5°
/
0-46±5°
Weight
Safe working load
kg
200
This bed is not suitable for patients weighing less than 40 kg. The maximum
recommended patient weight is168 kg. The safe working load of the bed is 200
kg.
The safe working load is calculated as follows (in accordance with EN
60601-2-52):
ENGLISH
5
Maximum patient weight Mattress Accessories (including attached loads) TOTAL
168kg 7kg 25kg 200kg
2.1 CASTERS USAGE When there is a patient on the bed, raise the side rails before moving the bed.
Except for when the bed must be moved, the casters must be blocked via brake
switch. Each caster must be braked
2.1.1 After moving the bed to the desired place press the “ON” switch, the
“click” sound means the caster is braked, as shown below.
1
2
2.4 IV POLE USAGE
IV pole height adjustment: first unscrew the outer rotating sleeve,
counterclockwise, insert the inner tube of the IV pole into the rotating
sleeve, and then adjust the IV pole height as needed. After adjusting the
required height, tighten the outer rotating
sleeve, clockwise. Safe working load of each individual hook is 2kg, total safe working load is 8kg.
2.1.2 Lift the brake: press the “OFF” switch, the “click” sound means caster’s
brake is lifted.
2.2 SIDE RAIL USAGE
To raise side rail: hold the handle, raise it up, the “click” sound means the
rail is locked.
rotating sleeve
handle
(height adjustment range: 890-1540mm, the picture is for reference only). 2.5 HAND CONTROLLER USAGE (The picture is for reference only)
To lower side rail: squeeze the red handle, lower the side rail. handle
1
2
3
4
5
6
7
8
2.3 DRAINAGE HOOKS USAGE Drainage hooks as show below. (The picture is for reference only)
1 – back rest raise button 2 – back rest lower button 3 – knee rest raise
button 4 knee rest lower button 5 – back+knee rest raise button 6 – back_knee
rest lower button 7 – bed raise button 8 – bed lower button
3. SPARE PARTS REPLACEMENT AND MAINTENANCE 3.1 MOTOR REMOVAL AND REPLACEMENT
After removing the R-type lock pins at both ends of the motor, pull out the
lock pins. After motor replacement, install the bolts in the order shown
below. (the pic. is for reference only)
dr.hook
6
R-type lock pin
R-type lock pin
Lock pin
4. TROUBLESHOOTING
Lock pin
Common issues Difficulty in moving the bed Side rail easy to loosen
Loud noise when adjusting bed height
Electric bed functions failure The bed is not working
Possible reason 1. didn’t lift the brakes, locked casters 2. casters are
broken side rail isn’t snapped into place
rusted moving parts, not lubricated enough
Hand controller malfunction
The power plug is not inserted, connect it to the power grid
Troubleshooting 1. lift the brakes, see 2.1 2. replace casters, see 1.2 Raise
side rail up again, make sure “click” sound
Add appropriate amount of lubricating oil to the connecting places of moving
parts Replace the hand controller
Plug in, power on
5. BED MAINTENANCE
1. In order to use the bed safely, it is necessary to conduct regular safety
inspections on the bed, it’s recommended to conduct a comprehensive inspection
every six months to ensure the connection parts are not loose and the lifting
function of the bed is operating normally.
2. The connection of the frequently moving parts of the bed needs to be
lubricated with an appropriate amount of oil during daily inspection; is
serious wear is found, it should be replaced immediately to ensure safety. The
moving part in figure below:
1
2
Moving part
3. Avoid hitting the bed and scratching the coating on the bed surface with
sharp objects. 4. Please prevent the bed and mattress from absorbing water and
moisture, which may cause rust, abnormal noises and bacterial growth. If there
was water spilled, please wipe it off immediately. 5. When the bed ages and
reaches its service life time, the metal part and the plastic parts can be
recycled.
6. BED CLEANING PRECAUTIONS
1. When cleaning, please wring the cloth soaked in neutral detergent diluted
with water, wipe it dry, then wring the cloth soaked in clean water, wipe off
the residual detergent ingredients, and finally wipe with a dry cloth. 2. Do
not use volatile items (thinners, propellants, gasoline, etc) that may cause
chemical reactions and damage the bed. 3. When cleaning with a disinfectant,
be sure to use it after diluting it at the specified concentration. Depending
on the composition of the disinfectant, it may corrode metal parts, resin
parts, etc, causing discoloration, deformation and other undesirable
consequences. Therefore, the recommended content of the disinfectant is as
follows: 0.05~0.2% ammonium chloride 0.05~0.2% chlorinated phenyl 0.05%
dichlorobenzene biguanide ethane 0.05~0.2% sodium hypochlorite Do not use a
smoking sterilizer or autoclave, and do not use methyl phenol to clean the
head foot board, bed feet, etc, which may cause corrosion, discoloration,
deterioration. Note: When cleaning the bed or changing the bedding, pay
attention to the corners, edges and screw parts of the frame to prevent
scratches.
7. MATTRESS MAINTENANCE
ENGLISH
1. Avoid scratching the fabric with sharp-angled appliances or knives when
using it, cover it with sheets or cleaning pads, and clean it frequently to
keep it clean and dry.
2. Regularly clean the mattress cover with a vacuum cleaner, it can be washed
directly with water or detergent.
3. Air the mattress often to keep your sleep fresh and comfortable, but be
careful not to expose the mattress to the sun for too long.
4. If the fabric is accidentally stained, please refer to the corresponding
recommen-
ded cleaning method in the instructions. Do not use corrosive chemicals to
clean the fabric.
5. If you accidentally get chewing gum or other glue-like objects on the
fabric, do not wipe it forcibly. You can use am ice pack to allow it to harden
and then peel it off gently.
6. If the thread is found to be loose due to wear and tear, it cannot be
broken by hand, and it should be cut short with scissors.
- Your order may not have a mattress included, please refer to the contract configuration.
8. CASTERS INSPECTION AND MAINTENANCE
1. Check the casters regularly, avoid hitting heavy objects or overloading,
which will cause damage to the casters.
2. The casters should be applied with lubricating grease regularly, the
frequency of appliance should be determined according to the specific actual
use situation. Under normal conditions, the grease should be applied every six
month.
3. After cleaning the caster equipment, pay attention to adding lubricating
grease. The selection of lubricating grease should be based on the actual use
environment, special high low temperature environment, etc.
4. It’s forbidden to push the bed with force when the casters are braked,
which will cause damage to the braking system.
5. The wear of the tire tread of the caster can be detected by visual
inspection. Some accumulations such as thread and spun yarn may wrap around
the caster. Remove the bolts and nuts of thecaster, clean up the debris, and
check whether the bearing of the caster is damaged. If the parts are not
damaged, the caster can be reassembled and used.
6. Decision whether to replace entire caster is according to the wear
condition. After the caster is replaced or reinstalled, make sure that the
axle bolt and nut are connected reliably, and use lock washers or lock nuts as
much as possible.
7. For casters equipped with brakes, it is necessary to regularly check
whether the brakes work normally, and check the brakes every day or before
each use. For a hospital bed equipped with multiple brake casters, you can
lock only one brake caster at a time and try to push the bed to check whether
the braking performance of each caster is good. If the caster brake fails to
function due to wear or damage to the caster, replace the caster immediately
and retest the brake system.
8. If the mechanism of the brake system is damaged, and the brake needs to be
repaired or replaced, please contact our after- sales or an authorized dealer.
Every time the brakes arereplaced, the brake performance of the casters needs
to be retested.
9. SIDE RAILS MAINTENANCE
1. Please check whether the side rails are installed correctly to prevent the
lying patient from shifting and rolling off. 2. Please always check the screws
of the side rail and the side rail fittings, to prevent loosening and falling
off. 3. Please use the red part of the side rail handle correctly, do not use
brute force to pull hard, which will cause damage to the side rail.
10. LIST OF WEARING AND SPARE PARTS
Name
Material code
Picture
Specification
1
V21 head foot board
GPSL101TS648
2
Aluminum alloy side rail, left
3
Aluminum alloy side rail, right
BKLLGH105943 BKLLGH105944
Stainless Steel IV 4 pole with 4 plastic
hooks
S0FJSY010003
990460
Six sections / QB/2605/ 101170
Six sections/QB /2605/10*1170
16-19/off-white
5″covered caster 5 with brake,slide bar LGPLU101TS122
with bolt 23.6
M8(QB)
6 Drainage hook
BYKBGKM02806
Transparent 7 bedside card
holder(1 card)
GPSL101TS184
30-50 (plastic) /
ENGLISH
7
8 V21 bed board 1
GPSL101TS610
733*903/PP/white
9 V21 bed board 2
GPSL101TS611
10 V21 bed board 3
GPSL101TS613
333903/PP/white 373903/PP/white
11 V21 bed board 4
GPSL101TS614
12
PP side rail shock damper
LGPQT101TS02 2
13
HJ motor 200, horizontal
LGPDJ101TS573
14
HJ motor 150, vertical
LGPDJ101TS574
473*903/PP/white
18-60-170-35N
HJA58/stroke 200mm
HJA58/stroke 150mm
15
HJ motor 100, vertical
LGPDJ101TS575
HJA58/stroke 100mm
16
HJ controller, 3 functions
LGPDJ101TS576
/
HJT13E power on/
off/100-240V/50/
60Hz/white
17
HJ hand controller, 3 functions
LGPDJ101TS577
HJH18 hand
controller/8 buttons/white
18
V21 back rest rail sticker
LGPFU101TS41 6
780350.4
19
V21 knee rest rail sticker
LGPFU101TS41 7
750350.4
Note: Due to the renewal of product lineup, the parts of the original product
may not be available. (The company ensures availability of the parts for sold
products within 6 years after their production ends).
11. PACKAGING, TRANSPORTATION AND STORAGE CONDITIONS
1. The packaging of the manual bed is carried out according to the contract
or product standard. 2. During the transportation of the product, avoid
rushing out, violent vibration, prevent direct sun and rain access. 3. Store
conditions: a) Ambient temperature: -10°C~ +40°C b) Relative humidity: 85% c)
Atmospheric pressure: 70 kPa ~ 106 kPa
ELECTRIC VARIABLE HEIGHT BED V6k5c
Supply voltage
V
Network frequency
Hz
Operating voltage
VA
Class of electrical protection
–
Applied part
–
IP Protection
100-230V a.c. 60 Hz 200 VA I B IPX4
SYMBOL EXPLANATION:
Notice: Read attachment paper
Safe working load is 2000N
Max bear loading is 200KG
12. AFTER-SALES SERVICE
1. C1. Please preserve the attachment papers and invoice of this product, you
need to present those to enjoy after-sales maintenance service.
2. If you have any question during use, please contact us any time, so that
our company can provide you with accurate and efficient technical support and
maintenance services in a timely manner.
3. Do not disassemble the internal parts of this product by yourself to avoid
unnecessary damage. I you find that the use is affected by quality issues,
please contact the company’s after- sales service department or the authorized
distributor.
4. The warranty is guaranteed according to the contract requirements. Outside
the
warranty period, the company is responsible for maintenance, providing
accessories, and implementing life-long service for clients.
5. After-sales service unit: Jiangsu Saikang Medical Equipment Co., Ltd.
Notice!any serious incident that has occurred in relation to the device shall
be
reported to the manufacturer and the competent authority of the Member State
in which the user and/or the patient is established
EMC
- This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
-
- Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
- Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
-
- Caution: This machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be use.
Guidance and Manufacturer’s Declaration Electromagnetic Immunity
The V6k5c is intended for use in the following electromagnetic environment. The customer or user of the V6k5c should make sure it is used in such an environment
Emission test
Compliance Electromagnetic Environment – Guidance
RF Emission to CISPR 11
Group 1
The bed uses RF energy for its internal function
RF Emission to CISPR 11
Harmonics according to IEC 61000-3-2
Class B Class A
Voltage fluctuations/ flicker acc. To IEC 61000-3-3
Complies
The bed is intended for use in all types of establishment including residential and similar user that are directly connected to a public supply network that also serves buildings used for residential purposes
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The V6k5c is intended for use within healthcare facility environment. Users or customers must use bed in this environment
Immunity test
IEC 60601-12/ EN 60601-1-2
Level test
Compliance level Electromagnetic environment guidance
Electrostat
ic discharge
(ESD) IEC 61000-42 /EN 61000-4-2
Contact discharge: +8kV Air discharge: +2kV, +4kV, +8kV, +15kV
Contact discharge: +8kV Scarica in aria: +2kV, +4kV, +8kV, +15kV
Floor should be made of wood and
concrete or be tiled with ceramic tiles. If the floor is covered with
synthetic flooring material, the relative air humidity must be at least 30%
Can be used when higher ESD levels
are present
Electrical fast
transient/burst 61000-4-4 /EN 61000-4-4
+2kV Power
supply line +1kV Input/Output lines
Linea di alimentazione +2kV
The quality of the supply voltage should be equivalent to that of a typical business or hospital environment.
Surge
IEC 61000-4-5 /EN 61000-4-5
Line – line: +0.5kV, +1.0kV Line ground: +0.5kV,
+1.0kV, +2.0kV
Linea – linea: +0,5kV, +1,0kV
Linea – terra: +0,5kV, +1,0kV,
+2,0kV
The quality of the supply voltage should be equivalent to that of a typical business or hospital environment.
Power
Frequency
Magnetic Fields IEC 61000-4-8 /EN 61000-4-8
30 A/m 60Hz
30 A/m 60Hz
Magnetic fields with a network frequency should be equivalent to
those to be found in a typical business
or hospital environment.
Voltage Dips
IEC 61000-411
/EN 61000-4-11
0% UT: 0.5 cycles Phase: 00, 450, 900, 1350, 1800, 2250, 2700, 3150
0% UT: 1 cycle 70% UT: 25/30 cycles Single phase: 00 0% UT ; 300 cycle
0% UT: 0,5 cicli Fase: 00, 450, 900, 1350, 1800, 2250, 2700, 3150
0% UT: 1 ciclo 70% UT: 25/30 cicli Monofase: 00 0% UT; 300 cicli
The quality of the supply voltage should be equivalent to that of a typical business or hospital environment. If the person using the bed requires that the bed functions must continue despite any interruptions in the energy supply, it is recommended that the bed be connected to an uninterruptible electricity supply or a battery
Short
interruptions IEC 61000-411 /EN 61000-4-11
0% UT: 250/300 cycles
0% UT: 250/300 cicli
The quality of the supply voltage should be equivalent to that of a typical business or hospital environment. If the person using the bed requires that the bed functions must continue despite any interruptions in the energy supply, it is recommended that the bed be connected to an uninterruptible electricity supply or a battery
Note: UT is the AC Power voltage before applying in the test level
8
Guidance and Manufacturer’s Declaration Electromagnetic Immunity
The V6k5c is intended for use in an electromagnetic environment that used wireless RF transceiver. Electromagnetic interference can be prevented by maintaining the minimum distance between RF transceiver and V6k5c depend on the maximum output and frequency of the communication device as recommended below.
Immunity test
IEC 606011-2 /EN 60601-1 2 Test level
27 V/m 385 MHz
28 V/m 450 MHz
9 V/m
710 MHz Enclosure
port to RF 9 V/m
wireless 745 MHz
communication sequi-
pment IEC
9 V/m 780 MHz
61000-4- 28 V/m
3
810 MHz
/EN 610004-
3
28 V/m 870 MHz
28 V/m 930 MHz
Complian Band elevel (MHz) (V/m)
Service
Modulatio n Maximum Distance
power
(m)
(W)
27
380390
TETRA 400
Pulse modulatio
n18Hz
1,8
0,3
28
430470
GMRS 460, FRS 460
FM+5 kHz deviation
1 kHz sine
2
0,3
LTE Banda
Pulse
9
704787
13, 17
modulatio n217Hz
0,2
0,3
GSM
800/900,
TETRA 800,
Pulse
28
800960
iDen 820, CDMA 850
modulatio n18Hz
2
0,3
LTE Band 5
Enclosure 28 V/m port to RF 1720 MHz wireless communication sequipment IEC
61000-43 /EN 6100043
28 V/m 1845 MHz
28 V/m 1970 MHz
28 V/m 2450 MHz
9 V/m 5240 MHz
9 V/m 5500 MHz
9 V/m 5785 MHz
GSM 1800;
CDMA 1900;
GSM 1900;
Pulse
28
1700-
DECT;
modulatio
2
0,3
1990 LTE Band
n217Hz
1,3,
4,25; UMTS
Bluetooth,
Pulse
28
2400-
WLAN,
modulatio
2
0,3
2570 802.11 b/g/n n18Hz
RFID 2450,
band LTE 7
9
5100- WLAN 802,11 Pulse
0,2
0,3
5800
a/n
modulatio
n217Hz
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The V6k5c is intended for use in the following electromagnetic environment. The customer or user of the V6k5c should make sure it is used in such an environment.
Immunity test
IEC 60601-1-2 /EN 60601-1-2
Livello test
Livello di conformità
Electromagnetic Environment Guidance
RF Conductivity
IEC 61000-4-6 /EN 61000-4-6
3V 0,15 MHz-80 MHz
6V in ISM bands between 0.15MHz80MHz 80%AM at 1kHz
3V 0,15MHz80 MHz
6V in ISM bands between 0.15MHz-80MHz 80%AM at 1kHz
Field intensity from the fixed RF transmitter determined by a field
investigation of the magnetic field must be lower than the compliance level of
the respective frequencyband.
Near devices with the following symbol, interference could occur.
Radioactive RF
electromagnetic
field IEC 61000-4-3 /EN 61000-4-3
3V/M 80MHZ-2,7GHz
80%AM a 1kHz
3V/M 80MHZ-2,7GHz
80%AM a 1kHz
a. The intensity of a magnetic field from fixed transmitters such as a radio
(mobile/radio) telephone base station, land mobile radio, amateur radio, AM
and FM radio broadcasting, and TV broadcasting
cannot be theoretically predicted accurately. To determine the electromagnetic
environment for a fixed RF transmitter, consider investigating the
electromagnetic field in the location where the bed is used. If the intensity
of the magnetic field in the location where the V6k5c is used exceeds the
above RF compliance level, make sure tomonitor if the V6k5c operates
appropriately. If abnormal movement is
found, take additional measures as needed, such as changing the direction or
position of the V6k5c.
b.These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption andreflection from structures, objects,
and people.
ENGLISH
SYMBOL EXPLANATION:
Keep away from sunlight
Product code
Date of manufacture Caution: read instructions (warnings) carefully Imported
by
Medical Device compliant with Regulation (EU) 2017/745 Follow instructions for
use Lot number
Temperature limit
Serial number
Atmospheric pressure limit
Keep in a cool, dry place
Manufacturer
Authorized representative in the European community
Type B applied part
Medical Device
%
Humidity limit
WEEE disposal
Press the red handle to lower the side rail(unlock), to secure press the red handle and push forward until you hear click
Disposal: The product must not be disposed of along with other domestic waste.
The users must dispose of this equipment by bringing it to a specific
recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS The Gima 12-month standard B2B warranty applies.