GIMA V6k5c Electric Hospital Bed Instruction Manual

June 2, 2024
GIMA

V6k5c Electric Hospital Bed

Product Information

Specifications

Manufacturer: Jiangsu Saikang Medical Equipment
Co.,Ltd

Origin: Made in China

Model: V6k5c (Gima 44744)

Drainage Hook: 2pcs

Product Usage Instructions

1. General Information

Before using the product, read the user manual thoroughly and
ensure all parts are complete as per the specifications.

2. Warnings

Duty of the Users: Users must comply with all
usage and safety precautions outlined in the manual.

Electrical Cabling and Connections: Connect the
bed only to a mains electricity supply with an earth wire.

3. Installation Guide

3.1 Product Installation

Follow the step-by-step guide provided in the manual for proper
installation of the bed, including attaching side rails, casters,
IV pole, bed platform, and bed board.

4. Maintenance

4.1 Spare Parts Replacement and Maintenance

Refer to the manual for instructions on motor removal and
replacement, troubleshooting, bed maintenance, mattress
maintenance, casters inspection, side rails maintenance, and list
of wearing and spare parts.

4.2 Bed Cleaning Precautions

Follow recommended cleaning precautions to ensure hygiene and
longevity of the product.

5. Technical Specifications

The product comes with aluminum alloy side rails, covered
casters, IV pole, and specific bed board dimensions as per the
specifications provided.

FAQ

Question: What should I do if I encounter unexpected situations

while using the product?

Answer: In case of unexpected situations, refer
to the troubleshooting section in the manual or contact the
after-sale service center for assistance.

LETTO OSPEDALIERO ELETTRICO ELECTRIC HOSPITAL BED LIT MÉDICALISÉ ÉLECTRIQUE CAMA ELÉCTRICA DE HOSPITAL

MANUALE D’USO ­ USER MANUAL ­ MODE D’EMPLOI ­ MANUAL DE USO

· È necessario segnalare qualsiasi incidente grave verificatosi in relazione al dispositivo medico da noi fornito al fabbricante e all’autorità competente dello Stato membro in cui si ha sede.
· All serious accidents concerning the medical device supplied by us must be reported to the manufacturer and competent authority of the member state where your registered office is located.
· Il est nécessaire de signaler tout accident grave survenu et lié au dispositif médical que nous avons livré au fabricant et à l’autorité compétente de l’état membre où on a le siège social.
· Es necesario informar al fabricante y a la autoridad competente del Estado miembro en el que se encuentra la sede sobre cualquier incidente grave que haya ocurrido en relación con el producto sanitario que le hemos suministrado.

V6k5c (Gima 44744)
Jiangsu Saikang Medical Equipment Co.,Ltd No. 35 Lehong Road, Modern Agriculture Demonstration Park, Zhangjiagang City, Jiangsu Province, China Made in China
SUNGO Europe B.V. Fascinatio Boulevard 522, Unit 1.7, 2909VA Capelleaan den IJssel, The Netherlands
Gima S.p.A. Via Marconi, 1 – 20060 Gessate (MI) Italy gima@gimaitaly.com – export@gimaitaly.com www.gimaitaly.com

106kPa

85%

+40°C

%

70kPa

0%

-10°C

M44744-M-Rev.0-12.23

2
SUMMARY
USER MANUAL ……………………………………………………………………….. 2
GENERAL INFORMATION…………………………………………………………. 2
WARNINGS…………………………………………………………………………….. 2
TECHNICAL SPECIFICATION ……………………………………………………… 2
1.PRODUCT INSTALLATION GUIDE …………………………………………… 2 1.1 List of standard accessories……………………………………………………2 1.2 Caster installation……………………………………………………………….. 3 1.3 Bed board installation………………………………………………………….. 3 1.4 Side rails installation ……………………………………………………………………3 1.5 Head foot board installation …………………………………………………. 4
1.6 IV pole installation………………………………………………………………. 4
2.BED OPERATION………………………………………………………………….. 4 2.1 Casters usage……………………………………………………………………… 4 2.2 Side rail usage ……………………………………………………………………. 5 2.3 Drainage hooks usage …………………………………………………………. 5 2.4 IV pole usage ……………………………………………………………………… 5 2.5 Hand controller usage………………………………………………………….. 5
3.SPARE PARTS REPLACEMENT AND MAINTENANCE …………………. 5 3.1 Motor removal and replacement …………………………………………… 5
4. TROUBLESHOOTING…………………………………………………………….. 5
5.BED MAINTENANCE …………………………………………………………….. 6
6.BED CLEANING PRECAUTIONS………………………………………………. 6
7.MATTRESS MAINTENANCE …………………………………………………… 6
8.CASTERS INSPECTION AND MAINTENANCE ……………………………. 6
9.SIDE RAILS MAINTENANCE …………………………………………………… 6
10.LIST OF WEARING AND SPARE PARTS…………………………………… 6
11.PACKAGING, TRANSPORTATION AND STORAGE CONDITIONS … 7 Symbol explanation: …………………………………………………………………. 7
12.AFTER-SALES SERVICE ………………………………………………………… 7
ENGLISH
USER MANUAL
GENERAL INFORMATION Essential Performance: This bed is designed for positioning and transporting patients, as an aid to diagnosing, monitoring, treating and alleviating illnesses or compensating for injuries or disabilities. The bed itself is not life sustaining or life supporting. The bed has no medical indication. User Environment: hospitals and medical clinics in application environments 2 or 3. The installation room must be equipped with electrical installation in accordance with current standards. Personnel for use of the product: specialized operators and medical personnel. Supervision and responsibility: the bed should be used under the supervision of a doctor. Limits of use: the bed can be used only as described in this manual Essential Performance: This bed is designed for positioning and transporting m Note for Users: Please carefully read this manual and strictly follow the steps before installing and using. We require all users to have the knowledge of user’s manual and all the cautions. Before carrying out any operation on the device, qualified operators and technicians should read carefully the instruction contained in this publication. This manual is applicable to V6k5c electric hospital bed. It contains the installation method, operating instructions and maintenance check points. The graphics in this manual are for users’ reference only. In actual use, please refer to the real object. This user manual includes the description, instruction, maintenance and technical parameters of the product.
Any questions, please contact the after-sale service center or distributor. Anyone using or operating this product should read and comply with it. Except for some unexpected situations, users are required to be familiar withthe usage and precautions

ENGLISH

WARNINGS

  1. DUTY OF THE USERS
    – Before using a hospital bed, you, as the user, must check each time that the hospital bed is fully functional and in perfect working order and must observe the instructions in the instruction manual, particularly in the safety information, during operation and maintenance. This is the only way to prevent operating errors and ensure correct handling in order to prevent from injuries and damage from occurring. – Pay special attention here to the safe routing of all loose connector cables, tubing etc… if any. – Ensure that no obstacles such as bedside cabinets, supply rails or chairs could impede adjustments to the bed – Consult the manufacturer of the equipment in question, or the distributor, if there are any uncertainties.
  2. ELECTRICAL CABLING AND CONNECTIONS Connect the bed only to a mains electricity supply with an earth wire

TECHNICAL SPECIFICATION
Model:V6K5C External Size (LxWxH): 2085995(420-730)mm Mattress Platform: 1925x900mm Safe Working Load: 200kg Back-rest Adjustment: 0-75°(±5°) Knee-rest Adjustment: 0-46°(±5°) Technical Configuration: PP Platform with Integrated Mattress Retainer, removable for sufficient sterilization Linear Motor: 3pcs Hand controller: 1pc 5″ Covered Castors: 4pcs Foldable Side Rail: 1set PP Bed Ends plug-in type, with safe lock: 1set Bumper wheels diameter 65mm: 4pcs IV Pole 4 hooks: 1set IV Pole Prevision: 4pcs
Drainage Hook: 2pcs

1. PRODUCT INSTALLATION GUIDE 1.1 LIST OF STANDARD ACCESSORIES
After opening the packing case, please check whether the following parts are complete. Table a:

N° Name

Picture

Specification

Qty

Unit

1

Head foot board packing box

Check table a.1 below

1

box

2 Side rail (left) 3 Side rail (right)

Aluminum alloy side rails

1

pc

Aluminum alloy side rails

1

pc

4 Caster

5″ covered caster

4

pc

5 IV pole

16-19/offwhite

1

pc

6 Bed platform
7 Bed board packing box
8 Screw pack
Table a.1
N° Name 1 Head board
2 Foot board

/ /
Picture

2048*1047

1

pc

check table a.2

1

box

below

check table a.3

1

bag

below

Specification

Qty

Unit

990*460

1

pc

990*460

1

pc

ENGLISH

3

Table a.2
N° Name

Picture

Specification

Qty

Unit

1.3 BED BOARD INSTALLATION

1 V21 bed board 1

733*903/PP/ white

1

pc

2 V21 bed board 2
3 V21 bed board 3
4 V21 bed board 4
Table a.3
N° Name 1 Hexagon head bolt 2 Flat washer 3 Spring washer 4 Jam nut 5 30-50 drainage hook 6 Asta 154

Picture

333*903/PP/ white

1

pc

373*903/PP/ white

1

473*903/PP/ white

1

pc Put the bed frame on legs, install the bed board; (the picture does not exactly match the real object, for installation guide reference only) align the holes on the sides of the bed board with the corresponding positioning posts.
Board hole 1
pc
Positioning post

Specification

Qty

Unit

8

8

pc

8

16

pc

M8

12

pc

30-50 (plastic)

8

pc

Positioning post Board hole 2

Board hole 1 Board hole 2

21.7*154

2

pc

21.7*184

2

pc

Positioning post Board hole 3

Positioning post

Board hole 3

7 Pole 184

12

2

pc

8 Flat washer

M12

4

pc

9 Jam nut

M8*16

4

pc

10 Hexagon head bolt

M8*16

8

pc

1.2 CASTER INSTALLATION

Lay down a protective pad, and place the bed frame upside down; prepare the screws shown in the picture below, using 14″inch wrench

first, according to the pic(1), install the casters and fix bed leg, then

using 17″inch wrench

according to the pic(2), fix bed legs (for easy

observation the bed board is hidden here).

Positioning post Board hole 4

Positioning post Board hole 4

Positioning post

Foro bordo 4

Positioning post

Paletto di posizionamento

1.4 SIDE RAILS INSTALLATION

Put the bed upright, install side rails; (the picture does not exactly match the real

object, for installation guide reference only) prepare the screws shown in the picture

below, use 14″ inch wrench.

.

(1)

Spring

(2)

Flat washer washer 8 8

Bolt M8*16

Bolt M8*16

Spring washer 8

Flat washer 8

4

ENGLISH

IV pole prevision 1

IV pole prevision 3

Bullone M8*50 Shim 8
Controdado M8
1.5 HEAD FOOT BOARD INSTALLATION (the picture does not exactly match the real object, for installation guide reference only). Head board poles: 21.7×154, foot board poles: 21.7×184, the installation method is as shown in the figure below:

IV pole prevision 2

IV pole prevision 4
Secondary hook part Primary hook part

2. BED OPERATION

13 predisposizione
2

Pole
Shim 12 Jam nut M12

Unlocked state

Locked state

1.6 IV POLE INSTALLATION
Primary part installation: place the primary plastic hook part on top of the 16mm Ø stainless steel pole and press it down.
Secondary part installation: place the secondary plastic hook part on top of primary, align the secondary hook part’s snap with the primary’s socket and press it down, after you hear a “click” sound, check if the clip is tight, if so, the installation is completed, if not, continue to press down. Place IV pole vertically in one of the IV
pole previsions.

1- IV pole 2- Head board

3- Side rail 4- Bed platform

5- Foot board

6- Bed frame

7- Drainage hook

8- Caster

length

1925±10

Bed platform dimension

width

900±10

height

420±10

Overall dimension Tilt range

length width height back rest knee rest

2085±10

mm

995±10

(420-730)±10

0-75±5°

/

0-46±5°

Weight

Safe working load

kg

200

This bed is not suitable for patients weighing less than 40 kg. The maximum recommended patient weight is168 kg. The safe working load of the bed is 200 kg.
The safe working load is calculated as follows (in accordance with EN 60601-2-52):

ENGLISH

5

Maximum patient weight Mattress Accessories (including attached loads) TOTAL

168kg 7kg 25kg 200kg

2.1 CASTERS USAGE When there is a patient on the bed, raise the side rails before moving the bed.

Except for when the bed must be moved, the casters must be blocked via brake switch. Each caster must be braked
2.1.1 After moving the bed to the desired place press the “ON” switch, the “click” sound means the caster is braked, as shown below.

1

2

2.4 IV POLE USAGE
IV pole height adjustment: first unscrew the outer rotating sleeve, counterclockwise, insert the inner tube of the IV pole into the rotating sleeve, and then adjust the IV pole height as needed. After adjusting the required height, tighten the outer rotating

sleeve, clockwise. Safe working load of each individual hook is 2kg, total safe working load is 8kg.

2.1.2 Lift the brake: press the “OFF” switch, the “click” sound means caster’s brake is lifted.
2.2 SIDE RAIL USAGE
To raise side rail: hold the handle, raise it up, the “click” sound means the rail is locked.

rotating sleeve

handle

(height adjustment range: 890-1540mm, the picture is for reference only). 2.5 HAND CONTROLLER USAGE (The picture is for reference only)

To lower side rail: squeeze the red handle, lower the side rail. handle

1

2

3

4

5

6

7

8

2.3 DRAINAGE HOOKS USAGE Drainage hooks as show below. (The picture is for reference only)

1 – back rest raise button 2 – back rest lower button 3 – knee rest raise button 4 knee rest lower button 5 – back+knee rest raise button 6 – back_knee rest lower button 7 – bed raise button 8 – bed lower button
3. SPARE PARTS REPLACEMENT AND MAINTENANCE 3.1 MOTOR REMOVAL AND REPLACEMENT
After removing the R-type lock pins at both ends of the motor, pull out the lock pins. After motor replacement, install the bolts in the order shown below. (the pic. is for reference only)

dr.hook

6
R-type lock pin

R-type lock pin

Lock pin
4. TROUBLESHOOTING

Lock pin

Common issues Difficulty in moving the bed Side rail easy to loosen
Loud noise when adjusting bed height
Electric bed functions failure The bed is not working

Possible reason 1. didn’t lift the brakes, locked casters 2. casters are broken side rail isn’t snapped into place
rusted moving parts, not lubricated enough
Hand controller malfunction
The power plug is not inserted, connect it to the power grid

Troubleshooting 1. lift the brakes, see 2.1 2. replace casters, see 1.2 Raise side rail up again, make sure “click” sound
Add appropriate amount of lubricating oil to the connecting places of moving parts Replace the hand controller
Plug in, power on

5. BED MAINTENANCE
1. In order to use the bed safely, it is necessary to conduct regular safety inspections on the bed, it’s recommended to conduct a comprehensive inspection every six months to ensure the connection parts are not loose and the lifting function of the bed is operating normally.
2. The connection of the frequently moving parts of the bed needs to be lubricated with an appropriate amount of oil during daily inspection; is serious wear is found, it should be replaced immediately to ensure safety. The moving part in figure below:

1

2

Moving part
3. Avoid hitting the bed and scratching the coating on the bed surface with sharp objects. 4. Please prevent the bed and mattress from absorbing water and moisture, which may cause rust, abnormal noises and bacterial growth. If there was water spilled, please wipe it off immediately. 5. When the bed ages and reaches its service life time, the metal part and the plastic parts can be recycled.
6. BED CLEANING PRECAUTIONS
1. When cleaning, please wring the cloth soaked in neutral detergent diluted with water, wipe it dry, then wring the cloth soaked in clean water, wipe off the residual detergent ingredients, and finally wipe with a dry cloth. 2. Do not use volatile items (thinners, propellants, gasoline, etc) that may cause chemical reactions and damage the bed. 3. When cleaning with a disinfectant, be sure to use it after diluting it at the specified concentration. Depending on the composition of the disinfectant, it may corrode metal parts, resin parts, etc, causing discoloration, deformation and other undesirable consequences. Therefore, the recommended content of the disinfectant is as follows: 0.05~0.2% ammonium chloride 0.05~0.2% chlorinated phenyl 0.05% dichlorobenzene biguanide ethane 0.05~0.2% sodium hypochlorite Do not use a smoking sterilizer or autoclave, and do not use methyl phenol to clean the head foot board, bed feet, etc, which may cause corrosion, discoloration, deterioration. Note: When cleaning the bed or changing the bedding, pay attention to the corners, edges and screw parts of the frame to prevent scratches.
7. MATTRESS MAINTENANCE

ENGLISH

1. Avoid scratching the fabric with sharp-angled appliances or knives when using it, cover it with sheets or cleaning pads, and clean it frequently to keep it clean and dry.
2. Regularly clean the mattress cover with a vacuum cleaner, it can be washed directly with water or detergent.
3. Air the mattress often to keep your sleep fresh and comfortable, but be careful not to expose the mattress to the sun for too long.
4. If the fabric is accidentally stained, please refer to the corresponding recommen-

ded cleaning method in the instructions. Do not use corrosive chemicals to clean the fabric.
5. If you accidentally get chewing gum or other glue-like objects on the fabric, do not wipe it forcibly. You can use am ice pack to allow it to harden and then peel it off gently.
6. If the thread is found to be loose due to wear and tear, it cannot be broken by hand, and it should be cut short with scissors.

  • Your order may not have a mattress included, please refer to the contract configuration.

8. CASTERS INSPECTION AND MAINTENANCE
1. Check the casters regularly, avoid hitting heavy objects or overloading, which will cause damage to the casters.
2. The casters should be applied with lubricating grease regularly, the frequency of appliance should be determined according to the specific actual use situation. Under normal conditions, the grease should be applied every six month.
3. After cleaning the caster equipment, pay attention to adding lubricating grease. The selection of lubricating grease should be based on the actual use environment, special high low temperature environment, etc.
4. It’s forbidden to push the bed with force when the casters are braked, which will cause damage to the braking system.
5. The wear of the tire tread of the caster can be detected by visual inspection. Some accumulations such as thread and spun yarn may wrap around the caster. Remove the bolts and nuts of thecaster, clean up the debris, and check whether the bearing of the caster is damaged. If the parts are not damaged, the caster can be reassembled and used.
6. Decision whether to replace entire caster is according to the wear condition. After the caster is replaced or reinstalled, make sure that the axle bolt and nut are connected reliably, and use lock washers or lock nuts as much as possible.
7. For casters equipped with brakes, it is necessary to regularly check whether the brakes work normally, and check the brakes every day or before each use. For a hospital bed equipped with multiple brake casters, you can lock only one brake caster at a time and try to push the bed to check whether the braking performance of each caster is good. If the caster brake fails to function due to wear or damage to the caster, replace the caster immediately and retest the brake system.
8. If the mechanism of the brake system is damaged, and the brake needs to be repaired or replaced, please contact our after- sales or an authorized dealer. Every time the brakes arereplaced, the brake performance of the casters needs to be retested.

9. SIDE RAILS MAINTENANCE
1. Please check whether the side rails are installed correctly to prevent the lying patient from shifting and rolling off. 2. Please always check the screws of the side rail and the side rail fittings, to prevent loosening and falling off. 3. Please use the red part of the side rail handle correctly, do not use brute force to pull hard, which will cause damage to the side rail.

10. LIST OF WEARING AND SPARE PARTS

Name

Material code

Picture

Specification

1

V21 head foot board

GPSL101TS648

2

Aluminum alloy side rail, left

3

Aluminum alloy side rail, right

BKLLGH105943 BKLLGH105944

Stainless Steel IV 4 pole with 4 plastic
hooks

S0FJSY010003

990460
Six sections / QB/260
5/ 101170
Six sections/QB /260
5/10*1170
16-19/off-white

5″covered caster 5 with brake,slide bar LGPLU101TS122
with bolt 23.6

M8(QB)

6 Drainage hook

BYKBGKM02806

Transparent 7 bedside card
holder(1 card)

GPSL101TS184

30-50 (plastic) /

ENGLISH

7

8 V21 bed board 1

GPSL101TS610

733*903/PP/white

9 V21 bed board 2

GPSL101TS611

10 V21 bed board 3

GPSL101TS613

333903/PP/white 373903/PP/white

11 V21 bed board 4

GPSL101TS614

12

PP side rail shock damper

LGPQT101TS02 2

13

HJ motor 200, horizontal

LGPDJ101TS573

14

HJ motor 150, vertical

LGPDJ101TS574

473*903/PP/white
18-60-170-35N
HJA58/stroke 200mm
HJA58/stroke 150mm

15

HJ motor 100, vertical

LGPDJ101TS575

HJA58/stroke 100mm

16

HJ controller, 3 functions

LGPDJ101TS576

/

HJT13E power on/

off/100-240V/50/

60Hz/white

17

HJ hand controller, 3 functions

LGPDJ101TS577

HJH18 hand
controller/8 buttons/white

18

V21 back rest rail sticker

LGPFU101TS41 6

780350.4

19

V21 knee rest rail sticker

LGPFU101TS41 7

750350.4

Note: Due to the renewal of product lineup, the parts of the original product may not be available. (The company ensures availability of the parts for sold products within 6 years after their production ends).
11. PACKAGING, TRANSPORTATION AND STORAGE CONDITIONS
1. The packaging of the manual bed is carried out according to the contract or product standard. 2. During the transportation of the product, avoid rushing out, violent vibration, prevent direct sun and rain access. 3. Store conditions: a) Ambient temperature: -10°C~ +40°C b) Relative humidity: 85% c) Atmospheric pressure: 70 kPa ~ 106 kPa

ELECTRIC VARIABLE HEIGHT BED ­ V6k5c

Supply voltage

V

Network frequency

Hz

Operating voltage

VA

Class of electrical protection

Applied part

IP Protection

100-230V a.c. 60 Hz 200 VA I B IPX4

SYMBOL EXPLANATION:

Notice: Read attachment paper

Safe working load is 2000N

Max bear loading is 200KG

12. AFTER-SALES SERVICE

1. C1. Please preserve the attachment papers and invoice of this product, you need to present those to enjoy after-sales maintenance service.
2. If you have any question during use, please contact us any time, so that our company can provide you with accurate and efficient technical support and maintenance services in a timely manner.
3. Do not disassemble the internal parts of this product by yourself to avoid unnecessary damage. I you find that the use is affected by quality issues, please contact the company’s after- sales service department or the authorized distributor.
4. The warranty is guaranteed according to the contract requirements. Outside the

warranty period, the company is responsible for maintenance, providing accessories, and implementing life-long service for clients.
5. After-sales service unit: Jiangsu Saikang Medical Equipment Co., Ltd.
Notice!any serious incident that has occurred in relation to the device shall be
reported to the manufacturer and the competent authority of the Member State in which the user and/or the patient is established

EMC

  1. This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.
    • Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
  2. Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
    • Caution: This machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be use.

Guidance and Manufacturer’s Declaration ­ Electromagnetic Immunity

The V6k5c is intended for use in the following electromagnetic environment. The customer or user of the V6k5c should make sure it is used in such an environment

Emission test

Compliance Electromagnetic Environment – Guidance

RF Emission to CISPR 11

Group 1

The bed uses RF energy for its internal function

RF Emission to CISPR 11
Harmonics according to IEC 61000-3-2

Class B Class A

Voltage fluctuations/ flicker acc. To IEC 61000-3-3

Complies

The bed is intended for use in all types of establishment including residential and similar user that are directly connected to a public supply network that also serves buildings used for residential purposes

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

The V6k5c is intended for use within healthcare facility environment. Users or customers must use bed in this environment

Immunity test

IEC 60601-12/ EN 60601-1-2
Level test

Compliance level Electromagnetic environment guidance

Electrostat
ic discharge
(ESD) IEC 61000-42 /EN 61000-4-2

Contact discharge: +8kV Air discharge: +2kV, +4kV, +8kV, +15kV

Contact discharge: +8kV Scarica in aria: +2kV, +4kV, +8kV, +15kV

Floor should be made of wood and
concrete or be tiled with ceramic tiles. If the floor is covered with synthetic flooring material, the relative air humidity must be at least 30%
Can be used when higher ESD levels
are present

Electrical fast
transient/burst 61000-4-4 /EN 61000-4-4

+2kV Power
supply line +1kV Input/Output lines

Linea di alimentazione +2kV

The quality of the supply voltage should be equivalent to that of a typical business or hospital environment.

Surge
IEC 61000-4-5 /EN 61000-4-5

Line – line: +0.5kV, +1.0kV Line ground: +0.5kV,
+1.0kV, +2.0kV

Linea – linea: +0,5kV, +1,0kV
Linea – terra: +0,5kV, +1,0kV,
+2,0kV

The quality of the supply voltage should be equivalent to that of a typical business or hospital environment.

Power
Frequency
Magnetic Fields IEC 61000-4-8 /EN 61000-4-8

30 A/m 60Hz

30 A/m 60Hz

Magnetic fields with a network frequency should be equivalent to
those to be found in a typical business
or hospital environment.

Voltage Dips
IEC 61000-411
/EN 61000-4-11

0% UT: 0.5 cycles Phase: 00, 450, 900, 1350, 1800, 2250, 2700, 3150
0% UT: 1 cycle 70% UT: 25/30 cycles Single phase: 00 0% UT ; 300 cycle

0% UT: 0,5 cicli Fase: 00, 450, 900, 1350, 1800, 2250, 2700, 3150
0% UT: 1 ciclo 70% UT: 25/30 cicli Monofase: 00 0% UT; 300 cicli

The quality of the supply voltage should be equivalent to that of a typical business or hospital environment. If the person using the bed requires that the bed functions must continue despite any interruptions in the energy supply, it is recommended that the bed be connected to an uninterruptible electricity supply or a battery

Short
interruptions IEC 61000-411 /EN 61000-4-11

0% UT: 250/300 cycles

0% UT: 250/300 cicli

The quality of the supply voltage should be equivalent to that of a typical business or hospital environment. If the person using the bed requires that the bed functions must continue despite any interruptions in the energy supply, it is recommended that the bed be connected to an uninterruptible electricity supply or a battery

Note: UT is the AC Power voltage before applying in the test level

8

Guidance and Manufacturer’s Declaration ­ Electromagnetic Immunity

The V6k5c is intended for use in an electromagnetic environment that used wireless RF transceiver. Electromagnetic interference can be prevented by maintaining the minimum distance between RF transceiver and V6k5c depend on the maximum output and frequency of the communication device as recommended below.

Immunity test

IEC 606011-2 /EN 60601-1 2 Test level
27 V/m 385 MHz

28 V/m 450 MHz

9 V/m

710 MHz Enclosure

port to RF 9 V/m

wireless 745 MHz

communication sequi-
pment IEC

9 V/m 780 MHz

61000-4- 28 V/m

3

810 MHz

/EN 610004-
3

28 V/m 870 MHz

28 V/m 930 MHz

Complian Band elevel (MHz) (V/m)

Service

Modulatio n Maximum Distance

power

(m)

(W)

27

380390

TETRA 400

Pulse modulatio
n18Hz

1,8

0,3

28

430470

GMRS 460, FRS 460

FM+5 kHz deviation
1 kHz sine

2

0,3

LTE Banda

Pulse

9

704787

13, 17

modulatio n217Hz

0,2

0,3

GSM

800/900,

TETRA 800,

Pulse

28

800960

iDen 820, CDMA 850

modulatio n18Hz

2

0,3

LTE Band 5

Enclosure 28 V/m port to RF 1720 MHz wireless communication sequipment IEC 61000-43 /EN 6100043
28 V/m 1845 MHz
28 V/m 1970 MHz
28 V/m 2450 MHz
9 V/m 5240 MHz
9 V/m 5500 MHz
9 V/m 5785 MHz

GSM 1800;

CDMA 1900;

GSM 1900;

Pulse

28

1700-

DECT;

modulatio

2

0,3

1990 LTE Band

n217Hz

1,3,

4,25; UMTS

Bluetooth,

Pulse

28

2400-

WLAN,

modulatio

2

0,3

2570 802.11 b/g/n n18Hz

RFID 2450,

band LTE 7

9

5100- WLAN 802,11 Pulse

0,2

0,3

5800

a/n

modulatio

n217Hz

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

The V6k5c is intended for use in the following electromagnetic environment. The customer or user of the V6k5c should make sure it is used in such an environment.

Immunity test

IEC 60601-1-2 /EN 60601-1-2
Livello test

Livello di conformità

Electromagnetic Environment Guidance

RF Conductivity
IEC 61000-4-6 /EN 61000-4-6

3V 0,15 MHz-80 MHz
6V in ISM bands between 0.15MHz80MHz 80%AM at 1kHz

3V 0,15MHz80 MHz
6V in ISM bands between 0.15MHz-80MHz 80%AM at 1kHz

Field intensity from the fixed RF transmitter determined by a field investigation of the magnetic field must be lower than the compliance level of the respective frequencyband.
Near devices with the following symbol, interference could occur.

Radioactive RF
electromagnetic
field IEC 61000-4-3 /EN 61000-4-3

3V/M 80MHZ-2,7GHz
80%AM a 1kHz

3V/M 80MHZ-2,7GHz
80%AM a 1kHz

a. The intensity of a magnetic field from fixed transmitters such as a radio (mobile/radio) telephone base station, land mobile radio, amateur radio, AM and FM radio broadcasting, and TV broadcasting
cannot be theoretically predicted accurately. To determine the electromagnetic environment for a fixed RF transmitter, consider investigating the electromagnetic field in the location where the bed is used. If the intensity of the magnetic field in the location where the V6k5c is used exceeds the above RF compliance level, make sure tomonitor if the V6k5c operates appropriately. If abnormal movement is
found, take additional measures as needed, such as changing the direction or position of the V6k5c.
b.These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption andreflection from structures, objects, and people.

ENGLISH

SYMBOL EXPLANATION:
Keep away from sunlight
Product code
Date of manufacture Caution: read instructions (warnings) carefully Imported by

Medical Device compliant with Regulation (EU) 2017/745 Follow instructions for use Lot number
Temperature limit
Serial number

Atmospheric pressure limit

Keep in a cool, dry place

Manufacturer
Authorized representative in the European community
Type B applied part

Medical Device

%

Humidity limit

WEEE disposal

Press the red handle to lower the side rail(unlock), to secure press the red handle and push forward until you hear click

Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equipment by bringing it to a specific recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS The Gima 12-month standard B2B warranty applies.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>