reer 98050 3-In-1 Contactless Infrared Clinical Thermometer Instruction Manual
- June 6, 2024
- Reer
Table of Contents
- Device description
- Display
- Colour SoftTemp 3in1 contactless infrared clinical thermometer
- Technical data
- Safety instructions
- Scope of delivery
- Start-up
- Fever alarm
- Troubleshooting
- Cleaning and care advice
- Warranty
- Disposal instructions
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Colour SoftTemp
3in1 contactless infrared clinical thermometer
Instruction manual
item no.: 98050
Device description
Display
Colour SoftTemp 3in1 contactless infrared clinical thermometer
The innovative reer thermometer lets you determine your body temperature
reliably and within seconds by taking a measurement at the temple. Contactless
temperature measurement guarantees safe and hygienic measurement and reduces
the risk of transmission of viruses or other pathogens.
A backlit display and switchable signal tone allow you to measure your child’s
temperature at night in the dark without waking them.
IMPORTANT!
READ CAREFULLY BEFORE USING RETAIN FOR FUTURE REFERENCE.
Technical data
Dimensions: | 144 x 39 x 40 mm |
---|---|
Material: | Plastic |
Output/nominal voltage: | 0,27W / 3V |
Type of battery: | 2 x 1.5V AAA batteries (included in delivery) |
Measuring range:
Body temperature (temple): 32,0 °C – 42,9 °C (89,6 °F – 109,2 °F)
Object/ambient temperature: 0 °C – 99,9 °C (32,0 °F – 211,8 °F)
Measuring accuracy:
Body temperature: ±0,2°C (±0,4°F), 36,0°C (96.8°F) – 39,0 °C (102,2°F)
±0,3°C (±0,5°F), 32,0°C (89.6 °F) – 35,9°C (96.6°F)
±0,3°C (±0,5°F), 39,1°C (102.4°F) – 42,9°C (109.22°F)
Object temperature: ± 4°C (±7.2°F), 0°C (32°F) – 4,9°C (40.8°F)
± 1°C (± 2°F), 5°C (41°F) – 60°C (140 °F)
± 4°C (± 7,2°F), 60,1°C (140.1°F) – 100°C (212°F)
Measuring units: Centigrade (°C) and Fahrenheit (° F)
Measuring distance: 1 – 3 cm to measuring point
Measuring time: approx. 1 second
Measured value memory: 25 measurements
Measuring display : Liquid crystal display (LCD) with backlight
Acoustic signal: switchable on/off
Automatic switch-off : approx. 15 seconds following last measurement
Net weight: approx. 76g
Protection class: IP22
Operating conditions: 10 °C – 40°C (50°F – 104 °F) ambient temperature,
15 % – 95 % relative humidity,
86 – 106 KPa air pressure
Storage conditions : -25 °C – 55 °C (-3 °F – 131 °F) ambient temperature,
15% – 93% RH relative humidity
Safety instructions
Please read the following points carefully and follow the instructions before
using the device.
Danger
- Avoid bringing the battery and/or electrical connection contacts on the unit into contact with metal objects.
- Keep the battery out of the reach of children. Risk of swallowing!
- If batteries have leaked, use chemical-resistant protective gloves and protective glasses.
- In case of high fever or unclear measurement results, always seek advice from a doctor.
- The infrared technology your reer thermometer uses is perfectly safe. Although it’s not harmful to look through the lens while the temperature is being measured, this is not recommended.
Warning
- Check the unit for visible damage before each use. If you detect any damage, do not use the unit.
- Avoid extreme heat and direct sunlight.
- Do not subject the device to physical shocks.
- The unit must not be opened, as this may cause an electric shock. If a unit does not work correctly, please do NOT try to fix the problem yourself. Please contact your dealer regarding any problems.
- To avoid the risk of electric shock and/or fire hazard, do not submerge the unit or the plug in water, other fluids or flammable cleaning materials.
- The unit is not waterproof and must therefore not be submerged in water or directly exposed to any liquids.
Caution
- Please use the article only for its intended purpose.
- Use the device exclusively at home for its intended purpose. This device is not intended for commercial use.
- This product is not a toy! Children must not play with the appliance.
- Do not use the unit outdoors.
- Remove all packaging material before use.
- Avoid dirt and damage to the measurement sensor’s lens.
- The measured core body temperature (measured rectally) can be significantly higher than the surface temperature (measured at the temple), depending on the course of the fever.
- A measurement of the core body temperature using a rectal clinical thermometer is recommended:
- if the measurement result is surprisingly low
- for newborns within the first 100 days
- for children under the age of 3 who have a compromised immune system or if the fever does or does not occur where a fever would or would not be expected.
Scope of delivery
- 1 x thermometer
- 2 x AAA alkaline batteries
- 1 x instruction manual
Start-up
Before using your thermometer, make sure that its sensor is free of dirt. A
dirty sensor can lead to incorrect measurement results.
Inserting batteries
The batteries must be inserted into the thermometer before it can be used.
- Open the battery compartment on the underside of the unit.
- Insert the batteries.
- Close the battery compartment.
Note: Do not use different battery types or combine new and used batteries. Batteries must be inserted with the correct polarity.
Measurement of body temperature
The thermometer has been designed to measure the body temperature above the
eyebrows, measured at the point between the forehead and temple. Measuring
elsewhere on the body can result in inaccurate measurement data.
-
Before each measurement, make sure that the sensor is clean. If this is not the
case, please clean the sensor according to the instructions on page 31. -
Check if the forehead is wet (cool skin effect). It must be completely dry.
-
Make sure that the temples are free of any dirt and cosmetics. Avoid taking measurements on any scars that may be present.
-
Vasoconstrictive drugs can influence the measured value.
-
Portable and mobile communication devices can also affect the device.
-
Wait at least 30 minutes before using following any sporting activities, bathing
-
If possible, do not use the thermometer outdoors, but only in closed rooms. External influences such as wind and outdoor temperature can influence the measurement result.
-
If the device is to be used in an extreme-temperature environment, store it at room temperature for at least an hour before taking any measurements.
-
Holding the thermometer in the hand for too long before taking the measurement will cause the device to heat up. This can cause a false measurement result to be given.
- Set the regulator to body temperature measurement. Set the regulator to body temperature measurement. Switch the thermometer on by briefly pressing the On/Off button.
- The backlight and start sequence are displayed.
- The display shows a dashed line with the flashing temperature unit.
- Position the thermometer on your child’s temple at a maximum distance of 3 cm. The positioning LED helps you to determine the correct measuring point.
- Press the On/Off key as soon as the temperature sign °C or °F flashes. Read the temperature.
Fever alarm
The thermometer has a built-in optical indicator to warn you in case of elevated body temperature. The colored display shows you the temperature range that is to be estimated.
Temperature| Display| Measured body
temperature| Signal tone
(Beep)
---|---|---|---
<32° °C / 89.6°F| LO| too low| 1x long
32,1 °C – 37,4°C /
89,6 °F – 99,5 °F| green| normal| 1x long
37,5°C – 37,9°C /
99,5 °F – 100,2 °F| orange| elevated| 1x long
38 °C – 42,9°C /
100,4 °F – 109,22 °F| red| high| 6 x short
42.9°C / 109.22°F| Hi| too high| 6 x short
** If the device’s LED lights up red during temperature measurement, body
temperature should be kept under constant observation and medical advice
sought in case of any uncertainty.
The use of the device does not in any way replace an examination by a doctor.
In case of high fever or unclear measurement results, always seek advice from
a doctor.
Measurement mode: Object/ambient temperature
If possible, the measurement of liquids should take place from outside the
vessel to avoid any condensation forming on the sensor. The warning indicator
is switched off in this mode. The display is constantly lit in green. The
object/ambient temperature mode shows only the exposure temperature of the
surface/room temperature and does not provide any indication of body
temperature. This measurement mode cannot, therefore, be used to measure
fever.
Set the regulator to object measurement . Follow steps 1 to 3 as for body
temperature measurement. Point the thermometer at the desired object and press
the measure button. Read the measured temperature on the display.
Setting functions**
You can adjust the thermometer according to your own particular needs. To do
so, follow these instructions.
Memory function:
| 1. Make sure that the thermometer is switched off.
---|---
| 2. Press and hold the On/Off button for approx. 2 seconds, until the
display shows “MEM”.
| 3. Press the
On/Off button to scroll through the last 25 measurement values.
4. The thermometer starts with the last measured temperature at number 1.
5. When the memory is full, the oldest value is automatically deleted and the
newest value is saved.
6. The display lights up in the respective display colors of green, orange
and red, as long as it is a body temperature measurement that is being
displayed.
| 7. To exit the menu item, leave the thermometer alone for approx. 10
seconds, until it automatically turns itself off.
| 8. To erase the memory, remove the batteries from the unit.
9. Press and hold the On/Off button while simultaneously replacing the
batteries, until the display shows “- –”.
Change temperature scale
| 1. Make sure that the thermometer is switched off.
---|---
| 2. Press and hold the On/Off button for approx. 8 seconds, until the
display shows °C or °F.
3. Press the On/Off button again to choose between °C and °F.
4. Press and hold the On/Off button again for approx. 3 seconds to save your
selection.
Switch positioning LED on or off: The positioning LED helps you determine the measuring point more precisely.
| 1. Make sure that the thermometer is switched off.
---|---
| 2. Press and hold the On/Off button for approx. 10 seconds, until the
display shows “LED”.
3. Press the On/Off button again to turn the positioning LED on or off.
4. Press and hold the On/Off button again for approx. 3 seconds to save your
selection.
Sound mode: selecting signal tone or mute
| 1. Make sure that the thermometer is switched off.
---|---
| 2. Press and hold the On/Off button for approx. 12 seconds, until the
display shows ON or OFF.
3. Press the On/Off button again to choose OFF (mute) or ON (signal tone).
4. Press and hold the On/Off button again for approx. 3 seconds to save your
selection.
Troubleshooting
Problem/symbol | Possible cause | Action |
---|---|---|
Unit won’t switch on. | Batteries are dead. | |
Batteries are wrongly inserted. | Insert new batteries. |
Check the position and polarity of the batteries.
Incorrect or fluctuating
temperature values| Sensor dirty/ broken.| Check the measurement sensor and if
necessary clean it.
| Low battery level, measurements are still possible.| Replace the batteries
soon.
| The temperature is outside the
measurement range.| Check the controller settings
or . If the problem persists,
leave the thermometer alone for a few minutes before taking any further
measurements.
| Measured body temperature exceeds
42.9 °C (109.2 °F).
Measured object/ ambient temperature
exceeds 100 °C (212 °F).| Check whether you are using the correct measuring
mode (body
temperature/ object temperature). Repeat the measurement.
Make sure the measurement sensor is
positioned correctly.
| Measured body temperature is below
32 °C (89.6 °F).
Measured object/ ambient temperature
is below 0 °C (3 °F).| Check whether you are using the correct measuring mode
(body
temperature/ object temperature). Repeat the measurement.
Make sure the measurement sensor
is positioned correctly.
Cleaning and care advice
The sensor is the most sensitive part of the infrared thermometer. Be careful not to touch or scratch the surface of the measurement sensor. In case of dirt, carefully and gently rub the sensor using a cotton swab soaked in alcohol or a mild cleaning agent (e.g. strongly diluted washing-up liquid). Do not use any aggressive cleaning agents on the sensor. Clean the housing and display using a soft, dry cloth or an alcohol pad. The device is not waterproof and must therefore not be allowed to be immersed in water or directly exposed to moisture. Remove the batteries from the unit if it is to be out of use for a longer period. This product does not require any special maintenance. Any repairs made to the unit must be carried out by a qualified professional.
Metrological control
As a rule, checking the thermometer’s precision is not required. For
devices used in Germany for medical practice, the Medical Products Operator
Ordinance of 29 June 1998 (in the version announced on 21 August 2002, last
amended by Article 4 of the law of 29 July 2009) requires metrological control
after 2 years.
Legal basis for placing on the market in the EU
The manufacturer has been certified by the EU notified body TÜV SÜD Produkt
Service GmbH in line with the 93/42/EEC Directive on Medical Devices. This
device complies with the European standard EN60601-1-2 for electromagnetic
compatibility and ASTM E 1965 “Standard Specification for Infrared
Thermometers for Intermittent Determination of Patient Temperature”.
This product complies with the following standards for medical electrical equipment:
- IEC 60601-1-2:2014
- IEC 60601-1:2012
- ISO 80601-2-56: 2009
- IEC 61000-4-2
- IEC 61000-4-8
- IEC 61000-4-3
**** | Caution |
---|---|
Permitted storage and transport temperature and air humidity | |
Permitted operating temperature and air humidity | |
Manufacturer | |
Observe the user manual | |
Application part type BF | |
**** | Disposal according to the Waste Electrical and Electronic Equipment EC |
directive
LOT
| Batch designation
IP22
| Protection against entry of solid foreign bodies with a diameter of >12.5
mm. Protection against dripping water with a 15° incline.
| The CE marking certifies conformity with the basic requirements of the
93/42/EEC guidelines for medical products.
Warranty
Please retain your receipt, the user manual, and the part number. Warranty claims or claims from promises of guarantee must be processed by the dealer. The technical specification, instructions, and characteristics of the product described herein have been compiled to the best of our knowledge and belief and are correct at the time of printing. Any instructions provided by reer should be correct and authoritative. Nevertheless,reer does not accept liability for potential errors in this document and reserves the right to make changes to the product design and/or specifications without prior notice.
Disposal instructions
Batteries must be disposed of properly. Battery collection containers are available free of charge in stores that sell batteries and at municipal collection points. At the end of its useful life, please dispose of the device properly via a municipal collection point for electrical waste.
SHANGHAI INTERNATIONAL HOLDING CORP. GMBH (EUROPE) EIFFESTRASSE 80, 20537 HAMBURG, GERMANY
SHENZHEN AEON TECHNOLOGY CO., LTD. RM6H02, BLOCK 27-29, TIAN XIA IC INDUSTRIAL PARK, MAJIALONG, NO.133 OF YIYUAN ROAD, NANTOU STREET, NANSHAN DISTRICT, SHENZHEN, PEOPLE‘S REPUBLIC OF CHINA
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>