YSI 9800 Portable Multi Photometer User Guide

May 15, 2024
YSI

YSI 9800 Portable Multi Photometer

Specifications

  • Product Name: YSI 9800 Photometer
  • Manufacturer: YSI
  • Model Number: 9800
  • Contact Information: 1725 Brannum Lane Yellow Springs, OH 45387
  • Phone: 800-765-4974, 937-767-7241
  • Email: info@ysi.com
  • Website: YSI.com/9800
  • Revision: XA00269 0523 Rev A

Product Usage Instructions

Getting Started

  • Onscreen instructions are available for all test methods.
  • Follow the prompts and select the desired test.

Achieving Accurate Results

  1. Prepare a blank in a separate tube and perform blanking just before the measurement step.
  2. Rinse equipment thoroughly with deionized water or the water being tested.
  3. Measure volumes accurately using a pipette, syringe, or as shown in the instructions.
  4. Use the correct YSI reagents and follow all safety information provided with them.
  5. Check for bubbles and remove them as demonstrated.
  6. Align tubes as shown, always using the tube cap to exclude light from the optics.
  7. Keep the instrument clean and dry all tubes before placing them in the optical cell.
  8. Verify instrument calibration regularly using available standards for the YSI 9800.

FAQ

  • Q: How do I verify the calibration of the YSI 9800 Photometer?
  • A: Regularly check and use the available standards for the YSI 9800 to confirm that the instrument is within calibration.
  • Q: What is the recommended practice for preparing a blank?
  • A: Prepare a blank in a separate tube and perform blanking just before the measurement step to achieve accurate results.
  • Q: How should I handle bubbles during testing?
  • A: Check for bubbles and remove them as demonstrated to ensure precise measurements.

Documentation support

Thank you for choosing the YSI 9800 Photometer.
Please refer to YSI.com/9800 for the following:

  • User Manual
  • Supporting documents for all test methods
  • Firmware updates

Getting started

  • Onscreen instructions are available for all test methods.

Achieving accurate results

Blanking

  • Blanking is critical and sets the zero level for each test.
  • For most test methods, the blank is a sample without any reagents added.
  • The onscreen instructions show blanking as the first few steps, however, the “blank” can be prepared in a separate tube and blanking carried out just before the measurement step.

Best practice when testing

  • Rinse equipment thoroughly with deionised water or the water being tested.
  • Measure volumes accurately as shown (or use a pipette or syringe).
  • Use the correct YSI reagents and observe all safety information supplied with them.
  • Check for bubbles and remove them as shown.
  • Align tubes as shown. Always use the tube cap as it excludes light from the optics.
  • Keep the instrument clean. Dry all tubes before placing them in the optical cell.

Verify instrument calibration
Check Standards are available for your YSI 9800. Use these regularly to confirm that the instrument is within calibration.