TTICM1000V 600A AC Clamp Meter Auto Ranging Instruction Manual

June 5, 2024
TTi

TTICM1000V-600A-AC-Clamp-Meter-Auto-Ranging-LOGO

TTICM1000V 600A AC Clamp Meter Auto Ranging

TTICM1000V-600A-AC-Clamp-Meter-Auto-Ranging-PRO

INTRODUCTION

TTTICM1000V is a true RMS (6000 count) auto ranging digital clamp meter with simulation bar. This meter measures AC Current, AC/DC Voltage, Resistance, Capacitance, Diode Test, Continuity and Non Contact Voltage detection.
The instruction manual includes relevant safety information and warning indication, please read them carefully and strictly observes all warnings and notes.

INTENDED USE
The clamp meter is to be used only for electrical inspection within the specifications the device is rated for. Failure to use the device within the specifications may result in serious injury or death.

PRODUCT CONTENTS

Unpack ensure the following attachments are complete or intact.

  1.  Operating instruction manual
  2.  Test Leads
  3.  TTTICM1000V Clamp Meter

GENERAL SAFETY

Prior to using the clamp meter, please read the product manual and ensure you have a solid understanding of the device’s functions and features.

SAFETY INSTRUCTIONS

SAFE OPERATION AND STANDARDS
This Meter complies with IEC 61010-1, Pollution Degree 2, Over voltage Category (CATll 1000V, CATlll 600V) and Double Insulation standards.
Anyone using this instrument should be knowledgeable and trained about the risks involved with measuring voltage, especially in an industrial setting, and the importance of taking safety precautions and of testing the instrument before and after using it to ensure that it is in good working condition.
This product has been tested to the requirements of IEC 61010
CAT II: Applicable to test and measuring circuits connected directly to utilization points
(socket outlets and similar points) of the low-voltage MAINS installation.
CAT Ill: Applicable to test and measuring circuits connected to the distribution part of the building’s low-voltage MAINS installation.

  • Check the clamp meter and test leads before using, guard against any damage.
    If any abnormalities or damage are found, do not use the clamp meter. Carefully examine the insulation around the jaws.

  • Inspect the test leads for damaged insulation or exposed wires. Check test leads continuity. Replace damaged test leads with leads that are suitable to the meter that meet or exceed the safety requirements.

  • Comply with local and national safety codes. Use personal protective equipment
    (approved rubber gloves, face protection, and flame-resistant clothes) to prevent shock and arc blast injury where hazardous live conductors are exposed.

  • Do not use in CAT III environments without the protective cap of test probe, The protective cap decreases the exposed probe metal < 4mm. This decreases the possibility of arc flash from short circuits.

  • Measure a known voltage first to make sure that the meter operates correctly.

  • When using the test leads, ensure leads are fully seated and keep your fingers behind the finger guards. Do not contact the bare wire and connector, unused input terminal or the circuit being measured when clamp meter is in operation.

  • Do not use the clamp meter without having rear cap and battery cover in place, otherwise electric shock may occur.

  • To avoid false readings that can lead to electrical shock and injury, replace the batteries as soon as the low battery indicator appears. Remove all probes, test leads, and accessories before replacing batteries.

  • Hold the meter behind the clamp body as specified in “external overview”.

  • Connect the common test lead before the live test lead and remove the live test lead before the common test lead.

  • Function selector shall be set at the correct position prior to measurement.
    Do not change selector settings while measuring to guard against damage to the meter.

  • Refrain from applying voltage over DC1000V/AC750V between the clamp meter terminals and ground to guard against electric shock and clamp meter damage.

  • Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc.

  • When the meter is working at an effective voltage over 60V in DC or 30V rms in AC, special care should be taken as there is danger of electric shock.

  • Do not apply more than the rated voltage, between the terminals or between each terminal and earth ground.

  • Do not measure current in circuits carrying more than 600A with the meter jaw.

  • Do not measure voltage or current higher than the allowable input value.

  • Disconnect circuit power and discharge all high-voltage capacitors before you measure resistance, continuity, or capacitance.

  • Do not modify the internal wiring in the meter to guard against damage to the meter and danger.

  • Do not use the HOLD function to measure unknown potentials. When HOLD is turned on, the display does not change when a different potential is measured.

  • Do not store or use clamp meter in an explosive or flammable environment, high temperature, high humidity or strong electromagnetic field.

  • Clean the clamp meter case with soft cloth and neutral detergent. To prevent causing corrosion to the case, or damage to the instrument do not use abrasives or solvents. Make sure all water is removed before use.

ELECTRICAL SYMBOLSTTICM1000V-600A-AC-Clamp-Meter-Auto-
Ranging-1

GENERAL SPECIFICATIONS

  • Display count: approx. 6000
  • Sampling rate: 3/s approx.
  • Sensor type: coil induction
  • Maximum jaw opening: 30mm
  • Power supply: 3xAAA 1.5V batteries
  • Dimension: 228mm x77mm x41mm

EXTERNAL OVERVIEW

TTICM1000V-600A-AC-Clamp-Meter-Auto-
Ranging-2

  1.  Clamp
  2.  Clamp body: Designed to protect operator from touching the dangerous area.
  3.  Clamp Trigger: Press to open clamp
  4.  Function Selector
  5.  LCD display.
  6.  Function key: Select basic functions.
  7.  Terminal inputs: Input test leads to measure via test lead

FUNCTION DIAL OVERVIEW

TTICM1000V-600A-AC-Clamp-Meter-Auto-
Ranging-3

BUTTONS OVERVIEW

TTICM1000V-600A-AC-Clamp-Meter-Auto-
Ranging-4

LCD SCREEN OVERVIEW

TTICM1000V-600A-AC-Clamp-Meter-Auto-
Ranging-5

OTHER FUNCTIONS
Flash Light A press of this button will turn the flashlight on, helping illuminate the desired workplace. Pressing the light button will turn off the flashlight.

OPERATING INSTRUCTIONS

AC CURRENT MEASUREMENT

  1.  Set function selector to A , open the meter jaw and clamp the conductor to be measured. Be sure the conductor to be measured is clamped at the centre of the clamp.
  2.  Care should be taken to ensure that the clamp is completely closed.
  3.  The current measurement will be shown in the display.

NOTE:
Disconnect test leads when measuring with the clamp.TTICM1000V-600A-AC-
Clamp-Meter-Auto-Ranging-6

AC VOLTAGE MEASUREMENT

  1.  Connect common test lead to the COM terminal, then insert the red test lead to V Ω terminal.
  2.  Set the function selector to the V setting and connect the test lead in parallel with the power or load to be measured.
  3.  The meter will auto-range to display the most appropriate measurement.

NOTE:

  • Do not input voltage higher than AC 750V as damage to the meter may result.
  • After completing all measurement operations, disconnect the test lead from the measured circuit.
  • When measured voltage is higher than the safety voltage 30V AC, the meter displays the high voltage warning prompt ” “
  • When measured voltage is higher than 750V AC, the meter beeps and displays the high voltage warning prompt.TTICM1000V-600A-AC-Clamp-Meter-Auto-Ranging-7

DC VOLTAGE MEASUREMENT

  1.  Connect common test lead to the COM terminal, then insert the red test lead to V Ω terminal.
  2.  Set the function selector to V and connect the test lead in parallel with the power or load to be measured.
  3.  The meter will auto-range to display the most appropriate measurement.

NOTE:

  • Do not input voltage higher than 1000V as damage to the meter may result.
  • When measuring at mV range, relative mode must be used.
  • After completing all measurement operations, disconnect the test leads from the measured circuit.
  • When the measured voltage is higher than the safety voltage 30V DC, the meter displays high voltage the warning prompt ” “
  • When measured voltage is higher than 1000V DC, the meter beeps and displays the high voltage warning prompt.TTICM1000V-600A-AC-Clamp-Meter-Auto-Ranging-8

RESISTANCE MEASUREMENT

  1.  Connect common test lead to the COM terminal, then insert the red test lead to V Ω terminal.
  2.  Set the function selector to the measurement setting Ω, press SELECT until the resistance icon Ω appears on the display.
  3.  Ensure all power is off in the circuit being measured and that all capacitors have been discharged.
  4.  Measure resistance by connecting test leads to circuit.

NOTE:

If the resistance open circuit or resistance being measured exceeds the maximum range of the meter, display will show “OL”.TTICM1000V-600A-AC-Clamp-
Meter-Auto-Ranging-9

CONTINUITY TESTING

  1.  Connect common test lead to the COM terminal, then insert the red test lead to V Ω terminal. Set the function selector to the measurement setting
  2.  Press SELECT until the screen displays
  3.  Ensure all power is off in the circuit being measured and that all capacitors have been discharged.
  4.  Measure continuity by connecting test leads to circuit.
  5.  When the measured resistance between two terminals is <30 Ω, the circuit should be deemed conductive, the buzzer will sound continuously.TTICM1000V-600A-AC-Clamp-Meter-Auto-Ranging-10

DIODE TESTING

  1.  Connect common test lead to the COM terminal, then insert the red test lead to V Ω terminal. Polarity of red and black test leads is respectively “+” and “-”.
  2.  Set the function selector to the measurement setting
  3.  Press SELECT until the screen displays
  4.  Ensure all power is off in the circuit being measured and that all capacitors have been discharged.
  5.  Touch test leads to diode. A reading of 500-800mV on the display indicates forward bias, “OL” indicates reverse bias. An open device will show “OL” in both polarities and a shorted device will show approximately 0mV.TTICM1000V-600A-AC-Clamp-Meter-Auto-Ranging-11

CAPACITANCE MEASUREMENT

  1.  Connect common test lead to the COM terminal, then insert the red test lead to V Ω terminal. Measure by REL mode.
  2.  Push the range switch to the measurement setting
  3. Press SELECT to select , connect test leads in parallel with the two terminals of  measured capacitance.
  4.  Ensure all power is off in the circuit being measured and that all capacitors have been discharged.
  5.  Touch test leads to the capacitorTTICM1000V-600A-AC-Clamp-Meter-Auto-Ranging-12

NON-CONTACT AC VOLTAGE DETECTING (NCV)
In order to sense whether there is AC voltage or electromagnetic field in this space, place the front end of the clamp meter near the conductor to be tested (distance <10mm).
When an electric field voltage is detected (100vRMS), the meter will beep with a red light above the NCV dial.TTICM1000V-600A-AC-Clamp-Meter-Auto-
Ranging-13

TECHNICAL INDEX

ENVIRONMENTAL CONSTRAINT

WORKING HEIGHT Indoor altitude 2000m
POLLUTION GRADE 2

OPERATING HUMIDITY & TEMPERATURE

| 0°C~30 °C (not bigger than 80%RH) 30°C~40°C (not bigger than 75%RH) 40°C~50°C (not bigger than 45%RH)
STORAGE HUMIDITY & TEMPERATURE| -20°C~60°C (not bigger than 80%RH)

ELECTRICAL SPECIFICATION AC

RANGE RESOLUTION ACCURACY OVERLOAD PROTECTION
6A 0.001A ±(2.5%+30) 600A

AC VOLTAGE MEASUREMENT

RANGE RESOLUTION ACCURACY OVERLOAD PROTECTION
6V 0.001V ****

±(1.2%+5)

| ****


1000V DC 750V AC

60V| 0.01V
600V| 0.1V
750V| 1V| ±(1.5%+5)

  • Display: Voltage true value; applicable for range 10% to 100%
  • Voltage: Input impedance > 10 MΩ
  •  Frequency response: 40~400Hz

DC VOLTAGE MEASUREMENT

RANGE RESOLUTION ACCURACY OVERLOAD PROTECTION
600mV 0.1mV ±(1.0%+8) ****


1000V DC

750V AC

6V| 0.001V| ±(0.8%+1)
60V| 0.01V| ****

±(0.8%+3)

600V| 0.1V
1000V| 1V| ±(1.0%+3)

RESISTANCE MEASUREMENT

RANGE RESOLUTION ACCURACY OVERLOAD PROTECTION
600Ω 0.1Ω ±(1.2%+2) ****


1000V DC

750 V AC

6kΩ| 0.001kΩ| ****


±(1.0%+2)

60kΩ| 0.01kΩ
600kΩ| 0.1kΩ
6MΩ| 0.001MΩ| ±(1.2%+2)
60MΩ| 0.01MΩ| ±(1.5%+5)

CAPACITANCE MEASUREMENT

RANGE RESOLUTION ACCURACY OVERLOAD PROTECTION
99.99nF 0.01nF ±(4.0%+25) ****



1000V DC

750V AC

999.99nF| 0.1nF| ****


±(4.0%+5)

9.999nF| 0.001nF
99.99nF| 0.01nF
999.9nF| 0.1nF|
9.999nF| 0.001nF| ±(10%)
59.99nF| 0.01nF| For reference only

NCV MEASUREMENT

RANGE ACCURACY
NCV >100Vrms ; <10mm (LED/Buzzer indication)

CONTINUITY MEASUREMENT

RANGE RESOLUTION ACCURACY OVERLOAD PROTECTION

600Ω

| ****

0.1Ω

| Buzzer beeps when

<30Ω

| ****

1000V DC

750V AC

Open-circuit voltage is about 1.2V

DIODE MEASUREMENT

RANGE RESOLUTION OVERLOAD PROTECTION



6V

| ****



0.001V

| ****


1000V DC

750V AC

POWERING OFF
To power off the clamp meter, move the function selector to the off position. Power off the device after each use to conserve battery.
The clamp meter will auto power off if not used for 15min.
To wake up the clamp meter, press any button or move the function dial.

MAINTENANCE

  • Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
  • Turn off the power of the meter when it is not in use.
  • Remove the battery when the unit won’t be in use for extended periods.
  • Do not use or store the meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the meter may deteriorate after dampened.

INSTALLATION OR REPLACEMENT OF BATTERY
3 x AAA 1.5V batteries will provide power for this product. Replace the battery as follows:

  1.  Shut down the clamp meter, unplug all any leads from the device
  2.  Remove the screw from the battery cover on the rear of the device.
  3.  Replace the batteries currently installed and insert new AAA batteries noting polarity.
  4.  Replace the battery cover and re-install the battery cover screw.

WARRANTY INFORMATION

This warranty is provided by Total Tools (Importing) Pty Ltd of 20 Thackray Road, Port Melbourne VIC 3207. Phone: 03 9261 1900 (we, us, our).

Express Warranty
Subject to the exclusions set out below, we warrant that this product will be free from defects in materials or workmanship for 12 months from the date of purchase. The benefits conferred by this warranty are in addition to all rights and remedies which you may be entitled to under the Australian Consumer Law, and any other statutory rights you may have under other applicable laws. This warranty does not exclude, restrict or modify any such rights or remedies.

Warranty exclusions
This express warranty does not apply where a defect or other issue with the product is caused by normal wear and tear, misuse or abuse of the product.

Consumer guarantees
Our goods come with guarantees that cannot be excluded under the Australian Consumer Law. You are entitled to a replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage. You are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable quality and the failure does not amount to a major failure.

Warranty claims
To make a claim under this warranty, you must bring the product along with the proof of purchase and any other documentary evidence which you think is relevant to the Total Tools’ place of purchase where the claim will be handled on our behalf. Any cost incurred by you in bringing the product to the place of purchase will be borne by you. To make a claim under this warranty, the product and proof of purchase must be returned to the Total Tools place of purchase during the warranty period specified above. If your warranty claim is accepted, we (or the Total Tools store that handles the claim on our behalf) will, at our discretion, repair or replace the product, or refund money to you and take back the product.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals