YSI IoLine Micro 3 in 1 Electrochemical Electrode User Manual
- June 6, 2024
- YSI
Table of Contents
YSI IoLine Micro 3 in 1 Electrochemical Electrode
IoLine – 3 in 1
IoLine Micro – 3 in 1
For the most recent version of the manual, please visit www.ysi.com.
Contact
YSI 1725 Brannum Lane
Yellow Springs, OH 45387 USA Tel: +1
937-767-7241
800-765-4974
Email: environmental@ysi.com
Internet: www.ysi.com
Copyright © 2014 Xylem Inc.
Technical data
General data
Model| Reference electrolyte| Junction| NTC| Special
features
---|---|---|---|---
IoLine – 3 in 1| KCl 3 mol/l, Ag+ free| Platinum| Yes| Iodine/Iodide reference
systeme
IoLine Micro – 3 in 1| KCl 3 mol/l, Ag+ free| Platinum| Yes| Iodine/Iodide
reference systeme
Measurement and application characteristics
Model| pH measuring range| Allowed temperature| Membrane
resistance at 25 °C| Typical application
---|---|---|---|---
IoLine – 3 in 1| 0 … 14| -5 … 100 °C
23 … 212 °F
| < 200 MOhm| Measurement in tris buffers
IoLine Micro – 3 in 1| 0 … 14| -5 … 100 °C
23 … 212 °F
| < 400 MOhm| Measurement in tris buffers
Shaft dimensions, shaft material, electrical connection
Model | Shaft | Electrical connection |
---|---|---|
Length [mm] | Ø [mm] | Material |
connection** | Meter connection | Cable length |
IoLine – 3 in 1 | 120 | 12 |
IoLine Micro – 3 in 1 | 200 (70/130) | 12/6 |
Commissioning, measuring, calibration
Commissioning
Commissioning, measuring, calibration
The combination electrodes except the Micro are delivered with an irrigation
cap. The irrigation cap can also be used as storage equipment between
measurements. Due to its small pipe diameter, the micro model is stored in a
screw-top glass pipe. The electrodes are filled with reference electrolyte
solution in the factory. Prepare the combination electrode for measuring as
follows:
-
The closure of the refilling opening consists of a slide. Open the refilling opening for the reference electrolyte solution.
The refilling opening must always be open during calibration and measurement! -
The bridge electrolyte (3 mol/1 KCI) that fills the item during delivery can be replaced as needed with an electrolyte solution that matches the sample. For a stabile measurement, the conductivity of the individual bridge electrolyte should be at least 10 mS/cm.
-
The colored, iodine electrolyte solutions of the IoLine electrodes are inside a closed system and must therefore not be refilled. You must make sure that the bridge electrolyte is free of silver ions.
Note
The bridge electrolyte must be free of silver ions.
-
Connect the combination electrode to the meter
-
Calibrate the combination electrode according to the operating manual of the meter. When doing so, observe the general rules on the following page.
-
Make sure the refilling opening for the reference electrolyte solution is open.
-
Avoid the displacement of any solution (sample or buffer solution) from one measurement to the next by taking the following measures:
– Shortly rinse the calibration and sample beakers with the solution the beakers are to be filled with next.
– Between measurements, rinse the combination electrode with the solution that follows. Alternatively, you can also rinse the combination electrode with deionized water and then carefully dab it dry. -
Immerse the combination electrode in the solution in a vertical or slightly tilted position.
-
Make sure the immersion depth is correct. The junction must be completely submersed in the solution. At the same time, the level of the reference electrolyte must be at least 2 cm above the level of the solution.
-
Provide approximately the same stirring conditions for measuring as for calibrating.
The frequency of subsequent calibrations depends on the application. Many meters provide an option where you can enter a calibration interval. After the calibration interval has expired, the meter will automatically remind you of the due calibration.
Storage
With the refilling opening open, immerse the combination electrode in
in reference electrolyte (KCl 3 mol/L, Ag+ free). Observe the minimum depth of
immersion. Prior to the next measurement, shortly rinse the combination
electrode with the test sample or deionized water.
Note
Prevent contact of the pH membrane to the beaker bottom to avoid scratches on
the pH membrane.
Put the clean combination electrode in the watering cap that is filled with
reference electrolyte (KCl 3 mol/L, Ag+ free) and close the refilling opening.
Note – Overnight or longer
pH combination electrodes must not be stored dry or in deionized water. The
combination electrode could be permanently damaged by this. If the liquid in
the irrigation cap has dried up, condition the combination electrode in
reference electrolyte (KCl 3 mol/L, Ag+ + free) for at least 24 hours.
Note
During longer storing periods, salt sediments may develop on the watering cap.
They do not affect the measuring characteristics and can easily be re- moved
with deionized water when the combination electrode is put into operation
again.
Aging
Every pH combination electrode undergoes a natural aging process.
With aging, the responding behavior becomes slower and the electrode slope and
asymmetry change. Moreover, extreme operating conditions can considerably
shorten the lifetime of the combination electrode. These are:
- Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bromides, sulfides, iodides, proteins
- High temperatures
- High changes in pH and temperatures
- The warranty does not cover failure caused by measuring conditions and mechanical damage.
Maintenance and cleaning
Refilling the reference electrolyte
During operation, a small amount of reference electrolyte leaks through the
junction from the combination electrode into the test sample.
If the level of reference electrolyte becomes too low with time, refill it
through the refilling opening. Refilling is very easy with the enclosed
dropping bottle. Proceed as follows:
- Cut off the tip of the dropping bottle at a right angle until the opening in the tip can be seen
- Open the refilling opening of the combination electrode
- Press the tip of the dropping bottle into the refilling opening while turning it slightly
- Pump the reference electrolyte in the shaft using the dropping bottle
- Pull the dropping bottle out of the refilling opening while turning it slightly if necessary
Cleaning
Remove water-soluble contamination by rinsing with deionized water. Re- move other contamination as follows:
Contamination | Cleaning procedure |
---|---|
Fat and oil | Rinse with water containing household washing-up liquid |
Lime and hydroxide deposits | Rinse with citric acid (10 % by weight) |
Proteins | Immerse in pepsin cleaning solution PEP/pH for approx. 1 hour. |
Note: Make sure the level of the reference electrolyte is above that of the cleaning solution.
Note
Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy
the glass membrane.
After cleaning
Rinse the combination electrode with deionized water and condition it in
reference electrolyte solution (KCl 3 mol/l, Ag+ free) for at least 1 hour.
Then recalibrate the combination electrode.
Wear parts and accessories
Description | Model | Order no. |
---|---|---|
Reference electrolyte solution 250 ml (KCl 3 mol/l, Ag+ free) | KCl-250 | 109 |
705Y
Contact information
Ordering & Technical Support
Telephone : 800-897-4151 937-767-7241
Monday through Friday, 8:00 AM to 5:00 PM ET
Fax : 937-767-1058
Email :environmental@ysi.com
Mail : YSI Incorporated 1725 Brannum Lane Yellow Springs, OH 45387 USA
Internet : www.ysi.com
When placing an order please have the following information available:
- YSI account number (if available)
- Name and Phone Number Model number or brief description
- Billing and shipping address
- Quantity
- Purchase Order or Credit Card
Service Information
YSI has authorized service centers throughout the United States and Internationally. For the nearest service center information, please visit www.ysi.com and click ‘Support’ or contact YSI Technical Support directly at 800-897-4151.
When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at www.ysi.com and clicking on the ‘Support‘ tab.
What can Xylem do for you?
We’re 12,500 people unified in a common purpose: creating innovative solutions
to meet our world’s water needs. Developing new technologies that will improve
the way water is used, conserved, and re-used in the future is central to our
work. We move, treat, analyze, and return water to the environment, and we
help people use water efficiently, in their homes, buildings, factories and
farms. In more than 150 countries, we have strong, long-standing relationships
with customers who know us for our powerful combination of leading product
brands and applications expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com