Product Information
Specifications
- Product Name: Bacillus cereus Detection Kit
- Product Number: KIT230201
- Revision: A, December 2023
- Usage: PCR kit for the qualitative detection of groEL target using real-time PCR instruments
- Intended Use: For food testing purposes
- Kit Size: 50 reactions
- For In Vitro Use Only
Product Usage Instructions
Program Setup
1. Program your real-time PCR instrument before setting up the PCR reactions. Select the following channels: FAM (groEL) and VIC/HEX (Internal Control).
Data Interpretation
- Verify the results of positive (Control Template) and negative (H2O) controls before interpreting the sample results. Always compare samples to positive and negative controls.
- Review data from each channel and interpret results as described in the table provided.
Preparation of the RT-PCR Mix
- Take appropriate precautions to prevent contamination, e.g., by using filter tips and wearing gloves. Thaw reagents, mix (do not vortex!) and briefly spin vials before opening.
- Prepare PCR mix for each reaction to a suitable tube. Mix carefully but thoroughly by pipetting up and down.
- Add PCR mix, followed by samples and controls.
- Seal strips/plates accurately.
- Briefly spin strips/plate in a suitable centrifuge.
- Start real-time PCR run following the specified cycle for samples.
Frequently Asked Questions (FAQ)
Q: Can this kit be used for purposes other than food testing?
- A: No, this kit is intended for food testing purposes only as indicated in the product information.
Q: How many reactions can be performed with this kit?
- A: The kit allows for 50 reactions to be conducted.
- Product No. KIT230201
- PCR kit for the qualitative detection of groEL target using real-time PCR instruments.
- Before starting, it is strongly recommended to read the entire product manual available on our website.
PROGRAM SETUP
- Program your real-time PCR instrument before setting up the PCR reactions.
Select the following channels:
- FAM (groEL) and VIC/HEX (Internal Control).
- Pre-incubation: 1 cycle
- Step 1: 95 °C for 10 min
- Amplification: 40 cycles
- Step 1: 95 °C for 15 s
- Step 2: 60 °C for 60 s
- Fluorescence detection
DATA INTERPRETATION
- Verify the results of positive (Control Template) and negative (H2O) controls, before interpreting the sample results.
- Always compare samples to positive and negative controls.
- Review data from each channel and interpret results as described in the table.
FAM | VIC/HEX | Result Interpretation |
---|---|---|
+ | + or – | Positive for groEL |
– | + | Negative for groEL |
– | – | Invalid |
PREPARATION OF THE RT-PCR MIX
- Take appropriate precautions to prevent contamination, e.g., using filter tips and gloves.
- Thaw reagents, mix (do not vortex!), and briefly spin vials before opening.
- PREPARE PCR MIX
- Add 10 µL Master Mix (2XM) and 4 µL Primer/Probe Mix (P) for each reaction to a suitable tube.
- (n samples + 2 controls + at least one additional reaction to cover pipetting loss). Mix carefully but thoroughly by pipetting up and down.
- ADD PCR MIX
- Pipet 14 µL of prepared PCR mix into each strip or plate well.
- ADD SAMPLES AND CONTROLS
- Pipet 6 µL of samples, negative control (PCR-grade H2O, not included) or Control Template (C) into respective wells.
- SEAL
- Seal strips/plates accurately.
- CENTRIFUGE
- Briefly spin strips/plate in a suitable centrifuge.
- START REAL-TIME PCR RUN
- Cycle samples as described above.
- foodproof® SL Bacillus cereus Detection Kit
- KIT230201
- Kit for 50 reactions Store kit at -15 to -25 °C
- For food testing purposes FOR IN VITRO USE ONLY
- Hygiena®
- Camarillo, CA 93012 USA
- diagnostics.support@hygiena.com
- www.hygiena.com
Documents / Resources
|
hygiena KIT230201 Bacillus Cereus Detection
Kit
[pdf] User Guide
KIT230201 Bacillus Cereus Detection Kit, KIT230201, Bacillus Cereus Detection
Kit, Cereus Detection Kit, Detection Kit
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>