RICE LAKE MSI Dyna-Clamp Tension Meter User Manual
- June 5, 2024
- RICE LAKE
Table of Contents
MSI Dyna-Clamp Tension Meter
MSI Dyna-Clamp
Tension Meter
Operation Manual
September 14, 2021
PN 193853 Rev C
An ISO 9001 registered company © Rice Lake Weighing Systems. All rights
reserved.
Rice Lake Weighing Systems® is a registered trademark of Rice Lake Weighing
Systems.
All other brand or product names within this publication are trademarks or
registered trademarks of their respective companies.
All information contained within this publication is, to the best of our
knowledge, complete and accurate at the time of publication. Rice Lake
Weighing Systems reserves the right to make changes to the technology,
features, specifications and design of the equipment without notice.
The most current version of this publication, software, firmware and all other
product updates can be found on our website:
www.ricelake.com
Contents
Contents
1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . 1
1.1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . 1 1.2 Display . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . 2
2.0 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . 3
2.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . 3 2.1.1 Battery Installation . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Product Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . 3 2.3 Power . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 2.4
Wire Rope Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . 3 2.5 Zero. . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 2.6
Tension Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . 4 2.7 Average Function . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.7.1 Average Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . 5 2.7.2 Clearing Average . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . 5 2.8 Peak Function . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 2.8.1 Clearing Peak. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . 5 2.9 Safety Strap . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 2.10 Carry
Case Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . 5
3.0 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . 6
3.1 General Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . 6 3.2 Setup Menu. . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 3.3
Function Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . 7
3.3.1 Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . 7 3.3.2 Test . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 3.3.3 Unit. . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . 7 3.3.4 Scroll . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . 7 3.3.5 Active Function. . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 3.4
Auto-Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . 8 3.5 Range . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8 3.5.1 Number Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . 8 3.6 Filter . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 3.7
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . 9 3.8 Battery Life . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.0 Troubleshooting and Maintenance . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . 11 4.1.1 Error Codes . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.2 Battery Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . 12 4.2.1 Disposable Batteries . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . 12 4.2.2 Rechargeable
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.3 Drive Rod Lubrication . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . 13 4.4 Bobbin Replacement. . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.0 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . 14
Technical training seminars are available through Rice Lake Weighing Systems. Course descriptions and dates can be viewed at www.ricelake.com/training or obtained by calling 715-234-9171 and asking for the training department.
© Rice Lake Weighing Systems All Rights Reserved
i
MSI Dyna-Clamp Tension Meter
Rice Lake continually offers web-based video training on a growing selection of product-related topics at no cost. Visit www.ricelake.com/webinars
ii
Visit our website www.RiceLake.com
Introduction
1.0 Introduction
The Dyna-Clamp is a reliable, accurate, easy to operate tension meter that is
sealed for outdoor use. The two line LCD display allows for selecting the
correct wire rope diameter and displaying the current wire rope tension. The
Dyna-Clamp is factory calibrated for up to eight wire rope diameters.
Manuals and additional resources are available from Rice Lake Weighing Systems
at www.ricelake.com/manuals
Warranty information can be found on the website at
www.ricelake.com/warranties
1.1 Safety
Safety Signal Definitions:
DANGER
Indicates an imminently hazardous situation that, if not avoided, will result in death or serious injury. Includes hazards that are exposed when guards are removed.
WARNING
Indicates a potentially hazardous situation that, if not avoided, could result in serious injury or death. Includes hazards that are exposed when guards are removed.
CAUTION Indicates a potentially hazardous situation that, if not avoided, could result in minor or moderate injury.
IMPORTANT
Indicates information about procedures that, if not observed, could result in
damage to equipment or corruption to and loss of data.
General Safety
Do not operate or work on this equipment unless this manual has been read and all instructions are understood. Failure to follow the instructions or heed the warnings could result in injury or death. Contact any Rice Lake Weighing Systems dealer for replacement manuals.
WARNING
Failure to heed could result in serious injury or death.
Do not allow minors (children) or inexperienced persons to operate this unit.
Do not use for purposes other then measuring the tension of a wire rope. Do
not use any load bearing component that is worn beyond 5% of the original
dimension. Do not use the tension meter if any of the components of the load
train are cracked, deformed, or show signs of fatigue. Do not exceed the rated
load limit of the tension meter. Do not use the tension meter on wire rope
diameters over 1”. Do not make alterations or modifications to the tension
meter. Do not remove or obscure warning labels. Replace labels when worn.
Contact Rice Lake Weighing Systems for replacement labels. Keep hands, feet
and loose clothing away from moving parts and pinch points. Repairs must be
performed by qualified service personnel only.
IMPORTANT
Failure to heed could result in equipment damage and will void the warranty.
Do not over tighten the drive rod. Over tightening could cause damage to the
wire rope or the Dyna-Clamp. Regularly apply moly grease to the drive rod
threads to maintain proper lubrication to avoid premature wear and damage.
Remove the batteries prior to storing if the Dyna-Clamp is not going to be
used for and extended period of time. Do not use solvents or aggressive
substances to clean the Dyna-Clamp. Do not submerge.
© Rice Lake Weighing Systems All Rights Reserved
1
MSI Dyna-Clamp Tension Meter
1.2 Display
The front panel consists of a two-line, LCD display with four LED annunciators
and four keys. This section provides a detailed description of what is
featured on the display and the key functions.
14
SETPOINTS
F1 F2 MSI DYNA-CLAMP 1 2 13
3 F1
USER KEYS
4 F2
5
7
6
Measurement Systems International
Figure 1-1. Dyna-Clamp Front Panel
8 9
1
10 POWER
11
12
2
ZERO
Item No.
Description
1 Power Key powers the Dyna-Clamp on and off; in setup mode, it returns the
unit to tension display without saving settings 2 Zero Key zeros out the
residual tension on the link; in setup mode, it saves settings and drops back
one menu level 3 F1 Key programmable to user functions (Section 3.3 on page
7); default function is Active Function () 4 F2 Key programmable to user
functions (Section 3.3 on page 7); default function is Scroll () 5 Low Battery
displays when approximately 10% of battery life remains and blinks when
automatic shutdown is imminent 6 Primary display with six 0.7” (18 mm) numeric
digits; displays live wire rope tension and is sunlight visible 7 Secondary
display with eight 0.5” (13mm) numeric digits; displays wire rope diameter,
average tension and peak tension 8 Average Function indicates the average
function is currently displaying on the secondary (top) display line; displays
the
running average tension measurement (Section 2.7 on page 5) 9 Peak Function
indicates the peak function is currently displaying on the secondary (top)
display line; displays the greatest
tension that has been measured (Section 2.8 on page 5) 10 Kilonewton
indicates the tension display is in kilonewtons 11 Pound-force indicates the
tension display is in pounds 12 Kilogram-force indicates the tension display
is in kilograms 13 Setpoint/Range LEDs indicates if tension measurements are
outside of the programmable range 14 F-Key LEDs not currently supported
Table 1-1. Key Descriptions
Wire Rope Examples:
Below are two examples of how wire rope types appear on the Dyna-Clamp
display. The diameter of the wire rope type displays first, followed by the
wire rope class.
Wire Rope Diameter = 3/4” | Wire Rope Class = 6×36
Wire Rope Diameter = 1/2” | Wire Rope Class = 8×19
2
Visit our website www.RiceLake.com
Operation
2.0 Operation
This section includes the basic operation procedures to get the Dyna-Clamp working out of the box.
2.1 Unpacking
Immediately after unpacking the Dyna-Clamp from the shipping container,
visually inspect the product to ensure all components are included and
undamaged. If parts were damaged in shipment, notify Rice Lake Weighing
Systems and the shipper immediately. Retain the shipping container for future
shipping or transporting of the unit.
2.1.1 Battery Installation
Batteries do not come pre-installed in the Dyna-Clamp. See Section 4.2 on page
12 for initial installation of the two (2) included AA cell 1.5 V alkaline
batteries.
IMPORTANT
Remove the batteries prior to storing to prevent possible battery corrosion if the Dyna-Clamp is not going to be used for an extended period of time.
2.2 Product Dimensions
A
D
B
C
Figure 2-1. Product Diagram
A 25.7” (652.8 mm)
B
C
7.3” (185.5 mm) 13.7515.5” (349.3393.7 mm)
Table 2-1. Product Dimensions
D 3.7” (94.0 mm)
2.3 Power
Press
to turn on the Dyna-Clamp. The startup sequence displays all LCD segments, followed briefly by the software
version number and the battery volt level. After the startup sequence is complete, the primary display area of the Dyna-Clamp enters tension mode.
2.4 Wire Rope Selection
Press F1 ( by default) with the current wire rope value displaying in the
secondary display area to toggle through the available wire rope options. The
wire rope selection needs to be made before Zeroing the Dyna-Clamp.
The F1-key is setup as Active Function () by default. See Section 3.3 on page
7 to program the function keys.
Note
The F2-key is setup as Scroll () by default and scrolls through the secondary
(top) display area options.
The Dyna-Clamp is factory calibrated for up to eight (8) wire rope diameters.
Calibrations may only be performed by Rice Lake Weighing Systems. Please
contact Rice Lake Weighing Systems for calibration assistance or if additional
wire rope calibrations are required.
© Rice Lake Weighing Systems All Rights Reserved
3
MSI Dyna-Clamp Tension Meter
2.5 Zero
Press
when the Dyna-Clamp is not under tension to remove small deviations of the zero reading. Zero on location, just
prior to positioning the Dyna-Clamp on the wire rope. Re-zero the Dyna-Clamp each time the wire rope selection is changed.
Note The Dyna-Clamp must be held horizontally by the carrying handle. The load cell mount assembly must be fully
suspended and not in contact with any surrounding surfaces or objects before pressing
. The unloaded
indication on the display will vary slightly depending on orientation. This variation is negated once under tension.
2.6 Tension Measurement
The following procedure describes how to connect the Dyna-Clamp to a wire rope to take a tension reading. The Dyna-Clamp can be used to monitor the tension of a wire rope for maintenance and installation applications.
IMPORTANT
Always ensure that the wire rope is centered on both bobbins and never use the Dyna-Clamp on a wire rope diameter over 1”. Regularly apply moly grease to the drive rod threads to maintain proper lubrication.
1. Loosen the drive rod. 2. Zero the Dyna-Clamp (Section 2.5).
Wire Rope Bobbin
Mount Assembly Top
Drive Rod End Drive Rod
Figure 2-2. Dyna-Clamp On Wire Rope
3. Position the Dyna-Clamp on the wire rope (Figure 2-2).
Note Position Dyna-Clamp directly to the wire rope, at least two feet from any
attachments or the termination points.
4. Roll the Dyna-Clamp 6-8” back and forth along the wire rope to align the
unit to the center of the wire rope. 5. Tighten the drive rod assembly firmly,
torquing the drive rod to approximately 60-100 in-lb. Drive rod assembly
should
be tightened until the rope contacts the stop block and the handle becomes
hand tight; be careful not to “crush” the rope by over-tightening. The optimal
torque rating varies depending on the wire rope diameter and construction.
IMPORTANT Do not over tighten the drive rod. Over tightening could cause
damage to the wire rope or the Dyna-Clamp.
6. Take the tension reading from display. See Section 2.7 on page 5 for how
to use the Average Function. 7. Loosen the drive rod assembly and remove the
Dyna-Clamp from the wire rope. 8. Repeat Steps 3-7 to take additional tension
measurements as needed. For the best accuracy take at least three
tension readings at slightly different locations along the wire rope using the
Average Function (Section 2.7 on page 5).
4
Visit our website www.RiceLake.com
Operation
2.7 Average Function
The average function allows the operator to combine multiple tension
measurements and displays the average in the secondary (top) display area.
Note The following procedures assume that the F-Keys are still programmed to
the default settings with F1 set to
and F2 set to . See Section 3.3 on page 7 to program the F-Keys.
2.7.1 Average Procedure Follow this procedure to utilize the average function:
1. Press F2 until Avg displays to the right side of the secondary display
area.
2. Position the Dyna-Clamp on the wire rope and allow a few seconds for a
stable reading.
3. Press F1 . This takes the current display value to be included into the
running average.
4. Remove the Dyna-Clamp from the wire rope. Additional values can not be
added to the running average until the Dyna-Clamp has been removed from the
wire rope and the live tension reading passes below a low % of capacity. This
is by design to prevent averaging the same load more than once.
5. Repeat Steps 2-4 as needed.
2.7.2 Clearing Average Follow this procedure when the running average needs to
be cleared:
1. Position the Dyna-Clamp on the wire rope and tighten slightly, so a little
positive tension is measured by the device.
2. Press F1 , followed quickly with pressing
to clear the current average register.
3. Remove the Dyna-Clamp from the wire rope and re-zero (Section 2.5 on page 4).
2.8 Peak Function
The purpose of the peak function is to take and record the greatest tension
that is applied. Press the F-Key until Peak displays to the right side of the
secondary display area. As long as Peak is displayed, the secondary display
area continuously displays the greatest tension that has been measured.
2.8.1 Clearing Peak Press F1 ( by default) with the peak value displaying in
the secondary display area to reset the peak function.
Note Changing from one wire rope setting to another will automatically clear
both the average and peak values.
2.9 Safety Strap
The Dyna-Clamp has a slot and hole on either side of the display to
accommodate the attachment of a safety strap.
2.10 Carry Case Option
A hard-sided carry case (PN 198418) is available as an option to provide more
protection and security when transporting the Dyna-Clamp tension meter.
© Rice Lake Weighing Systems All Rights Reserved
5
MSI Dyna-Clamp Tension Meter
3.0 Configuration
This section describes parameter settings on the Dyna-Clamp. The Dyna-Clamp is factory calibrated for up to eight wire rope diameters. Calibrations may only be performed by Rice Lake Weighing Systems. Please contact Rice Lake Weighing Systems for calibration assistance or if additional wire rope calibrations are required.
3.1 General Navigation
Use the following keys for navigating through the menus when setting up the
Dyna-Clamp:
· Press F2 to scroll through the parameters or settings
· Press F1 to enter a selected parameter or to save a selection and return to previous menu
· Press
to save a setting and return to previous menu or to tension mode ( displays)
· Press
to return to tension mode without saving settings ( displays)
3.2 Setup Menu
To enter the setup menu, press
6HWYS
and F2 at the same time.
)YQF
2II WHVW 8QLW SULQW 6FUR// $FWI9Q
IYQF
2II WHVW 8QLW SULQW 6FUR// $FWI9Q
D2II
2II
UDQJH
2II
21
/R
+L
value
value
Figure 3-1. Setup Menu
ILOWHU
2II /2 +L +L
XQLW
/EI ,*) ,P
E/LIH
VWDQG /RQJ
Parameter
Description
Func1 Func2
A-OFF Range
Filter Unit b.Life
F1 key configurable to listed parameters; ActFun (default) F2 key
configurable to listed parameters; Scroll (default)
· Off key is disabled · Test runs display and system test (Section 3.3.2
on page 7) · Unit ability to change between available units lbf, kgf and kN
(Section 3.3.3 on page 7) · Print not currently implemented · Scroll
scrolls through the available wire rope types and the average and peak values
in the secondary (top) display area
(Section 3.3.4 on page 7) · ActFun performs action based on the active
function that is displayed in the secondary (top) display; adds an additional
value for average or clears the current peak value (Section 3.3.5 on page 7)
Auto-Off sets the amount of standstill time (in minutes) before the Dyna-
Clamp automatically turns off (Section 3.4 on page 8); settings: Off
(default), 15, 30, 45, 60
Range sets a low (Lo) and high (Hi) setpoint value; provides an indication
of when the tension measurement is within this specified range (Section 3.5 on
page 8); settings:
· Off indicates setpoints are disabled · On indicates setpoints are
enabled · Lo prompts user to enter setpoint value for low range setpoint ·
Hi prompts user to enter setpoint value for high range setpoint
Tension Filter allows the scale to adjust to situations where there may be
movement or vibration (Section 3.8 on page 10); settings: Off, LO, Hr-1, Hr-2
Tension Units sets the units displayed (Section 3.7 on page 9); settings:
Lb.F (lbf), I G.F (kgf), I N (kN)
Battery Life sets the options for standard or extended battery life (Section
3.8 on page 10); settings: Stand, Long
Table 3-1. Setup Menu Parameter Descriptions
6
Visit our website www.RiceLake.com
Configuration
3.3 Function Keys
There are two function keys, F1 and F2 on the front panel of the Dyna-Clamp. The F-keys can be programmed to several different functions. To assign a function to an F-key:
1. Press
and F2 simultaneously to enter the setup menu. displays.
2. Press F1 to enter the setup of the F1 key.
3. Press F2 to scroll through the available settings.
4. Press F1 to select the desired setting and return to the setup menu. displays.
5. Repeat Step 2 through Step 6 for the F2 key if needed.
6. Press
to save. displays briefly and the Dyna-Clamp returns to tension mode.
3.3.1 Off When an F-Key is set to , the F-Key is disabled.
3.3.2 Test See Section 3.3 to set an F-Key as . When an F-Key is set to ,
pressing the F-Key prompts the display to scroll through the following:
· Lights all LCD segments and the LEDs · followed by the version number ·
followed by the battery level in volts · followed by counting from to ·
followed by the c-cal value displaying in the secondary display area
The test can be single stepped:
1. Press F-key set to , then immediately press F2 to stop the auto scroll.
2. Press F2 to scroll through each step and press F1 to view the step value.
3. Press
to stop the test.
IMPORTANT Internal tests are also performed and if the test fails, an error code displays (Section 4.1.1 on page 12).
3.3.3 Unit See Section 3.3 to set an F-Key as . When an F-Key is set to ,
pressing the F-Key scrolls through the available unit options: lbf, kgf, and
kN.
3.3.4 Scroll See Section 3.3 to set an F-Key as . When an F-Key is set to ,
pressing the F-Key scrolls through the secondary (top) display area options,
which include available wire rope types, the average value and the peak value.
The F2-key is setup as Scroll () by default.
3.3.5 Active Function See Section 3.3 to set an F-Key as . When an F-Key is
set to , pressing the F-Key performs an action based on the active function
that is displayed in the secondary (top) display area. If the average function
is displayed, then adds an additional value to the average. If peak function
is displayed, then clears the current peak value. The F1-key is setup as
Active Function () by default.
© Rice Lake Weighing Systems All Rights Reserved
7
MSI Dyna-Clamp Tension Meter
3.4 Auto-Off
The Auto-Off feature prolongs the life of the battery by turning power off when Dyna-Clamp is not in use. A key press or detected tension in motion exceeding 10 d, resets the Auto-Off time count.
When set to , the Dyna-Clamp only turns off by pressing
or if the battery is depleted.
Use the following steps to set the Auto-Off parameter:
1. Press
and F2 simultaneously to enter setup menu. displays.
2. Press F2 to scroll to .
3. Press F1 . The current auto-off setting displays.
4. Press F2 to scroll to the desired time (in minutes).
5. Press F1 to accept the desired time setting. displays.
6. Press
to save and return to tension mode.
3.5 Range
The purpose of Range is to provide an indication of when the tension
measurement is within a specified range. Range consists of a low and high
setpoint value. If the tension is below the low setpoint value (), then the
setpoint 1 LED turns on. If the tension is above the high setpoint value (),
then the setpoint 2 LED turns on. If the tension measurement is within the
specified range, then neither setpoint LED is on.
Use the following steps to set the Range setpoint values:
1. Press
and F2 simultaneously to enter the setup menu. displays.
2. Press F2 until displays.
3. Press F1 . The current setting flashes in the secondary (top) display area.
4. Press F2 to toggle from to .
5. Press F1 with flashing. displays with the current value flashing in the secondary display area.
6. Press F1 to accept the current value. – OR Press F2 to enter a new value.
See Section 3.5.1 for the number entry procedure. Press F1 to save the new
value. displays with current value flashing in the secondary display area.
7. Repeat Step 6 for setpoint value.
8. Press
to save and return to tension mode.
3.5.1 Number Entry Use the following keys to enter a number value:
· Press F2 to increment the number value of the flashing digit or to add an additional digit
· Press F1 to select the number value of the flashing digit or to save value when no digits are flashing
· Press
to add a decimal point after the last digit displayed
· Press
to clear the last made selection
8
Visit our website www.RiceLake.com
Configuration
3.6 Filter
Change the filter settings to adjust to situations where vibration in the wire
rope or the structure. If the reading is not stable, it can be improved by
increasing the filter setting. Settling time is longer when the filter setting
is increased. The Dyna-Clamp uses algorithms that speed up large tension
changes while still controlling vibration even with higher filter settings.
Parameter
Description
Off
Disables filtering function
Lo
Low Filter
Hi-1
High Filter
Hi-2
Very High Filter
Table 3-2. Filter Settings
Use the following steps to set the Filter parameter:
1. Press
and F2 simultaneously to enter setup menu. displays.
2. Press F2 to scroll to .
3. Press F1 . The current setting flashes in the secondary (top) display area.
4. Press F2 to scroll through the available filter settings.
5. With desired filter displayed, press F1 to accept. displays.
6. Press
to save and return to tension mode.
3.7 Units
Units can be changed in two ways. By programming an F-key to (Section 3.3 on
page 7) or by changing the units within the setup menu using the following
procedure:
1. Press
and F2 simultaneously to enter setup menu. displays.
2. Press F2 to scroll to .
3. Press F1 . The current setting flashes in the secondary (top) display area.
4. Press F2 to scroll through the available units.
5. With desired unit displayed, press F1 to accept. displays.
6. Press
to save and return to tension mode.
© Rice Lake Weighing Systems All Rights Reserved
9
MSI Dyna-Clamp Tension Meter
3.8 Battery Life
The battery life parameter () offers a standard mode () and a long battery
life mode (). When in the long battery life mode, the system is placed into a
sleep state if there is no change in tension after a period of time. This
disables the display in order to reduce power consumption and increase the
battery life. The unit does not enter the sleep state if it is in the setup
menu. The auto-off parameter (Section 3.4 on page 8) also conserves battery
life by turning the unit off if no activity is sensed for a set time.
Note
The standard mode ( ) is recommended for best performance. Long battery life mode ( ) does not affect accuracy, but may impact performance and usability.
To set battery life mode:
1. Press
and F2 simultaneously to enter setup menu. displays.
2. Press F2 to scroll to .
3. Press F1 . The current setting flashes in the secondary (top) display area.
4. Press F2 to toggle between and .
5. With desired unit displayed, press F1 to accept. displays.
6. Press
to save and return to tension mode.
10
Visit our website www.RiceLake.com
Troubleshooting and Maintenance
4.0 Troubleshooting and Maintenance
4.1 Troubleshooting
Problem
Possible Cause
Solution
Display is blank when POWER Discharged battery
key is pressed
Defective battery
Replace cells, or if using NiMH, recharge Replace
Corroded battery or battery contacts
Clean contacts
Defective switch or circuit board
Requires authorized service
Display does not function properly or front panel keys do not function normally or Dyna-Clamp not powering off
Improperly loaded software Faulty circuit board Loose connectors
Reinstall software Requires authorized service Requires authorized service
Dyna-Clamp does not respond Out of calibration
to tension changes
Faulty load cell
Requires authorized service Requires authorized service
Worn out or damaged drive rod
Requires authorized service
Display over ranges below 100% Zero requires adjustment
of capacity
Too much tension/load has been zeroed
Rezero the Dyna-Clamp Rezero the Dyna-Clamp
Display drifts
AZM (Auto0) is turned off
Turn AZM on
Rapid temperature changes such as moving Wait until the Dyna-Clamp temperature has stabilized the Dyna-Clamp from indoors to outdoors
Displayed tension shows larger Dyna-Clamp not zeroed before load is lifted Zero the Dyna-Clamp with no load attached
error
lbf/kgf/kN units causing confusion
Select proper units
Requires recalibration
Requires authorized service
Display reading not stable
Excessive vibration in wire rope
Increase filtering
Load cell faulty
Check LC connections
Display toggles between Error Tension exceeds capacity
and Load
Faulty load cell or wiring
Immediately reduce tension Check LC and LC wiring
Display toggles between Error A key is stuck or is being held down and Key
Check switches for damage Ensure that a remote is not continuously transmitting
Low battery is blinking
Battery is low
Replace (alkaline) or recharge batteries
Unit turns on then immediately off Battery is low
Replace (alkaline) or recharge batteries
Tension does not zero
System not stable
Wait for standstill annunciator to turn on
Increase filtering for more stability
Zero out of range
Zero range might be limited; only zero while Dyna-Clamp is not clamped to a wire rope
Tension does not average
System is not stable
Wait for standstill annunciator to turn on, or see Display reading not stable above for an additional solution
Low threshold % of capacity not met between Remove Dyna-Clamp from the rope being tested, the device must drop
average measurements
below the low % of capacity before a new average can be taken
Setpoint LEDs blink
Range is enabled and a trigger point has been reached
Disable range if it is not needed
Table 4-1. Troubleshooting
© Rice Lake Weighing Systems All Rights Reserved
11
MSI Dyna-Clamp Tension Meter
4.1.1 Error Codes The ScaleCore processor of the Dyna-Clamp detects errors and generates error codes to aid in troubleshooting.
Error Code
Description
LC Disabled A load cell was not enabled
Not Calibrated System has not been calibrated
Overload Tension exceeds set capacity +8 d; load Cell damaged or
disconnected
Underload Tension is more than 10% negative; load cell damaged or
disconnected
Table 4-2. Error Codes
4.2
1. 2. 3.
Battery Replacement
Unscrew the battery cover on the side of the handle. Insert two AA batteries
positive end first (can be replaced with rechargeable NiMH cells). Reinstall
the battery cover securely.
AA Batteries Battery Cover
Figure 4-1. Battery Installation
Part No.
Description
Qty
191147 Battery Cap Assembly
1
16431 AA Cell 1.5 V Alkaline Battery
2
Table 4-3. Battery Parts List
4.2.1 Disposable Batteries
Low Battery displays when the battery is getting low. When Low Battery starts
to blink, the batteries are close to being drained. For maximum life, use the
batteries until the system shuts off.
4.2.2 Rechargeable Batteries When using Nickel-Metal-Hydride (NiMH) cells, it
is recommended that the cells are recharged immediately after Low Battery
starts to blink. Do not allow the batteries to discharge completely, it may
compromise the recharge life of the battery.
NiMH Cells have a lower capacity than Alkaline. Having two sets of NiMH
batteries is recommended, so one set can be charging while the other is in
use.
IMPORTANT
The use of NiCad batteries is not recommended.
Remove the batteries prior to storing to prevent possible battery corrosion if
the Dyna-Clamp is not going to be used for an extended period of time.
12
Visit our website www.RiceLake.com
Troubleshooting and Maintenance
4.3 Drive Rod Lubrication
Make sure the drive rod threads are free of any dirt or debris. Regularly
apply moly grease to the drive rod threads to maintain proper lubrication.
Drive Rod Threads
Figure 4-2. Grease Drive Rod Threads IMPORTANT Do not use solvents or
aggressive substances to clean the Dyna-Clamp. Do not clean by submersion.
Make sure to lubricate the drive rod threads after cleaning.
4.4
1. 2. 3.
Bobbin Replacement
Remove the bolt securing the wire rope bobbin to the frame and discard old
bobbin. Align new wire rope bobbin with hole in frame.
Torque bolt to 30 ft-lb (40 N-m) with blue Loctite® 242 to reinstall new wire
rope bobbin. The bobbin should spin freely with very little play along the
bolt axis.
Bobbin
Bolt Figure 4-3. Bobbin Installation
Part No.
Description
Qty
191155 Wire Rope Bobbin
2
192660 Shoulder Bolt, 1/2-13 x 1-1/4
2
Table 4-4. Bobbin Parts List
© Rice Lake Weighing Systems All Rights Reserved
13
MSI Dyna-Clamp Tension Meter
5.0 Specifications
Power Battery Operated
Two AA Cell 1.5 V Alkaline Batteries (can be replaced with rechargeable NiMH cells)
Operating Time Alkaline AA Cells
24 hours
Wire Rope Diameter
3/16” (5 mm) 1” (25 mm)
Accuracy Rated Accuracy
±3% of capacity
Capacity Maximum
12,500 lb (5,670 kg)
Units lbf, kgf, kN
Resolution Standard Internal A/D
2500 divisions 24 bits
Operator Interface Two-Line LCD Display: Primary Display (bottom) Secondary
Display (top)
Function Keys: Programmable Keys Permanent Keys
Six, 0.7” (18 mm) numeric digits Eight, 0.5” (13 mm) numeric digits
Two (user programmable) F1, F2 (TEST, UNIT, PRINT, SCROLL, ACTFUN) Two
(fixed) POWER, ZERO
Setpoints Two dedicated setpoints for minimum and maximum allowable wire rope tension (LEDs for measurements outside of allowable range)
Auto-Off Mode Prolongs battery life by turning POWER off after 15, 30, 45 or 60 minutes of no scale activity (operator determined)
Load Cell and Enclosure
Load Cell
350 Bridge
Enclosure
IP65 Powder Coated Aluminum Frame
Calibration Up to eight (8) factory programmed wire rope calibrations
Environmental Operating Temperature Operator Use Storage
-40°F122°F (-40°C50°C) Outdoor Indoor
Warranty Two-year limited warranty
14
Visit our website www.RiceLake.com
© Rice Lake Weighing Systems Specifications subject to change without notice. Rice Lake Weighing Systems is an ISO 9001 registered company.
230 W. Coleman St. · Rice Lake, WI 54868 · USA U.S. 800-472-6703 · Canada/Mexico 800-321-6703 · International 715-234-9171 · Europe +31 (0)26 472 1319
September 14, 2021
www.ricelake.com
PN 193853 Rev C
References
- Rice Lake Weighing Systems
- Videos and Webinars | Rice Lake Weighing Systems
- Rice Lake Weighing Systems
- Rice Lake Weighing Systems
- Rice Lake Weighing Systems
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>