cardio ATX2410V4P1 remote patient monitor Instructions
- June 5, 2024
- cardio
Table of Contents
INSTRUCTIONS FOR USE
[kar-dee-o] (remote patient monitor)
CARDI/O® for touchless, no-wearable use
Device Description The Cardi/o ® Sensor device is a component of the Cardi/o ® remote patient monitoring (RPM) platform. The Cardi/o Sensor device is a wireless, no-touch/no-wearable vital signs sensor, set on a desktop or mounted to a wall to record heart rate, respiratory rate, and stress-related to heart rate variability, and motion. The Cardi/o ® Sensor continuously gathers physiological data from the person being monitored and then transmits encrypted data via WiFi to the Cardi/o ® Cloud platform. The encrypted wireless data provided by the Cardi/o ® Sensor may be accessed and viewed via Cardi/o ® App for mobile phones or the Cardi/o ® Central Dashboard web app for desktop computers. The data from the Cardi/o Sensor may be downloaded from the Cardi/o ® Cloud for storage or integrated into a Third-Party Relay Application via the APIs of the Cardi/o ® Software Library. The data provided by the Cardi/o Sensor is intended to aid caregivers in making diagnoses by providing additional information. During normal operation, data is collected via the touchless radar sensing capability of the Cardi/o ® Sensor and transmitted immediately to the Cardi/o ® Cloud via WiFi. A continuous connection is needed between the Cardi/o ® Sensor and the Internet to facilitate continuous physiological data transmission. The continuous wireless transmission of data occurs with an update rate of 2 seconds between data collection and transmission. Patients may opt-in to request assistance from healthcare providers through predefined alarm thresholds that, when exceeded, send alerts via a wall-mounted transceiver that is connected to the cloud.
Indications for Use
The Cardi/o® RPM is a wireless “no-wearable” remote monitoring system intended
for use by healthcare professionals for continuous collection of physiological
data in healthcare settings as well as in environments where healthcare is
administered, such as home healthcare services, assisted living, and skilled
nursing facilities. This physiological data can include heart rate,
respiratory rate, heart rate variability, and motion. The RPM may be used for
synchronous or asynchronous telemedicine consultations.
The Cardi/o®
The sensor is intended for use on general care patients who are 18 years of
age or older who may be susceptible to decubitus ulcers of the skin, such as
the elderly, for use as a general patient monitor, to provide physiological
information. The collection and transmission of data are performed on the
device.
The data from the device is intended for use by healthcare professionals as an
aid to remote diagnosis and treatment while preventing the spread of
communicable diseases such as COVID-19. The device can be used on patients
unable or averse to wearable patient monitors, such as patients living with
dementia or autism. The device is not intended for use on critical care
patients.
The data provided by the Cardi/o® Sensor may also be intended for trained
technicians at a remote site to review the data and determine if they concur
with the analysis made by the algorithm in the Cardi/o® Sensor.
The data collected by the Cardi/o® Sensor is an adjunct to clinical assessment; clinician review of the data should precede any diagnosis or therapeutic intervention.
Contraindications
- The Cardi/o® Sensor is not intended as a stand-alone diagnostic monitor for vital signs, but the data may be applicable for use in diagnosis.
Warnings
- Depending on wireless connectivity, a temporary interruption of data transmission is possible, which may impact continuous or real-time monitoring. Limited data will be stored on the device for transfer once connectivity is reestablished.
- The Cardi/o ® Sensor is designed to monitor one individual at a time. Having more than one individual in the field of view of the sensor may produce skewed data that may not represent the parameters of the intended subject.
- Disturbances to cardiac-respiratory signals include distortions due to foreign interference and noise artifacts that may occur when using the Cardi/o ® Sensor. Do not rely entirely upon the heart rate signals of the Cardi/o ® Sensor. Keep patients under close surveillance.
- Automated interpretation of heart rate variability by the Cardi/o ® Sensor is a valuable tool when used properly. However, no automated interpretation is completely reliable, and a qualified physician shall review the interpretations before treatment, or non-treatment, of any patient.
- Heart rate, respiratory rate, stress level (HRV), and motion data will not be transmitted or stored if the Sensor is not powered.
- The patient must be in the field of view of the Cardi/o ® Sensor to collect physiological data.
- Clinical validation has not been performed on patients who are pregnant or breastfeeding.
Precautions
-
For data to be sent to a healthcare professional for review:
o The Cardi/o ® Sensor must be properly used by the patient.
o The patient must remain within the range of their Sensor device.
o The Cardi/o ® Sensor must have adequate power for data transmission. Notification of interruption in data collection of the Cardi/o ® Sensor will be indicated on the Cardi/o ® App when the Sensor device is offline and may also be sent via SMS Text if this feature is enabled.
o The Sensor device must remain connected to the Internet and functional for data transmission. Wireless connectivity must be active for the transmission of data from the Sensor device to the Cardi/o ®. -
The healthcare professional can be notified when there are interruptions in data connectivity if this feature is enabled.
-
Physiological data collected by the Cardi/o ® Sensor for patients experiencing stress may indicate slightly higher or lower respiratory rate values, compared to visual observation.
-
The Cardi/o ® Sensor is for Individual Use Only. Do not monitor more than one patient at a time.
-
Wireless electronic devices may cause signal interference during data transmission. Avoid close proximity with interfering devices.
-
Do not use the Cardi/o ® Sensor if the package has been opened, or appears used, damaged, or expired.
• Incorrect handling, excessive force, or dropping the Cardi/o ® Sensor may cause malfunction or permanent damage. -
If the Cardi/o Sensor fails to operate, contact your healthcare provider immediately.
-
Dispose of the Cardi/o ® Sensor per local laws, care facility laws, or hospital laws for routine/non-hazardous electronic waste.
Additional Warnings and Precautions for Emergency Use Authorization
(EUA)
The following additional warning and precautions apply when using the Cardi/o®
Sensor under the Emergency Use Authorization for general remote patient
monitoring:
- The Cardi/o ® Remote patient Monitor has neither been cleared nor approved for remote monitoring and detection of vital signs such as heart rate, respiratory rate, and heart rate variability in patients who are undergoing treatment for COVID-19 or who have pre-existing conditions (such as hypertension or heart failure). Such remote monitoring may reduce health care personnel to exposure to SARS-CoV-2, the virus that causes COVID-19;
- The Cardi/o ® Remote patient Monitor has been authorized for the above emergency use by the FDA under a EUA;
- The Cardi/o ® Remote patient Monitor has been authorized only for the duration of the declaration that circumstances exist justifying the authorization of the emergency use of medical devices under section 564(b)(1) of the Act, 21 U.S.C. § 360bbb-3(b)(1), unless the authorization is terminated or revoked sooner.
- WARNING: When interpreting the Cardi/o ® heart rate, respiratory rate, and heart rate variability measurements, data recordings should be reviewed for interference artifacts. Heart rate, respiratory rate and heart rate variability measurements with the Cardi/o ® Sensor may be unreliable in the presence of significant noise or artifacts. Relevant changes in these measurements should be further investigated using a standard 3-lead or 8-lead ECG.
- WARNING: Heart rate, respiratory rate, and heart rate variability measurements with the Cardi/o ® Sensor may be unreliable in cases of significant motion. Consider taking physiological measurements when the patient is at rest.
- The accuracy of the Cardi/o ® Remote Patient Monitor for measuring heart rate, respiratory rate, heart rate variability, and motion has only been tested for accuracy over 30-minute intervals.
- Using the Cardi/o ® Remote Patient Monitor to measure heart rate, respiratory rate, heart rate variability and motion has only been tested with the recommended Sensor placements. The accuracy of this measurement with nonstandard Sensor placement is unknown.
Product Storage
- Storage temperature range: (-15) – 85°C
- Storage relative humidity range: 10 – 95% RH
System Interoperability
The Cardi/o® Remote Patient Monitor is compatible with Sensor devices and software developed with the Cardi/o® Application Programming Interface (API). Please contact Advanced Telesensors Inc. to obtain implementation information, including the Cardi/o® Application Programming Interface (API) Instructions.
Components Included with the Cardi/o Remote Patient Monitor System
The Cardi/o® Remote Patient Monitor System contains:
-
The Cardi/o ® Sensor Device – a 3 x 3-inch device (desktop or wall-mounted) that detects and transmits the vital signs measured by the device, presenting this data locally or remotely to the patient and to healthcare personnel. The Cardi/o ® device integrates remote radar cardio-pulmonary sensing technology and Wi-Fi data connection to collect physiological data. The patient keeps the Cardi/o ® device on a desktop during remote telemedicine consultations or mounted in rooms where continuous remote monitoring is to occur, allowing continuous upload of vital signs data to the Cardi/o ® Cloud platform while the patient remains within sensing range;
o The Sensor device is designed to record physiological data of the patient sitting, lying prone or standing within the Sensor’s 10-foot (3 meters) range and 30º field of view. -
Cardi/o ® App Application – a software Graphical User Interface on mobile phones and tablets intended for use by HCP to display physiological data collected by the Cardi/o ® Sensor device;
-
The Cardi/o ® Central Dashboard Application – a web application that runs in a computer web browser or mobile device. This allows healthcare personnel to monitor patients from multiple Internet-connected devices in a single user interface;
-
Cardi/o ® Cloud – stores data uploaded by the Cardi/o ® Sensor and makes it available to cloud-enabled monitoring and data analytics applications, such as the Cardi/o ® Central Dashboard application.
CARDI/O® USE INSTRUCTIONS
Note: It is recommended that healthcare providers advise users on the correct placement of the Cardi/o® Sensor. To collect data the Cardi/o® Sensor must be connected to the Internet.
Cardi/o ® Overview
Cardi/o ® Remote Patient Monitor (RMP) may be used for synchronous telehealth
consultations where real-time intermittent spot-checking of vital signs may be
desired, and for asynchronous telehealth consultations where historical data
can be viewed and downloaded. Follow Steps 1 through 2.B to configure the
Sensor device for WiFi connection and proper setup/positioning of the Sensor
device for usage. Follow Steps 3 through 5 to access patient data via the
Cardi/o ® mobile app and Cardi/o ® Central Dashboard web app.
Step 1. Connect the Cardi/o® Sensor to the Internet via WiFi
Scan the barcode containing a unique ID that is located on the sensor device
with a smartphone, then input the local WiFi passcode to complete the Sensor
connection.
Determine whether the Cardi/o® RMP will be for intermittent use or long-term
use and position the Sensor device according to Figure A or Figure B below.
Step 2.A Positioning Cardi/o ® for Intermittent Use – tabletop option
Note: Orientation of the Logo Side is important when positioning the
device to collect data on the patient.
See Figure A below for a view of the Cardi/o ® Sensor set on a tabletop,
showing the logo side in relation to the patient.
Figure A: Use Cardi/o® for sterile, no-contact vital signs capture. The data will automatically upload to the Cardi/o® patient record.
Step 2.B Positioning Cardi/o® for Long Term Use – wall-mount option
See Figure B below for a view of the Cardi/o® Sensor wall placement in
relation to the patient. The sensor is placed between 6 to 8 feet vertical
from the floor facing the bed where the patient spends most of their time at
rest. The Sensor device is designed to record physiological data of the
patient sitting, lying prone or standing within the Sensor’s 10-foot range and
30º field of view.
Figure B: Position the Cardi/o Sensor device on a wall facing the bed for nighttime monitoring. Or, determine the best position for daytime monitoring using the 30° field of view reference point. Call your healthcare provider or Cardi/o support for assistance with proper placement and installation.
In Figure C below, follow steps 1 through 4 to install the Sensor on a wall for long-term patient monitoring.
Figure C: The dual-mount wall bracket can be used in a corner or flat wall for optimal positioning.
Step 3. Accessing Patient Data
See Figure D below for five (5) steps to access patient data collected by the Cardi/o® remote monitor system.
Figure D represents the top-level navigation screens. After login, navigate between the Patients list, Stats, and Settings.
Note:
Step 1: “Download Phone App is performed once
Step 2: “Login to the app may accept biometric fingerprint login if the
feature is enabled on the phone.
Step 3: Displays the scrollable list of patients who have been prescribed
Remote Patient Monitoring and their respective vital signs.
Step 4: Click on a bar to access individual patient data and histogram
information.
Step 5: Displays a menu of settings for viewing preference and app
support.
See Figure E below for five (5) primary steps to navigate patient data collected by the Cardi/o® Sensor.
Figure E: Investigate patient data by clicking the respective bars. Access historical data trends for analysis by clicking the down arrow located on the bar of each parameter.
Note:
Step 1: Displays a patient dashboard
Step 2: Click the heart rate bar to view a graph indicating data trending
over time.
Step 3: Click the respiratory rate bar to view a graph indicating data
trending over time.
Step 4: Click the stress bar to view a graph indicating data trending
over time.
Step 5: Click the notes icon [ ] in the main patient dashboard (seen in
Step 1) to type notes pertaining to patient data or consultation.
Step 4. Managing Patient Alerts See Figure F to the right to locate
alerts on the main Residents Dashboard screen of the Cardi/o ® App. The number
on the bell indicates the number of patients that have exceeded one or more
predefined thresholds related to heart rate, respiratory rate, and stress
level. Clear alerts by swiping left on each highlighted bar on the Patient
list and clicking “clear alert”. You may also enable the mobile phone SMS Text
Alerts feature in the Cardi/o ® Central Dashboard administrator portal by
assigning telephone numbers to respective caregivers.
Figure F: The number of active alerts is indicated in the red dot.
Personalize Alarm Thresholds via the Cardi/o® Central Dashboard Administrator Portal Adjust the high and low thresholds of the alarm parameters in the Cardi/o ® Central Dashboard administrator portal to personalize alert settings for individual patients. Table 1 below indicates the default values for the high and low thresholds for each parameter.
PARAMETERS | Range | Low Threshold | High Threshold |
---|---|---|---|
Heart Rate | 20 ̶ 180 BPM | 50 | 100 |
Respiratory Rate | 0 ̶ 60 bpm | 9 | 29 |
Stress Score (HRV-related) | 0 ̶ 10 score | 0 | 5 |
Table 1: Default thresholds for Alarm settings are indicated above, and may be adjusted according to guidance by the healthcare professional
Step 5. Downloading and Reporting Data
Note: Recorded patient data is intended to automatically upload data via
the cloud to the unique patient Cardi/o ® record on a continuing basis.
Download data via the Cardi/o ® Central Dashboard web application installed on
a desktop computer. Select date ranges and time periods to view data trends
and values on a pdf report. Email the report to an email you designate; set
reporting for as many times a day or week as needed for treatments. Refer to
the application guide for the Cardi/o ® Central Dashboard web application to
navigate reporting features.
Cardi/o ® Sensor Return or Disposal
Follow instructions given by your healthcare provider to return the Cardi/o ® sensor. If directed to dispose of the Sensor, please observe local laws for the disposal of electronic products.
Note: The Cardi/o® Sensor is a re-usable device. Use standard sterilization practices before re-issuing the device to another user. The device will need to be reconfigured for WiFi connectivity to the Internet if reissued to a user at a different location.
Troubleshooting
For issues related to a user interface application and for additional troubleshooting guidance, refer to separate Instructions for Use for the interface.
Contact Information
Advanced Telesensors, Inc.
3815 Jarrett Way, Suite A100
Austin, TX 78728 USA
Phone: 888-292-2208
www.cardio.io
Customer Support
support@cardio.io
888-292-2208, option 4
Product Specifications
Measurements| Heart Rate Dynamic Range| Not available for patients with
pacemakers. 20 – 180 Beats Per Minute (BPM) Accuracy tested +/- 2 BPM of
3-lead ECG
---|---|---
Respiration Rate Dynamic Range| Not available for patients with pacemakers. 0
– 60 breaths per minute (bpm) Accuracy tested +/- 2 BPM of 3-lead ECG
Stress – Heart Rate Variability (stationary and ambulatory)| Not available for
patients with pacemakers.
Ratio of Low Frequency to High Frequency (LF/HF); also known as “HRV Score”
Dynamic sensing range| 10-feet (3-meter) max separation between subject and
Cardi/o® Sensor
o 2-inch (5 cm) minimum separation
Communications| WiFi WLAN Standard| IEEE 802.11b/g/n, Wi-Fi compliant
Serial WiFi (eS-WiFiTM)| ISM43362-M3G-L44
Radio Frequency Range| 24.0 – 24.497 GHz ISM
Modulation| 802.11 g/n : OFDM /64-QAM,16-QAM, QPSK, BPSK
802.11b : CCK, DQPSK, DBPSK
Max Power Output| 802.11n /72Mbps: 18* dBm ± 1.5 dB
Security| Wi-Fi authentication WEP-128, WPA-PSK (TKIP), WPA2-PSK.
Power| AC Adapter Input| 110V – 240V – 50/60 Hz 0.4A max
Output| 5.9V 2.0A max
Operating Conditions| Humidity| 10 – 95% RH (noncondensing)
Altitude| <3000 m
Barometric Pressure| 70 kPa to 102 kPa
Ambient Temperature| 0 – 70°C
Electromagnetic Emission Declaration
The Cardi/o® Sensor is intended for use in the electromagnetic environment
specified below. The end-user of the device should assure that it is used in
such an environment.
Emission test | Compliance | Electromagnetic environment |
---|---|---|
RF emissions CISPR 11 | Group 1 | The Cardi/o® Sensor uses RF energy only for |
its internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The Cardi/o® Sensor is suitable for use in all
establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
FCC Compliance (FCC ID: To be determined)
The Cardi/o ® Sensor complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation (FCC Title 47, Subpart C, Part 15.249). Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment (FCC Title 47, Subpart A, Part 15.21)
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures (FCC Title 47, Subpart B, Part 15.105(b)):
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Guidance and Declaration – Electromagnetic Immunity
(For ME equipment ME systems that is not life-supporting)
The Cardi/o® The sensor is intended for use in the electromagnetic environment specified below. The end-user of the device should assure that it is used in such an environment.
Immunity test| Compliance level| Electromagnetic environment-
guidance
---|---|---
Electrostatic discharge (ESD) IEC 61000-4-2| ± 8 kV contact ± 15 kV air|
Floors should be wood, concrete, or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Radiated RF [24.0 — 24.25 GHz] IEC 61000-4-3| 5 mW/cm2 (uncontrolled) )
1 mW/cm2 (uncontrolled) )| Portable and mobile RF communications equipment
should be used no closer to any part of the Cardi/o® Sensor device than the
recommended separation distance calculated from the equation applicable to the
frequency of the transmitter. Recommended separation distance: Pd = (PoutG) /
(4pi*r2
Where: Pd = power density in mW/cm2
Pout = output power to antenna in mW G = gain of antenna in linear scale
Pi = 3.1416
R = distance between observation point and center of the radiator in cm)
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey should be less than the compliance level in each
frequency range. Interference may occur in the vicinity of equipment marked
with the following symbol:
Electrical fast transient/burst IEC 61000-4-4| + 2kV for power supply lines|
Mains power quality should be that of a typical home healthcare environment.
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8| 30 Alm| Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or hospital
environment.
Voltage Dips, short interruptions, and voltage variations on power supply
input lines
IEC 61000-4-11| Voltage dips:
0 % UT; 0,5 cycle 0/0
UT; 1 cycle 0 70 % UT; 25/30 cycles Voltage interruptions: 0 % UT; 250/300
cycle| Mains power quality should be that of a typical home healthcare
environment. If the user of the N507 requires continued operation during power
mains interruptions, it is recommended that the Cardi/o® Sensor device be
powered from an uninterruptible power supply.
Note 1: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
Note 2: Over the frequency range of 1,500 MHz to 1,500-100,000 MHz, field
strengths should be less than 5 mW/cm² for uncontrolled environments and less
than 1 mW/cm²
for controlled environments.
Note 3: UT is the a.c. mains voltage prior to application of the test
level. (a) Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcasts,s and TV broadcasts cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the device.
The recommended separation distance between portable and mobile RF communications equipment and Cardi/o® Sensor. (For ME equipment ME systems that is not life-supporting)
The Cardi/o® Sensor is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The end-user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Cardi/o® Sensor as recommended below, according to the maximum output power of the communications equipment.
Maximum Permissible Exposure (MPE) Limits (§1.1310)
Frequency
range (MHz)| Electric field
strength| Magnetic field
strength| Power density| Averaging
time
---|---|---|---|---
(V/m)| (A/m)| (mW/cm 2)| (minutes)
(i)Limits for Occupational/Controlled Exposure
0.3-3.0| 614| 2.| (100)|
3.0-30| 1842/f| 4.89/f| (900/f 2)| <6
30-300| 61.| 0.163| 1| <6
300-1,500| | | f/300| <6
1,500-100,000| | | 5| <6
(ii)Limits for General Population/Uncontrolled Exposure
0.3-1.34| 614| 2.| (100)| <30
1.34-30| 824/f| 2.19/f| (180/f 2)| <30
30-300| 28.| 0.073| 0.2| <30
300-1,500| | | f/1500| <30
1,500-100,000| | | 1| <30
f = frequency in MHz. * = Plane-wave equivalent power density.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
Note 2: Cardi/o is intended to operate at ≥ 20 cm from persons during
normal use.
Note 3: Over the frequency range of 1,500 MHz to 1,500-100,000 MHz, field
strengths should be less than 5 mW/cm 2 for uncontrolled environments and less
than 1 mW/cm 2 for controlled environments.
General Symbols
Symbol | Title | Symbol | Title |
---|---|---|---|
**IP24** |
Protected against splashing water | Read usage instructions | |
Properly dispose of EEE (Electrical and Electronic Equipment) | Non- |
ionizing radiation
| Prescription-only| | FCC Marking
| Manufacturer| | Manufacture Date YYY-MM-DD
| Caution, consult documents| | Not to be used in case package is damaged
**REF**
| Catalog number| **LOT**
| Batch code
**SN**
| Serial number| | Temperature limits (Storage)
| Humidity limits (Storage)| | Contents (Numeral represents the number of
units inside)
| Underwriters Laboratories
MEDICAL — PATIENT MONITORING EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE, AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH ANSI/AAMI ES60601-1 (2005),
“Medical Electrical Equipment – Part 1: General Requirements for Basic Safety
and Essential Performance; ANSUAAMVIEC 60601-2-25, “Medical Electrical
Equipment – Part 2-25: Particular Requirements for the Basic Safety and
Essential Performance of Electrocardiographs” E358758
Document History
Revision | Description |
---|---|
Rev. 01 | The initial release of this document. |
This “Instructions for Use” has been approved by the U.S. Food and Drug Administration.
IFU-EUA Rev. 01
Date: 01Sep2020