  • Product Name: InnovateTM Rapid Microbial Screening System
  • Model Number: INN-INS-4081-REV_C
  • Reagent Shelf Life: Reconstituted reagents are good for 5 days in cooling unit

FAQ

  • Q: How long are reconstituted reagents good for?
    • A: Reconstituted reagents are good for 5 days when stored in the cooling unit.
  • Q: What should I do at the end of the day with Injectors 1 and 3?
    • A: Run Wash for Injectors 1 and 3 at the end of the day or every 8 hours. Do not leave Injectors 1 and 3 in reagents for more than 8 hours.

Daily Setup

Incubate

  • Incubate sample according to SOP.

Daily Setup

hygiena-INN-INS-4081-REV-C-Innovate-Rapid-Microbial-Screening-System-
FIG-2

  • Turn on Innovate instrument.
    • Launch Innovate Software.
    • Make sure LED light turns green.
    • Run Wash for Injectors 1 and 3.
  • Reconstitute reagents and/or check volume.
    • Reconstituted reagents are good for 5 days in cooling unit.hygiena-INN-INS-4081-REV-C-Innovate-Rapid-Microbial-Screening-System-FIG-3
  • Insert Injector 1 into ATX and Injector 3 into Sensilux before priming. Then set-up plate prep with daily controls and run.
    • See Customer Guide for Details.
  • Use Plate Prep to create a RS_Dairy Run and enter sample information* Then click Run Plate.
    • Multiple product types can be tested in the same run.

Daily Maintenance

Run Test

hygiena-INN-INS-4081-REV-C-Innovate-Rapid-Microbial-Screening-System-
FIG-4

  • Pipette 50 microliters of samples into microtiter plate wells.
  • Click Start to begin run, then load sample plate and click next.hygiena-INN-INS-4081-REV-C-Innovate-Rapid-Microbial-Screening-System-FIG-5
  • Review results on screen.
    • See Customer Guide for details.
  • Remove microtiter plate.
  • Run Wash for Injectors 1 and 3 at end of day or every 8 hours.
    • Do not leave Injectors 1 and 3 in reagents for more than 8 hours. Power down Innovate System.