WT326A Infrared Thewintact rmometer Instruction Manual

June 5, 2024
wintact

Infrared thermometer
**Instruction manual

MODEL: WT326B/WT326C/WT326D/WT326A
Standaed: Q/HTY 004-2017
Version: WT326-EN-00**

Introduction

This infrared thermometer is used for measuring the temperature of the object’s surface, which is applicable for various hot, hazardous or hard- toreach objects without contact safely and quickly.
This unit consists of Optics, Temperature Sensor Signal amplifier, Processing circuit, and LCD Display.
The Optics collected the infrared energy emitted by an object and focus on the Sensor. Then the sensor translates the energy into an electrical signal. This signal will be turned out to be digital and shown on the LCD after the signal amplifier and processing circuit.

Warning & Cautions

  1. Warning:
    To avoid the potential situation that may cause harm or damage to people, please pay attention to the following items:
    1). Do not point the laser directly at the eye or indirectly off reflective surfaces.
    2). The unit cannot measure through transparent surfaces such as glass or plastic. It will measure the surface temperature of these materials instead.
    3). Steam, dust, smoke, or other particles can prevent accurate measurement by obstructing the unit’s optics.

  2. Cautions:
    The infrared thermometer should be protected for the following:
    1). EMF (electro-magnetic fields) from arc welders and induction heaters.
    2). Thermal shock (caused by large or abrupt ambient temperature changes allow 30 minutes for the unit to stabilize before use).
    3). Do not leave the unit on or near objects of high temperature.

Distance to spot size

WT326A Infrared Thewintact rmometer - Figure 1

  1. When taking measurements, pay attention to the Distance to Spot Size. As the Distance (D) from the target surface increases, the spot size (S) of the area measured by the unit becomes larger.
    The Distance to a Spot size of the unit is 12:1. (Figure 1)

  2. Field of view:
    Make sure the target is larger than the unit’s spot size.
    The smaller the target the closer the measure distance.
    When accuracy is critical, make sure the target is at least twice as large as the spot size.

Emissivity
Emissivity: Most organic materials and painted or oxidized surfaces have an emissivity of 0.95(preset in the unit).
Inaccurate readings will result from measuring shiny or polished metal surfaces.
To compensate for this, adjust the unit’s emissivity reading or cover the surface to be measured with masking tape or flat black paint.
Measure the tape or painted surface when the tape or painted reaches the same temperature as the material underneath.

Material

| Emissivity| Material|

Emissivity

---|---|---|---
Aluminum| 0.30| Iron| 0.70
Asbestos| 0.95| Lead| 0.50
Asphalt| 0.95| Limestone| 0.98
Basalt| 0.70| Oil| 0.94
Brass| 0.50| Paint| 0.93
Brick| 0.90| Paper| 0.95
Carbon| 0.85| Plastic| 0.95
Ceramic| 0.95| Rubber| 0.95
Concrete| 0.95| Sand| 0.90
Copper| 0.95| Skin| 0.98
Dirt| 0.94| Snow| 0.90
Frozen food| 0.90| Steel| 0.80
Hot food| 0.93| Textiles| 0.94
Glass(plate)| 0.85| Water| 0.93
Ice| 0.98| Wood| 0.94

Operation

  1. Operating the unit:
    1). Open the battery door and load two 1.5V AAA batteries.
    2). Pull the trigger to turn on the unit.
    3). Aim at the target surface and pull the trigger, then the temperature will be shown on the LCD.
    This unit is equipped with a laser, which is only used for aiming.

  2. Locating a Hot Spot:
    To find a hot spot, aim the thermometer outside of interest, then scan across with an up and down motion until you locate the hot spot. (Figure 2).

WT326A Infrared Thewintact rmometer - Figure 2

LCD display & buttons

  1. LCD: as shown in Figure 3.
    a.Temperature reading
    b.Temperature units
    c.Turning on state indicator of positioning laser
    d.Backlight on an icon
    e.Low battery indicator
    f.Environmental temperature icon
    g.Data reading indicator
    h.Data holding the indicator
    i.emissivity icon
    j.self-calibration icon
    k.minimum icon
    l.maximum icon
    m. Low-temperature alarm icon
    n. High-temperature alarm icon
    WT326A Infrared Thewintact rmometer - Figure 3

  2. Names and functions of parts: as shown in Figure 4

WT326A Infrared Thewintact rmometer - Figure 4

  1. Trigger: press it to display temperature value with SCAN appearing in the meantime. Release the trigger and enter into HOLD mode to save the data automatically, and the unit turns off automatically if there is no further operation. Built-in 7 sec auto power-off function.

  2. Switch key between Celcius degree and Fahrenheit degree. This key can also be used for increasing the value set.

  3. Mode switch key: press the Mode key to switch modes in turn among MAX→MIN→AT→EMS→CAL→Hi-Low→MEASURING INTERFACE
    a. MAX: measuring maximum temperature
    b. MIN: measuring minimum temperature
    Note: In measuring, hold on to the Mode key to switch to Max or Min review.
    c. AT: current environment temperature
    d.EMS: emissivity that can be set between 0.10 and 1.00 with the ℃/℉ key and laser locating key
    e.CAL: Under self-calibration mode, calibrate the unit between -5.0℃ and +5.0℃
    f. Hi and Low: high-temperature alarm and Low-temperature alarm
    ①Switch to Hi or Low mode and adjust the setting of high or Low alarm points by using the key of ℃/℉ and the positioning laser key;
    ②When the measured temperature is greater than or less than the setpoint, the corresponding Hi symbol or Low symbol is displayed, and the “BI BI” sound is issued.

  4. Positioning laser switch: press the positioning switch to close and open the positioning laser. In the Settings of EMS, CAL, and high and low-temperature alarm points, it can be used as a down-regulation function and the value can be rapidly reduced by long press.

  5. Battery door

  6. Infrared temperature sensing area

  7. Laser (assisted positioning)

Maintenence

  1. Lens Cleaning:
    Blow off loose particles using clean compressed air.
    Gently brush the remaining debris away with a moist cotton swab. The swab may be moistened with water.

  2. Case cleaning:
    Clean the case with a damp sponge/cloth and mild soap.
    Note:

  1. Do not use solvent to clean the plastic lenses.
  2. Do not submerge the unit in water.

Specification

Temperature range| WT326A : -50~400℃(-58~752℉) WT326B : -50~600℃(-58~1112℉) WT326C : -50~800℃(-58~1472℉) WT326D : -50~950℃(-58~1742℉)
---|---
Temperature measurement error| <0℃(32℉): ±3℃(±5℉)

0℃(32℉): ±1.5℃(±2.7℉) or±1.5%,
Whichever is greater
Repeatability| 1% rdg or 1℃ whichever is greater
Response time| 0.5s, 95% Response
Emissivity| 0.10~1.00 Adjustable(preset as 0.95)
D:S| 12:1
Response wavelength| 5μm~14μm
No operation shutdown| About 7 seconds
Batteries| 1.5V AAAbattery*2(No.7 battery)
Low power indicator| Low power indicator for power below 2.5V
Overload indicator| “Hi”/“Lo” displayed on the LCD
Exceed upper/lower limits of work environment| “AH”/“AL” displayed on the LCD
Working environment temperature| 0℃~40℃(32℉~104℉)
Storage temperature| -10℃~60℃(14℉~140℉)

Specific Declarations:
Our company shall hold no responsibility resulting from using output from this product as direct or indirect evidence.
We reserve the right to modify product design and specifications without notice.