FOREO ESPADA Range User Manual
- June 6, 2024
- FOREO
Table of Contents
FOREO ESPADA Range User Manual
GETTING STARTED
This booklet contains all essential instructions; for additional product resources, visit foreo.com.
WARNING: No modification of this equipment is allowed.
Please READ ALL INSTRUCTIONS BEFORE USE. Use this product only for its
intended use as described in this manual.
INTENDED USE
ESPADA™ has been developed to be used as an at-home device for the
treatment of mild to moderate inflammatory acne on the facial area. The device
is safe to use on all skin tones. ESPADA™ quickly, noninvasively and
effectively eliminates acne-causing bacteria.
Suitable for:
-
Mild acne (whiteheads, blackheads and papules)
-
Moderate inflammatory acne (inflamed, sensitive bumps filled with pus)
ESPADA™ FEATURES- Treatment surface
- Antibacterial silicone
- Start/stop button
- Charging port
- Indication light
-
Blue (LED) light – For killing acne bacteria. Activates when device touches the skin. Flashes 3 times to indicate the start of each routine. Each time ESPADA™ is switched on, it performsa self-check to ensure blue LED emissions remain at the appropriate level for use. If moved from the skin the light will automatically turn off.
-
T-Sonic™ pulsations – when the device is turned on, its pulsations will be in a lower intensity and when it touches the skin, the intensity will increase so more blue light can penetrate the pores for increased effectiveness. After the device is moved from the skin, the pulsations will return to a lower intensity setting automatically.
-
Targeting light (red cross-shaped light) – Targets the exact spot you wish to apply the treatment. Activates when device is away from the skin and turns off when the device touches the skin.
-
Built-in timer – Every 30 seconds the device will pause its pulsations and flash to indicate the end of each routine. After indication, the device should be moved to the next treatment spot.
-
If the device does not make contact with the skin when it is turned on, it will automatically switch off after 3 minutes.
USING YOUR ESPADA
Cleanse and dry your face before using the device. Press the universal
button once to activate. Direct it over the blemish you wish to treat with the
targeting light for ultimate precision. Gently place the device onto the
blemish and hold it there for 30 seconds and wait for the built-in timer to
indicate when the treatment is finished. Blemishes can be subjected to more
than one 30 second routine, depending on severity. For mild acne, apply two 30
second routines. For moderate acne, 3 to 4 routines. Move the device to
another blemish you wish to treat, if any, and repeat the procedure. There is
no need to switch off the device between treatments.
NOTE: SKIN WARMTH: The skin may feel warm during treatment, which is
normal and not cause for any concern. TEMPORARY SKIN REDNESS: The skin may
develop a slight redness that fades within 24 hours. TEMPORARY SKIN
DISCOLORATION: If the skin develops a slight discoloration that lasts more
than 24 hours, stop using ESPADA™.
Before your 1st treatment perform a simple compatibility test. Test the device
on a less sensitive area of the skin, such as the inner arm, before beginning
treatment on the face following the HOW TO USE instructions exactly. Wait 24
hours and check the treated area. If you don’t notice any difference (redness,
discoloration or irritation) between the treated and untreated skin, you may
begin the Blue Light Acne Treatment.
RISKS
Hazards | Effects to user | Caution |
---|---|---|
Overheating | May cause injury, discomfort or interrupted treatment | Stop use |
immediately, allow device to cool.
Inadvertent light contact with eyes| May cause discomfort and possibly
headache| Use caution to prevent light source from shining directly into the
eyes.
Photosensitivity| May cause redness, inflammation and irritation| Contact your
physician if you think you may be having a photosensitive reaction.
Adverse reaction with certain medications or supplements| May cause redness,
inflammation and irritation| Please read Contraindications before use.
Contact with neck
in users with thyroid conditions
| Possible stimulation of growth| Please read Contraindications before use.
Incorrect identification of “mild to moderate inflammatory acne”| Attempting
to use on more severe forms of acne or other skin conditions may cause adverse
results| Please read Intended Use before starting
WARNINGS
-
Consult your dermatologist prior to using ESPADA™ if you have medical concerns.
-
Do NOT use the device for severe, nodular or cystic acne.
-
Do NOT use the device on or come in contact with injured skin or skin that has undergone surgery.
-
Do NOT use the device on beauty spots/moles, warts, ingrown hairs or irritated skin.
-
Do NOT use device in and around the eye and lip area.
-
Do NOT use the device if you are particularly sensitive to light.
-
Do NOT use the device if you have had sunburn within the past 2 weeks.
-
Do NOT apply the treatment to a spot more than 3 times daily. Any more often than this could lead to skin burns or injuries. Leave a minimum time of 4 hours between each treatment.
-
Do NOT use device longer than 2 minutes on one blemish. Overuse can lead to skin injury.
-
Excessive use beyond the recommended treatment time may cause the device to become warm.
The device’s temperature will not go beyond 40°C if used at room temperature, but may cause discomfort or prolonged redness. -
If after using device you experience a worsening of the skin (skin feels tense, uncomfortable, irritated, red, burning, etc.), undesired side effects or prolonged redness, stop using the device and consult your doctor. However, a temporary, mild reddening after application is normal.
-
Never look directly at the blue light as this could lead to eye injury.
-
Do NOT let children use or play with ESPADA™. ESPADA™ has not been tested on children and may cause skin injury such as temporary redness or pigment changes (hypo- and hyperpigmentation) or affect their eyes.
-
Proper supervision and instructions for safe use should be given to those with reduced physical and mental abilities.
-
Do NOT share ESPADA™ with other people. Sharing the device could transfer bacteria from one person to another.
-
Do NOT use the device if it is visibly damaged, and never attempt to open or repair the device.
Doing this could result in damage or malfunction to the device, serious eye or skin injury, fire, or possibly fatal electric shock. -
Pregnant or breastfeeding women should not use this device.
-
Those who have had a migraine in the past two years should not use this device.
-
Do NOT use over a thyroid gland as it may result in growth stimulation.
-
If you are taking drugs or steroids for pain or skin conditions that cause light sensitivity, using this device may result in a photosensitive reaction.
-
Do not use over or near a steroid injection received within the last six months. Doing so may result in a photosensitive reaction.
-
In order to prevent fire or serious burns, immediately turn off device if it begins to overheat.
Keep away from aerosol products.
Contraindications
- Do NOT use if you have a history of light triggered seizures.
- Do NOT use this device over a suspicious or cancerous lesion as it can delay proper medical treatment
Typical characteristics of potentially cancerous lesions include:
- Asymmetry: one half of the abnormal skin area is different than the other half
- Borders: irregular borders
- Color: varies from one area to another with shades of tan, brown, or black (sometimes white, red, blue)
- Diameter: usually (but not always) larger than 6 mm in size (diameter of a pencil eraser)
- Bleeding: Any skin growth that bleeds or will not heal
Emissions test | Emissions test | Emissions test |
---|---|---|
Actinic UV | Risk Group 2 | UV emitted from this device |
May result in eye or skin irritation
Near UV| Risk Group 2| UV emitted from this device
May result in eye irritation
Blue-light| Risk Group 1| Not required
Retinal Thermal| Exempt Group| Not required
Retinal Thermal, weak visual stimulus| Exempt Group| Not required
Corneal/ Lens IR| Exempt Group| Not required
Spectral irradiance: 0.5lm. Maximum output of optical radiation: 75mW
ASSEMBLY INSTRUCTIONS
CHARGING
Your FOREO device is rechargeable with its included USB cable. It is
recommended that IEC60950 standard power suppliers are used to charge the
ESPADA™. 1.5 hours of charging delivers a full charge that will provide up to
50 30-second treatments.
While charging, the indicator will flash repeatedly. When the battery is full,
the indicator will emit a steady glow. When the battery is low, the indicator
will start to flash repeatedly.
CAUTION: Before charging, make sure the charging port and charger are
free from water. Do NOT use your FOREO device while charging, and do NOT
charge for more than 24 hours. Never use the device or charger if they are
damaged, and only use the charger supplied by FOREO with the device. To reduce
risk of electric shock, do not place the device where it can fall into water
or other liquid. If device accidentally falls into water when charging, unplug
immediately.
The charging chord should be kept away from all heated surfaces. Never insert
any object into any openings of the device. Never operate this device if its
chord or plug has been damaged or is working improperly.
MAINTENANCE AND SAFETY
To keep your FOREO device in optimum condition, we recommend cleaning ESPADA™’s treatment surface after each use. Clean the device with soap, rinse with warm water and dry with a soft, lint-free cloth. We recommend spraying the device with FOREO’s Silicone Cleaning Spray for optimal results. NEVER use cleaning products that are silicone-based or containing alcohol, petrol or acetone as they may damage silicone.
ENVIRONMENTAL REQUIREMENTS
Emissions test | Operation | Storage and Transport |
---|---|---|
Temperature | 5 to 40 °C | -10 to 50 °C |
Humidity | 15 to 93% RH, non-condensing | <=93% RH, non-condensing |
Pressure | Storage and Transport | 700 – 1060hPa |
WARRANTY TERMS AND CONDITIONS
REGISTER WARRANTY: To activate your 2-Year Limited Warranty, register the number provided on the FOREO Magnetic Card at www.foreo.com/support.
2-YEAR LIMITED WARRANTY: FOREO warrants this device for a period of TWO (2) YEARS after the original date of purchase against defects due to faulty workmanship or materials arising from Normal Use of the device. The warranty covers working parts that affect the function of the device. It does NOT cover cosmetic deterioration caused by fair wear and tear, or damage caused by accident, misuse or neglect. Any attempt to open or take apart the device (or its accessories) will void the warranty. If you discover a defect and notify FOREO during the warranty period, FOREO will, at its discretion, replace the device free of charge. Claims under warranty must be supported by reasonable evidence that the date of the claim is within the warranty period. To validate your warranty, please keep your original purchase receipt together with these warranty conditions for the duration of the warranty period. To claim your warranty, you must log in to your account at www.foreo.com then select the option to make a warranty claim. Shipping costs are nonrefundable. This undertaking is in addition to your statutory rights as a consumer and does not affect those rights in any way.
FOR AUSTRALIA-BASED CUSTOMERS ONLY: Our goods come with guarantees that cannot be excluded under the Australian Consumer Law. You are entitled to a replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage. You are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable quality and the failure does not amount to a major failure. Australia Office, FOREO Oceania, Level 1, Suite 10, 1014 Doncaster Road, Doncaster East, VIC, 3109, Australia. TEL +61 (0) 448 885 779
DISPOSAL INFORMATION: Battery must be recycled or disposed of properly.
Disposal of old electronic equipment (applicable in the EU and other European
countries with separate waste collection systems). The crossed-out dustbin
symbol indicates that this device should not be treated as household waste,
but rather be brought to the appropriate collection point for recycling of
electrical and electronic equipment. Please see back page for instructions on
how to remove the battery.
BATTERY CARE: It is recommended to regularly use the device to ensure proper performance. ESPADA™ should be fully charged at least every 3 months to maintain the battery.
GUIDANCE AND MANUFACTURER’S DECLARATION: ELECTROMAGNETIC EMISSIONS & IMMUNITY: Medical electrical equipment requires special precautions regarding EMC and must be installed and put into service according to EMC information provided in this document. Portable and mobile RF communications equipment can affect the medical electrical equipment. This declaration currently applies to ESPADA™ device
GUIDANCE AND MANUFACTURER’S DECLARATION: ELECTROMAGNETIC EMISSIONS: The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment
Emissions test| Compliance| Electromagnetic environment-
guidance
---|---|---
RF emissions CISPR 15| Group 1| The device is supplied by very low voltage Li-
on battery.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR 11| Group 2| The device does not need to use an AC adapter
to charge, so it’s radiated power is very low.
WARNINGS:
- This device should not be used adjacent to other high field equipment or areas. If adjacent use is necessary, the device should be observed to verify normal operation in the configuration in which it will be used.
- This device should not be used adjacent to high-frequency equipment. If the device is adjacent to a short wave device or microwave device, the device’s output may be unstable.
- The use of accessories other than those specified for the device is not recommended as it may result in increased emissions or decreased immunity of the device.
- The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that the device is used in such an environment.
Immunity test| IEC60601-1-2 test level| Compliance
level| Electromagnetic environment guidance
---|---|---|---
Electrostatic Discharge(ESD) IEC 61000-4-2| ± 6kV contact ± 8kV air| ± 6kV
contact ± 8kV air| Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity should be at least
30%.
Electromagnetic field susceptibility IEC61000-4-3| 3V/m| 3V/m| Because it is
not life-supporting equipment, it does not use only shield location.
Power frequency (50/60 HZ Magnetic field IEC 61000-4-8| 3A/m| 3A/m| Power
frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial
or hospital environment.
Radiated emission| 40 dBµV/M (3M)
30MHZ ~ 230MHZ
47 dBµV/M (3M) 230MHZ~ 1000MHZ| 40 dBµV/M (3M) 30MHZ ~ 230MHZ 47 dBµV/M (3M)
230MHZ~ 1000MHZ| In all frequencies, the radiated strength is less than
limited value.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless), telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which ESPADA™ device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.
Notes: These guidelines may not apply in all situations. Electromagnetic propagation is affected.
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE DEVICE
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communication equipment.
Rated maximum output Power of transmitter (w)|
Separation distance according to frequency of transmitter (m)
---|---
30MHZ to 80MHZ| 80MHZ to 800MHZ
d = 1.2√p
| 800MHZ to 1000MHZ
d = 2.3√p
0.01| d = 1.2√p| 0.12| 0.23
0.1| d = 1.2√p| 0.38| 0.73
1| d = 2.3√p| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output not listed above, the recommended separation distance (d) in meters (m) can be estimated using the equation application to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacture.
Notes: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
ESPADA™ complies with regulatory requirements of the following European Directive 93/42/EEC concerning medical devices and all the latest standards listed below.
N60601-1/IEC60601-1, EN60601-1-2/IEC60601-1-2, IEC60601-1-11, EN60601-2-57/IEC60601-2-57, EN62471/IEC62471, EN ISO10993-5, EN ISO10993-10
SUPPORT
Suzhou Sventech Co., Ltd.
Address: No.71, Xinqing Road, Suzhou Industrial Park, Jiangsu, China
215123,
Foreo AB, Karlavägen 41, 114 31, Stockholm, Sweden
customercare@foreo.com