Kamoer DIPump550 intelligent Peristaltic Pump User Manual

May 15, 2024
Kamoer

Kamoer-LOGO

Kamoer DIPump550 intelligent Peristaltic Pump

Kamoer-DIPump550-intelligent-Peristaltic-Pump-PRODUCT

Product Information

Specifications

  • Product Name: DIPump550 intelligent peristaltic pump
  • Speed Range: 0.1 – 400 RPM
  • Speed Resolution: 0.1 RPM
  • Flow Range: 450 ml/min
  • Direction Control: Support clockwise and counterclockwise operation
  • Display Mode: LED 4-digit digital tube
  • Control Mode: Encoder, toggle switch, external analog signal control, RS485 communication control, foot switch control
  • Analog Signal Control: 4-20mA, 0-5V support
  • Working Modes: Speed mode, Flow mode
  • Support Function: Fully automatic circulation, semiautomatic circulation, manual Start and stop, forward and reverse, speed adjustment, etc.
  • Power Supply Mode: External power adapter
  • Power: 50W

Product Usage Instructions

Working Conditions

Working Environment: Temperature 0-60°C, relative humidity

Usage Steps

  1. Connect the peristaltic pump to a power source using the external power adapter.
  2. Set the desired speed or flow mode using the control options available (encoder, toggle switch, analog signal control).
  3. Select the direction of operation (clockwise or counterclockwise) as needed.
  4. Start the pump operation and monitor the LED display for speed and other relevant information.
  5. To stop or change the operation mode, use the corresponding controls provided.

FAQ

  • Q: What should I do if the pump does not start?
  • A: Check the power connection and ensure the power adapter is properly plugged in. Also, verify that the settings are correct for the desired operation mode.
  • Q: Can I adjust the speed while the pump is running?
  • A: Yes, you can adjust the speed using the control options even while the pump is in operation.

Warranty service

Warranty Conditions
The warranty service is only valid for users who have correctly installed, used, and maintained in accordance with the manual, and all man-made faults or damages are not covered by the warranty.

Warranty coverage
Within one year from the date of purchase, if there is any damage caused by the manufacturing process or components, the company will provide free warranty service. The free maintenance service provided during the warranty period includes free repair, free replacement of faulty spare parts, and replacement of products that cannot be repaired with products of the same model. Free service does not include shipping costs incurred for product repairs.

Non-warranty coverage
The following factors are not covered by the free warranty, and customer repairs are subject to fees.

  1. Product appearance (please confirm when purchasing);
  2. Improper use, maintenance, or storage (please use, maintain, and store correctly according to the user manual);
  3. Access to inappropriate power supply;
  4. Damage to the components caused by the short circuit of the circuit board caused by the entry of various insects into the machine;
  5. Losses caused by accidents;
  6. Use inappropriate spare parts (non-company spare parts are not applicable);
  7. Persons not authorized by the company negligently handle, modify or repair (please do not dismantle or repair without authorization);
  8. Failure or damage caused by use outside the applicable occasion;
  9. Damage caused by force majeure, etc.;
  10. Consumable and wearing parts (pump tube, connector, connecting tube, etc.);
  11. The warranty period has expired.

Main features

Speed range 0.1- 400RPM
Speed resolution 0.1RPM
Flow range ≤450ml/min
Direction control Support clockwise and

counterclockwise operation

Display mode| LED 4-digit digital tube
Control mode| Encoder, toggle switch, external analog

signal control, RS485 communication control, foot switch control

External analog

signal control

| 4-20mA, 0-5V
para Power failure meter memory| support
Working mode| Speed mode, Flow mode
Way of working| Fully  automatic    circulation,  semi-

automatic circulation, manual

Support function| Start and stop, forward and reverse,

speed adjustment, etc.

Power supply mode| External power adapter
Power| <50W

Working conditions

  • Working Environment: Temperature 0℃~60℃, relative humidity <80%RH without condensation
  • Rated voltage: DC24V
  • Maximum current: 2A

Appearance Description

Kamoer-DIPump550-intelligent-Peristaltic-Pump-FIG-1

Digital display

  • Used to display information such as speed, flow and time;
  • When all are numbers: the current speed/currently set speed;
  • When all are numbers and all numbers are beating: The number indicates the currently set operating flow
  • When the first digit is displayed and is beating: the last three digits indicate the currently set working time
  • When the first digit shows no beating: the last three digits indicate the current remaining working time
  • When the first digit is displayed and is beating: the last three digits indicate the currently set stop duration
  • When the first digit is displayed without bounce: the last three digits indicate the current remaining stop duration
  • When the first digit is displayed without bounce: the last digit indicates the selection of setting function

Rotary encoder
Used to adjust the speed and time; increase clockwise and decrease counterclockwise. The faster the rotation speed, the larger the increment of adjustment; The speed can be adjusted during the operation of the peristaltic pump.

Direction control switch
Used to control the direction of rotation of the peristaltic pump, when dialed to the left, the pump rotates counterclockwise; when it is dialed to the right, the pump rotates clockwise. The switch is ineffective during the operation of the peristaltic pump, and will only switch to the post-switching state when the peristaltic pump stops running and runs again.

Kamoer-DIPump550-intelligent-Peristaltic-Pump-FIG-4

Power socket

  • External DC power, 24VDC 2A.

Combination control port

  1. 485 communication (pin 2, 3)
  2. External start and stop control (pin 4, 5)
  3. Analog signal control 0 ~ 5V (pin 6, 7)
  4. Analog signal control 4 ~ 20mA (pin 8, 9)

Working mode setting—Speed Mode

Fully automatic cycle mode setting

  1. Press and hold the encoder in the speed setting state, the digital tube switches to the time setting mode, and the first upper port jumps (red in the figure), at this time is set the running time of the pump;

  2. The rotary encoder can set the time, and the settable time range is 0 to 999 seconds.

  3. Press the encoder (not long press) to switch to the pump stop time setting, and the first position of the digital tube will jump (red in the picture);

  4. The rotary encoder can set the time, and the settable time range is 0 to 999s;

  5. Press and hold the encoder to return to the speed display interface. The running time and stop time will be saved automatically.

  6. Press the encoder (not long press), the pump will run according to the set running time and stop time;
    Note: When the pump starts to cycle, the digital tube displays the time countdown, and follows 1.3. The upper and lower ports are displayed;

  7. Press the encoder (not long press) and the pump stops cycling. The digital tube switches to the display speed interface.

Semi-automatic cycle mode setting
Simply set the time to 0 in step 4 of the fully automatic cycle mode setting mode;
At this time, after the pump stops running according to the set working time, it will not start automatically; if it starts again, the user needs to press the encoder or step on the foot switch.

Manual mode setting

  1. Set the time to 0 in step 2 of the automatic cycle mode setting;
  2. Set the time to 0 in step 4 of the automatic cycle mode setting;
    At this point, the start and stop of the pump is controlled by the encoder or foot switch.

Working mode setting—Flow Mode

Fully automatic cycle mode setting

  1. In the flow mode, long press the encoder to switch to the operating flow setting. The Nixie tube displays all numbers and flickers. At this time, it is to set the operating flow of the pump;
  2. The rotary encoder can set the Operating flow, and the setting range is 0 ~ 9999 ml.
  3. Press the encoder (not long press) to switch to the pump stop time setting, and the first position of the digital tube will jump (red in the picture);
  4. The rotary encoder can set the time, and the settable time range is 0 to 999 seconds;
  5. Press and hold the encoder to return to the flow rate display interface, and the running flow and stop time will be saved automatically;
  6. Press the encoder (not long press), the pump will run according to the set Operating flow and stop time;
  7. Press the encoder (not long press) and the pump stops cycling. The digital tube switches to the display flow rate interface.

Semi-automatic cycle mode setting
Simply set the time to 0 in step 4 of the fully automatic cycle mode setting mode. At this time, after the pump stops running according to the set Operating flow, it will not start automatically; if it starts again, the user needs to press the encoder or step on foot switch.

Manual mode setting

  1. Set the Operating flow to 0 in step 2 of the automatic cycle mode setting;
  2. Set the time to 0 in step 4 of the automatic cycle mode setting;

At this point, the start and stop of the pump is controlled by the encoder or the foot switch.

Enter menu settings page

  1. On the operation preparation page, long press the rotary encoder to switch to the working mode setting interface (If the operation mode is speed mode , The first digit of the nixie tube jumps (Red in the picture ); If the operation mode is flow mode, the digital display on the nixie tube is all digital, and it jumps and flashes ).
  2. In the working mode setting interface, long press the rotary encoder again to switch to the menu setting interface (the first digit of the nixie tube jumps , the middle two digits are 0, and the last digit is the corresponding function selection).
  3. In the menu setting interface, the rotary encoder can select the functions to be set. The last 8 digits of “1-8” correspond to 8 different functions. Select the function to be set and press the rotary encoder to enter the detailed setting interface of the function.

Specific function settings

Note: In the setting interface, only long press the rotary encoder to save the settings and make the settings take effect. Short pressing the rotary encoder will not save the settings.

“C001” Mode switching
In the function selection interface, the rotary encoder, when it displays “C001”, the corresponding function is “mode switching”, short press the rotary encoder to enter the setting interface, you can switch between “speed mode” and “flow mode”, the last digit represents Mode, select the rotary encoder, short press the rotary encoder to return to the function selection interface; long press the rotary encoder (the setting takes effect) and return to the home page.

Interface Setting Introduce

Kamoer-DIPump550-intelligent-Peristaltic-Pump-FIG-5

| | Speed mode
| Flow mode

“C002” Direction adjustment

In the function selection interface, the rotary encoder, when it displays “C002”, the corresponding function is “direction level mode setting”, short press the rotary encoder to enter the setting interface, and you can set the direction level mode. The last digit represents the direction level, which can be used to match the direction of the direction button and the label on the surface.
Tip: If the running direction is the same as the direction of the label on the surface, there is no need to set this item.

Interface Setting Introduce

|

| Low level mode (When the

direction button is turned counterclockwise but the pump actually rotates clockwise, select

0 to adjust the pump to counterclockwise)

| High-level mode (When the direction button is turned counterclockwise rotation and the pump actually rotates counterclockwise, select 1 to adjust the pump to clockwise)

“C003” Foot switch setting
In the function selection interface, the rotary encoder, when it displays “C003”, the corresponding function is “Foot Switch Setting”, short press the rotary encoder to enter the setting interface, and you can set the foot switch usage mode. The last digit represents how the foot switch is used.

Interface Setting Introduce

|

| Edge trigger (press the foot switch

once, the pump starts to work, press again, and the pump stops)

| Low level trigger (long press the

foot switch, pump work, release the foot switch, pump stop)

| High level trigger (releasing the

foot switch, the pump works, long press the foot switch, the pump stops working)

“C004” Calibration
In the function selection interface, rotary encoder, when it displays “C004”, the corresponding function is “flow calibration”, short press the rotary encoder to enter the setting interface, and you can perform pump head flow calibration.
After entering the setting interface, the calibration speed and calibration time will switch back and forth every 2 seconds; the first digit shows 4, the middle one space, and the last two digits represent the calibration time; the rotary encoder sets the calibration speed and calibration time, after completing the setting Short press the rotary encoder to start flow calibration.
The flow calibration is accompanied by a countdown. After the countdown is over, the volume input interface is displayed, enter the actual volume, press and hold the rotary encoder (the setting takes effect) and return to the home page.

Interface Setting Introduce

| | Set the calibration speed
| Set calibration time
| Enter actual volume

“C005” Running status setting
In the function selection interface, for the rotary encoder, when the corresponding function of “C005” is displayed as “saving in running state”, short press the rotary encoder to enter the setting interface, and you can set whether to save the current running state. The last digit represents whether to save.

Interface Setting Introduce

| | Not save the running state
| save the running state

“C006” Analog quantity selection
In the function selection interface, rotary encoder, when the corresponding function of “C006” is displayed as “analog selection”, short press the rotary encoder to enter the setting interface, and you can set the analog selection. The last digit represents the analog option.

Interface Setting Introduce

| | 0-5V
| 4-20mA
| 485 mode

“C007” Suction back function
In the function selection interface, the rotary encoder, when “C007” is displayed, the corresponding function is “Suction Setting”, short press the rotary encoder to enter the setting interface, the settings of “Suction Time” and “Suction Speed” in the interface are set every time Circulating scrolling every 2 seconds, the pump head can be set to suck back after running. “Suction time” setting, the first digit of the interface digital tube is “7”, the second space, and the last two digits represent the suction time. At this time, turn the encoder to adjust the suction time (0-99 seconds) ; “Suction speed” setting, there is a decimal point display between the third and fourth digits of the interface. At this time, turn the encoder to adjust the speed of suction (0.1-999.9ml/min).suction time. At this time, turn the encoder to adjust the suction time (0-99 seconds) ; “Suction speed” setting, there is a decimal point display between the third and fourth digits of the interface. At this time, turn the encoder to adjust the speed of suction (0.1-999.9ml/min).

Interface Setting Introduce

| | Suck back time
| Suction speed

“C008” Drive enable
In the function selection interface, for the rotary encoder, when the corresponding function of “C008” is displayed as “drive enable the setting”, short press the rotary encoder to enter the setting interface, and the drive enable setting can be performed. The last digit represents the enable option.

Interface Setting Introduce

| | Low level enable
| High level enable

“C009” Maximum speed setting
This function is used by the manufacturer to debug the equipment. The internal display is the equipment speed. The minimum speed is 150 rpm and the maximum speed is 1500 rpm. Please do not change the option, otherwise it will easily lead to abnormal phenomena such as pump head jamming.

Interface Setting Introduce
The default speed is 400 rpm

“C010” Restore factory defaults

Interface Setting Introduce


| | Do not restore factory defaults
| Restore factory defaults

External control

485 mode
The board is controlled by 485 communication mode; please connect to the 485 communication wire.

Analog signal control mode

  • Control the board by analog signal mode (0~5V, 4~20mA);

  • Please connect the signal line according to the actual control mode.

  • Pin 1: reserved

  • Pin 2: 485 B

  • Pin 3: 485 A

  • Pin 4: Foot switch +

  • Pin 5: Foot switch –

  • Pin 6: Analog signal 5V-

  • Pin 7: Analog signal 5V+

  • Pin 8: Analog signal mA-

  • Pin 9: Analog signal mA+

Appearance dimension drawing

Kamoer-DIPump550-intelligent-Peristaltic-Pump-
FIG-38Pump tube flow reference

Kamoer-DIPump550-intelligent-Peristaltic-Pump-FIG-39

Code| Pump tube

material

| *ID OD**

mm

| Number

of rotors

| Max flow

ml/min

---|---|---|---|---
B146| BPT| 1.64.8| 6| 64
B163| BPT| 3.2
6.4| 3| 252
B403| BPT| 4.07.2| 3| 336
B253| BPT| 4.8
8.0| 3| 452

Note: the maximum flow in the above chart is obtained by testing water at 400 rpm with a new pump pipe aged for 30 minutes at room temperature (about 25 ℃) for reference only. Ambient temperature, material and elasticity of pump pipe, viscosity of test liquid, and other factors will affect the actual flow. The thickness of the pump pipe will affect the maximum speed of actual stable operation.

CONTACT

Kamoer Fluid Tech (Shanghai) Co., Ltd.

  • Add: 4th Building, No 49 Xiangjing Road, Songjiang District, Shanghai, China
  • Phone: +86-021-67742578
  • Fax: +86-021-67741776
  • Website: www.kamoer.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Kamoer User Manuals

Related Manuals