NEUATION iFuge L400P 8x15ml Fixed Angle Rotor User Manual

June 5, 2024
NEUATION

iFuge L400P 8x15ml Fixed Angle Rotor
User Manual

INTRODUCTION

This centrifuge is equipped with a maintenance-free drive, a large display & simple interface for efficient operation for daily lab usage. The programmable centrifuge can deliver up to 4500 RPM and can accommodate different types of rotors. It features various programmable modes to save time & add convenience.

INTENDED USE

This is a large capacity centrifuge designed to separate, sediment, spin down aqueous solutions & solvent suspensions of differing densities incompatible sample containers.
NOTE: Before using the centrifuge, please read this user manual carefully. This user manual is intended to assist with the operation and care of the unit & is not a document that aids in repair. For repair please contact the supplier.

SALIENT FEATURES

The centrifuge has the following features:

  • Delivers up to 4500 RPM for all compatible rotors
  • BLDC maintenance-free motor drive
  • Imbalance detection safety with auto cutoff feature
  • Lid lock safety feature: The lid does not open during operation
  • Program mode for customized operation
  • Speed setting by RPM/RCF mode
  • Countdown timer range from 1 to 99 minutes
  • Last run memory feature
  • Convenient and easy user interface
  • Emergency lid release during power cutoff
  • Automatic internal diagnosis & error display

STANDARD ACCESSORIES

  • power cord
  • T-Allen key
  • User manual
  • Warranty card

TECHNICAL SPECIFICATIONS

Motor Type Brushless DC Motor
Max capacity 400ml (4x100ml)
Speed Setting Variable 500 – 4500 rpm (Step of 100 rpm)
Speed Accuracy ± 100 rpm
Run Time 1 min to 99 mins & infinite mode
Min. Acceleration Time 30 seconds
Min. Deceleration Time 30 seconds
Noise Level <65 d8
Ambient Temperature 5-40C
Permissible Relative Moisture <80%
Size (Lx Bx H) 475 x 585 x 325 mm
Weight 23 Kg (without rotor)
Input Power 110V
Power Consumption 300W

SAFETY PRECAUTIONS

  • Never use the centrifuge in any manner not specified in this manual.
  • Always use recommended original rotors and spare parts for the best result & product safety.
  • The rotors must be loaded symmetrically. Each tube should be counterbalanced by another tube of the same weight.
  • Do not use a centrifuge or rotor that has not been correctly installed or shows any sign of damage.
  • The rotor must always be securely fastened. If the centrifuge makes an unusual noise during operation, the rotor fitment needs to be checked. Switch OFF the device immediately by pressing STOP, check fitment & fasten it well.
  • Never move the centrifuge during its operation.
  • Prior to centrifugation, the tubes should be visually inspected for material damage. Damaged tubes must not be centrifuged. This is because broken tubes can result in sample  loss and can create an imbalance which can result in
  • Further damage to the centrifuge and accessories.
  • Do not fill tubes while they are in the rotor. Liquid spillage may harm the device. If liquids are spilled on the rotor or rotor chamber, the centrifuge must be cleaned carefully and properly before being used again.
  • The centrifuge may be used for the specified applications only. It must not be operated in a hazardous or flammable environment and must not be used to centrifuge explosive or highly reactive substances. Also do not place the potentially hazardous material within the clearance area/envelope.

• Equipment if used in any manner not specified in this manual or by the manufacturer can result in the lapse of the product warranty.
• Repairs must only be performed by an authorized service technician.
• Do not lean on the equipment. It may damage the equipment or even harm the operator.
• In the event of contamination caused by aggressive agents, the rotor must be cleaned immediately using a natural cleaning liquid. This is particularly important for the bores of the tubes. If any damage is seen, contact the service technician.
• Before using cleaning or decontamination methods other than those mentioned by the manufacturer, contact the manufacturer to ensure that the intended method will not damage the centrifuge.
• For safety we have provided protective earthing with a power supply. Make sure the power supply is earthed.

INSTALLATION

The Laboratory centrifuge is supplied in the packaging box. Open the box, remove the packaging and gently place the centrifuge out of the box. Before 1st time usage, open the centrifuge & ensure to remove all packaging from the rotor chamber.
Please keep all packaging in safe storage for at least 2 years for warranty purposes.

LOCATION & MOUNTING

Place the centrifuge on a flat, solid, and leveled surface and ensure that all four feet of this centrifuge stand on the surface firmly. Avoid installing on slippery or surfaces prone to vibration.

• Ideal ambient temperature is 25° C ± 5C, avoid placing the centrifuge in direct sunlight.
• Keep a clearance of at least 30 cm from all sides for ease of usage.
• Keep away from heat or water to avoid sample temperature issues or centrifuge failures.
• Do not place the centrifuge such that it becomes difficult to operate the device.

USER INTERFACE & DISPLAY

tem Button Name Function
1 Start Single press, the motor will start as per selected rpm and time if

the lid is closed. The start button works only if the lid is closed.  Motor start or stop indicates with LED
2| Stop/open| A single press stop/open button will stop the motor as per the defined DCC time if the motor is running. Second pressing the stop/ open button will open the lid if the motor is stopped. Also, use as lid open function if centrifuge motor is not running.
3| RPM/RCF| Single press, RPM will be displayed like 4500 and RPM LED will blink. Long Press, RPM/RCF will toggle the display to RPM to RCF. RPM and RCF modes are also indicated via LED on the panel.
4| Time| Single press, Time will be displayed like 99.00 The timer can be set from to 99 min.
5| Increment/
Decrements| A short press of {+}Plus and {-}Minus buttons will increase and decrease speed, time, ACC, and DCC  values normally.
A long press of these keys would change the parameters quicker
6| Acceleration| First Single press, Acceleration will display on time segment like ACC. We can change acceleration  using {+} and {-} buttons. (1 to 9).
Deceleration| Second Single press, Deceleration will display on time
segment like DCCl. We can change Deceleration using
{+} and {-} buttons. (1 to 9).
7| Rotor| To select among multiple rotors listed below
8| Edit| Single press Pedit, SPEED/ TIME/ ACC/ DCC will blink one after one 5 times on the respective display, and the user can change blinking parameters by using {+}Plus and {-}Minus key. In edit mode user can also press SPEED/TIME/ACC/ DCC and change its value.
9| PROG| Press the PRG button to select the program from 1-99.
The different parameters can be set in each program for quick selection.
10| Quick Spin| Press hold for immediate centrifugation up to last set RPM

OPERATING THE CENTRIFUGE

SWITCH ON THE CENTRIFUGE
After connecting the power cord. Switch ON the main power supply & then Switch on the power switch located on the right side of the instrument. Make sure to check the rotor fitment before use. The centrifuge will not operate with an open lid.
Note: Maintain a gap of 3 seconds between switching OFF and switching ON again. DO NOT switch OFF  and ON again instantly.

Speed Setting:
Press RPM/RCF button to set the speed from 500 to 4500 RPM. Single press to set speed, long press to toggle mode RPM/RCF. The parameter can be set when the speed screen blinks by using+ the button for increment &”- “button for decrement.
The values will automatically save after the screen stops blinking.

Timer Setting:
Press the timer button to set a timer. Time can be set by + for increment &- for decrement. The timer will save automatically once the blinking stops. The timer can be set in minutes only while seconds are for the countdown timer for running operation.

Acc/Dcc:
This button will set the parameters of acceleration or deceleration of the rotor ramp. Press the ace/dee button once to set the acceleration from 1-9 by using the +/­ button for increment/decrement respectively. Similarly, press once again the ace/dee button to set deceleration from 1-9 by using the +/- button for increment/decrement respectively. The values of acceleration & deceleration timings are listed below.

PRG:
Press the PRG button to select the program from 1-99. The different parameters can be set in each programm for quick selection. To set parameters in each program follow the parameters (Speed, time & Acc/Dcc) as described above. To select from programs press using the +/- button for increment or decrement. All the parameters in the program save automatically once the blinking stops. The saved programs also can be edited by following the instruction of setting Speed, time & Acc/Dcc as described above.

Quick Spin:
Press hold the quick spin button for immediate centrifugation of the samples up to the previous run RPM. Releasing the hold will stop the rotor gradually.

Petit:
Press the edit button to enter all the different parameters – Speed, Timer, acceleration & deceleration automatically one after the other. This parameter will be asked by the instrument sequentially to be entered if required to be changed (if not edited the display value will be set). These parameters will save automatically.

Rotor:
Press the rotor button to select rotors as listed below. The rotor numbers can be changed by using+/- for increment or decrement of the rotor number.

Swing Out Rotors

Rotor| Rotor No.| Max. Volume| Max. RPM| Max. RCF
LC01-L30-130| Rotor 1| 4 x 100 ml| 4500| 3164
LC01-L30-129| Rotor 2| 4 x 50 ml| 4500| 3485
LC01-L30-128| Rotor 3| 6 x 50 ml| 4500| 3530
LC01-L30-125| Rotor 4| 16 x 15 ml| 4500| 3485
LC01-L30-126| Rotor 5| 12 x 15 ml| 4500| 3485
LC01-L400-140| Rotor 25| 32 x 6 ml| 4500| 3164
LC01-L400-124| Rotor 26| 2x Microplates| 4500| 2377
Fixed Angle Rotors

Rotor| Rotor No.| Max. Volume| Max. RPM| Max. RCF
LC01-L30-136| Rotor 7| 4 x 100 ml| 4500| 2559
LC01-L30-135| Rotor 8| 4 x 50 ml| 4500| 2644
LC1-L30-134| Rotor9| 6x50ml| 4500| 2644
LC01-L30-137| Rotor 10| 24x 15 ml| 4500| 2852
LC01-L30-131| Rotor 11| 16×15 ml| 4500| 3144
LC01-L30-132| Rotor 12| 12 x 15 ml| 4500| 3144
LC01-L30-133| Rotor 13| 8×15 ml| 4500| 2644

START &STOP
The centrifugation can be started by pressing the start button once the desired operation is set. The operation can be stopped by pressing the stop button once. To open the lid press the stop/open key once after the rotor comes to stop.

MAINTENANCE AND CLEANING

  • The rotor and the outside of the centrifuge should be cleaned regularly with a mild wet (with water) cloth.

  • Ensure that while cleaning the unit is not plugging in.

  • Wear protective gloves & safety glass while operating & cleaning the device.

  • The brushless motor in the centrifuge requires no routine maintenance. Any required service should be performed by authorized, qualified personnel only. Repairs performed by unauthorized personnel may void the warranty.

  • Always keep the centrifuge housing, rotor chamber, and rotor clean. All parts should be wiped down periodically with a soft cloth. For more thorough cleaning, use a neutral cleaning agent (Ph between 6 and 8) and clean with a soft cloth. Exclusive amounts of liquid should be avoided.
    Note: Liquid should not come into contact with the motor.

  • After cleaning, ensure that all parts are dry before re-use.

  • Regularly cleaning the rotor is important.

  • If the rotor chamber needs cleaning, clean with a cloth or sponge moistened with a neutral detergent solution.

  • Do not place the rotor into the cleaning solution.

  • If corrosive, toxic, or pathogenic bacteria are accidentally spilled in the rotor or rotor chamber the centrifuge must be decontaminated thoroughly.

TROUBLESHOOTING

This centrifuge has a self-diagnostic function. If a problem occurs, an error/warning code will be displayed on the display screen and the operator can determine the malfunction with the warning code below.

ERROR PROBLEM SOLUTION
No display No main power
connection. Power check & proper plug-in of

mains cable at both ends.
Power failure| Check the main fuse of the lab.
Improper connection.| Connect the adaptor properly.
| The lid is not closed correctly.| Close the lid correctly.
Error with lid closing and
opening mechanism.| Contact service.
Err 55| Rotor not loaded
symmetrically.| Load rotor symmetrically & restart
centrifuge.
The centrifuge lid
cannot be opened| The rotor is still spinning.| Wait for the rotor to come to a stop.
Power failure| Emergency lid release after
rotor stops
Centrifuge shakes
during
acceleration &
exceptional
running noise| Rotor not loaded
symmetrically.| Load rotor symmetrically &
restart operation
Either a broken tube, damage to the rotor, or motor is cause for run noise.| Replace broken tube. For damaged rotor/motor contact service representative.
Rotor damaged.| Remove & change the rotor
Display error| Loose connection of display.| Contact service representative.
Err 1| Latch damaged, Latch jammed| Contact service representative.
Err 52| Motor stuck or incorrect operating voltage| Turn OFF the centrifuge, Check the rotor fitment or apply the correct 230VAC ± 10VAC operating voltage
Power tripping| The cable is not fit properly.| Remove the cable and connect properly.
Last run memory
not displayed| Turing ON centrifuge immediately after turning it OFF.| Maintain 3 seconds gap Between switching OFF and switching ON again.
System gets hang| Electronics error.| Switch off the centrifuge and then switch it ON again. If the error still shows, contact the service representative.

IMPORTANT NOTE:

  • If the system gets hung or gets heated due to overcurrent, switch OFF & switch ON (restart) the centrifuge and check it again.
  • Maintain 3 seconds gap between switch OFF and switch ON. Instant ON-OFF can lead to a reset, erasing last run memory.
  • If the motor gets hot due to which there will be fluctuation in speed value then allow centrifuge to get cool for at least 30 minutes. Do not do any operation for 30 minutes.

WARRANTY STATEMENT

This product is warranted to be free from defects in material and workmanship for a period of two (2) years from the date of purchase. Your product will be duly repaired upon  prompt notification in compliance with the following conditions:
This warranty is valid only if the product is used for its intended purpose and within the guidelines specified in this instruction manual. This warranty does not cover damage caused by accident, neglect, misuse, improper service, natural forces, or other causes not arising from defects in original material or workmanship. This warranty does not cover any incidental or consequential damages, commercial loss, or any other damages from the use of this product.
The warranty is invalidated by any non-factory modification, which will immediately terminate all liabilities on us for the products or damages caused by its use. The buyer and its customer shall be responsible for the product or use of products as well as any supervision required for safety. If requested the products must be returned to the distributor in a well-packed and insured manner and all shipping charges must be paid.
Some states do not allow limitations on the length of implied warranties or the exclusion or limitation of incidental or consequential damages. This warranty gives you specific legal rights. This warranty is given expressly in lieu of all other warranties, expressed or implied.
The purchaser agrees that there is no warranty of merchantability or of fitness for any intended purpose and that there are no other remedies or warranties, expressed or implied, which extend beyond the description on the face of the agreement. This warranty is only applicable to the original purchaser.
Products received without proper authorization will not be entertained. All items returned for service should be sent postage prepaid in the original packaging or
another suitable carton, padded to avoid damage. We will not be responsible for damage incurred by improper packaging.
All items returned for service should be set postage prepaid in the original packaging or another suitable carton, added to avoid damage.
This warranty is valid only if the warranty is registered with the supplier within 30 days from the date of purchase.
In case the product is to be disposed of, the relevant legal regulations are to be observed.

PRODUCT DISPOSAL

Information on the disposal of electrical and electronic devices in the European Community
The disposal of electrical devices is regulated within the European Community by national regulations based on EU Directive 2012/19/EU on waste electrical and electronic equipment (WEEE). According to these regulations, any devices supplied after 13.06.05 in the business to business sphere, to which this product is assigned, may no longer be disposed of in the municipal or domestic waste. They are marked with the following symbol to indicate this.

As disposal regulations within the EU may vary from country to country, please contact your supplier if necessary.

For your reference, make a note of the serial number, date of purchase, and supplier here.

Serial No.| Purchase Date
Supplier

Negation Technologies Pvt. Ltd.
Plot No. 15, GIDC Electronic Park SEZ Kolavada Road, Gandhinagar – 382026, Gujarat, India
Website: www.neuation.com