INTRODUCTION
This instrument is a digital thermometer for use with any K-type thermocouple as a temperature sensor. Temperature indication follows the National Bureau of Standards and IEC584 temperature/voltage table for K-type thermocouples.
IMPORTANT INFORMATION
Date and Time must be set correctly for the data logger function to work.
SPECIFICATIONS
NUMERICAL DISPLAY: 4 digital liquid crystal display
MEASUREMENT RANGE:
-200°C ~ 1370°C -328°F ~ 2498°F
RESOLUTION:
-200°C~ 200°C 0.1°C; 800°C ~1370°C 1°C
-200°F~ 200°F 0.1°F; else 1°F
MAXIMUM VOLTAGE AT THERMOCOUPLE INPUT:
60V DC, or 24Vrms AC
ENVIRONMENTAL:
- Operating Temperature and Humidity: 0°C ~50°C (32°F ~ 122°F); 0 ~ 80% RH
- Storage Temperature and Humidity: -10°C to 60°C (14°F ~ 140°F); 0 ~ 80% RH
- Altitude up to 2000 meters.
SPECIFICATIONS
ACCURACY: AT (23 ± 5°C)
Range | Accuracy |
---|---|
-200°C — 200°C | ±(0.3% reading + 1°C) |
200°C — 400°C | ±(0.5% reading + 1°C) |
400°C-1370°C | ±(0.3% reading + 1°C) |
-328°F — -400°F | ±(0.5% reading + 2°F) |
-200°F — 200°F | ±(0.3% reading + 2°F) |
200°F — 400°F | ±(0.5% reading + 2°F) |
400°F — 2498°F | ±(0.3% reading + 2°F) |
For T1-T2 Measurement, the accuracy is ±( 0.5% T1-T2 reading + 2°C) or ±( 0.5% T1-T2 reading + 2°F)
TEMPERATURE COEFFICIENT:
For ambient temperatures from 0°C ~ 18°C and 28°C ~ 50°C, for each °C ambient
below 18°C or above 28°C add the following tolerance into the accuracy spec.
0.01% of reading + 0.03°C ( 0.01% of reading + 0.06°F )
NOTE:
The basic accuracy Specification does not include the error of the probe
please refer to the probe accuracy specification for additional details.
SAMPLE RATE : 0.6 times per second
DIMENSION: 184×64×30mm
WEIGHT: 210g Approx.(7.4oz)
ACCESSORY: K Type Bead Probe, Battery, Carrying Case, Instruction Menu.
OPTION: Soft Ware Package ( Program, RS232 Connection Cable) , AC
Adapter.
POWER REQUIREMENT : 9 Volt Battery, NEDA 1604 or JIS 006P or IEC6F22
BATTERY LIFE : Approx. 100hrs with alkaline battery
AC ADAPTER : 9VDC ±15% 100mA ; Plug Diameter: 3.5×1.35mm
SYMBOL DEFINITION AND BUTTON LOCATION
– : This indicates that the minus temperature is sensed.
°C °F : Centigrade and Fahrenheit indication.
K : Thermocouple Type Indication
HOLD : This indicates that the display data is being hold.
MAX : The Maximum value is now being displayed.
MIN : The Minimum value is now being displayed.
AVG : The Average value is now being displayed.
: The reading is now under Relative Mode.
: The Battery is not sufficient for proper operation.
- T1 K type temperature sensor connector
- T2 K type temperature sensor connector
- LCD display
- ON/OFF button
- HOLD button
- T1,T2,T1-T2 control button
- Relative readout button
- MAX MIN Average control button
- °C, °F control button
- Offset calibration screw
- Digital output connector
- AC power adapter connector
- Tripod connector
- Battery cabinet cover
OPERATION INSTRUCTIONS
4.1 POWER-UP
Press the key to turn the thermometer On or OFF.
4.2 CONNECTING THE THERMOCOUPLES
For measurement, plug the thermocouple into the input connectors.
4.3 SELECTING THE TEMPERATURE SCALE
When the meter was first powered on, the default scale setting is set at
Celsius (°C) scale. The user may change it to Fahrenheit (°F) by pressing “
°C/°F ” button and vice versa to Celsius.
4.4 DATA-HOLD OPERATION
The user may hold the present reading and keep it on the display by pressing
the “HOLD” button. When the held data is no longer needed, one may release the
data-hold operation by pressing “HOLD” button again. When the meter is under
Data Hold operation, the “▲REL” , and “ °C/°F ” button are disabled.
4.5 T1,T2,T1-T2 DISPLAY CONTROL
One may select T1,T2 or T1-T2 to show on the main display by pressing
button. When T1 or T2 is select to show on the main display, the other
temperature will be shown on the second display. When one select T1-T2 to show
on the main display, T1 and T2 will be shown on the second display
alternately.
4.6 RELATIVE OPERATION FOR MAIN DISPLAY
When one press the “▲REL” button, the meter will memorize the present reading
and the difierence between the new reading and the memorized data will be
shown on the display. Press the “▲REL” button again to exit the Relative
operation.
4.7 MAX/MIN/AVG OPERATION FOR MAIN DISPLAY
When one presses the button the meter will enter the MAX/MIN mode. Under this
mode the maximum value, minimum value and average value of latest 8 data is
kept in the memory simultaneously and updated with every new data.
When the MAX symbol is display, the Maximum is shown on the display.
Press again, then the MIN symbol is on the display and also the minimum
reading.
Press again, the AVG symbol is on the display and also the average reading.
Press again, MAX, MIN and AVG will blink together. This means that all these
data is updated in the memory and the reading is the present temperature.
One may press to circulate the display mode among these options.
When the meter is under operation, “▲REL” and “ °C/°F ” are disabled.
To exit the MAX/MIN mode, one may press and hold for two seconds.
4.8 AUTO POWER OFF
By default, when the meter is powered on, it is under auto power off mode. The
meter will power itself off after 30 minutes if no key operation or RS232
communication. Key combination at power on or RS232 communication can disable
auto power off. One may press and hold “HOLD” button and then power on the
meter and there will be two successive beeps to indicate that auto power off
is disabled.
4.9 LOW BATTERY CONDITION
When the battery voltage is under proper operation requirement, the symbol
will show on the LCD and the battery need to be replaced with new one.
4.10 CALIBRATION POINT
Room Temperature 23 ± 3°C
input | Adjust VR | Tolerance |
---|---|---|
0 °C | VR1 | ± 0.1 °C |
190 °C | VR2 | ± 0.1 °C |
1 000 °C | VR3 | + 1 °C |
1900 °F | VR4 | + 1 °F |
Normally, performing offset Calibration with thermal stabled ice water through VR1 will give a very good calibration result.
4.11 DIGITAL OUTPUT
The Digital Output is a 9600bps N 81 serial interface.
The RX is a 5V normal high input port.
The TX is a 5V normal high output port.
The command of Digital Output is list below.
RS232 command
| Function|
Remarks
---|---|---
K(ASC 4BH)| Ask for model No.| Send 4 bytes
D(ASC 44H)| Ask for main display Range, Data, Unit| Send 22 bytes
B(ASC 42H)| Ask for secondary display Range, Data, Unit| Send 22 bytes
S(ASH 53H)| Ask Status| Send 13 bytes
H(ASC 48H)| Hold button|
T(ASC 54H)| TIMER button|
M(ASC 4DH)| AVG/MAX/MIN button|
N(ASC 4EH)| Exit AVG/MAX/MIN mode|
R(ASC 52H)| REL button|
C(ASC 43H)| C/F button|
A(ASC 41 H)| Inquire all encoded data| Send encoded 8 byte
• COMMAND K
Return 4 bytes. For example, when sends command “K” to meter, it will return
“3”,”0″,”1″, ASCII(13) .
• COMMAND D
Return data of main window.
Range: T1, T2, T1-T2 (7bytes), the unused characters is left as space ASC(13).
Data: ±9999.9,-OL,OL ( 7bytes include polarity and decimal point),
Unit: C,F(5bytes)
When the meter receive the D command, it will send :
Range□Data□Unit.
( where □ represent space (ASC(20H) )
For example:
T1□□□□□□-199.9□□C□□□□
(0x13) represent T1,- 199.9°C, The total byte number should be
7+1+7+1+5+chr(13)=22Bytes
• COMMAND B
Return the operation mode HOLD□MAX□REL, if the mode is not entered, the
related characters will be left as space.
For example: when the meter is under MAX display, the meter will return: □□□□□
MAX□□□□
• COMMAND T
Equivalent to one pushing on the HOLD button.
• COMMAND M
Equivalent to one pushing on the HOLD AVG/MAX/MIN button and no message is
returned.
• COMMAND R
Equivalent to one pushing on the REL button and no message is returned.
• COMMAND C
Equivalent to one pushing on the C/F button and no message is returned.
• COMMAND A
1nd BYTE:
The first byte is the start byte , it value is 2.
2nd BYTE:
bit7 | bit6 | bit5 | bit4 | bit3 | bit2 | bitl | bit0 |
---|---|---|---|---|---|---|---|
C/F | Low Bat | Hold | REL | K/J | MAX/AVG/MIN |
bit 2 bit 1 bit0
0 0 0 → normal mode
0 0 1 → MAXIMUM mode
0 1 0 → MINIMUM mode
1 0 0 → AVG mode
1 1 1 → calculate MAX/MIN/AVG in back-ground and lcd “MAX””AVG””MIN” will
flash
bit3:1→0→K TYPE 1→J TYPE
bit4:1→REL
bit5:1- HOLD 0→not HOLD
bit6:1→LOW BATTERY 0→BATTERY NORMAL
bit7:1→C 0→F
3nd BYTE:
bit7 | bit6 | bit5 | bit4 | bit3 | bit2 | bitl | bit() |
---|---|---|---|---|---|---|---|
Point | Minus | OL | Point | Minus | OL |
bit0:1→main window value is OL 0→not OL
bit1:1→main window value is minus, 0→main window value is plus.
bit2:1→4th byte and 5th byte represent #### 0→4th byte and 5th byte represent
.#
bit3:1→sub window value is OL 0→not OL
bit4:1→sub window value is minus, 0→sub window value is plus.
bit5:1→6th byte and 7th byte represent #### 0→6th byte and 7th byte represent
.
bit7 bit6:
00→Main window is T1-T2,sub window is T1
01→Main window is T1-T2, sub window is T2
10→Main window is T1, sub window is T2
11→Main window is T2, sub window is T1
4th BYTE:
rst two BCD code of main window value.
5th BYTE:
last two BCD code of main window value
6th BYTE:
rst two BCD code of sub window value.
7th BYTE:
last two BCD code of sub window value.
8th BYTE:
The last byte is the end byte , it value is 3, First and last byte are used to
check frame error
DOWNLOAD THE PROTOCOL OF SERIAL INTERFACE
HERE:https://cdn.shopify.com/s/files/1/1291/1589/files/TC0301_Protocol.pdf?v=1614323438
4.12 SETUP SOFTWARE
OPERATING SYSTEM REQUIRED: WINDOWS XP/ VISTA/ WINDOWS 7/ WINDOWS 8/ WINDOWS
10 MINIMUM HARDWARE REQUIRED:
- At least 50 MB hard disk space available to install the software
- Download software here: https://perfectprime.com/pages/manual ,locate your product under Thermocouple Thermometer and download the software
TUTORIAL – QUICK START TO USE SOFTWARE:
- Turn on the device
- Load the software and it should be connected automatically with the device. You can confirm by seeing values on the image of the thermocouple thermometer in the newly opened window, representing the temperature it is detecting.
- For live temperature reading click the “Start graphing real time data” button on the tool bar”
- To save live data, click stop and click on the graph and click save
CUSTOMER SERVICE INQUIRIES
Your emails are important to us so we strive to reply all inquiries and emails
within 24 hours. In exceptional cases, we may require more time to respond.
Thank you for your understanding.
For more information about our products and services, please send us an email:
cs@perfectprime.com
For B2B or project-based application, please send an email:
sales@perfectprime.com
FOR MORE INFORMATION ABOUT PERFECTPRIME PLEASE VISIT OUR ABOUT US PAGE AND
FEEL FREE TO BROWSE.
http://qr.w69b.com/g/r4oLxa2l2
|
http://qr.w69b.com/g/qRw248shO
|
---|---|---
Retailer Tyche Smart Limited
Email cs@perfectprime.com
2nd Floor, 107 Charterhouse Street, EC1M 6HW, London.
Address
England UNITED KINGDOM
Telephone +44 203 7695377
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>