CEMARLINE MAXX-606 Patient Positioning Laser
WARNING
This manual contains the latest information at the time of publication, Cemar
Electro Inc. reserves the right to revise this manual without notice.
The MAXX-606/607 Patient Positioning Laser is intended for use only by physicians qualified in diagnostic radiology and radiation therapy and experienced in marking patients utilizing patient positioning lasers or by therapists at the specific direction of such a qualified physician. It is the sole responsibility of the physician to judge whether the use of the MAXX-606/607 Patient Positioning Laser is clinically appropriate.
INTRODUCTION
This Installation and User’s Guide includes all the information you need to install, operate, and maintain your MAXX-606/607 Patient Positioning Laser safely and efficiently.
The MAXX-606/607 is a manual patient positioning system. The system is based on the original CL-606 series but takes advantage of a streamlined and economic design along with the latest laser technology to offer a simple laser with minimal maintenance. Easy to install, the MAXX-606/607 embodies the essence of simplicity.
With a small footprint, the MAXX-606/607 is a laser diode modular design that allows for low-cost maintenance and replacement components. The unit features 635nm red and 520nm green lasers. Simple controls permit easy use. The MAXX-606/607 includes a two-year warranty from Cemar Electro Inc.
The purpose of this manual is to provide instructions on the installation and setup of the patient positioning lasers within the treatment room.
Note: This Installation and Set-up Manual pertains to the MAXX-606/607 laser, which is similar, but not identical in configuration to the CL-606 model or the MAXX-700 series. Please ensure to follow these instructions for the MAXX-606/607 Patient Positioning Laser only.
INTENDED USAGE
The intended use for the MAXX-606/607 patient positioning laser is to align patients to medical diagnostic imaging and radiation therapy equipment utilizing ionizing radiation. MAXX-606/607 lasers are intended to be used in a medical clinic or hospital settings by trained medical staff. The lasers are not designed to be used around flammable anesthetics, which present a risk of explosion or fire. The lasers are not intended to be used in a sterile or corrosive environment. No other uses are intended or implied.
Note: Use of the MAXX-606/607 laser system other than its intended use is carried out entirely at the user’s risk.
LASER SAFETY CONSIDERATIONS
To help ensure safe operation of your MAXX-606/607 system, please read this section carefully and follow the instructions and procedures in this manual before installing and operating the units.
Signal Words
Signal words are used according to international standards throughout this
manual. The meanings of these signal words are:
- DANGER: Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. This signal word is to be limited to the most extreme situations.
- WARNING: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
- CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. Permitted for property-damage only accidents.
General Safety
The MAXX-606/607 is a Class 2M laser product which meets the requirements of
IEC/EN 60825-1:2007. The American National Standards Institute (ANSI) notes
that Class 2M lasers emit laser pulses in the visible portion of the spectrum
(wavelengths of 0.4 µm to 0.7 µm) and are potentially hazardous when viewed
with an optical instrument. Any emissions outside this wavelength region must
be below the Class 1M AEL.
WARNING: Risk of incorrect diagnosis or treatment.
- Isocentre of alignment lasers must coincide with isocentre of diagnostic or treatment equipment. Testing the alignment of lasers to isocentre of diagnostic or treatment equipment is based on hospital procedures.
- Newly installed equipment can possibly shift as building walls and floor settle and as building materials dry out. Realign lasers to isocentre before use if they have moved.
CAUTION: Laser light, risk of eye damage.
- To prevent eye damage, physicians, patients, and/or operators should not stare into the laser beam. The normal blink reflex will prevent eye damage; however, if a patient is unable to normally blink the laser operator must prevent the laser from shining in the patient’s eye.
- Replace laser diode only with assembly provided by Cemar Electro Inc. to ensure that laser power levels do not exceed Class 2M limits.
Protective Eyewear
Protective eyewear is not necessary for typical applications where direct
papillary impingement by the beams is a random momentary event. Under typical
conditions requiring more protracted ocular exposure, protective eyewear or
other protective measures may be required. Consult the user standards of the
ANSI for further guidance
(https://www.rli.com/resources/articles/classification.aspx).
Laser Specifications
Specifications for MAXX-606/607 are as follows:
External Housing Specifications
Housing Dimensions| 210.0 mm x 118.0 mm x 100.0 mm (1.81 kg)
Housing Material
| Aluminum
Hard black anodized T6061 Electrically isolated
Exit Aperture Protection| Window
Electrical Specifications
Operating Voltage| 90-240V AC
Load Current| 1.0A
Power Consumption| 1 W
Power Supply| 110/240V AC
50-60Hz 0.8A
9.0V DC-1.7A
Interconnect Cable| 6ft DC input
Environmental Specifications
Operating Temperature| 15°C to 35°C (59°F to 95°F)
Storage Temperature| -40°C to 80°C (-40°F to 176°F)
Dust/Water Resistant| IP50
Optical Specifications
Wavelengths| 635nm (red) and 520nm (green) wavelengths
Wavelength Stability| 0.3nm/°C nominal
Optical Output Power| 1 mW to 10mW (at the exit aperture)
Power Output Stability at 25°C| ± 1%
Line Generator| Customized, fan angle 60°
Beam Divergence| < 5 mrad
Focal Distance| User defined
Line Thickness
| 1 mm @ 3 m distance
3/16” @ 10’ (@1/e²)
Line Thickness Divergence| < 1 mRad
Line Intensity Profile| Gaussian
Lens Material| Glass
Pointing Specifications
Accuracy| < 1 mrad
Stability vs. Temperature| < 10 µrad/°C
Reliability/Regulatory Specifications
Laser Lifetime Expected| > 10.000 hours @ 25°C
Certification/Regulatory| FDA/CDRH
Warranty| 2 years
Safety Labels for Maxx-606/607 Units
The labels on all MAXX-606/607 units are required for compliance with federal
regulations. Do not remove these labels.
Safety labels are located on the laser unit:
Figure 3.3 Safety labels as displayed on MAXX-606/607 unit.
Storage
To ensure optimal functioning of the MAXX-606/607 unit, ensure the following:
- Do not drop
- Store all components in a non-corrosive environment
- Maintain storage temperature between -20°C and +75°C (-4° F and 167° F)
- Maintain storage humidity between 20% and 85% relative humidity. Do not permit dew condensation
POWER SUPPLY
Each MAXX-606/607 laser includes a universal power supply (transformer) with
an input of 100V to 240V at 50/60Hz-0.8A.
A 1.2-meter (4 foot) power cord supplies 9V DC 1.7A power to the laser unit
with a screw-on power plug. For servicing switch off the unit and remove power
from the supply using the appliance inlet.
GENERAL ROOM LAYOUT
Before the MAXX-606/607 can be correctly installed in the treatment room, the system isocentre must first be identified. Once identified, appropriate reference marks should be placed on the wall to determine the precise location of the mounting, as well as aid in the alignment of the laser lines once the lasers have been installed.
Accurate positioning is crucial; a failure in installation can result in inaccurate laser readings. Please contact Cemar Electro Inc. if you wish to coordinate installation by one of our professionals.
Treatment Room Preparation
Correct treatment room set-up is crucial to the successful installation of a
patient positioning laser system.
Preparation work on the treatment room should be completed before the system
components are shipped to the installation site. Necessary preparation work
comprises:
- Establishing where various system components will be located within the room
- Determining the mounting configuration for each laser unit
- Providing an adequate structure to mount all the laser units in the system
- Determining routings for cables
- Establishing necessary power for the system
In the case where the treatment room is not yet constructed, Cemar Electro Inc. recommends that a rigid steel support frame for the moving lasers is incorporated into the room construction.
Room Specifications
The treatment room must meet the following specifications to permit for
installation of the MAXX-606/607 unit.
Treatment Room Specifications
Category| Feature| Specification
Electrical| Power supply| 90-240V AC, 0.8A, 50-60Hz
Output voltage: 9.0V DC-1.7A
Mechanical| Laser unit dimensions| Length: 210.0 mm (8.2 in)
Width: 118.0 mm (4.6 in)
Height: 100.0 mm (3.9 in)
Weight| 1.81 kg (4 lbs)
Environmental| Room size| Minimum: 10’ by 16’ by 7.5’
Maximum: 40’ by 60’ by 20’
Operating temperature| 15°C to 35°C (59°F to 95°F)
Others| No flammable anaesthetics
No corrosives
LASER INSTALLATION
Pre-Installation
The MAXX-606/607 Patient Positioning Laser System provides an optical aid for
rapid, accurate patient positioning and repositioning for therapy and
simulator machines.
Various combinations of the MAXX-606 (crosshair) and the MAXX-607 (sagittal)
are used with different systems.
Option 1. Two MAXX-606 crosshair lasers—mounted on each side—beam crosshairs
along the X-axis.
Option 2. One MAXX-606 crosshair laser—mounted on the ceiling—beams a
crosshair vertically along the Z-axis.
Option 3. The MAXX-607 sagittal laser beams a straight line in the Z-Y plane.
Installation Instructions
The following instructions do not supersede hospital procedure for room laser
setup and are suggested procedure only.
Note: Manually align the lasers as close as possible when mounting the
laser to the wall, then fine tune using controls.
- Scan a phantom to determine isocentre.
- Install MAXX-607 (sagittal) laser according to the phantom location and indication marks.
- Rotate head and project to wall locations to mount MAXX-606 (crosshair) lasers.
- Note: Optional laser level or water levels may be used for accuracy.
- Verify all laser intersections.
- Lasers should be installed as co-planer with no parallax.
Mounting Options
Steel or aluminum base plates must be installed in the walls and ceiling where
laser units are to be mounted. The location, size, and stability of these base
plates are critical to a satisfactory installation of a patient positioning
laser system.
Lasers can be installed to either mounting plates (Figure 6.1) or adjustable
mounting brackets (Figure 6.2).
For the MAXX-606 laser, mount plates (Figure 6.1) to walls first, then attach the MAXX-606 laser units to the plates.
Adjustable mounting brackets can be used to install all units. Cemar Electro Inc. highly recommends using these specific brackets for ceiling-mounted crosshair laser (MAXX-606-CHG) as well as sagittal laser (MAXX-607-SAG).
Note: The mounting bracket can be adjusted to a maximum of 45º.
Operator Controls
Operator controls facilitate the movement and operation of the patient positioning laser.
Option 1. All lasers are connected to one wall switch; lasers can be powered
on and off simultaneously from a single convenient location.
Option 2. Lasers are connected to a timer so that the lasers are automatically
turned on or off at a set time.
Figure 6.3. Overview of operator controls.
SYSTEM MAINTENANCE
Repair and Servicing
WARNING: Only certified Cemar Electro Inc. technicians should carry out any repairs to your laser units other than changing the laser module; if authorized, note replacement of the laser module will affect the 90º angle in the cross.
Note: Without written permission of Cemar Electro Inc., any unauthorized repairs to the laser unit can cause further damages and will result in the warranty extended by Cemar Electro Inc. to be rendered null and void.
Maintenance, Cleaning, and Disposal
Once installed, the MAXX-606/607 laser requires no maintenance other than a
light cleaning. All cleaning efforts should be undertaken with a soft lint-
free cloth and, if necessary, mild soap containing no ammonia or abrasives.
- An occasional outside dusting of the antireflective window in the cover may be necessary. Use only standard lens cleaner with clean wipes.
- Clean other outside surfaces of the laser unit with a soft cloth slightly moistened with water or a mild detergent solution and then wipe the surface with a dry cloth.
- Do not use any type of abrasive pad, scouring powder, or chemical solvents such as thinner, benzene, alcohol, or disposable wipes as these may damage the finish.
- Do not mix this device with general waste upon disposal. For proper treatment, recovery, and recycling, please contact your disposal local supplier for further information.
In case that a laser fails to function properly, please contact Cemar Electro Inc. Customer Service as soon as possible (see Section 8 for Contact Information).
TECHNICAL SUPPORT AND CUSTOMER SERVICE
Cemar Electro Inc. is committed to satisfying our customer’s needs. If you have any questions, comments or suggestions regarding our products and service, please call or email us.
Please contact a Cemar Electro Inc. distributor or our representative for a quotation or for a detailed description of our ordering policies, warranties, delivery policy, conditions of sale, damaged goods policy, and returned goods policy.
Cemar Electro Inc. Service Representatives are available to work with facility
planners, architects, and hospital personnel throughout the planning and
installation process.
Service Department hours are: 8.00 a.m. through 4.30 p.m. (Eastern Time).
Phone: 514-631-5807
Fax: 514-631-7505
Toll free: 1-800-298-5273
Phone: 518-298-3065
Fax: 518-298-2927
Toll free: 1-800-298-5273
REGULATORY INFORMATION
Manufacturer Information
Manufacturer:
Cemar Electro Inc. 1370 55th Ave, Lachine, QC H8T 3J8 Canada
Authorised Representative (Europe):
P. Spirig GmbH
Baunstrasse 23
CH-9435 Heerbrugg,
Switzerland
WARRANTY
Cemar Electro Inc. warrants its patient positioning equipment to be free from defects in materials and workmanship under normal use and service for two years from the date of shipment. The sole obligation of Cemar Electro Inc. under this warranty is to repair or replace without charge or to refund the purchase price, at the option of Cemar Electro Inc., of any parts which its examination shall have disclosed to be defective, provided that buyer shall have given to Cemar Electro Inc. a written notice of the claimed defect no later than seven days after the end of the warranty period (two years of the date of shipment of such equipment to buyer). At the request of Cemar Electro Inc., the buyer, at their expense, shall return the claimed defective part to Cemar Electro Inc.
Disclaimer of Other Warranties
The aforesaid warranty rights are buyer’s exclusive remedies and are in lieu
of any other remedies, obligations, or rights, including, without limitation,
any other warranties, expressed or implied (e.g., implied warranties of
merchantability or fitness for a particular purpose).
Under no circumstances shall Cemar Electro Inc. be liable for any incidental, indirect, special, or consequential damages or for any other loss, damage, penalty, or expense of any kind including, without limitation, loss of profits or overhead, reimbursement, personal injury, or property damage. The aforesaid warranty obligation of Cemar Electro Inc. constitutes its sole liability and under no circumstances shall the maximum liability of Cemar Electro Inc., under any legal theory (e.g., contract, warranty, negligence, promissory estoppel, strict liability, misrepresentation, tort) and for any reason whatsoever (e.g., defect, delay, or otherwise) exceed the purchase price of the defective part regardless whether the claim is asserted by the buyer or any other person or entity.
The liabilities of Cemar Electro Inc. as above set forth, shall not be extended because of advice given by it in connection with the design, installation, or use of the equipment or parts, therefore.
GLOSSARY OF TERMS & LIST OF SYMBOLS
Glossary of terms:
American National
Standards Institute (ANSI)
| The primary organization for fostering the development
of technology standards in the United States. ANSI facilitates the development of American National Standards (ANS) by accrediting the procedures of standards developing organizations (SDOs). Accreditation by ANSI signifies that the procedures used by the standards body in connection with the development of American National Standards meet the Institute’s essential requirements for openness, balance, consensus, and due process.
---|---
Anodized T6061| A type of aluminum alloy whose properties include its
structural strength and toughness. It offers good finishing characteristics
and responds well to anodizing, including clear and colored dye, and hardcoat.
T6061 aluminum alloy is also easily welded and joined. T6061 alloy conducts
and dissipates heat well. Since it can be extruded, it can be formed into
almost any imaginable shape.
Crosshair laser| Crosshair lasers mark the point where a tool will start
working or an angle for alignment in two directions.
Co-planer| Three or more points which lie in the same plane.
Diagnostic imaging| Diagnostic or medical imaging is the technique and
process of creating visual representations of the interior of a body for
clinical analysis and medical intervention, as well as visual representation
of the function of some organs or tissues (physiology).
Ionizing radiation| A form of radiation consisting of particles, X-rays,
or gamma rays with sufficient energy to cause ionization in the medium through
which it passes.
International Electrotechnical Commission (IEC)| The International
Electrotechnical Commission is the leading global organization that publishes
consensus based International Standards and manages conformity assessment
systems for electric and electronic products, systems and services,
collectively known as electrotechnology. IEC publications serve as a basis for
national standardization and as references when drafting international tenders
and contracts.
Isocentre| The convergence of the three axes of rotation in radiation
therapy; the intersecting point of the axis of rotation of the gantry, the collimator, and the treatment couch.
---|---
Laser (diode) modular design| Modular laser design are modules
specifically designed for applications where space is a key consideration. The
laser modules are compact in size while offering the same high reliability and
constant power output of larger lasers.
Mounting bracket| A mounting bracket is support, composed of metal,
which projects from a wall with the purpose of holding or bearing weight.
Mounting plate| A mounting plate is the part of a hinge, handle, or
latch that mounts an object to a wall. The plate is attached directly to the
wall, which the object mounted thereupon attaches to the plate.
Parallax| The effect whereby the position or direction of an object
appears to differ when viewed from different positions, ( e.g., through the
viewfinder and the lens of a camera).
Phantom| A phantom is a specially designed object that is scanned or
imaged in the field of medical imaging to evaluate, analyze, and tune the
performance of various imaging devices. A phantom is more readily available
and provides more consistent results than the use of a living subject or
cadaver, and likewise avoids subjecting a living subject to direct risk.
Sagittal laser| Ceiling or wall-mounted single laser, preferably
moveable. The scanner couch can move up/down and in/out but cannot move
left/right, therefore the sagittal laser should move left/right to allow
marking away from patient mid- line. Ceiling lasers that can move in lateral
as well as longitudinal direction can improve the efficiency of CT simulations
and potentially reduce errors due to manual control errors.
Treatment room| Room in a patient care unit ( e.g., a hospital,
medical center, clinic, etc.) in which treatments, procedures, or examinations
are conducted. The room houses specific equipment and/or machinery to execute
the necessary procedures.