tyromotion R82 Pondus Instruction Manual
- June 9, 2024
- tyromotion
Table of Contents
PONDUS® – INSTRUCTIONS FOR USE
Issue Date of User Guide: 2023-03-01, Rev. 1.0
R82 Pondus
Contact information:
Manufacturer:
TYROMOTION GmbH
Bahnhofgürtel 59
8020 Graz
AUSTRIA
PHONE +43 316 908 909
EMAIL office@tyromotion.com
WEB www.tyromotion.com
SUPPORT support@tyromotion.com
Distributor:
R82 A/S
Parallelvej 3
8751 Gedved
Denmark
PHONE +45 79 68 58 88
FAX +45 75665192
EMAIL R82@etac.com
WEB etac.com
Declaration of Conformity
This product conforms to the requirements of the Medical Devices Regulation
(EU) 2017/745. The CE mark must be removed when rebuilding the product or when
using other than original PONDUS® accessories.
PONDUS® is a registered trademark owned by R82 A/S.
http://pondus-download.tyromotion.com
Introduction
We appreciate your choice of a new product from TYROMOTION. To fully benefit from the options offered by this product please read these instructions for use and save it for future use.
1.1 Symbols
1.1.1 Symbols in the instructions for use
****| Warnings: This symbol indicates a possible risk of injury to your own
health or to the health of others. Be especially mindful of these notices!
---|---
1.1.2 Symbols on PONDUS®
**** | Follow the instructions for use |
---|---|
Applied part, type BF | |
**** | Do not discard with household waste |
**** | CE mark |
IP22 | Protection class against foreign objects and water: |
– 2 means protection against penetration of solid objects with a diameter ≥
12,5 mm
– 2 means protection against ingress of dripping water (15° tilt)
| Information about the manufacturer of PONDUS®, including the manufacturer’s
full mailing address is displayed next to the factory symbol.
| Date of manufacture
| Serial number of the device
| Medical device
| Upper and lower limit of temperature Indicates the upper and lower limit of
temperature to which the medical device can be safely exposed.
| Humidity limitation Indicates the range of humidity to which the medical
device can be safely exposed.
| Atmospheric pressure limitation Indicates the range of atmospheric pressure
to which the medical device can be safely exposed.
1.2 System content
PONDUS® consists of the following components, included in the delivery:
- 1x PONDUS® set (size 2, 3 or 4) consisting of two plates
- 1x Software Application
- 1x power supply
- 1x USB charging cable
- 1x Instructions for use
- 1x hook and loop adhesive pad set (for use with R82 Caribou and R82 Gazelle PS)
1.3 Intended use
PONDUS® is a medical device to support the positioning of the user in a
standing frame.
1.4 Warranty and legal disclaimer
TYROMOTION GmbH issues a warranty to the original system purchaser that the
system shall be free of material and qualitative processing defects for a
period of 12 months under normal usage conditions from the date of
installation on the owner’s premises and that the system complies with the
mechanical and electrical specifications published by TYROMOTION. This
warranty is granted under the provision that the system is installed, operated
and maintained in accordance with the instructions for use. The customer must
submit all warranty claims to TYROMOTION in written form within 60 days of the
occurrence of the problem and before the expiry of the warranty. TYROMOTION is
exclusively obligated to repair, exchange or correct faulty or non-compliant
parts at its own discretion in accordance with the warranty. TYROMOTION has no
further obligations to the owner in regard to these parts after the repair or
exchange of faulty or non-compliant parts. All repairs or maintenance work
must be performed by an authorized TYROMOTION service representative in
accordance with this warranty. The above mentioned warranty becomes null and
void if repairs, maintenance or other work is performed by third parties.
Moreover, problems resulting from accidents, improper use, incorrect
application, storage damage, negligence as well as system or component
modifications are excluded from the warranty.
The above mentioned warranty is granted in place of all other warranties,
rights or conditions, and the system is delivered “without deficiency
warranty” apart from the limited warranty. TYROMOTION and its third-party
suppliers specifically and unreservedly reject all other explicit or implicit
warranties held by the owner, his personnel and patients, customers, users and
any third parties, unreservedly including all warranties for marketability,
applicability for a specific purpose, non-infringement and any warranties
resulting from performance development, business trends or commercial customs.
TYROMOTION and its third-party suppliers do not provide declarations or
warranties for system compliance with the owner’s requirements or for
functionality without interruption, errors or deficiencies.
TYROMOTION is in no way liable for indirect, incidental, specific or
consequential damage or for punitive damage compensation including, among
other things, the loss or absence of profits, yield, goodwill or usage, which
the owner or third parties may incur or for damage to connected equipment,
costs for replacement products, installations, servicing, exchange elements or
idle time or for claims from patients, customers, visitors, the owner’s
employees or other persons, regardless whether submitted within the context of
a contractual claim, due to unauthorized behavior, strict liability or imposed
by law or otherwise even when TYROMOTION has been informed about the
possibility of such damages. TYROMOTION’s liability for damages resulting from
or in connection with this contract may not in any event exceed the purchasing
price of the system.
Some jurisdictions limit or exclude the extent of restrictions, the exclusion
of legal means, compensation or liability, such as liability for gross
negligence or willful misconduct according to or in the abovementioned extent
or do not permit the exclusion of implicit warranties. In such jurisdictions,
the restriction or exclusion of warranties, legal means, compensations or
liabilities described above may not be valid for the owner. Such restrictions
or exclusions apply according to the highest legally permitted extent even if
they are not valid according to the legally prohibited extent. The owner may
also have other rights that vary depending on the specific country or other
jurisdictions.
Technology
2.1 Overview
Device name: | PONDUS® |
---|---|
Classification: | PONDUS® is a medical device class I according to medical |
device regulation (EU) 2017/745, rule 13.
Type of applied part:| Typ BF
Protection against electric shock:| Internally powered medical device
Electromagnetic compatibility:| Class B device (CISPR 11) PONDUS® is suitable
for usage in all establishments including residential areas and areas that are
directly connected to the PUBLIC SUPPLY GRID, which also supplies residential
buildings.
---|---
Country of origin:| AUSTRIA
Power supply voltage:| 100 – 240V alternating current
Supply frequency:| 50/60Hz
Electricity/Power consumption:| 5V DC / 100mA charging
Battery:| Polymer-lithium-ion battery, 3.7V, 470mAh The battery must not be
replaced.
Radio transmission frequency:| ISM Band (Bluetooth BLE)
Radiated transmission power:| Max. 10mW Operational area: 10 meters given
uninterrupted view between PONDUS® plates and used Tablet/Smartphone
Operating type:| Continuous operation
2.2 Area of application
The product is for indoor use.
** PONDUS® is classified as a medical electronic device and therefore
subject to specific precautionary measures relating to electromagnetic
compatibility (EMC). It is absolutely imperative to observe the stated
indications for EMC. Portable and mobile HF communication devices may affect
the PONDUS®.
Guidelines and MANUFACTURER‘s declaration – ELECTROMAGNETIC EMISSIONS**
PONDUS® is designed for operation in an ELECTROMAGNETIC ENVIRONMENT as
indicated below.
The customer or user of PONDUS® must ensure that it is used in such an
environment.
Guidelines and MANUFACTURER‘s declaration – ELECTROMAGNETIC EMISSIONS
PONDUS® is designed for operation in an ELECTROMAGNETIC ENVIRONMENT as
indicated below.
The customer or user of PONDUS® must ensure that it is used in such an
environment.
Interference emission measurements| Agreement| ELECTROMAGNETIC ENVIRONMENT –
Guidelines
---|---|---
HF emmissions according to CISPR 11| Group 1| PONDUS® exclusively uses HF
energy for its internal FUNCTIONS. HF emissions are very low and unlikely to
disrupt electronic devices within range.
HF emissions according to CISPR 11| Class B| PONDUS® is suitable for usage in
all establishments including residential areas and areas that are directly
connected to the PUBLIC SUPPLY GRID, which also supplies residential
buildings.
Harmonics emissions according to IEC 61000-3-2| Not applicable
Emissions of voltage fluctuations/flicker according to IEC 61000-3-3| Not
applicable
Table 1: Guidelines and manufacturer‘s declaration – electromagnetic emissions
Guidelines and MANUFACTURER‘s declaration – ELECTROMAGNETIC IMMUNITY
PONDUS® is designed for operation in an ELECTROMAGNETIC ENVIRONMENT as
indicated below. The customer or user of PONDUS® must ensure that it is used
in such an environment.
Immunity Test| IEC 60601-Test Level| Compliance Level| ELECTROMAGNETIC
ENVIRONMENT – Guidance
Electrostatic Discharge (ESD) EC 61000-4-2| ± 8 kV contact ± 2 kV, ± 4 kV, ±
8 kV, ± 15 kV air| ± 8 kV contact ± 15 kV air| Floors should be wood,
concrete, or ceramic tile. If floors are covered with synthetic material,
relative humidity should be at least 30%.
Electrical Fast Transient/Burst IEC 61000-4-4| ± 2 kV for power supply lines
Not applicable| ± 2 kV for power supply lines Not applicable| Mains power
quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5| ± 1 kV differential mode Not applicable| ± 1 kV
differential mode Not applicable| Mains power quality should be that of a
typical commercial or hospital environment.
---|---|---|---
Voltage dips, short interruptions, and voltage variations on power supply
input lines IEC 61000-4-11| < 5 % UT (> 95 % dip in UT) für 0.5 cycles 70 %
UT (30 % dip in UT) for 25 cycles < 5 % UT (> 95 % dip in UT) for 5 s| < 5 %
UT (> 95 dip in UT) for 0.5 cycles 70 % UT (30 % dip in UT) for 25 cycles
< 5 % UT (> 95 % dip in UT) for 5 s| Mains power quality should be that of a
typical commercial or hospital environment.
Power Frequency (50/60 Hz) Magnetic Field IEC 61000-4-8| Not applicable| Not
Applicable| Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital environment.
NOTE: UT is the AC mains voltage before application of the test level.
Table 2: Guidelines and manufacturer‘s declaration – electromagnetic immunity
Guidance and Manufacturer’s Declaration— ELECTROMAGNETIC IMMUNITY
PONDUS® is designed for operation in an ELECTROMAGNETIC ENVIRONMENT as
indicated below. The customer or user of PONDUS® must ensure that it is used
in such an environment.
Portable and mobile RF communications equipment should be used no closer to
any part of the device, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter.
Immunity Test| IEC 60601-Test Level| Compliance Level| ELECTROMAGNETIC
ENVIRONMENT – Guidance
| | | Recommended Separation Distance:
Conducted RF IEC 61000-4-6| 6 Vrms 150 kHz to 80 MHz| 6 V| = 0.58 √
---|---|---|---
Radiated RF IEC 61000-4-3
| 10 V/m 80 MHz to 2.7 GHz
| 10 V/m
| = 0.35 √ 80 MHz to 800 MHz
= 0.7 √ 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site surveya, should be less than the compliance level in each
frequency range b.
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTES:
– At 80 MHz and 800 MHz, the higher frequency range applies.
– These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the PONDUS® system is used
exceeds the applicable RF compliance
level above, the device should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the PONDUS® system.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than [4] V/m.
Table 3: Guidelines and manufacturer‘s declaration – electromagnetic immunity
Recommended separation distances between portable and mobile RF communications equipment and PONDUS®
PONDUS® is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. Users of this device can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communication equipment (transmitters) and the device
as recommended below, according to maximum output power of the communications
equipment.
Rated Maximum Output Power of Transmitter W| Separation Distance According to
Frequency of Transmitter
__
__
| __ 150 kHz to 80 MHz = 0.58 √| 80 MHz to 800 MHz = 0.35√| 800 MHz to
2.5 GHz = 0.7√
0.01| 0.058| 0.035| 0.07
0.1| 0.18| 0.11| 0.22
1| 0.58| 0.35| 0.70
10| 1.83| 1.11| 2.21
100| 5.80| 3.50| 7.00
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTES:
– At 80 MHz and 800 MHz, the higher frequency range applies.
– These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
Table 4: Recommended separation distances between portable and mobile RF communications equipment and PONDUS®
Clinical Application
3.1 Indications/Contraindications
In principle, for PONDUS® the same indications and contraindications apply as
for the standing frames it is used in combination with. Knowledge of the
contraindications is essential in order not to put the patient at risk. Before
applying PONDUS® in combination with a standing frame to a patient, check
carefully if one or more contraindications exist. Also, be aware that your
patient may have additional indications and/or contraindications that have not
been listed here but may be relevant. The following listings have no claim to
completeness.
Indications
PONDUS® can be used by children and adults with disabilities using a standing
frame. The target group for the PONDUS® is based on functional ability and
anthropometry. It is not based on a specific health condition or age. The user
is the person standing in the frame in which PONDUS® plates are mounted. The
caregiver is the person operating the frame and PONDUS®. Never leave the user
unattended in a standing frame. Ensure permanent supervision by a caregiver.
The use of the PONDUS® and standing frame will be discontinued more times
during the day where the user will be lying down or sitting in another
assistive device or a conventional product.
Contraindications
There are no known contraindications. However, you must familiarize yourself
with the contraindications of the specific standing frame model the PONDUS® is
used in combination with.
Precautions
Be aware of the height of the PONDUS® plates. When placed in the foot holds of
a standing frame, it positions the user approx. 20 mm higher, and thus the
mounted standing frame accessories should be adjusted accordingly to support
the user properly. Special attention should be given to the considerations
about the configuration of the standing frame and the use of standing frame
accessories for users with limitation in their:
- sensory functions and the sensation of pain, consider extra checks of the pressure caused by the support provided by the standing frame and its accessories.
- movement function (e.g., experience involuntary movements), pay special attention when using a standing frame
- ability to maintain a standing position, use the product with standing frame accessories (e.g., chest or back support).
- ability to maintain the position of their head, consider using standing frame accessories or to use a standing frame that provide support to the positioning of the head.
3.2 Safety
3.2.1 Safety concept
In order to prevent problems with the use of the device the following
instructions for proper use are also to be adhered to.
** Caregivers are obligated to familiarize themselves with these safety
instructions and avoid conditions that may lead to injuries or damages.
Caregiver
**
- Every caregiver must have read the instructions for use prior to using PONDUS®. TYROMOTION GmbH rejects any liability for damages to persons or material if safety regulations and instructions relevant to the usage of PONDUS® are not observed!
- Only use original accessories from the manufacturer that are supplied in the system contents.
- The caregiver must take appropriate measures to stabilize the user during use.
- PONDUS® must not be charged during use.
- If there is any doubt as to the continued safe use of PONDUS® or if any parts should fail or become defective, stop using the product immediately and contact TYROMOTION GmbH or your local dealer.
- The cleaning instructions stated in these instructions for use must be strictly observed.
Environment
- Do not use PONDUS® in combination with standing frames that have not been recommended by TYROMOTION GmbH. Unforeseen risks could arise for people and equipment.
PONDUS® may only be used in combination with the following standing frames:
o R82 Meerkat
o R82 Rabbit Up
o R82 Toucan
o R82 Caribou
o R82 Gazelle PS
Product
- No modifications of PONDUS® are allowed.
- Always contact the manufacturer for repairs!
3.2.2 Residual risk
An unpredictable residual risk remains despite all safety precautions. In rare
cases, the user may experience minor pinching or crushing injuries even during
proper operation. However, the probability of such injuries is very low, and
the injuries should not be severe as long as all safety instructions in the
present instructions for use are observed. TYROMOTION GmbH can provide a
detailed risk analysis upon request.
If you think that you or someone who used PONDUS® has experienced an injury
due to the use of PONDUS® please report this to the manufacturer and to your
national health authority.
PONDUS®
4.1 First steps
** The time required for PONDUS® to warm/cool from the min./max. storage
temperature between uses until the system is ready for its intended use at
ambient temperature is one hour. Please wait for this period of time before
operating the unit.
4.1.1 Charging PONDUS®
The PONDUS® system is equipped with a rechargeable battery. Before using
PONDUS® for the first time, fully charge the system using the supplied USB
cable and power supply unit (figure 4). The charging time is 3 hours when the
battery is completely empty.
4.1.2 Installation PONDUS® App**
Install the PONDUS® app directly from the respective app store or scan the QR
code (figure 2) using your smartphone or tablet.
4.2 Use of PONDUS®
To use PONDUS®, carry out the following steps:
- Switch on the two plates of the PONDUS® system by pushing the LED button on the front panel of the plates (figure 3). They should now start flashing yellow, which means they are in paring mode.
- Place the plates in the sandals provided on the standing frames. Make sure that the plates are engaged in the designated indentations of the sandals (figure 6). If you use PONDUS® in combination with the standing frames R82 Gazelle PS or R82 Caribou, you can also use the additional hook and loop adhesive pad set (figure 8).
- Now put the user on the plates and fix the feet in the sandals (figure 7).
- The plates must be paired to the PONDUS® App. For this, the Bluetooth function of your smartphone or tablet must be activated. For a successful Bluetooth connection, the plates must be positioned within a radius of 10m (in the same room) from the smartphone/tablet.
- Start the PONDUS® App on your smartphone/tablet (figure 3).
- The App should automatically find the two plates. If not, drag down from the top in the list view.
- Select the two devices matching the colors of your plates.
- Now the LED button on the front side of the plates should change color to green and blue, indicating left and right and the PONDUS® plates (figure 3).
- The plates are now connected to the PONDUS® App and PONDUS® is ready to use.
- Follow the instructions in the PONDUS® App.
Error messages are displayed directly in the app and errors can be corrected
interactively. If necessary, restart the app and plates and follow the steps
described above again.
If the PONDUS® plates are not connected to the PONDUS® App, they switch off
automatically after 10 minutes.
4.3 Shutdown of PONDUS®
Follow these steps to shut down PONDUS®:
- Release the patient from the sandals of the standing frame.
- End the Bluetooth connection by closing the PONDUS® App.
- Take the plates out of the sandals and switch off the PONDUS® plates by pushing the LED button on the front panel for at least 3 seconds (figure 3).
4.4 LED Status PONDUS®
Observe the LED button on the front side of the plates (figure 3). This
provides information on the status of PONDUS®.
LED-Display | Meaning |
---|---|
Flashing YELLOW | Waiting for a connection |
Solid COLOR | A smartphone/tablet was connected and set a color for |
identification
Flashing in-between ORANGE| Battery below 10%
Solid GREEN| PONDUS® plate is connected to a smartphone/tablet
Solid BLUE| PONDUS® plate is connected to a smartphone/tablet
Flashing GREEN| PONDUS® plate is sending data to a connected smartphone/tablet
Flashing BLUE| PONDUS® plate is sending data to a
connected smartphone/tablet
Solid ORANGE| PONDUS® plate is charging or fully charged
No light| PONDUS® plate is turned OFF. Click the button to switch it ON
4.5 Battery
Charging battery:
The integrated battery can be charged via power supply included in the system
content (figure 4). PONDUS® switches into the charging mode automatically.
Duration of Charging:
The duration of charging depends, among other things, upon whether the battery
was fully uncharged at the start of the charging process or not. The charging
only lasts at most 3 hours and thus should be done within 3 hours.
Charging Cycles and Lifespan:
Due to the use of high-quality lithium-polymer batteries, it is ensured that
the battery running times still meet the requirements even after many charging
cycles. Nevertheless, no guarantee can be given for the lifespan of batteries
going beyond 12 months after the date of purchase.
If you adhere to the following advice, you can certainly save the battery of
your device and thereby extend its lifespan:
- Avoid direct exposure on PONDUS® to sunlight in order to prevent the battery inside the device from becoming too hot.
- Use exclusively the power supply provided by TYROMOTION GmbH for the recharging of the battery.
Battery life:
In use, the battery operating time is minimum 8 hours (when fully charged).
Service information
Maintenance of a medical device is wholly the responsibility of the owner of
that device. Failure to maintain a device in accordance with the instructions
for use may invalidate the device’s warranty. Furthermore, failure to maintain
a device may compromise the clinical condition or safety of users.
5.1 Monthly functionality check
The functionality checks described here must be performed monthly. Perform the
checks even if PONDUS® indicates a malfunction (e.g., in case of unusual
sounds, elementary damages, etc.).
Inspection: | Malfunction: | Resulting measure: |
---|---|---|
Protective covers | • Covers shake |
• Covers missing
• Covers damaged| • Further use is prohibited
• Contact TYROMOTION GmbH
Externally visible deformations| • Parts bent out of shape
• Parts asymmetrical
• Parts defective| • Further use is prohibited
• Contact TYROMOTION GmbH
Table 5: Inspection points
5.2 Periodic check
The periodic test differs from the tests in chapter 5.1 in that the test
described here can be required by law, whereas the tests specified in chapter
5.1 serve, amongst other things, to determine acute damage or the need for
replacement of wear parts. The operating company of the device itself is
responsible for carrying out both tests.
An interval of one year is defined by TYROMOTION GmbH for carrying out the
periodic tests.
5.3 Lifetime
The lifetime of this product in normal use is 5 years if all maintenance and
servicing is carried out in accordance with the manufacturer’s instructions.
5.4 Cleaning instructions
The following instructions apply to the manual cleaning of medical devices by
Tyromotion GmbH.
Thorough cleaning and wiping is essential for the first time and reuse of
reusable medical devices. Effective cleaning must be performed to achieve
adequate decontamination.
The goal of cleaning is to remove any visibly sticky soil and reduce the
number of particles and microorganisms.
Cleaning must be carried out in a manner that minimizes the risks posed by
pathogens.
The devices of the Tyromotion GmbH must be cleaned and disinfected after
delivery before the first and any further use on the user.
5.4.1 Disinfection
The product can be disinfected with a 70% disinfectant IPA solution. It is
recommended to wipe off any residue and dirt from the product, using a cloth
with warm water and a mild detergent/soap without chlorine and let it dry
before disinfection.
5.4.2 Cleaning process
To minimize the risk of germ transmission, all surfaces that are touched by
the user should be periodically cleaned and disinfected and at least before
any further use on the user.
- If the user perspires heavily during use, dry the PONDUS® components after use before disinfecting them.
- Moisten the disposable cloth, according to the product information leaflet, only slightly with disinfectant. Wipe the PONDUS® components with the clean, soft and lint-free cloth. Observe the contact time of the disinfectant used according to the label and product information.
- Dry the area with a clean, non-abrasive, soft, lint-free cloth.
- Always store the PONDUS® components in clean and dry rooms or facilities after use.
5.5 Repair
Always contact the manufacturer for repairs!
5.6 Disposal
The PONDUS® must not be disposed of as household waste according to the
Directive 2012/19/EU on waste electrical and electronic equipment (WEEE-RL)
and the respective national legislation. The product must be disposed of at
the intended collection point or at a collection point approved for the
recycling of waste electrical and electronic equipment. It can also be
returned to TYROMOTION GmbH.
TYROMOTION GmbH
Bahnhofgürtel 59
8020 Graz, Austria
PHONE +43 316 908 909
EMAIL office@tyromotion.com
WEB www.tyromotion.com
SUPPORT support@tyromotion.com