SENKO T2J 0M3 Docking Station II for SGT User Guide

June 6, 2024
SENKO

SENKO T2J 0M3 Docking Station II for SGT

SENKO T2J 0M3 Docking Station II for SGT

Power On & Off

  • To turn on, press and hold the Calibration or Bump Test button for 2 seconds.
  • Automatic power save mode after 5 minutes of inactivity.
  • To turn off, press and hold the Calibration and Bump Test button for 2 seconds.SENKO T2J 0M3 Docking Station II for SGT fig 2

Bump Test

  • Insert the monitor(s) into the bay facing down.
  • Press the bump test button for 2 seconds to execute. Bump Test
  • Up to 1,000 bump tests on a single charge

Calibration

  • Insert the monitor(s) into the bay facing down.
  • Press the calibration button for 2 seconds to execute Calibration test.

Test Results

  • All (six) blue Bay LEDs illumination indicates if the SGT has passed the test.
  • The third blue Bay LED illumination indicates if the SGT has failed the test.

Set up Calibration Gas Cylinder

  • Remove thumbscrews and lift the handle
  • Install the cylinder to the gas regulator
  • Fasten the cylinder straps and close the cylinder compartment
  • Check the pressure gauge on the front panel
  • Re-install thumbscrews

Trouble Shooting

  • Bay Light does not turn on /test not performed
    • Clean Monitor IR window
    • IR interference (bright lights/sunlight)
    • Monitor in gas alarm state
  • All tests fail
    • Gas cylinder empty
    • Gas Expired
    • Wrong gas concentration(s)
  • Test results are not stored
    • USB not attached or full storage
  • If you’re unable to resolve technical issues, please contact the product’s authorized reseller or SENKO service center.

Important Notes

  • USB drive must be installed to store the test results
  • The latest SGT firmware will be updated automatically
  • For bump testing, ensure the cylinder gas concertation is above the “Low Alarm” setting of the SGT
  • For calibration, ensure the SGT “Cal Concentration” setting is the same as the cylinder gas concentration
  • Any monitor that fails the bump test or calibration must be re-calibrated and perform the bump test and calibration again. If the attempt fails continuously, please do not use the detector and contact the product’s reseller or SENKO service center.
  • When connecting SENKO IR Link, make sure “Write” button has clicked to save customized settings.

For more information

SENKO Canada
548 Lake Bonavista Drive, Calgary,
Alberta T2J 0M3, Canada
Email:[email protected], [email protected]
Web:www.senkocanada.com
Tel:587-322-1616

Read User Manual Online (PDF format)