INSTRUCTION MANUAL
INFRARED THERMOMETER
Model No.: JXB-315
PRODUCT INTRODUCE
Thank you for purchasing our Kachi Infrared Thermometer JXB-315! All products
have been manufactured and tested according to rigorous procedures to ensure
the safety of the user.
The device is an infrared thermometer that measures the forehead temperature
of infants andadults without contact with the human body.
The JXB-315 non-contact infrared thermometer is developed using the latest
infrared technology. This technology allows for the temperature measurement of
the temporal artery (TA) taken at a distance of 0~3cm (0~1.2 inches) from the
forehead. Thanks to its accuracy, instantaneous, and contactlessness, the
Kachi JXB-315 thermometer is by far the most suitable thermometer for zero
risk of temperature measurement. This TA temperature measurement has been
shown to be more accurate than atrial thermography and better tolerated than
rectal thermography. However, as with other thermometers, it is most essential
to use the JXB-315 thermometer correctly for reliable and stable results.
Therefore, you should carefully read this manual and the safety precautions
before using it.
PRODUCT FEATURE
- Specially designed to help measure human body temperature with a distance of 0 ~ 3 cm (0 ~1.2 inch) from the forehead.
- Stable and accurate measurement by the Infrared Temperature Sensor System.
- Vibration alarm if the temperature is higher than 38°C (100.4°F).
- Up to 32 times memory, allows storage of the nearest measured temperatures.
- White color LED digital display screen.
- Temperature units can be displayed in degrees Celsius or Fahrenheit.
- Automatic power-off (<30 seconds) to save energy.
- Longevity use (100.000 readings).
- Convenient and easy to use.
The JXB-315 thermometer can also be used to measure the temperature of baby bottles or baths (using the Surface Temperature Mode) or room temperature (using the Room Temperature Mode).
IMPORTANT SAFETY INSTRUCTION
- Read the instructions carefully before using the product.
- This device can be used for clinics or for personal use.
- The device is only allowed to be used within the ambient temperature range between 10°C (50°F) and 40°C (104°F).
- Do not expose the thermometer to an environment that causes an electric shock.
- Do not store thermometers in extreme temperatures > 55°C (131°F) or < -20°C (-4°F).
- Do not use the device in environments with humidity > 85%.
- Do not touch the infrared lens on the back of the thermometer with your fingers. To clean the glass, please use a cotton pad lightly dampened with 95° alcohol.
- Do not use or store thermometer under sunlight, water or any other liquid.
- If the device is damaged or shows signs of damage, please contact our service center for warranty and repair advice. Do not attempt to disassemble, repair or replace any part of the device or the warranty will no longer be valid.
- Do not use thermometers near active HF surgical equipment and magnetic resonance imaging devices, which may cause a high intensity of EM disturbance.
- Avoid using the infrared thermometer adjacent to or stacked with another device as it may lead to improper operation. If such use is required, infrared thermometers and other devices should be closely monitored to ensure they are working properly.
- The use of accessories, probes, and cables that are not specified or supplied by the manufacturer may increase the electromagnetic radiation or reduce the electromagnetic immunity of the thermometer and lead to the failure of the device.
- Mobile RF communication equipment (including peripherals such as antenna cables and external antennas) should not be used near any part of the thermometer, including cables specified by the manufacturer. Otherwise, the performance of the device will be reduced. The minimum distance between the thermometer and the antenna is 12 inches (30 cm).
BEFORE USING PRODUCT
- It is not necessary to calibrate the device when starting up.
- To get stable and reliable results, you should know that there is a significant change in temperature every time due to environmental changes, allowing the JXB-315 to adapt to the ambient temperature within 15 to 20 minutes before use. It is important to allow 3-5 seconds between 2 times measurements.
OPERATING PRINCIPLE
All objects, solid, liquid, or gas emit energy by radiation. The intensity of this energy depends on the temperature of the object. Therefore, the infrared thermometer JXB-315 can measure a person’s temperature with the energy that he or she emits. This measurement can be performed using the temperature probe on the device. Therefore, when the user holds the thermometer close to the body and activates the radiation sensor, the measurement will be made immediately by detecting the infrared heat generated by the arterial blood flow. From there, the body temperature can be measured without any interference from the ambient heat.
THE DIFFERENT METHODS OF TEMPERATURE MEASUREMENT
Core temperature
Core temperature is the most precise measurement and involves measuring the
temperature in the pulmonary artery by means of a catheter equipped with a
thermal probe that can read the temperature in the situation. The same method
is employed for probes measuring the oesophageal temperature. However, such
invasive temperature measurement methods require specific equipment and
expertise.
Rectal thermometry
Rectal temperature adjusts slowly in comparison to the evolution of the body’s
internal temperature.
It has been demonstrated that rectal temperature remains raised long after the
internal temperature of the patient has started to drop and vice versa.
Furthermore, rectal perforations have been known to occur as a result of this
method and without appropriate sterilization techniques, rectal thermometry
can spread germs often found in feces.
Oral thermometry
Oral temperature is easily influenced by recent ingestion of food or drinks
and by breathing through the mouth. To measure oral temperature, the mouth
must remain closed and the tongue lowered for three to four minutes which is a
difficult task for young children to accomplish.
Axillary (armpit) temperature
Although it may be easy to measure axillary temperature, it has been proven
that it does not provide an accurate measurement of the child’s internal
temperature. To make this type of temperature, the thermometer must be wedged
tightly over the axillary artery. Despite the low sensitivity and relative
inaccuracy of axillary temperature in detecting fever, this method is
recommended by The American Academy of Pediatrics as a screening test for
fever in newborns.
Tympanic thermometry
In order to obtain a precise temperature reading, good command of the
measurement technique is required. The thermometer probe must be placed as
close as possible to the warmest part of the external ear canal.
Normal temperatures according to measurement method
MEASUREMENT METHOD RECTAL
ORAL
AXILLARY
AURICULAR
TEMPORAL| NORMAL TEMPERATURE
36.6°C – 38°C (97.9°F 100.4°F)
35.5°C – 37.5°C (95.9°F – 99.5°F)
34.7°C – 37.3°C (94.5°F 99.1°F)
35.8°C – 38°C (96.4°F 100.4°F)
35.8°C – 37.8°C (96.4°F – 100°F)
---|---
The temperature of the human body varies throughout the day. It can also be
influenced by numerous external factors: age, sex, type, and thickness of
skin,…
Advantages of temporal artery (ta) temperature
The temporal artery temperature can be measured using a device placed on the
forehead, in the temporal artery area. It has been demonstrated that this
relatively new method of temperature measurement is more accurate than the ear
temperature and rectal temperature measurement method. The JXB-315 thermometer
has been designed to read temporal artery temperatures without any physical
contact. Because this artery is quite close to the skin’s surface and is
easily accessible from there for the most accurate temperature measurements.
The normal temperature according to age
Age | °C | °F |
---|---|---|
0-2 years | 36.4-38.0 | 97.5-100.4 |
3-10 years | 36.1-37.8 | 97.0-100.0 |
11-65 years | 35.9-37.6 | 96.6-99.7 |
>65 years | 35.8-37.5 | 96.4-99.5 |
-
In order to ensure that precise and accurate temperature measurements are obtained, it is essential that each user has received adequate information on and training in the temperature measurement technique when using such a device.
-
The temperature should be taken in a neutral context. The patient must not have undertaken vigorous physical activity prior to taking his/her temperature and the room temperature must be moderate.
-
Be aware of physiological variations in temperature which must be taken into consideration when evaluating the results: temperature increases by 0.5°C (0.9°F) between 6 am and 3 pm.
Women have a temperature that is higher, on average, by around 0.2°C (0.4°F). Their temperature also varies in accordance with their ovarian cycle. It rises by 0.5°C (0.9°F) in the second half of the cycle and at the early stages of pregnancy. -
When sitting, the temperature is lower by about 0.3°C (0.6°F) to 0.4°C (0.8°F) than when standing.
HOW TO TAKE A TEMPERATURE
Aim at the middle of the forehead, from a distance of about 0~3cm(0~1.2 in),
Press the thermometer’s measurement button, and the temperature is instantly
displayed. The reliability of the measurement cannot be guaranteed if the
temperature is measured over another part of the body (e.g.arm, torso…).
Note! Please observe the following before any temperature measurement to
ensure a stable and reliable result:
- Push back the hair from the forehead.
- Wipe away any perspiration from the forehead..
- Avoid any drafts (e.g. from nasal specs, air conditioning…)
- Allow a 3~5 seconds interval between two times measurement.
- Each time there is a significant change in the ambient temperature due to a change in environment, to allow the JXB-315 to acclimatize to this ambient temperature for at least 15 minutes before using it.
PRODUCT STRUCTURE
Screen
OPERATE INSTRUCTION
- Remove the packaging. Open the battery box cover on the back of the thermometer and insert 2 AAA batteries.
- For the first use or when inserting a new Battery, wait from 10~15 minutes for the warm-up of the unit. This will allow the unit to become acclimated to the temperature of the room.
- When measuring body temperature, the sensor of the device should be pointed to the center of the forehead, holding the device upright and above the eyebrows. The distance between the thermometer and the forehead should be kept within 3 cm (1.2 inches).
Note: Do not let the thermometer touch your forehead directly.
In the switch-on state, press the “ ” button o take the temperature, the
device will vibrate and the measurement result will be displayed on the
screen. At the same time, the display will show a measure of deviation. If the
distance between the device and the forehead is greater than 3 cm (1.2
inches), the device will not vibrate, the display interface will blink and the
thermometer will be put to standby mode after 5 seconds.
Tip: Do not remove the thermometer from the measuring position when the
measurement is not completed.
The measurement distance will be different for different skin colors.
Setting and function of menu
Switch on the device
Press the “ ” button in one second after the screen panel in full display, it
will enter the standby mode with the sign”— °C” hoặc “— F”. Then press the “ ”
again, ou will get the measuring result in 1 second. But if there is no more
operation, it will turn off in 30 seconds automatically, or pressing the “ ”
button for more than 3 seconds will be turned off.
Temperature mode setting
In the switch-on state, press the “MODE” button, the screen will display “ ”
it is “Body temperature mode”. When press the “MODE” button again, the screen
will display “ ”, it is “Surface temperature mode”. Press the “MODE” button
again, the screen will display “ ”, it is “Room temperature mode”.
Note: The thermometer default is set to Body temperature mode.
Temperature adjustment setting
In the event that the temperature of the infrared thermometer measured is
different from that of a mercury thermometer and you want to reset the
temperature, please refer to the following instructions:
In the standby state, press and hold the “MODE” button for 2 seconds. After
the display shows “F4”, enter the overall temperature compensation setting.
When the icon “ ” and “0.0°C (°F)” flashes, press the “MEM” button to increase
0.1°C, press the “ ” button to decrease 0.1°C (°F). The offset value can be in
the range of -3ooC ~ 3C (-5.4°F ~ 5.4°F) by the internal setting, press the
“MODE” button twice, and the setting data will be returned to the standby
state.
Note: The thermometer default is set to + 0.0°C (0°F).
Temperature unit setting
In the switch-on state, press and hold the “MODE” button for 2 seconds, the
screen will display “F4”, press the “MODE” button again, the screen will
display “F5”, when entering the temperature unit setting, then “F5” and the
unit symbol “°C” or °F” will be flashed continuously, press the “MEM” button
to select Celsius “°C”, press the “ ” button to select Fahrenheit °F, then
press the “MODE” button to save and return to the standby state.
Memory query and clear
In the switch-on state, press the “MEM” button, the display will show the last
measured data and you can review up to 32 groups of data stored.
To clear data, press the “MEM” button for 5 seconds, all current memory data
will be erased and restored to factory settings.
Over-temperature prompt function
Under the body temperature mode, if the measured body temperature is higher
than or equal to 38°C(100.4°F), the product will be vibrated three times.
Changing the batteries
When the icon “ ” flashes the battery needs to be changed.
Note: Do not use rechargeable batteries for this device.
The battery terminals must be properly inserted in the thermometer (see
picture below).
EMC DECLARATION
The Medical equipment or system is suitable for home healthcare and clinic
environments.
Warning
- Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
- Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
- Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
- Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment ET001, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Portable and mobile RF communication equipment may affect the performance of the ET001, avoiding strong electromagnetic interference when used, such as near mobile phones, microwave ovens, etc.
Guidance and manufacturer’s declaration -electromagnetic emissions and Immunity
Emissions test | Compliance |
---|
RF emissions
CISPR 11| Group 1
RF emissions
CISPR 11| Class B
Harmonic emissions
IEC 61000-3-2| Not application
Voltage fluctuations/ flicker emissions
IEC 61000-3-3| Not application
Table 1
Immunity Test | IEC 60601-1-2, Test level | Compliance level |
---|---|---|
Electrostatic discharge (ESD) IEC 61000-4-2 | ±8 kV contact | |
±2 kV, ±4 kV, ±8 kV, ±15 kV air | ±8 kV contact |
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst IEC 61000 4- 4| Not application| Not
application
Surge
IEC 61000-4-5| Not application| Not application
Voltage dips, short interruptions, and voltage variations on power supply
input lines
IEC 61000-4-11| Not application| Not application
Power frequency magnetic field
IEC 61000-4-8| 30 A/m 50H7160Hz| 30 Alm 50Hz/60Hz
---|---|---
Conducted RF
IEC61000-4-6| Not application| Not application
Radiated RF IEC61000-4-3| 10 V/m
80 MHz – 2.7 GHz 80 %AM at 1 kHz| 10 V/m
80 MHz – 2.7 Glitz 80 %AM at 1 kHz
NOTE UT is the AC mains voltage prior to application of the test level.
Table 2
Radiated RF
IEC61000-4-3
(Test specifications
for ENCLOSURE
PORT IMMUNITY to
RF wireless
communications
equipment)| Test Frequency
(MHz)| Band (MHz)| Service| Modulation| Modulation|
Distance(m)| IMMUNITY TEST LEVEL (V/M)
---|---|---|---|---|---|---|---
385| 380-390| TETRA 400| Pulse modulation18 Hz| 1.8| 0.3| 27
450| 430-470| GMRS 460 FRS 460| FM ±5 kHz deviation
1 kHz sine| 2| 0.3| 28
710| 704-787| LTE Band 13,17| Pulse
modulation
217 Hz
| 0.2| 0.3| 9
745
780
810| 800-960| GSM800/900,TETRA800,
CDMA850,
LTE Band 5| Pulse
modulation
18 Hz
| 2| 0.3| 28
870
930
1720| 1700-1990| GSM 1800; CDMA 1900;
GSM 1900; DECT; LTE Band 1,3,4,25; UMTS| Pulse modulation
217 Hz| 2| 0.3| 28
1845
1970
2450| 2400-2570| Bluetooth,WLAN, 802.11 b/g/n, RFID2450, LTE Band 7| Pulse
modulation217 Hz| 2| 0.3| 28
5240| 5100-5800| WLAN 802.11 a/n| Pulse modulation
217 Hz| 0.2| 0.3| 9
5500
5785
Table 3
CLEANING AND MAINTENANCE
The protective glass over the lens is the most important and fragile part of the thermometer, please take great care of it.
- Clean the glass with cotton fabric, wet it
- with 95° alcohol.
- Do not use batteries other than AAA batteries, or use rechargeable batteries.
- Remove the batteries when the thermometer is not used for an extended period of time.
- Do not expose the thermometer to sunlight or water.
PRODUCT SPECIFICATION
Product name | Infrared thermometer |
---|---|
Model | JXB-315 |
Power | DC 3V (2 pin AAA) |
Normal using condition | Ambient temperature: 10°C – 40°C (50°F – 104°F) |
Relative humidity: s85%
Storage and moving condition| Ambient temperature: -20°C – 55°C (-4°F – 131°F)
Relative humidity: s95%
Measuring range| In body mode: 32°C – 43°C (89.6°F – 109.4°F)
In surface temp mode: 0°C – 60°C (32°F – 140°F) In room mode: 0°C – 40°C (32°F
– 104°F)
Precision| 32.0°C-34.9°C (89.6°F – 94.8°F) ±0.3°C (±0.6°F)
35.0°C-42.0°C (95°F – 107.6°F) ±0.2°C (±0.4°F)
42.1°C-43.0°C (107.8°F – 109.4°F) ±0.3°C (±0.6°F)
Measuring distance| 5 3 cm
Automatic power-off| After 30 second
Product size| 14.5 x 4 x 3.2 cm
Product weight| 66 Gram
TROUBLESHOOTING
If you have problems while using your thermometer, please refer to this guide
to help resolve the problem. If the problem persists, please contact our
customer service.
The screen displays temperatures higher than 43.0°C
Temperatures are displayed in units of Fahrenheit ( °F). Change the
measurement to Celsius (°C) before measuring.
The screen displays the message “HI”
When using the JXB-315 thermometer, the message “HI” may be displayed on the
display. In this case, the temperature of the measuring object is above the
measuring range of the thermometer, > 43.0°C (109.4°F) while in body
temperature measurement.
The screen displays the message “LO”
When using the JXB-315 thermometer, the message “LO” may be displayed on the
screen. In this case, the analyzed temperature is within the selected
measuring range, or below 32°C (89.6°F) while in the body temperature
measurement mode.
Reasons for LO message display | Solution |
---|---|
Temperature reading hampered by hair or perspiration | Remove hair and wipe off |
sweat before taking measurements
Measurement temperature is influenced by the air conditioner. fan…| Do not
take the temperature in an environment where air conditioning is used
The temperature readings are too close together and there is no rest time
between measurements| Let the device rest for 3-5 seconds after each use
Distance between thermometer and forehead is too far| The maximum distance
between the forehead and thermometer required is 3cm (1.2 in).
INSTRUCTION OF SCANNING QR CODE
Method 1: Scanning QR code by Zalo (Vietnam app)
Method 2: Scan QR code on iPhone with iOS11.
– Activate scanning QR code by the camera: Go to Settings > Camera > Scan QR
Code).
– After completing the install, open the camera and move the camera to QR code, then you will receive a notification, please click “Connect” and follow the instructions.
Method 3: Scan QR code on Android with Google Assistant
Instructions for using Google Assistant on Android.
Hold the Home button to open the Google Assistant application and follow the
illustration| After clicking on the icon above the application to open the
camera, move it to QR code and click on the link in the image| Click the
“Start” and follow the instructions
---|---|---
| |
TERMS AND CONDITIONS OF PRODUCT RETURN
CONDITIONS OF RETURN – RENEWAL
I. Regulations on return
Customers need to check the status of the goods and can exchange/return the
goods in the following cases:
- The goods are not of the correct type or model in the ordered order or as on the website at the time of order.
- Not enough quantity, not enough sets as in the order.
- The product is not working.
Customers are responsible for submitting relevant documents proving the above deficiencies to complete the return/exchange of goods.
II Regulations on the time to notify and return products
- Time of return notice: within 7 days after receiving the product in case the product lacks accessories, gifts, or is broken.
- Product delivery time: within 14 days after receiving the product.
- Product return location: Customers can bring products directly to our office/store or by mail.
In case you have comments/complaints related to product quality, please contact our hotline.
III. Conditions of return – renewal
The product is intact, not scratched, scratched, distorted, broken, cracked,
impacted, flammable, moldy, or chemical.
- The product has not been opened for repair by the user or by a third party.
- Products with full documents: delivery note, warranty card, VAT invoice (if any), manual, catalog …)
- Goods are confirmed to have technical defects checked by the company’s technician.
- Product intact box (no distortion, no tearing box, packaging, foam,..) full of original accessories, full promotional gifts (if any).
- Accessories must be complete and intact.
- If lost accessories or promotions or cases are scratched, fees will be charged depending on the case of the product.
Note: Our company does not accept returns in the following cases.
- You want to change the category for your liking, but the product really does not have any defects.
- You do not operate properly, causing damage to the goods.
- You do not comply with the regulations as required to enjoy the warranty (for example, do not present the warranty card, the company’s vouchers when exchanging).
- You self-affect external conditions such as packaging tear, scratches, scratches, breakage, …
- The goods have been checked, agreed to buy by you, but then you asked to return the goods due to the wrong form (scratches, distortions, yellowing, broken …).
- Please check the goods and sign the receipt of the status with the Delivery Officer immediately upon receiving the goods. When detecting one of the above cases, you can discuss directly with the Delivery Officer or immediately respond to the customer receiving the customer order.
CONDITIONS FOR HOME SUPPORT
I. Free case
- Massage chair products are supported for free at home nationwide
- If the product is newly received and defective because the manufacturer is bulky, the warranty department will let the technician handle it at home for free in Hanoi and Ho Chi Minh City.
- New products received defective parts, the warranty department will send free parts compensation to customers.
II. Charge case
- Error caused by the user
- Home installation service
Travel and accommodation costs, if any + labor costs 250.000 vnd/ half a day – 500.000/day.
WARRANTY CONDITIONS
I. Warranty conditions:
- A purchase invoice or electronic warranty registration is required
- All products must be intact and free from modified erasers.
- Damage to the product due to manufacturer error.
- For innovative non-electric products within 7 days of purchase, if there is a manufacturer fault
- For electrical products, an 18-month warranty is free of charge if there is a manufacturer’s fault
- For health care products such as scales, blood pressure monitors, blood glucose meters, oil-free fryers with a 24-month warranty
II. Warranty refusal cases:
- Product is out of warranty or without the warranty card of the Company.
- The product is not physically intact, dropped, collided, exploded, and traces of user repairing voluntarily.
- Products are damaged by natural disasters, lightning, floods, fires, computer viruses, insects, animals, oxidized humid environment, sand, and soil, due to improper transportation.
- Misuse of power source or improperly
- No warranty accessories such as gifts, accessories, and accessories come with the product depending on the product.
- Products, components, accessories that are easily scratched, peeling, … such as fryer, stainless steel pot, rice cooker, …
- For high-pressure washers, components, accessories such as spray gun, spray body, hose, connector … will not be covered under warranty.
-
Attention: Customers must pay shipping costs if any.
III. Warranty period- Time to receive warranty: from 8H to 17H from Monday to Saturday (except Sunday and holidays, New Year).
- Hotline: 1900.1738
IV. Warranty refund period
- All products of the Company provide free warranty when customers bring to the Company to ensure that there are full warranty cards of the machine still within the warranty period as prescribed.
- Time to resolve warranty from 5-7 days from the date of receipt except for Sunday, holidays, and depending on the case of damage.
1. In the case of minor equipment failure, the return time is between 5 and 7 working days from the date of warranty receipt.
2. In case the device is severely defective, it is necessary to wait for the parts to be replaced, the return time of the device should not exceed 4 weeks (28 days) from the date of receipt of warranty.