SENKO SGT Docking Station Version II User Manual
- June 6, 2024
- SENKO
Table of Contents
SENKO SGT Docking Station Version II
WARNING
- Any unauthorized attempt to repair or modify the product, or any other cause of damage beyond the range of the intended use including damage by fire, lightning, or other hazard, voids the liability of the manufacturer.
- Do not use if the device appears to be damaged.
- Use the device in a clean environment, which has no hazardous gases.
- Do not use the expired gas cylinders.
- Ensure to use the certified gas cylinders.
- Ensure that the gas types and the concentrations match with manuals and devices.
- Severe vibration or shock to the device may cause a sudden reading change and cause the device to malfunction.
- Do not leave or use the device at a watery place or at where there is any liquid.
- Do not use a device that has been failed the test.
OVERVIEW
Read the operation manual before use
- Before the operation, read the manual carefully and follow all the instructions.
- IR link communication works only when it is placed at bay 1 on the docking station.
- The device only operates when the power LED turns green.
- Before upgrading firmware, connect the plug to avoid any data loss and malfunction by unexpected power off. And the file names are fixed as below.
Docking F/W File Name : DK_FW.bin,
SGT F/W Name : SGT_FW.bin
SGT_CO.bin, SGT_H2S.bin, SGT_O2.bin
Description
A Bump test and calibration are the key features to ensure the safety of
devices and users. The SGT docking station version II provides bump testing,
event management, and calibration from a simple-to-use multi-unit station and
gives fleet or safety managers the ability to update configurations on large
fleets. It also maintains all the important event logs that demonstrate users
are working within company requirements.
Instructions and Parts
- Bay LED
- Bay 1
- Handle
- Thumbscrews
- Power LED
- Bump Test button g Calibration button h Bay 2, 3, 4
- Pressure gauge
- Cylinder 58L/116L
- Cylinder Straps
- Gas regulator
- Cylinder compartmen
- Exhaust port
- Fresh air Inlet
- LAN port
- Charging port
- USB port
Key Features
- Zero and Span Calibration, Bump Test Function
- The rechargeable internal battery lasts up to 1000 bump tests on a single charge
- Indicate the current status of the operation by the Power LED and Unit LED color
- Configure settings and automatic update firmware of up to four SGT devices
- Gas Ventilation system to remove a remained gas
- Computer not required for normal operation
- Save the log events after calibration and bump test
- Enable users to adjust the setting points of the Docking station and SGT via IR Link or USB
Specifications
Size | 52.5 x 43.7 x 21.3cm (20.66” x 17.20” x 8.40”) |
---|---|
Weight (Without cylinder) | 11.0kg |
Temperature | 41 to +104˚F (5 to +40˚C) |
Warranty | Full 2 years |
Battery Type | Rechargeable Lithium-Ion |
Battery Life | Up to 1,000 Bump Tests |
LEDs | 6 Blue LEDs for each bay, 3-color LED for power, Switch Back Light |
LED (Orange, Green)
Memory| USB 8GB Standard
Log Capacity| Approximately Up-to 5 million tests (8GB, removable USB
memory)
Tests Performed| Bump Test and Calibration
Information Stored| Bump/Calibration Logs, Individual Monitor Event
Logs, Firmware, and Unit Configurations
Gases Available| CO, H2S, O2
Compatibility| Works with SGT & SGT-P (O2, CO, H2S)
Calibration Gas| Compatible with 58L & 116L cylinders with C10 fitting
Interface| Ethernet RJ-45(TCP/IP), 10Mb/s
Charging Adaptor| DC12.0V, 3A/h
Set up and Operation
Power turns on/off
The SGT Docking Station version II operates using only 2 (two) buttons for
calibration and Bump Test.
To turn on the power, follow the below:
- Press and hold either the Calibration or Bump Test button for 2 seconds.
- The Power LED turns on to orange and blinks during the warming up and after a while turns to green when it is ready. When the unit has a low battery, the LED turns to red.
- To turn off the power, follow the below:
- Press and hold both Calibration and Bump Test buttons simultaneously for 2 seconds.
- Note: If the unit is inactive for 300 seconds, it will automatically off (save power mode) The button/unit only works when the power LED turns to green.
Bay LEDs and meanings
Each Bay has six blue LEDs (see Figure 2) which indicates the process status.
Each of them will turn on by order from left to right while testing. After
testing is finished, make sure all six bays are turned on otherwise, 3rd bay
LED will be shown if the test is failed.
2.3 Charging battery
The unit`s Power LED turns red when the battery is very low as shown above.
However, it comes with a DC adapter with 5plugs which can be used to charge
the unit using a 110V power supply. A full charge will take approximately 3
hours and will last for approximately 1000 bump tests.
Note: Refer to Figure 1 for charging location.
Exhaust Port & Fresh Air Inlet
There are two small ports on the left side of the charging port (see Figure
1). The Fresh air inlet port will draw clean air in during an active test. The
incoming air automatically counts as fresh air (O2, 20.9%vol). The Exhaust
Port is the inverse of the Fresh air inlet port. During an active test, the
used test gas will exit the unit through the exhaust port. Optionally, tubing
can be attached to the exhaust port using the included twist connector to
direct the used test gas to a suitable location.
Installation of Gas Cylinder
The gas Cylinder which fits Docking Station II must have the C10 fitting on
it. The size of 58L and 116L are both suitable in-cylinder compartments.
Before installing the cylinder for testing SGT, ensure that the Calibration
concentration of gases is the same as SGT.
To install a cylinder, follow the below:
- Remove thumbscrews and lift the handle up.
- Install the cylinder to the gas regulator.
- Fasten the cylinder straps and close the cylinder compartment.
- Check the pressure gauge on the front panel.
- Re-install thumbscrews.
- After installing a gas cylinder, ensure the USB is attached.
Testing
Bump test and calibration are the key features to ensure the safety of devices
and users. The SGT Docking Station version II provides bump testing, event
management and calibration from a simple to use the multi-unit station.
Calibration
Calibration is the comparison of measurement values delivered by a device
under test with those of a calibration standard of known accuracy. It is very
important to calibrate a device periodically for better performance of the
device and sensor accuracy, especially, when it is safety equipment.
To perform a calibration, follow the below:
- Insert the monitor(s) into the bay facing down.
- Slightly press down on the monitor until it locks into place.
- Press the calibration button for 2 seconds to execute the Calibration test.
- During the 90seconds, bay LEDs begin to light as the test progress.
- When the test is finished, all (six) blue Bay LEDs illumination indicates if the SGT has passed the test (see Figure 3).
- Otherwise, the third blue Bay LED illumination indicates if the SGT has failed the test (see Figure 3).
Note: Do not use the monitors that failed the calibration test.
Bump Test
A bump test is the process that checks and verifies the performance of the
monitors and ensures that sensors are correctly responding to the target gas.
To perform a Bump Test, follow the below:
- Insert the monitor(s) into the bay facing down.
- Slightly press down on the monitor until it locks into place.
- Press the Bump Test button for 2 seconds to execute the Bump test.
- During the 20~30seconds, bay LEDs begin to light as the test progress.
- When the test is finished, all (six) blue Bay LEDs illumination indicates if the SGT has passed the test (see Figure 3).
- Otherwise, the third blue Bay LED illumination indicates if the SGT has failed the test (see Figure 3).
Trouble Shooting
Bay Light does not turn on /test not performed
- Clean Monitor IR window.
- IR interference (bright lights/sunlight).
- Monitor in gas alarm state.
All tests fail
- Gas cylinder empty – check pressure gauge if it indicates at least 100psi/10bar.
- Gas Expired – check the gas expiration date located on the cylinder.
- Wrong gas concentration(s) – gas cylinder and SGT concentrations are not the same.
- Wrong gas type – check the gas type.
Test results are not stored
- USB not attached or full storage.
- Uploaded Settings are not saved
- The “Write” button has not clicked.
- The connection has been lost.
SGT Docking Station II Manager Software
The purpose of SGT Docking Station II Manager Software is to customize and
configure the Stations settings and SGT devices
internal configurations. The
software can adjust and make changes to the Docking Station and the SGT units
using SENKO IR Link and/or partly USB Disk.
Software Installation
-
After downloading the file, navigate to the saved location.
-
Double click for running an installer.
-
Click “Read More” and “Run Away” if you receive a warning message.
-
The Docking Manager Setup window will open. Click “Next”.
-
The “Select Installation Folder” comes up.
-
To install to a different folder, click “Browse” or enter the link.
-
To install in the default folder, click “Next”
-
The Installation will be ready to begin.
-
Click “Install”.
Click “Yes” if you get a warning message.
- The Docking Station Setup Wizard is complete.
- Click the “Finish” button to exit the Setup Wizard.
Connecting to the SGT Docking Station II
When the Docking Manager software opened, the fields are grayed out and before it can be used, the IR Link must be connected to the Docking Station II. There are 2 ways of connection to the Docking Station II: IR Link and USB Disk.
Note: The IR Link connection works only when it is placed into Bay 1. The USB drive must be installed to store the test results. Do not open the software if IR Link is not plugged into the PC.
Connecting with SENKO IR Link
SENKO IR Link is a device that conveniently configures and communicates with
SGT Docking Station II using the IR communication in Bay Unit 1.
To connect the IR Link, follow the below:
- Plug the IR link into a USB port on a PC.
- Power on (green LED) the Docking Station II.
- Place the IR Link facing up into the Bay Unit 1 (see Figure 4).
- Double click and open the installed software.
- Click the “OPEN” button at the top right corner of the software interface.
- The IR Link starts to flash rapidly.
- After a few seconds, the “Read complete” message will be displayed, and the current settings of Docking Station II will be shown on the interface.
Note: If the “Send timeOut” message displays, the IR Link has failed to communicate with Docking Station II. In this case, make sure the IR Link screen is aimed at the IR screen of Bay Unit 1 and check Power of Docking Station II is on.
Connecting with USB
To connect the IR Link, follow the below:
- Open the Cylinder compartment.
- Find the USB port (see Figure 1) and remove the USB Disk.
- Plug the USB into to PC.
- Double click and open the installed software.
- Select the “USB Disk” in the upper middle-left corner.
- Select the location of the USB Disk and click the “OK” button.
Using Software
Overview of Software
When IR Link is connected to the Docking Station II, Figure 5 will be displayed. But if it is connected using a USB Disk, the “Logs” tab will not indicate.
Docking Station Tab
- Serial Number – Serial number of Docking Station II (cannot be changed).
- Version – The current firmware version.
- Auto Power-Off – The time when the unit automatically goes to Save Power Mode. Min/max time can be set 1 ~ 3,600 seconds.
- Maximum Bump Time – The maximum and minimum time spent for the Bump Test process. If the unit cannot respond correctly before time expires, the process will stop and the result will fail. Can e adjust the min/max of 20 ~ 120 seconds?
- Location – The location where the test has been done. It is beneficial when downloading the data logs. Can add 12alphanumeric characters.
- Gas Expiration Date – Informs the date of expiration of gas in the installed cylinder. Must be updated after installing a new one.
- Gas Bottle Lot # – Displays lot number of installed cylinders. Must be updated after installing a new cylinder.
- Info Read/Write – By pressing the “Read” it displays the current setting of the Docking Station II and will discard any unsaved changes in the Tab.
- Calibrate – If the box is not selected, the calibration process will not proceed by pressing the Calibration button. However, it will still update the firmware and configuration of SGT.
- Bump Test – If the box is not selected, the Bump test process will not proceed by pressing the Bump Test button. However, it will still update the firmware and configuration of SGT.
- Bump Result Decision – The dropdown list is selectable for “Beep”, “Concentration”, or “Beep + Concentration”. The “Beep” option will run the bump test until the alarm condition is met and the audible alarm is detected. The “Concentration” option will run the bump test until the test gas concentration is verified. If the units are in Stealth Mode, the Concentration option will be required for a successful bump test. The “Beep + Concentration” option will run the bump test until the audible alarm is detected and the test gas concentration is verified.
- Browse & Write – To upgrade the latest firmware of Docking Station II, click “Browse”, navigate to the firmware location and click “Open”. Click “Write” to begin the upgrading process.
Note:
Without clicking the “Write” button, configured and customized settings
will not be applied and neither be saved.
SGT Tab
- User ID – The User ID can be used to add a usage message.
- Calibration Interval (Days) – The calibration reminder informs every fixed day (can adjust 0 (n/a) ~ 365).
- Bump Interval (Days) – The Bump test reminder informs every fixed day (can adjust 0 (n/a) ~ 365).
- Self-Test Interval (Days) – The Self-Test reminder informs every fixed hour (can adjust 0 (n/a) ~ 20).
- Low Alarm & High Alarm – The 1st and 2nd alarm setpoints for each gas.
- Display mode– The unit can display the remaining lifetime of the unit or measure gas concentration.
- Gas Concentration – This allows a user to enter/amend the correct concentration of the installed gas cylinder on Docking Station II. It is necessary to adjust for all tests.
- SGT Option Info Read/Write – By pressing the “Read” it displays the current setting of the SGT and will discard any unsaved changes in the Tab. By clicking the “Write”, all configured and customized settings will be saved.
- Browse & Write – To upgrade the latest firmware of SGT, click “Browse”, navigate to the firmware location and click “Open”. Click “Write” to begin the upgrading process.
Logs Tab
All logs will be stored on the USB Disk (logs are not available if USB is not
plugged) and can be downloaded via SENKO IR Link using Docking Manager
software. The log files will be created for each device by their Serial
numbers and will be in “.csv” format.
Note:
The USB drive must be installed to store the test results (logs). By
connecting USB Disk, the “Logs” tab is not available on the software
interface. But logs are still available to download from the “Events” file in
a USB Disk. Click the “Refresh” button to view and download the latest logs.
After clicking the “Refresh” button, all logs will be available to view (see Figure 8).
To download the logs, follow the below:
- Plug the IR Link into a USB port on the PC and place it into Bay Unit 1.
- Double click and open the installed software.
- Click the “OPEN” button at the top right corner of the software interface.
- Click the “Logs” Tab and Click the “Refresh” button.
- Select needed log by indicating SGT`s serial number.
- Click the “Local Save” button and a Browse window will open.
- Select a location to save a file.
- Click the “OK” button to begin the downloading process.
- When downloading is finished, the “Log data save success” message will come up.
Certification Tab
A calibration certificate can be issued based on the latest calibration data.
The latest calibration data will be stored on the USB Disk and can be printed
using Docking Manager software. The certi file name will be created for each
device by their Serial numbers and will be in “.csv” format.
Note:
A USB drive must be installed to store the test results. By connecting USB
Disk, the “Certification” Tab is not available on the software interface.
Click the “Refresh” button to view and print the calibration certificate.(only
can print latest calibration data)
To print the latest calibration certificate, follow the below:
- Plug the IR Link into a USB port on the PC and place it into Bay Unit 1.
- Double click and open the installed software.
- Click the “OPEN” button at the top right corner of the software interface.
- Click the “Certification” tab and Click the “Refresh” button.
- Select the needed device by indicating the device serial number in the Certi File Name section.
- Click the “Load” button and information will be shown automatically on the right side.
- Select address section between Senko, Senko Canada, Senko EU, No address.
- Write the authorized person’s name in the “Calibrated by” & “Approved by”
- Click the “Print” button to print the calibration certificate.
- When printing is finished, a “success” message will come up.
Limited Warranty
SENKO warrants this product to be free of defects in workmanship and materials
under normal use and service for two years from the date of purchase from the
manufacturer or from the product’s authorized reseller. The manufacturer is
not liable (under this warranty) if its testing and examination disclose that
the alleged defect in the product does not exist or was caused by the
purchaser’s (or any third party’s) misuse, neglect, or improper installation,
testing, or calibrations. Any unauthorized attempt to repair or modify the
product, or any other cause of damage beyond the range of the intended use,
including damage by fire, lightning, water damage, or another hazard, voids
the liability of the manufacturer.
In the event that a product should fail to perform up to manufacturer
specifications during the applicable warranty period, please contact the
product’s authorized reseller or SENKO service center at 82-31-492-0445 to
repair/return information.
SENKO Canada Ltd.
Address: Calgary, Alberta, Canada Tel:
587-322-1616
Email : ben@senkocanada.com
Web: www.senkocanada.com
References
-
[ Senko Canada Products - Senko Canada](http://www.senkocanada.com "Senko Canada Products
- Senko Canada")
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>