ACON SARS-CoV-2 IgG/IgM Rapid Test Instructions
- June 6, 2024
- ACON
Table of Contents
- ACON SARS-CoV-2 IgG/IgM Rapid Test
- INTENDED USE
- PRINCIPLE
- PRECAUTIONS
- SPECIMEN COLLECTION AND PREPARATION
- MATERIALS
- DIRECTIONS FOR USE
- INTERPRETATION OF RESULTS
- QUALITY CONTROL
- LIMITATIONS
- CONDITIONS OF AUTHORIZATION FOR THE LABORATORY
- PERFORMANCE CHARACTERISTICS
- INTENDED USE
- WARNINGS AND PRECAUTIONS
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ACON SARS-CoV-2 IgG/IgM Rapid Test
REF L031- 11711
A rapid test for the qualitative detection of IgM and IgG antibodies to the
SARS-CoV-2 virus in serum, plasma, or venous whole blood.
For prescription use only. For in vitro diagnostic use only. For Emergency Use
Authorization only.
INTENDED USE
The ACON SARS-CoV-2 IgG/IgM Rapid Test is a lateral flow chromatographic immunoassay intended for the qualitative detection and differentiation of IgM and IgG antibodies to SARS-CoV-2 in human serum, plasma (K2-EDTA, sodium heparin and sodium citrate) and venous whole blood (K2-EDTA, sodium heparin and sodium citrate). The ACON SARS-CoV-2 IgG/IgM Rapid Test is intended for use as an aid in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection. The ACON SARS-CoV-2 IgG/IgM Rapid Test should not be used to diagnose acute SARS-CoV-2 infection. At this time, it is unknown how long antibodies persist following infection and if the presence of antibodies confers protective immunity. Testing is limited to laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C 263a, that meet requirements to perform moderate or high complexity tests.
Results are for the detection of IgM and IgG SARS-CoV-2 antibodies. IgM and IgG antibodies to SARS-CoV-2 are generally detectable in blood several days after initial infection, although the duration of time antibodies are present post-infection is not well characterized. Individuals may have detectable virus present for several weeks following seroconversion.
Laboratories within the United States and its territories should report all
results to the appropriate public health authorities.
The sensitivity of the ACON SARS-CoV-2 IgG/IgM Rapid Test after infection is
unknown. Negative results do not preclude acute SARS-CoV-2 infection. If acute
infection is suspected, direct testing for SARS-CoV-2 is necessary.
False positive results for the ACON SARS-CoV-2 IgG/IgM Rapid Test may occur
due to cross-reactivity from pre-existing antibodies or other possible causes.
Due to the risk of false positive results, confirmation of positive results
should be considered using a second, different IgG or IgM assay.
The ACON SARS-CoV-2 IgG/IgM Rapid Test is only for use under the Food and Drug
Administration’s Emergency Use Authorization.
SUMMARY
Coronaviruses (CoV) are a large family of viruses that cause illness ranging
from the common cold to more severe diseases such as Severe Acute Respiratory
Syndrome (SARS-CoV). The novel coronaviruses belong to the β genus. COVID-19
is an acute respiratory infectious disease. People are generally susceptible.
Currently, the patients infected by the novel coronavirus are the main source
of infection; asymptomatic infected people can also be an infectious source.
Based on the current epidemiological investigation, the incubation period is 1
to 14 days, mostly 3 to 7 days. The main manifestations include fever, fatigue
and dry cough. Nasal congestion, runny nose, sore throat, myalgia and diarrhea
are found in a few cases.
Standard recommendations to prevent infection spread include regular hand
washing, covering mouth and nose when coughing and sneezing, and socially
distancing as recommended by your local jurisdictions.
PRINCIPLE
The ACON SARS-CoV-2 IgG/IgM Rapid Test is a qualitative membrane-based immunoassay for the detection of IgG and IgM antibodies to SARS-CoV-2 in human serum, plasma or whole blood collected using K2-EDTA, sodium heparin or sodium citrate anticoagulants. The membrane is pre-coated with anti-human IgM antibody on the IgM Test Line region (M) and anti-human IgG antibody on the IgG Test Line region (G). During testing, SARS-CoV-2 antibodies, if present in the specimen, will react with the SARS-CoV-2 antigen-coated particles, which have been pre-coated on the test strip. The mixture then migrates upward on the membrane by capillary action, reacting with anti-human IgM antibody on the IgM Test Line region (M) and/or with anti-human IgG antibody on the IgG Test Line region (G), forming a colored line in IgM line region (M) and/or IgG line region (G). To serve as a procedure control, a colored line should appear in the control line region indicating that proper volume of specimen has been added and membrane wicking has occurred. The presence of the control line together with the absence of the colored lines in IgM line region (M) and IgG line region (G) indicates that the specimen does not have any SARS-CoV-2 antibodies.
REAGENTS
The test cassette contains SARS-CoV-2 recombination antigens coated particles
and rabbit IgG coated particles; anti-human IgM, anti-human IgG and anti-
rabbit IgG are coated on the membrane.
PRECAUTIONS
- For Emergency Use Authorization only
- For Prescription Use only
- For professional in vitro diagnostic use only.
- Do not use after the expiration date.
- Do not eat, drink or smoke in the area where the specimens or kits are handled.
- Do not use the test if the pouch is damaged.
- Handle all specimens as if they contain infectious agents. Observe established precautions against biological hazards throughout testing and follow the standard procedures for proper disposal of specimens.
- Wear protective clothing such as laboratory coats, disposable gloves and eye protection when specimens are being tested.
- The used test should be discarded according to local regulations.
- Humidity and temperature can adversely affect results.
- This package insert must be read completely before performing the test. Failure to follow directions in insert may yield inaccurate test results.
- Caution: Federal Law restricts this device to sale by or on the order of a licensed practitioner.
- This test has not been FDA cleared or approved, but has been authorized for emergency use by FDA under an EUA for use by laboratories certified under CLIA that meet requirements to perform high-complexity tests.
- This test has been authorized only for the presence of IgG and IgM antibodies against SARS-CoV-2, not for any other viruses or pathogens.
- The emergency use of this test is only authorized for the duration of the declaration that circumstances exist justifying the authorization of emergency use of in vitro diagnostic tests for detection and/or diagnosis of COVID-19 under Section 564(b)(1) of the Act, 21 U.S.C. § 360bbb-3(b)(1), unless the declarations is terminated or authorization is revoked sooner.
STORAGE AND STABILITY
- The kit can be stored at 2-30°C.
- The test is stable until the expiration date printed on the sealed pouch.
- The test must remain in the sealed pouch until use.
- DO NOT FREEZE.
- Do not use after the expiration date.
SPECIMEN COLLECTION AND PREPARATION
- The ACON SARS-CoV-2 IgG/IgM Rapid Test can be performed using serum, or plasma or whole blood specimen collected using K2-EDTA, sodium heparin or sodium citrate anticoagulants.
- Testing should be performed immediately after specimen collection. Store specimens right after collection, if not tested immediately. Serum and plasma specimens may be stored at 2-8°C for up to 3 days. For long-term storage, specimens should be kept below -20°C. Whole blood collected by venipuncture should be stored at 2-8°C if not tested immediately. The specimens must be tested within 2 days of collection. Do not freeze whole blood specimens.
- Bring specimens to room temperature prior to testing. Frozen specimens must be completely thawed and mixed well prior to testing. Specimens should not be frozen and thawed more than once.
- If specimens are to be shipped, they should be packed in compliance with local regulations covering the transportation of etiologic agents.
- Anticoagulants sodium heparin, K2-EDTA and sodium citrate do not affect the test result.
MATERIALS
Materials Provided
- Test Cassettes (25/kit)
- Buffer (2 bottles/ 2.5mL each)
- Droppers (25/kit)
- Package insert (1)
Materials Required But Not Provided
- Pipette and disposable tips
- Timer
- ACON External Positive and Negative Controls (Catalog # L021-1011)
- Centrifuge to isolate plasma and serum
DIRECTIONS FOR USE
Allow the test, specimen and buffer to equilibrate to room temperature (15-30°C) prior to testing.
-
Remove the test from the foil pouch and use it as soon as possible. Best results will be obtained if the assay is performed within one hour.
-
Place the test on a flat and clean surface. Transfer the specimen by a Pipette or a Dropper:
To use a Pipette for Serum, Plasma (sodium heparin, K2-EDTA and sodium citrate) or venous whole blood (sodium heparin, K2-EDTA and sodium citrate):
Transfer 10 μL of Serum, Plasma, or 15 μL of venous Whole blood specimen into the Sample Well (S), then add 2~3 drops of buffer into the Buffer Well (B) and start the timer. Avoid air bubbles in the Sample and Buffer well. See illustration below.
To use a Dropper for Serum or Plasma (sodium heparin, K2-EDTA and sodium citrate): Hold the dropper vertically and fill the capillary part of the dropper (not to exceed the capillary part) with Serum or Plasma (approximately 10 μL), then carefully dispense the specimen into the Sample Well (S), immediately add 2~3 drops of buffer into the Buffer Well (B), and start the timer. Avoid air bubbles in the Sample and Buffer well. See illustration below.
To use a Dropper for venous whole blood (sodium heparin, K2-EDTA and sodium citrate):: Hold the dropper vertically, draw the specimen about 2-3mm above the capillary part and then transfer 1 full drop (approximately 15 μL) of specimen into the Sample Well (S). Immediately add 2~3 drops of buffer into the Buffer Well (B) and start the timer. Avoid air bubbles in the Sample and Buffer well. See illustration below. -
Wait for the colored line(s) to appear. The result should be read at 15 minutes. Do not interpret the result after 20 minutes.
INTERPRETATION OF RESULTS
Results | Interpretation of results (Please refer to the illustration above) |
---|---|
IgM (-) IgG (-) | Only one colored control line appears in the control region. |
The result is negative for SARS-CoV-2 virus specific IgM and IgG antibodies.
IgM (+) IgG (-)| Colored control line appears in the control region and one
colored line appears in the IgM line region (M). The result is positive for
IgM antibodies and negative for IgG antibodies to SARS-CoV-2.
IgM (+) IgG (+)| Colored control line appears in the control region, one
colored line appears in the IgG line region (G), and one colored line appears
in the IgM line region (M). The color intensities of the lines do not have to
match. The result is positive for IgG and IgM antibodies to SARS-CoV-2.
IgM (-) IgG (+)| Colored control line appears in the control region and one
colored line appears in the IgG line region (G). The result is positive for
IgG antibodies and negative for IgM antibodies to SARS-CoV-2.
INVALID| Control line fails to appear. Insufficient specimen volume
or incorrect operation are the most likely reasons for control line
failure. Review the procedure and repeat testing with a new test
cassette. If the problem persists, do not use the test kit and contact
ACON Laboratories Inc.
QUALITY CONTROL
Internal procedural controls are included in the test. The control line is coated with anti-rabbit IgG antibody on the membrane which will bind to the pre-coated gold-conjugated rabbit IgG when an adequate volume of test specimen is applied into the sample well and adequate buffer volume is applies to the corresponding well on the test cassette. In the control region of the membrane, a colored line appears regardless of the presence of SARS-CoV-2 IgG or IgM antibodies in the specimen. This confirms sufficient specimen and buffer volume, and correct procedural technique. Absence of this line indicates an invalid result.
Positive and negative control standards are not supplied with this kit, however they should be purchased separately: ACON SARS-CoV-2 IgG/IgM Rapid Test Control Solution (Catalog # L021-1011). Please contact ACON Laboratories, Inc. for purchasing information. The positive and negative controls should be tested as a good laboratory practice to confirm the test procedure and to verify proper test performance. Follow the test procedure provided here and in the control solution package insert. It is recommended that both controls be processed under the following circumstances:
- A new operator uses the test kits for the first time.
- A new shipment of test kits is received.
- Device storage falls out of the 2-30°C range.
- To verify a higher or lower than expected frequency of positive or negative results.
- To investigate the cause of repeated invalid results.
- A new test environment is used.
LIMITATIONS
- This test is only to be used in CLIA certified laboratories that meet requirements to perform moderate or high complexity testing and not in point-of-care or at-home testing settings.
- The test should be used for the detection of SARS-CoV-2 antibodies in serum, or plasma, or whole blood specimens collected with anticoagulants K2-EDTA, sodium heparin and sodium citrate. Other specimens have not been evaluated and should not be used with this assay. Do not use with fingerstick (capillary) whole blood samples.
- The test is limited to the qualitative detection of IgM and IgG antibodies against the SARS-CoV-2 virus. The intensity of the test line does not necessarily correlate to SARS-CoV-2 antibody titer in the specimen.
- It is unknown at this time if the presence of antibodies to SARS-CoV-2 confers immunity to re-infection
- This test should only be used for testing samples collected 8 days after symptom onset. The performance of this test in samples collected less than 8 days after symptom onset has not been established
- Results from antibody testing should not be used to diagnose or exclude acute SARS-CoV-2 infection or to inform infection status. An assay that directly detects the virus should be used to evaluate symptomatic patients for acute COVID-19.
- A positive result may not indicate previous SARS-CoV-2 infection. Consider other information, including clinical history, physical findings, local disease prevalence, and other diagnostic procedures in assessing the need for a second but different serology test to confirm an immune response.
- Positive results may be due to past or present infection with non-SARS-CoV-2 coronavirus strains, such as coronavirus HKU1, NL63, OC43, or 229E, OR if appropriate: Pedigreed specimens with direct evidence of antibodies to non-SARS-CoV-2 coronavirus (common cold) strains such as HKU1, NL63, OC43, or 229E have not been evaluated with this assay.
- A negative result for an individual subject indicates absence of detectable anti-SARS-CoV-2 antibodies. Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient management decisions. The sensitivity of this assay early after infection is unknown.
- A negative result can occur if the quantity of the anti-SARS-CoV-2 antibodies present in the specimen is below the detection limits of the assay, or the antibodies that are detected are not present during the stage of disease in which a sample is collected.
- Results from immunosuppressed patients should be interpreted with caution.
- Reading test results earlier than 15 minutes or later than 20 minutes after the addition of Buffer may yield erroneous results.
- This test should not be used for screening of donated blood.
- The performance of this test has not been established in individuals that have received a COVID-19 vaccine. The clinical significance of a positive or negative antibody result following COVID-19 vaccination has not been established, and the result from this test should not be interpreted as an indication or degree of protection from infection after vaccination.
- The performance of this test was established based on the evaluation of a limited number of clinical specimens collected between July 2016 and September 2020 in Hefei, Anhui, China; Zhejiang, P.R. China; National City, California; La Jolla, California; San Diego, California, and Brooklyn, New York. Clinical performance has not been established with all circulating variants but is anticipated to be reflective of the prevalent variants in circulation at the time and location of the clinical evaluation. Performance at the time of testing may vary depending on the variants circulating, including newly emerging strains of SARS-CoV-2 and their prevalence, which change over time.
CONDITIONS OF AUTHORIZATION FOR THE LABORATORY
The ACON SARS-CoV-2 IgG/IgM Rapid Test Letter of Authorization, along with the
authorized Fact Sheet for Healthcare Providers, the authorized Fact Sheet for
Recipients, and authorized labeling are available on the FDA website:
https://www.fda.gov/medical-devices/coronavirus-disease-2019-covid-19
-emergency-use-authorizations-medical-devices/vitro-diagnostics-euas.
Authorized laboratories using the ACON SARS-CoV-2 IgG/IgM Rapid Test (“your
product” in the conditions below), must adhere to the Conditions of
Authorization indicated in the Letter of Authorization as listed below:
- Authorized laboratories* using your product will include with test result reports, all authorized Fact Sheets. Under exigent circumstances, other appropriate methods for disseminating these Fact Sheets may be used, which may include mass media.
- Authorized laboratories will use your product as outlined in the authorized labeling. Deviations from the authorized procedures, including the authorized clinical specimen types, authorized control materials, authorized other ancillary reagents and authorized materials required to use your product are not permitted.
- Authorized laboratories that receive your product will notify the relevant public health authorities of their intent to run your product prior to initiating testing.
- Authorized laboratories using your product will have a process in place for reporting test results to healthcare providers and relevant public health authorities, as appropriate.
- Authorized laboratories will collect information on the performance of your product and report to DMD/OHT7-OIR/OPEQ/CDRH (via email: CDRH-EUA-Reporting@fda.hhs.gov) and to you (support@aconlabs.com) any suspected occurrence of false positive or false negative results and significant deviations from the established performance characteristics of your product of which they become aware.
- All laboratory personnel using your product must be appropriately trained in immunochromatographic techniques and use appropriate laboratory and personal protective equipment when handling this kit, and use your product in accordance with the authorized labeling. All laboratory personnel using the assay must also be trained in and be familiar with the interpretation of results of the product.
- ACON Laboratories, Inc, authorized distributors, and authorized laboratories using your product will ensure that any records associated with this EUA are maintained until otherwise notified by FDA. Such records will be made available to FDA for inspection upon request.
*The letter of authorization refers to, “Laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. §263a, that meet requirements to perform moderate or high complexity tests” as “authorized laboratories.”
PERFORMANCE CHARACTERISTICS
Clinical Performance
Study 1
The ACON SARS-CoV-2 IgG/IgM Rapid Test has been tested with serum collected in
the U.S. during the pandemic. All patients were confirmed positive or negative
with an EUA authorized RT-PCR test. A total of 127 unique serum samples were
collected. Seventy-three samples were collected from RT-PCR negative
individuals, and 54 from RT-PCR positive individuals. The results are
summarized in the tables below:
SARS-CoV-2 IgM results from RT-PCR confirmed positive subjects
| ACON SARS-CoV-2IgG/IgM Rapid Test IgM Results
---|---
Days from Symptom Onset| Number of Subjects Tested| IgM Positive results|
IgM PPA
| ****
95% CI
0-7 days| 3| 3| 100% (3/3)| 29.2-100%
8-14 days| 4| 3| 75% (3/4)| 19.4-99.4%
≥15 days| 47| 38| 80.9% (38/47)| 66.7-90.9%
Total| 54| | |
SARS-CoV-2 IgG results from RT-PCR confirmed positive subjects
| ACON SARS-CoV-2 IgG/IgM Rapid test IgG Results
---|---
Days from Symptom Onset| Number of Subjects Tested| IgG Positive results|
IgG PPA
|
95% CI
0-7 days| 3| 3| 100% (3/3)| 29.2-100%
8-14 days| 4| 4| 100% (4/4)| 39.8-100%
≥15 days| 47| 47| 100% (47/47)| 92.5-100%
Total Subjects| 54| | |
SARS-CoV-2 IgG/IgM combined results from RT-PCR positive confirmed subjects
| ACON SARS-CoV-2 IgG/IgM Rapid test IgG/IgM combined Results
---|---
Days from Symptom Onset
|
Number of Subjects Tested
| IgG/IgM Combined Antibody Positive results| IgG/IgM Combined Antibody PPA|
95% CI
0-7 days| 3| 3| 100% (3/3)| 29.2-100%
8-14 days| 4| 4| 100% (4/4)| 39.8-100%
≥15 days| 47| 47| 100% (47/47)| 92.5-100%
Total Subjects| 54| | |
Overall NPA of IgG/IgM results with RT-PCR confirmed negative samples
| ACON SARS-CoV-2 IgG/IgM Rapid test Results
---|---
Number of Subjects Tested| Overall (IgG+IgM) Negative Results| Overall NPA
(95% CI)
73| 70| 95.89% (70/73)
(95% CI: 88.5 – 99.1)
Study 2
The ACON SARS-CoV-2 Rapid Test was also tested with plasma samples collected
in the U.S. before SARS-CoV-2 pandemic. A total of 368 unique plasma samples
were collected from 4 sites in the U.S. before November 2019. Sodium heparin
(167 samples) and sodium citrate (201 samples) were used as anticoagulants.
NPA calculations were very similar in all matrices evaluated. The overall NPA
performance of the test is summarized in the table below:
Overall negative percent agreement (NPA) from plasma samples collected in the US prior to COVID-19 pandemic.
| ACON SARS-CoV-2 IgG/IgM Rapid test Results (all samples combined)
---|---
Number of Subjects Tested| Overall (IgG+IgM) Negative Results|
Overall NPA (95% CI)
368
|
364
| 364/368 (98.9%);
(95% CI: 97.2-99.7%)
Study 3
The ACON SARS-CoV-2 IgG/IgM Rapid Test was tested with 112 unique serum
samples collected from RT-PCR positive individuals in China during the
pandemic. The results are summarized in the tables below:
SARS-CoV-2 IgM results from RT-PCR confirmed positive subjects
| ACON SARS-CoV-2IgG/IgM Rapid Test IgM Results
---|---
Days from Symptom Onset| Number of Subjects Tested| IgM Positive results|
IgM PPA
| ****
95% CI
0-7 days| 0| N/A| N/A| N/A
8-14 days| 36| 36| 36/36 (100%)| 90.3 – 100%
≥15 days| 76| 61| 61/76 (80.3%)| 69.5 -88.5%
Total| 112| | |
SARS-CoV-2 IgG results from RT-PCR confirmed positive subjects
| ACON SARS-CoV-2 IgG/IgM Rapid test IgG Results
---|---
Days from Symptom Onset| Number of Subjects Tested| IgG Positive results|
IgG PPA
|
95% CI
0-7 days| 0| N/A| N/A| N/A
8-14 days| 36| 35| 35/36 (97.2%)| 85.5 -99.9%
≥15 days| 76| 75| 75/76 (98.7%)| 92.9 – 100%
Total Subjects| 112| | |
SARS-CoV-2 IgG/IgM combined results from RT-PCR positive confirmed subjects
| ACON SARS-CoV-2 IgG/IgM Rapid test IgG/IgM combined Results
---|---
Days from Symptom Onset
| ****
Number of Subjects Tested
| IgG/IgM Combined Antibody Positive results| IgG/IgM Combined Antibody PPA|
95% CI
0-7 days| 0| N/A| N/A| N/A
8-14 days| 36| 36| 36/36 (100%)| 90.3 -100%
≥15 days| 76| 75| 75/76 (98.7%)| 92.9 – 100%
Total Subjects| 112| | |
Study 4
The ACON SARS-CoV-2 Rapid Test was also tested with 686 unique serum (300
samples) and plasma samples (191 K2-EDTA, 155 sodium heparin, and 40 sodium
citrate) collected in China before November 2019. The overall NPA performance
of the test is summarized in the table below:
Overall negative percent agreement (NPA) from plasma samples collected in the China prior to COVID-19 pandemic.
| ACON SARS-CoV-2 IgG/IgM Rapid test Results (all samples combined)
---|---
Number of Subjects Tested| Overall (IgG+IgM) Negative Results|
Overall NPA (95% CI)
686
|
674
| 674/686 (98.3%);
(97 – 99.1%)
Study 5: Seroconversion Study
A longitudinal study was conducted to evaluate the performance of the test in
samples collected over time and detect seroconversion. In this study multiple
serum samples were collected over time from 10 RT-PCR positive individuals,
and the first bleed result from each subject was analyzed in each of the time
bins post symptom onset as summarized in the tables below.
SARS-CoV-2 IgM longitudinal study results
| ACON SARS-CoV-2IgG/IgM
Rapid Test IgM Results
---|---
Days from Symptom Onset| Number of Subjects Tested| IgM Positive results| IgM
PPA
0-7 days| 10| 6| 60%
8-14 days| 10| 10| 100%
≥15 days| 10| 10| 100%
SARS-CoV-2 IgG longitudinal study results
| ACON SARS-CoV-2IgG/IgM
Rapid Test IgG Results
---|---
Days from Symptom Onset| Number of Subjects Tested| IgG Positive results|
IgG PPA
0-7 days| 10| 4| 40%
8-14 days| 10| 9| 90%
≥15 days| 10| 10| 100%
Independent Clinical Agreement Validation.
The SARS-CoV-2 IgG/IgM Rapid Test from ACON was tested on November 02, 2020 at
the Frederick National Laboratory for Cancer Research (FNLCR) sponsored by the
National Cancer Institute (NCI). The test was validated against a panel of
previously frozen samples consisting of 30 SARS-CoV-2 antibody-positive serum
samples and 80 antibody-negative serum and plasma samples. Each of the 30
antibody-positive samples were confirmed with a nucleic acid amplification
test (NAAT) and both IgM and IgG antibodies were confirmed to be present in
all 30 samples. The presence of antibodies in the samples was confirmed by
several orthogonal methods prior to testing with the ACON SARS-CoV-2 IgG/IgM
Rapid Test. The presence of IgM and IgG antibodies specifically was confirmed
by one or more comparator methods. Antibody-positive samples were selected at
different antibody titers.
All antibody-negative samples were collected prior to 2020 and include: i) Seventy (70) samples selected without regard to clinical status, “Negatives” and ii) Ten (10) samples selected from banked serum from HIV+ patients, “HIV+”. A summary of the results of this study are shown in the tables below.
A C ON S A R S -CoV-2
Ig G/IgM R a pi d T es t
| Co mparator M e thod|
To tal
---|---|---
P o s i tive (IgM+,IgG+)| N e g a tive
(IgM/IgG)-| N e g a tive, HIV+
P o s i tive
| Ig M + / Ig G+| 29| 0| 0| 29
IgM + , Ig G –| 0| 1| 0| 1
IgM -, Ig G+| 1| 2| 0| 3
N e g a tive| IgM – / Ig G-| 0| 67|
10| 77
To tal (n=110)| 30| 70| 10| 110
M eas u r e| Estimate| Con fidence In
terval
---|---|---
IgM Sensitivity| 96.7% (29/30)| (83.3%; 99.4%)
IgM Specificity| 98.8% (79/80)| (93.3%; 99.8%)
IgG Sensitivity| 100% (30/30)| (88.7%; 100%)
IgG Specificity| 97.5% (78/80)| (91.3%; 99.3%)
Combined Sensitivity| 100% (30/30)| (88.7%; 100%)
Combined Specificity| 96.2% (77/80)| (89.5%; 98.7%)
Combined PPV for prevalence =5.0%| 58.4%| (30.9%; 80.4%)
Combined NPV for prevalence =5.0%| 100%| (99.3%; 100%)
---|---|---
Cross-reactivity with HIV+| 0.0% (0/10),
not detected
|
Potential Cross-Reactivity
The ACON SARS-CoV-2 Rapid Test was evaluated for potential cross-reactivity
testing 5 samples containing antibodies to the following underlying
conditions: HIV, Hepatitis B virus, Hepatitis C virus, Treponema pallidum, EB
virus, Mycoplasma pneumoniae, Varicella-zoster virus, Influenza A/B,
Rhinovirus, Haemophilus influenza, other coronaviruses (229E, NL63, OC43,
HKU1), Chlamydia pneumonia, Legionella pneumophila, Adenovirus, Measles virus,
Cytomegalovirus, Herpes simplex virus-1/2 or Respiratory syncytial virus,
Rheumatoid factor positive and ANA positive samples. No false positive results
were observed.
Potential Interference
Potential interference with the ACON SARS-CoV-2 Rapid Test was evaluated by
assessing the potentially interfering substances listed in the table below in
SARS-CoV-2 antibody negative and IgG and IgM low positive specimens. No false
results were observed with the ACON SARS-CoV-2 IgG/IgM Rapid Test at the
concentrations of the potential interference substances shown in the table
below.
E ndog e nou s In terfering S ub s
tances| Con centration
---|---
Creatinine hydrochoride| 5mg/dL
Urea| 257mg/dL
Hemoglobin| 9g/L
Glucose| 1000mg/dL
Glynbenclomide| 1mg/dL
Dopamine| 0.09mg/dL
Albumin Human| 6mg/dL
Bilirubin| 37.5mg/dL (342µmol/L)
Uric acid| 25mg/dL
Biotin| 1200ng/mL
Triglyceride| 1329mg/dL (15mmol/L)
Rheumatoid factor| 80IU/mL
Antinuclear antibody (ANA)| Titer 1:240
Anti-mitochondrial antibody (AMA)| 80U/mL
Ex og e nou s In terfering S ub s
tances| Con centration
Ibuprofen| 50mg/dL
Sulfamethoxazole| 40mg/dL
Benzafibrate| 10mg/dL
Acetylsalicylic acid| 65mg/dL
Sodium Pyroracemic| 2.7mg/dL
Lactate dehydrogenase| 2000mg/dL
Methydopa| 250U/L
Indomethacin| 1.5mg/dL
Nicotinic acid| 3.6mg/dL
Dextran| 0.1mg/dL
Quindine hydrochloride mono hydrate| 6g/dL
Probenecid| 1.2mg/dL
Acetaminophen| 60mg/dL
Ascorbic acid| 20mg/dL
Glycerinum| 30mg/dL
Oxalic acid| 6.4mg/dL
Furosemide| 1mg/dL
Mouse IgG| 1000µg/mL
Within Laboratory Precision Study
Total 90 replicates were tested by three operators in five days using negative
specimen, SARS-CoV-2 IgM positive specimen, and SARS-CoV-2 IgG positive
specimen. The specimens were correctly identified >99% of the time.
BIBLIOGRAPHY
[1] P. Zhou et al. A pneumonia outbreak associated with a new coronavirus of
probable bat origin, Nature, 579, 270-273.
Manufacturing site:
ACON Biotech (Hangzhou) Co., Ltd. No.210 Zhenzhong Road, West Lake District,
Hangzhou, P.R.China, 310030
ACON Laboratories, Inc. 5850 Oberlin Drive, #340 San Diego, CA 92121, USA
www.aconlabs.c om
Phone: 1-858-838-9502
Number: 1151216406 Effective date: 2021-12-15
SARS-CoV-2 IgG/IgM Rapid Test
Control Solution Package Insert
REF L021-1011
INTENDED USE
The SARS-CoV-2 IgG/IgM Control Solutions are quality control reagents for use
with the ACON SARS-CoV-2 IgG/IgM Rapid Test. It is intended to ensure the test
performance and verify the user’s ability to properly perform the test and
interpret the results.
SUMMARY
The SARS-CoV-2 IgG/IgM Control Solution contains two vials – one positive
control solution vial and one negative control solution vial. They are
specifically formulated and manufactured to ensure performance of the test,
and are used to verify the user’s ability to properly perform the test and
interpret the results. The SARS-CoV-2 IgG/IgM Positive Control Solution is
recombinant Ab, spiked to human negative serum and will produce IgG and IgM
Positive test results. The SARS-CoV-2 IgG/IgM Negative Control Solution
contains human serum and will produce a Negative test result. Both positive
and negative controls tested negative for antibodies against HIV, HCV,
syphilis and HBsAg.
Use of Control Reagents manufactured by another source may not produce the
required results, and therefore, will not meet the requirements for an
adequate quality assurance for ACON SARS-CoV-2 IgG/IgM Rapid Test.
WARNINGS AND PRECAUTIONS
- For Emergency Use Authorization only
- For in vitro diagnostic use only.
- This product has not been FDA cleared or approved, but has been authorized for emergency use by FDA under an EUA for use by laboratories certified under CLIA, that meet requirements to perform moderate or high complexity tests.
- The emergency use of this product is only authorized for the duration of the declaration that circumstances exist justifying the authorization of emergency use of in vitro diagnostic tests for detection and/or diagnosis of COVID-19 under Section 564(b)(1) of the Federal Food Drug and Cosmetic Act, 21 U.S.C. § 360bbb3(b)(1), unless the declaration is terminated or authorization is revoked sooner.
- Do not use after the expiration date.
- Mix well and let the control solution reach room temperature (15-30°C) prior to testing.
- The SARS-CoV-2 IgG/IgM Control Solution contains human and/or potentially infectious components. No known test method can offer complete assurance of the biosafety of the derivatives. Therefore, all human-sourced materials should be considered potentially infectious. It is recommended that these reagents should be handled using established good laboratory working practices.
- Handle the Control Solution and materials as if they are capable of transmitting infectious agents.
- Do not eat, drink or smoke in the area where specimens and the Control Solution are handled. Avoid any contact with bare hands, eyes or mouth during collecting and testing specimen.
- Wear protective clothing such as laboratory coats, disposable gloves and eye protection while handling the Control Solution and specimens. Wash hands thoroughly when finished.
- Handle and dispose all specimens and materials used to perform the test as if they contained infectious agents. Observe established precautions against biological hazards throughout all the procedures and follow the standard procedures for proper disposal of specimens.
- Observe Good Laboratory Practices when handling potentially infectious material. Discard all contaminated materials, specimens and reagents of human origin after proper decontamination and by following local, state and federal regulations.
- This product is not intended for use as a standard.
- The SARS-CoV-2 IgG/IgM Control Solution is only intended to be used with the ACON SARS-CoV-2 IgG/IgM Rapid Test.
- Use of Control Solution manufactured by another source may not produce the required results, and therefore, will not meet the requirements for an adequate quality assurance for ACON SARS-CoV-2 IgG/IgM Rapid Test.
STORAGE AND STABILITY
- Store the control solution at 2-8°C.
- Do not use the control solution until it is warmed to room temperature.
- Use the control solution before the expiration date showed on the control vials.
- The control solution can be used for two months after the vials are opened and stored at 2-8°C.
- Recap and store the control vials in their original container at 2-8°C after use.
MATERIALS PROVIDED
- SARS-CoV-2 IgG/IgM Positive Control Solution (1 vial/0.1mL)
- SARS-CoV-2 IgG/IgM Negative Control Solution (1 vial/0.1mL)
- Package Insert
MATERIALS REQUIRED BUT NOT PROVIDED
- SARS-CoV-2 IgG/IgM Rapid Test/Cassette
- Timer
- Buffer
DIRECTIONS FOR USE
Allow all test materials (e.g. test strip/cassette, control solution, etc.) to
reach room temperature (15-30°C) prior to testing. Refer to the SARS-CoV-2
IgG/IgM Rapid Test Package Insert for detailed instructions. It is recommended
that both controls be processed under the following circumstances:
- A new operator uses the test kits for the first time.
- A new shipment of test kits is received.
- Device storage falls out of the 2-30°C range.
- To verify a higher or lower than expected frequency of positive or negative results.
- To investigate the cause of repeated invalid results.
- A new test environment is used.
- Open the control solution vial by unscrewing the cap. Mix well before testing. Hold the dropper provided in the test kit vertically and fill the capillary part of the dropper (not to exceed the capillary part) with the control (approximately 10 μL), then carefully dispense it into the Sample Well (S), immediately add 2~3 drops of buffer into the Buffer Well (B), and start the timer. Avoid air bubbles in the Sample and Buffer well.
- Wait for the colored line(s) to appear. The result should be read at 15 minutes. Do not interpret the result after 20 minutes.
INTERPRETATION OF RESULTS
NEGATIVE: Only colored control line appears in the control region (C).
POSITIVE:** Colored control line appears in the control region (C) and
colored test line(s) appears in the test line region (G) and (M).
NOTE: The color intensity of the IgM and IgG test line(s) may vary
depending on the concentration of the SARS-CoV-2 IgM antibodies and SARS-CoV-2
IgG antibodies present in the specimen.
INVALID:** Control line fails to appear. Insufficient specimen volume or
incorrect operation are the most likely reasons for control line failure.
Review the procedure and repeat the test with a new test cassette. If the
problem persists, do not use the test kit immediately and contact ACON
Laboratories Inc.
EXPECTED RESULTS
- The SARS-CoV-2 IgG/IgM Rapid Test Control Solution will produce examples of the color response to be expected for positive and negative specimens when tested with the ACON SARS-CoV-2 IgG/IgM Rapid Test.
- The SARS-CoV-2 IgG/IgM Positive Control Solution will produce both IgG and IgM Positive test results on ACON SARS-CoV-2 IgG/IgM Rapid Test
- The SARS-CoV-2 IgG/IgM Negative Control Solution will produce both IgG and IgM Negative test result on ACON SARS-CoV-2 IgG/IgM Rapid Test.
- The failure to obtain Negative results for both IgG and IgM with the Negative Control or Positive results for both IgG and IgM with the Positive Control indicates that the test is not performed properly or that the test reagents were not functioning properly.
- if an inappropriate result is obtained for either the Negative or Positive Control, retest the control sample using a new test unit. If the problem persists, please contact ACON Laboratories Inc.
LIMITATION
- The SARS-CoV-2 IgG/IgM Rapid Test Control Solution contains qualitative reagents, which are not to be used as quantitative calibrators. The controls should not be diluted and may be incompatible for use with other assays.
- The SARS-CoV-2 IgG/IgM Rapid Test Control Solution must be used at room temperature 15-30°C. Performance of the assay at other temperatures may yield invalid results.
Index of Symbols
ACON Laboratories, Inc. 5850 Oberlin Drive, #340 San Diego, CA 92121, USA
www.aconlabs.com
Phone: 1-858-838-9502
Number:
Effective day: 2020-12-15