HIOKI BT3554 Battery Tester Instruction Manual

June 17, 2024
HIOKI

HIOKI BT3554 Battery Tester

HIOKI BT3554 Battery Tester

Introduction

  1. This manual provides information necessary to calibrate the applicable product. Information for adjustment is not included.
  2. For details about operating procedures, operation, specifications, and safety information, refer to the instruction manual.
  3. Perform the calibration with great care so as to avoid any risk of electric shock due to voltage or current.
  4. If your product cannot be calibrated following the instructions in this calibration manual, please contact Hioki for calibration services.

Status

Ambient temperature and humidity 23ºC ±5ºC, 80% RH or less
Power supply voltage and frequency Power supply voltage and

frequency specified on the main unit
Warm-up time| None
Unit settings| Noise frequency avoidance function OFF (normal start), no zero adjustment

Recommended Equipment and Accessories

Name Manufacturer, model name, specifications Remarks
Variable resistor Alpha Electronics ADR-3204GR or equivalent Accuracy test

(temperature measurement)
Voltage calibrator| ADCMT 6166 or equivalent| Accuracy test (voltage measurement)
Resistor| Effective resistance at 980 Hz 0 mΩ (0.002 mΩ or less)| Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 1 mΩ (±0.0045 mΩ, 51.2 µW or more)| Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 3 mΩ (±0.0095 mΩ, 153.6 µW or more)| Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 10 mΩ (±0.035 mΩ, 512 µW or more)| Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 30 mΩ (±0.075 mΩ, 1.536 mW or more)| Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 100 mΩ (±0.35 mΩ, 51.2 µW or more)| Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 300 mΩ (±0.75 mΩ, 153.6 µW or more)| Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 1 Ω (±0.0035 Ω, 5.12 µW or more)| Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 3 Ω (±0.0075 Ω, 15.36 µW or more)| Accuracy test (resistance measurement)
Connection cable for accuracy test (temperature measurement)| Jack-Banana Use a cable equivalent to the cable supplied with the product to connect the equipment and main unit.| Accuracy test (temperature measurement)
Various other connection cables| Use a cable equivalent to the cable supplied with the product to connect the equipment and main unit.| Various tests

Calibration Items

No.| Item| Function| Range| Input| Stability time| Tolerance
---|---|---|---|---|---|---
1| Accuracy| Resistance Ω| 3 mΩ| 0 mΩ| 3.0 s| Refer to the Test Report.1
1 mΩ| 3.0 s
3 mΩ| 3.0 s
30 mΩ| 0 mΩ| 3.0 s
10 mΩ| 3.0 s
30 mΩ| 3.0 s
300 mΩ| 0 mΩ| 3.0 s
100 mΩ| 3.0 s
300 mΩ| 3.0 s
3 Ω| 0 Ω| 3.0 s
1 Ω| 3.0 s
3 Ω| 3.0 s
Voltage V Conduct the test in the 3 Ω range.| 6V| 0V| 3.0 s
3V| 3.0 s
5.95V| 3.0 s
60V| 30V| 3.0 s
-30V| 3.0 s
59.5V| 3.0 s
Temperature| When the temperature unit setting is Fahrenheit °F, change it to Celsius °C.
2
−| 0°C (459.29 Ω) 3| 1.0 s| Refer to the Test Report.1
25°C (500.00 Ω) 3| 1.0 s
50°C (540.37 Ω)
3| 1.0 s
If the temperature unit setting is Fahrenheit °F before calibration, set the unit to Fahrenheit °F again.

  1. For tolerance, download the Test Report from https://www.hioki.com/global/support/download.
  2. Changing the temperature display unit ( Fahrenheit °F ↔ Celsius °C ).
  3. Turn off the instrument.
  4. Display the setting of the temperature unit.
  5. Press the key for at least 3 seconds.
  6. Confirm the setting.
    The instrument is restarted.
    The settings will not change if the power is turned off before the settings are applied.
    *3 Include the resistance value of the connecting cable.
    Calibration Items

Connection Diagrams

Figure 1 Wiring example for accuracy test (resistance measurement)

Connection Diagrams

Figure 2 Wiring example for accuracy test (voltage measurement)

Connection Diagrams

Figure 3 Wiring example for accuracy test (temperature measurement)

Connection Diagrams

Customer Support

www.hioki.com/

HEADQUARTERS
81 Koizumi
Ueda, Nagano 386-1192 Japan

Edited and published by HIOKI E.E. CORPORATION

  • Contents subject to change without notice.
  • This document contains copyrighted content.
  • It is prohibited to copy, reproduce, or modify the content of this document without permission.
  • Company names, product names, etc. mentioned in this document are trademarks or registered trademarks of their respective companies.

Logo

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

HIOKI User Manuals

Related Manuals