hygiena INN-INS-4081-REV_C Innovate Rapid Microbial Screening System User Guide
- June 1, 2024
- Hygiena
Table of Contents
hygiena INN-INS-4081-REV_C Innovate Rapid Microbial Screening System
Specifications
- Product Name: InnovateTM Rapid Microbial Screening System
- Model Number: INN-INS-4081-REV_C
- Reagent Shelf Life: Reconstituted reagents are good for 5 days in cooling unit
FAQ
- Q: How long are reconstituted reagents good for?
- A: Reconstituted reagents are good for 5 days when stored in the cooling unit.
- Q: What should I do at the end of the day with Injectors 1 and 3?
- A: Run Wash for Injectors 1 and 3 at the end of the day or every 8 hours. Do not leave Injectors 1 and 3 in reagents for more than 8 hours.
Daily Setup
Incubate
- Incubate sample according to SOP.
Daily Setup
- Turn on Innovate instrument.
- Launch Innovate Software.
- Make sure LED light turns green.
- Run Wash for Injectors 1 and 3.
- Reconstitute reagents and/or check volume.
- Reconstituted reagents are good for 5 days in cooling unit.
- Insert Injector 1 into ATX and Injector 3 into Sensilux before priming. Then set-up plate prep with daily controls and run.
- See Customer Guide for Details.
- Use Plate Prep to create a RS_Dairy Run and enter sample information* Then click Run Plate.
- Multiple product types can be tested in the same run.
Daily Maintenance
Run Test
- Pipette 50 microliters of samples into microtiter plate wells.
- Click Start to begin run, then load sample plate and click next.
- Review results on screen.
- See Customer Guide for details.
- Remove microtiter plate.
- Run Wash for Injectors 1 and 3 at end of day or every 8 hours.
- Do not leave Injectors 1 and 3 in reagents for more than 8 hours. Power down Innovate System.