SGT Disposable Personal Monitor
User Manual
Product Overview
SGT is a portable single gas detector designed to detect the presence of oxygen, toxic and combustible gases in the ambient environment. When activated, SGT continuously monitors ambient air for the presence of a specific gas and alerts the user to potentially unsafe exposure with LED, vibrating, and audible alarms in the event that gas concentration exceeds alarm set points. And the alarm set point, calibration range, and display configuration can be changed via SENKO-IR Link (Optional).
WARNING
Any unauthorized attempt to repair or modify the product, or any other cause
of damage beyond the range of the intended use, including damage by fire,
lightning, or another hazard, voids the liability of the manufacturer.
Activate this product only if a sensor, visual, detection, and audible cover
are clear from contaminants such as dirt and debris that could block the area
where gas is to be detected.
Do not clean and rub the LCD screen of the products with a dry cloth or hands
in a hazardous environment to prevent static electricity.
Perform cleaning and maintenance of the products in the fresh air that is free
of hazardous gases
Test the response of a sensor regularly by the gas concentration exceeding the
alarm set point.
Test LED, audio and vibration manually.
Gas concentration measurements by the sensor can vary based on the environment
(temperature, pressure, and humidity). Therefore, calibration of SGT should be
performed in the same (or similar) environment of the device’s actual use.
If the temperature changes sharply during the use of the device (e.g., indoors
vs outdoors), the value of the measured gas concentration can suddenly change.
Please use the SGT after the gas concentration value has stabilized.
Severe vibration or shock to the device may cause a sudden reading change.
Please use SGT after the value of gas concentration has stabilized. Excessive
shock to SGT can cause the device and/or sensor to malfunction.
All alarm value is set based on the alarm standard that is required by
international standard.
Therefore, alarm values should be changed only under the responsibility and
approval of the administration of the worksite where the instrument is used.
Use IR communications in the safety zone which is free of hazardous gases.
Do not attempt to replace the battery and sensor as SGT is designed to be
disposable. Changing the battery and sensor may impair intrinsic safety and
the attempt will void the warranty.
CAUTION
Before operating this device, please read the manual carefully.
This device is not a measurement device, but a gas detector.
If calibration and self-test fail continuously, please do not use the device.
For the O2 detector, perform calibration every 30 days in the fresh air
environment.
Before use, please check the activation date, and if the activation date is
passed, please do not use the device.
Clean detectors with a soft cloth and do not use chemical substances for
cleaning.
To maintain 24 months lifetime, avoid the below activities except the
necessary cases to check events(Max/Min), lifetime/concentration, and alarm
setpoints. Otherwise, the frequent use of the button will deplete the battery
life of fewer than 24 months.
- Push the button frequently without valid reasons.
- Frequent alarm operations or alarms remain for a long time. *Normal Alarm Use: 1 time and 2 minutes per day.
- Connect with SGT IR Link frequently except for the bump testing.
View a serial number on the label at the backside of the device. (ex, 20170101)
- The serial number indicates below.
Product Overview
LCD Display Symbols
****
| Alarm condition|
| Remaining Month(Month)
---|---|---|---
****
| Low Alarm Display|
| Remaining Day
****
| High Alarm Display|
| Remaining Time(Hour)
****
| Stabilization Success|
| Max Peak Value
****
| Stabilization Failure|
| Min Peak Value
****
| Fresh Air Calibration|
| Measurement Unit
| Standard Gas Calibration|
| Lifetime less than 30 days
Or Low Battery
Activation
< CAUTION >
Before use, check the manufacturing date on the box. Please do not open if it
is over the shelf life below.
Shelf Life
SGT-Oxygen: 3 months from the manufacturing date
SGT-Toxic: 6 months from the manufacturing date
In a safe environment, when pressing the button for 3 seconds, gas type and firmware version (ex. v2.2) will be displayed. For 10 seconds countdown, the device will be stabilized. After stabilization is completed, “V” icon will appear on the display screen and the device will move to Measure mode.
In the event that stabilization of the device fails, ****will appear on the display, and Measuring mode will not be entered. In this case, contact the authorized reseller or SENKO service center at 82-31-492-0445 for repair/return information.
< CAUTION >
Appropriate calibration of the device is required prior to operation. Always
ensure that the device makes the proper detection response to the pertinent
gas. Verify that debris that could interfere with the detection of gas is not
blocking the area where gas is to be detected.
Mode
Measuring mode
- When activated, in Measuring mode, gas concentration or remaining life (Option) appears on the screen.
- Oxygen concentration is displayed in percent by volume (%Vol) and toxic concentration is displayed in parts per million. (PPM)
Display mode
In Measuring Mode, by pressing the button for one second, the following ICONs
will appear in order.
Min (only for oxygen) -> Max -> CLR -> Remaining Lifetime( , , ) -> 1st alarm
set point ->
2nd alarm set point -> Firmware version -> Calibration Concentration
At the last step, if you press a button or do not push any button for a
second, the device will return to Measuring Mode.
Alarms / Battery / Test Failure Display
When the gas concentration exceeds alarm setpoints or will be
displayed and the device will vibrate, flash (LED), and beep. To remove
alarms, move to a clean air location. and then a gas concentration will
decrease, and the alarm will stop.
Failure of Test and Calibration: Display icon ****with beeping
Alarm| Alarm Standard| LCD Display| Alarm and Vibration
Display
---|---|---|---
1st Alarm| Exceeds 1st alarm| Icon & concentration|
2nd Alarm| Exceeds 2nd alarm| Icon &concentration|
Remaining life| Below 30 days| Icon|
Lifetime Expiration| Past 24 months
(End Of Life)| | The lifetime is over.
(Replace with a new SGT.)
Test Failure| Failure of sensor
test / Failure of
calibration| ****Icon displays|
Battery Test| Low Power| |
Bump Test| Bump Test Period| | Press the button to activate the bump test
Calibration| Calibration Period| | Press the button to activate the
calibration
Alarm SetPoint
Gas | O2 | CO | H2S | H2 | SO2 | Cl2 | NH3 | NO2 |
---|---|---|---|---|---|---|---|---|
1st | 19% | 30ppm | 10ppm | 100ppm | 2ppm | 0.5ppm | 25ppm | 3ppm |
2nd | 23% | 60ppm | 15ppm | 500ppm | 5ppm | 1ppm | 35ppm | 5ppm |
< CAUTION >
All alarm values are set based on the alarm standard that is required by
international standards. Therefore, alarm values must be changed only under
the responsibility and approval of the administration of the worksite where
the instrument is used.
Event Log
The last 30 events are stored on a device. Once more than 30 events are
stored, the log events are removed automatically in the order starting at
Event 1. And, the stored log events data can be transferred to SENKO-IR LINK.
Each alarm event records followings:
– Types of alarms (1st or 2nd) / Alarm concentration in ppm or % / Peak
concentration
Calibration
< CAUTION >
Initial calibration is performed on all devices prior to shipment. Once
received, calibration should be performed monthly (or quarterly) depending on
the frequency of use.
Fresh Air Calibration
When pressing and holding the button for 5 seconds in the calibration mode( ),
an icon and ‘CAL’ mark will appear on the LCD.
And, press the button to initiate calibration for three seconds. When
calibration begins, a countdown (starting at 10) will appear on the screen.
Once completed, the icon will appear on the LCD.
If calibration fails, the ****icon will appear on the display. If this continues, please contact the sales representative or SENKO service center.
< CAUTION >
Calibration should be performed in a fresh-air environment that is free for
any influence of other gases (since calibration is assumed to be performed in
an environment with 20.9% of Oxygen). It is also recommended that calibration
should be performed in a space that is not confined.
Standard Gas Calibration
When pressing and holding the button for 5 seconds in the calibration mode ,
an icon and ‘CAL’ mark will appear, and when pressing the button once again,
the icon will appear. And then, press and hold the button for three seconds to
initiate calibration. When calibration begins, a countdown (60 seconds or more
depending on sensor types) will appear on the screen.
Once completed, the icon will appear several seconds on the display. Then, the device will return to Measuring mode.
Once calibration fails, ****the icon will appear on the display. If this continues, please contact the sales representatives or SENKO service center.
Return to Measuring Mode.
In the standard calibration mode , by pressing the button for a second, the
fresh air calibration, standard calibration, and ESC will appear on the
display consecutively. In the ESC mode, press the button for 3 seconds, the
device will be getting out of the calibration mode. And press the button one
time, it will return Measuring mode.
Calibration concentration.
Gas | O2 | CO | H2S | H2 | SO2 | Cl2 | NH3 | NO2 |
---|---|---|---|---|---|---|---|---|
Concentration | 1 | 100 ppm | 50 ppm | 500 ppm | 10 ppm | 10 ppm | 100 ppm | 10:00 PM |
Self Test & Bump Test
Self Test
The default of Self-Test is N/A, and the interval is 8hr~20hr, or N/A. To initiate the self-test, please set the self-test interval via the IR link. After the setting interval is activated, STS message will flash. (The message will flash until users perform the Self-test.) Once you press the button, it will test buzzer, LED, Vibration, LCD, 1ST alarm, and 2ND alarm. After the test is completed, END message with icon will be displayed. (Users are required to check the test processes manually.)
Bump Test
The interval of the Bump-test is 1~365days, and the default is N/A. To initiate the bump test, set the bump test interval. Once the bump test time reaches, its message will flash. Once you press and hold the button for 3 seconds, the test message will be displayed for 45 seconds (To cancel, press the button for one second). Within 45 seconds, apply a test gas (If no gases are applied, the best message will flash again). After the selected gas is applied, if the test is successful, the SUC message will be displayed after 30 seconds. And then, remove the calibration cap and gas tube. If the test fails, an FA message ****will be displayed and its message will be flashing until the test is successful.
Specification
Model | SGT |
---|---|
Measure Gas | 2 |
Range | 0.30%vo, |
0-20Xen
Sensor Type| Electrochemical
Measurement| Diffusion type
Display| LCD display
Audible| 90dB at 10cm
Warning
Lamp| Red Flashing LEDs (Light-Emitting Diode)
Vibration| Vibration Alarm
Battery| Manufacture Vitzrocell / P/N SB-AA02(P) / System: Lithium Primary
Battery Nominal voltage 3.6V / Nominal capacity 1 2Ah
-40°C – +50°C(for Toxic) / -35°C – +50°C(for 02)
5% – 95% RH (non-condensing)
Temperature
& Humidity
Case| Rubber Enclosure
Accessories| Calibration Cap. Manual. Test Report
Option| External Sampling Pump (SP-Pump101). SENKO-IR Link. Docking Station
Size 8 Weight| Size 54mm(W) x 91mm(H) x 32mm(D)./
Weight 93g(Toxic). 1049(02) (Battery. clip included)
Operating Life| 24 months. based on 2 minutes of alarm per day
Event Log| Recent 30 alarms
Approval| Ex o IIC 74 Gal IP67 (Korea Ex. IECEx. ATEX. CSA & UL. INMETRO
Certificates
Intrinsic Safety:
The detector is in conformity of the following standards
IECEx:
IECEx KTL 15.0018
- Explosion Protected
- Protection Concept
- Gas Group
- Temperature Classification
- Equipment Protection level
Ex ia IIC T4 Ga
Class I, Zone 0, AEx ia IIC T4 Ga
Class I, Division 1, Groups A, B, C, D, T4
C22.2 No. 60079-0:2015; C22.2 No. 60079-11:2014;
C22.2 No. 61010-1-12:2010; UL 61010-1,
Ed. 3; UL 913, Ed. 8; UL 60079-0, Ed. 6; UL 60079-11, Ed. 6
ATEX: 2198 II 1 G Ex ia IIC T4 Ga IP67
KRH16ATEX1048
Directive 2014/34/EU
KCS: Ex ia IIC T4
KTL 16-KA2BO-0457
INMETRO Ex ia IIC T4 Ga
BVC16.5919
Compliance: Electromagnetic Compatibility Directive 2014/30/EU
Standards:
The electrical apparatus and any acceptable variations to it specified in the
schedule of this certificate and the identified documents was found to comply
with the following standards:
- IEC 60079-0: 2011 Ed. 6
- IEC 60079-11: 2011 Ed 6
- UL 61010-1, Ed. 3
- UL 913, Ed. 8
- UL 60079-0, Ed. 6
- UL 60079-11, Ed. 6
- C22.2 No. 60079-0:2015
- C22.2 No. 60079-11:2014
- C22.2 No. 61010-1-12:2012
- EN 60079-0: 2012+A11:2013
- EN 60079-11: 2012
Manufacturing Approval: The detector manufacturer is certified compliant with ISO 9001:2000 provisions
Limited Warranty
SENKO warrants this product to be free of defects in workmanship and materials
under normal use and service for two years from the date of purchase from the
manufacturer or from the product’s authorized reseller.
The manufacturer is not liable (under this warranty) if its testing and
examination disclose that the alleged defect in the product does not exist or
was caused by the purchaser’s (or any third party’s) misuse, neglect, or
improper installation, testing, or calibrations. Any unauthorized attempt to
repair or modify the product or any other cause of damage beyond the range of
the intended use, including damage by fire, lightning, water damage, or other
hazard, voids liability of the
manufacturer.
In the event that a product should fail to perform up to manufacturer
specifications during the applicable warranty period, please contact the
product’s authorized reseller or SENKO service center at 82-31-492-0445 to
repair/return information.
SENKO Canada Ltd.
Address: Calgary, Alberta, Canada
Tel: 587-322-1616
Email: ben@senkocanada.com
Web: www.senkocanada.com
73, Oesammi-ro 15 beon-gil, Osan-si, Gyeonggi-do, 18111, South Korea
Tel : 82-31-492-0445
Email : sales@senko.co.kr
Fax : 82-31-492-0446
Web : www.senko.co.kr