Status
Ambient temperature and humidity | 23ºC ±5ºC, 80% RH or less |
---|---|
Power supply voltage and frequency | Power supply voltage and |
frequency specified on the main unit
Warm-up time| None
Unit settings| Noise frequency avoidance function OFF (normal start), no
zero adjustment
Recommended Equipment and Accessories
Name | Manufacturer, model name, specifications | Remarks |
---|---|---|
Variable resistor | Alpha Electronics ADR-3204GR or equivalent | Accuracy test |
(temperature measurement)
Voltage calibrator| ADCMT 6166 or equivalent| Accuracy test (voltage
measurement)
Resistor| Effective resistance at 980 Hz 0 mΩ (0.002 mΩ or less)| Accuracy
test (resistance measurement)
Resistor| Effective resistance at 980 Hz 1 mΩ (±0.0045 mΩ, 51.2 µW or more)|
Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 3 mΩ (±0.0095 mΩ, 153.6 µW or more)|
Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 10 mΩ (±0.035 mΩ, 512 µW or more)|
Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 30 mΩ (±0.075 mΩ, 1.536 mW or more)|
Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 100 mΩ (±0.35 mΩ, 51.2 µW or more)|
Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 300 mΩ (±0.75 mΩ, 153.6 µW or more)|
Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 1 Ω (±0.0035 Ω, 5.12 µW or more)|
Accuracy test (resistance measurement)
Resistor| Effective resistance at 980 Hz 3 Ω (±0.0075 Ω, 15.36 µW or more)|
Accuracy test (resistance measurement)
Connection cable for accuracy test (temperature measurement)| Jack-Banana Use
a cable equivalent to the cable supplied with the product to connect the
equipment and main unit.| Accuracy test (temperature measurement)
Various other connection cables| Use a cable equivalent to the cable supplied
with the product to connect the equipment and main unit.| Various tests
Calibration Items
No.| Item| Function| Range| Input| Stability
time| Tolerance
---|---|---|---|---|---|---
1| Accuracy| Resistance Ω| 3 mΩ| 0 mΩ| 3.0 s| Refer to the Test
Report.1
1 mΩ| 3.0 s
3 mΩ| 3.0 s
30 mΩ| 0 mΩ| 3.0 s
10 mΩ| 3.0 s
30 mΩ| 3.0 s
300 mΩ| 0 mΩ| 3.0 s
100 mΩ| 3.0 s
300 mΩ| 3.0 s
3 Ω| 0 Ω| 3.0 s
1 Ω| 3.0 s
3 Ω| 3.0 s
Voltage V Conduct the test in the 3 Ω range.| 6V| 0V| 3.0 s
3V| 3.0 s
5.95V| 3.0 s
60V| 30V| 3.0 s
-30V| 3.0 s
59.5V| 3.0 s
Temperature| When the temperature unit setting is Fahrenheit °F, change it to
Celsius °C.2
−| 0°C (459.29 Ω) 3| 1.0 s| Refer to the Test Report.1
25°C (500.00 Ω) 3| 1.0 s
50°C (540.37 Ω) 3| 1.0 s
If the temperature unit setting is Fahrenheit °F before calibration, set the
unit to Fahrenheit °F again.
- For tolerance, download the Test Report from https://www.hioki.com/global/support/download.
- Changing the temperature display unit ( Fahrenheit °F ↔ Celsius °C ).
- Turn off the instrument.
- Display the setting of the temperature unit.
- Press the key for at least 3 seconds.
- Confirm the setting.
The instrument is restarted.
The settings will not change if the power is turned off before the settings are applied.
*3 Include the resistance value of the connecting cable.
Connection Diagrams
Figure 1 Wiring example for accuracy test (resistance measurement)
Figure 2 Wiring example for accuracy test (voltage measurement)
Figure 3 Wiring example for accuracy test (temperature measurement)
Customer Support
HEADQUARTERS
81 Koizumi
Ueda, Nagano 386-1192 Japan
Edited and published by HIOKI E.E. CORPORATION
- Contents subject to change without notice.
- This document contains copyrighted content.
- It is prohibited to copy, reproduce, or modify the content of this document without permission.
- Company names, product names, etc. mentioned in this document are trademarks or registered trademarks of their respective companies.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>