SplusS TF65 Temperature Sensor Instruction Manual
- May 15, 2024
- SplusS
Table of Contents
- Operating And Mounting Instructions
- Basic device with accessories
- Mounting Flange, Plastic
- TECHNICAL DATA
- General notes
- Installation and Commissioning
- Resistance characteristics of passive temperature sensors
- CUSTOMER SUPPORT
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
SplusS TF65 Temperature Sensor
Operating And Mounting Instructions
Duct temperature sensor, including mounting flange, with passive output
High-performance encapsulation against vibration, mechanical stress and
humidity
Basic device with accessories
Mounting Flange, Plastic
Resistance thermometer with a passive output, housing made from impact-
resistant plastic with quick-locking screws, and straight protective tube,
incl. mounting flange.
These duct temperature sensors are electric contact thermometers for
temperature measurement in liquids and gases, which are installed for example
in piping systems and vessels. For aggressive media, stainless steel immersion
sleeves must be used.
Applications of these temperature sensors in piping systems, in heating
technology, in storage tanks, in district heating compact stations, in hot and
cold-water systems, in oil and lubricant circulation systems, in mechanical,
apparatus and plant engineering as well as in the entire industrial sector.
TECHNICAL DATA
Measuring range: | –30…+150 °C (T max NTC = +150 °C, T max LM235Z = +125 °C) |
---|---|
Sensors ⁄ output: | Pt1000 (according to DIN EN 60 751, class B), passive |
(Perfect Sensor Protection)
Connection type:| 2-wire connection
Testing current:| < 0.6 mA (Pt1000)
Insulating resistance:| ≥ 100 MΩ, at +20 °C (500 V DC)
Ambient temperature:| –20…+100 °C
Housing| plastic, UV-resistant, material polyamide, 30 % glass-globe
reinforced, with quick-locking screws (slotted ⁄ Phillips head combination),
colour traffic white (similar to RAL 9016)
Housing dimensions:| 72 x 64 x 37.8 mm (Tyr 1)
Cable connection:| cable gland, plastic (M 16 x 1.5; with strain relief,
exchangeable, max. inner diameter 10.4 mm)
0.14 – 1.5 mm², via terminal screws| Electrical connection:
Protective tube:| stainless steel, V4A (1.4571), Ø = 6 mm, inserted length
(EL) = 300 mm
Process connection:| by mounting flange, plastic
Permissible humidity:| < 95 % RH, non-precipitating air
Protection class:| III (according to EN 60 730)
Protection type:| IP 67 (according to EN 60 529) Housing in the built-in state
Housing tested, TÜV SÜD, Report No. 713139052 (Tyr 1)
ACCESSORIES
Mounting flange, plastic, 56.8 x 84.3 mm, Ø = 15.2 mm tube gland, Tmax= +100
°C (included in the scope of delivery)| MF – 15 – K
TF 65| Temperature sensor, passive, Premium
---|---
Type| Inserted Length (EL)| Type| Danfoss Part Number
Type TF 65| 300 mm| Pt1000 (according to DIN EN 60 751, class B)| 134B9412
General notes
In order to avoid damages ⁄ errors, preferably shielded cables are to be used.
Laying measuring cables parallel with current-carrying cables must in any case
be avoided. EMC directives shall be observed!
These instruments must be installed by authorised specialists only!
Note!
Select immersion depth for built-in sensors so that the error caused by heat
dissipation stays within the admissible error margins. A standard value is: 10
x diameter of protection tube + sensor length. In connection with housing-type
sensors, particularly with outdoor sensors, please consider the influence of
thermal radiation. If the sensor is used in refrigeration circuits, it must be
insulated together with the housing to reduce the temperature potential
between the device and the medium to a minimum and thus prevent condensation
damage.
Maximum thermal load on components:
On principle, all temperature sensors shall be protected against unacceptable
overheating!
Standard values for individual components and materials selected are shown for
operation under neutral atmosphere and otherwise normal conditions (see table
to the right).
For combinations of different insulating materials, the lowest temperature
limit shall always apply.
Limiting deviation according to classes
Tolerances at 0 °C:
Platinum sensors (Pt1000):
DIN EN 60751, class B :± 0,3 K
ATTENTION, NOTE !
Testing current influences the thermometer‘s measuring accuracy due to
intrinsic heating and therefore, should never be greater than as specified
below:
Sensor current, maximum I max
Pt1000 (thin-layer) :< 0.6 mA
Installation and Commissioning
Devices are to be connected under dead-voltage condition. Devices must only be
connected to safety extra-low voltage. Consequential damages caused by a fault
in this device are excluded from warranty or liability.
These devices must be installed and commissioned by authorised specialists.
The technical data and connecting conditions shown on the device labels and in
the mounting and operating instructions delivered together with the device are
exclusively valid. Deviations from the catalogue representation are not
explicitly mentioned and are possible in terms of technical progress and
continuous improvement of our pro ducts. In case of any modifications made by
the user, all warranty claims are forfeited. Operating this device close to
other devices that do not comply with EMC directives may influence
functionality. This device must not be used for monitoring applications, which
serve the purpose of protecting persons against hazards or injury, or as an
EMERGENCY STOP switch for systems or machinery, or for any other similar
safety-relevant purposes.
Dimensions of housing or housing accessories may show slight tolerances on the
specifications provided in these instructions.
Modifications of these records are not permitted.
In case of a complaint, only complete devices returned in original packing
will be accepted.
Our “General Terms and Conditions for Business“ together with the “General
Conditions for the Supply of Products and Services of the Electrical and
Electronics Industry“ (ZVEI conditions) including supplementary clause
“Extended Retention of Title“ apply as the exclusive terms and conditions“.
Notes regarding mechanical mounting and attachment: Mounting shall take
place while observing all relevant regulations and standards applicable for
the place of measurement (e.g. such as welding instructions, etc.).
Particularly the following shall be regarded:
-
VDE ⁄ VDI directive technical temperature measurements, measurement set – up for temperature measurements.
-
The EMC directives must be adhered to.
-
It is imperative to avoid parallel laying of current-carrying lines.
-
We recommend to use shielded cables with the shielding being attached at one side to the DDC ⁄ PLC.
-
If the sensor is used in refrigeration circuits, it must be insulated together with the housing to reduce the temperature potential between the device and the medium to a minimum and thus prevent condensation damage.
Before mounting, make sure that the existing thermometer‘s technical parameters comply with the actual conditions at the place of utilization, in particular in respect of: -
Measuring range
-
Permissible maximum pressure, flow velocity
-
Installation length, tube dimensions
-
Oscillations, vibrations, shocks are to be avoided (< 0.5 g)
Attention! In any case, please observe the mechanical and thermal load limits of the protective tubes according to DIN 43763 or according to specific S+S standards!
Flange mounting:
In case of flange mounting, screws in the flange part must be equally tightened. The lateral pressure screw must clamp securely, otherwise the feeler shaft might slip through.
Commissioning is mandatory and may only be performed by qualified personnel!
Resistance characteristics of passive temperature sensors
Pt 1000
°C| Ω
– 50| 803
– 40| 843
– 30| 882
– 20| 922
– 10| 961
0| 1000
+10| 1039
+20| 1078
+30| 1117
+40| 1155
+50| 1194
+60| 1232
+70| 1271
+80| 1309
+90| 1347
+100| 1385
+110| 1423
+120| 1461
+130| 1498
+140| 1536
+150| 1573
© Copyright by S+S Regeltechnik GmbH
Reprint in full or in parts requires permission from S+S Regeltechnik GmbH.
Subject to errors and technical changes. All statements and data herein represent our best knowledge at date of publication. They are only meant to inform about our products and their application potential, but do not imply any warranty as to certain product characteristics. Since the devices are used under a wide range of different conditions and loads beyond our control, their particular suitability must be verified by each customer and/or end user themselves. Existing property rights must be observed. We warrant the faultless quality of our products as stated in our General Terms and Conditions.
CUSTOMER SUPPORT
S+S REGELTECHNIK GMBH THURN-UND-TAXIS-STR. 22 90411 NÜRNBERG ⁄ GERMANY FON +49
(0) 911 ⁄ 5 19 47- 0
mail@SplusS.de
www.SplusS.de
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>