HIOKI SS7012 DC Signal Source Instruction Manual

June 1, 2024
HIOKI

HIOKI SS7012 DC Signal Source

HIOKI-SS7012-DC-Signal-Source-product-image

Product Information

Specifications :

  • Product Name: SS7012 DC Signal Source
  • Model: SS7012
  • Compatibility: RS-232
  • Baud Rate: 9600bps (fixed)
  • Data Bit: 8bit (fixed)
  • Parity: None (fixed)
  • Stop Bit: 1 (fixed)
  • Flow Control: None (fixed)

Product Usage Instructions

  1. SS7012 Remote Interface Overview:
    The SS7012 DC Signal Source can be operated remotely using communication functions. The remote interface utilizes virtual RS-232 serial communication with the USB interface acting as the transmission route

  2. Connect a USB cable:
    Before connecting the USB cable, ensure that the USB driver from the accompanying CD is installed on the personal computer. Connect the USB cable as shown in Figure 2 to establish communication between the PC and SS7012.

  3. Setting Remote Interface Parameters:
    In order to send communications successfully, ensure that the serial interface parameters of the SS7012 and personal computer match. The fixed settings required for communication are:

    • Baud rate: 9600
    • Data bit: 8
    • Parity: None
    • Stop bit: 1
    • Flow control: None

Frequently Asked Questions (FAQ)

  • Q: How do I install the USB driver for the SS7012?
    A: The USB driver can be found on the CD that accompanies the SS9000 Communication Package. Refer to the instruction manual of the SS9000 Communication Package for installation instructions.

  • Q: What should I do if communication is not possible?
    A: Ensure that the serial interface parameters of the SS7012 and personal computer match the fixed settings mentioned in Table 1. Any deviation from these settings may result in communication issues.

Remote Operation Manual
SS7012

DC SIGNAL SOURCE

Introduction

This manual details the methods to operate the SS7012 DC Signal Source using communication functions, as well as all the commands used by the SS7012. When the communication function is used, the SS7012 can be manually oper-ated using communication terminal software. Furthermore, application software to automatically control the SS7012 can be run.

Safety Information

  • The following symbols in this manual indicate the relative importance of cautions and warnings.
  • Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument.
    Indicates advisory items related to performance or correct operation of the instrument.

SS7012 Remote Interface Overview

HIOKI-SS7012-DC-Signal-Source-fig- \(2\) Operating principle

  • Communication between a personal computer and the SS7012 is virtual RS-232 serial communication. While the USB interface is installed as the transmission route, actual communication uses RS-232.
  • It is necessary to issue RS-232 commands via the virtual COM port.
  • When RS-232 commands are issued to the virtual COM port, commands are sent to the SS7012 in the following order. (For the receipt of data, the order is reversed.)HIOKI-SS7012-DC-Signal-Source-fig- \(3\)

Connect a USB cable

  • Before connecting the USB cable, install the USB driver on the personal computer. The USB driver is located on the CD that accompanied the SS9000 Communication Package.
  • For installation instructions, please refer to the instruction manual of the SS9000 Communication Package.
  • After installing the USB driver, connect the personal computer and SS7012 as shown in Figure 2. The virtual RS-232 COM port will be automatically assigned to the USB port.

Setting Remote Interface Parameters

  • In order to send communications, the serial interface parameters of the SS7012 and personal computer must match.
  • Table 1 shows the settings required for communication. These parameters are fixed. If any other settings are used, communication is not possible.

Table 1: Remote Interface Parameters (All fixed values)

Item Parameter Fixed settings
1 Baud rate 9600
2 Data bit 8
3 Parity None
4 Stop bit 1
5 Flow control None

Enabling key operation

  • While the SS7012 is communicating, key operation of the SS7012 cannot be used. (On the LCD of the SS7012, is on.)
  • To enable key operation of the SS7012, push the LOCAL key of the SS7012. goes out.

Commands Summary

  • Overview of Command Type and Format
    All commands can be entered in upper and lowercase.

  • Command Format and Query Format
    The commands control instrument functions.
    The following example shows a command and its associated query:HIOKI-SS7012
-DC-Signal-Source-fig-07

  • Query Command
    The query command ends with a question mark “?”.

  • Terminator (delimiter)
    A terminator is a character, which identifies the end of a command string. A valid terminator consists of two-byte data:

    (Carriage Return, ASC(&H0D)) (Line Feed, ASC(&H0A))

Response Messages

  • Return Result

    • After the instrument executes a query command the return of the result will be in the following format:

    • The format of measurement data will be shown as following:

    • The format of monitor data will be shown as following:

    • The format of setting command will be shown as following:
      (Response 1) OK

    • (When the setting command was pro- cessed normally)
      (Response 2) CMD ERR (When there is an error in the transmission command)

  • Data Types

    • Returned message is the ASCII string from the instrument responding to a query.
    • A query is a command followed by a question mark.
    • The following table 2 is explanation for data types.

Table 2: Data Types of Responding Message and Parameter

Data Type Explanation Example
NR1 An integer 0, 1, 2, 3, etc.
NR2 This numeric representation has an explicit radix point. +13.234, 400.0,

etc.
ASCII| ASCII string| XXXXXXXXXXXX (device ID etc.)

Note
When key operation is used with the SS7012 and memory write mode or memory clear mode is enabled, commands are disabled and “CMD ERR” is returned. Before sending a command, always exit memory write mode or memory clear mode.
See: Memory write mode  Instruction manual of the SS7012 (p. 53) Memory clear mode  Instruction manual of the SS7012 (p. 61)

Command Message List

Sourcing
Normal source

Command Description Formats
FCC Set the source function.

The setting value will be 0, the output will be “OFF.” d1 (Setting of source function: 0 to 4)

  • 0: CV:2.5V
  • 1: CV:25V
  • 2: CC:25mA
  • 3: TC:0°C
  • 4: TC:RJ

| [Format] FCC d1
d1: NR1 format [Response] OK or CMD ERR
FCC?| Query the setting of source function.
The contents of responses are the same as the settings.| [Format] FCC? [Response] d1: NR1 format
OUT| Set the output ON or OFF d1 (Setting of output: 0 to 1)
0: OFF
1: ON

A CMD ERR response is issued in the following situations. Furthermore, bit2 of the ERR? Query is 1.

  1. When all memory is SKIP in recall source mode or scan source mode
  2. In the battery low status
  3. In TC:RJ function, when no temperature probe is connected or the measurement temperature is outside the -25°C to 80°C range
  4. In TC:RJ function, when the thermocouple type is B and the measurement temperature is less than 0 ºC

| [Format] OUT d1
d1: NR1 format
[Response] OK or CMD ERR
OUT?| Query the output setting The contents of responses are the same as the settings.| [Format] OUT?
[Response] d1: NR1 format
CVV| Set the source voltage value
This command is valid only in CV:2.5V or CV:25V. d1 (Source voltage value (in CV:2.5V): unit; V) -2.5000 to 2.5000
d1 (Source voltage value (in CV:25V): unit; V) -25.000 to 25.000| [Format] CVV d1 d1: NR2 format [Response] OK or CMD ERR
CVV?| Query the source voltage value
This command is valid only in CV:2.5V or CV:25V.
The contents of responses are the same as the settings.| [Format] CVV? [Response
d1: NR2 format or
d1: SKIP (ASCII format, when a skip setting in memory source)
Command| Description| Formats
---|---|---
CCA| Set the source current value
This command is valid only in CC:25mA. d1 (Source current value: unit; mA) 25.000 to 25.000| [Format] CCA d1 d1: NR2 format
[Response] OK or CMD ERR
CCA?| Query the source current value
This command is valid only in CC:25mA.
The contents of responses are the same as the settings.| [Format] CCA? [Response] d1: NR2 format or
d1: SKIP (ASCII format, when a skip setting in memory source)
TCC| Set the source thermoelectromotive force value This command is valid only in TC:0°C or TC:RJ. d1 (Thermocouple type: K, E, J, T, R, S, B, N) K(K thermocouple) temperature range: -174.0 to 1372.0 E(E thermocouple) temperature range: -220.0 to 839.0 J(J thermocouple) temperature range: -208.0 to 1108.0 T(T thermocouple) temperature range: -169.0 to 400.0 R(R thermocouple) temperature range: -50 to 1768
S(S thermocouple) temperature range: -50 to 1768 B(B thermocouple) temperature range: 300 to 1820
N(N thermocouple) temperature range: -113.0 to 1300.0 d2 (Temperature: unit; °C)
The temperature value of the above mentioned thermo- couple ranges| [Format] TCC d1, d2 d1: ASCII format d2: NR2 format [Response] OK or CMD ERR
TCC?| Query the source thermoelectromotive force value This command is valid only in TC:0°C or TC:RJ.
The contents of responses are the same as the settings.| [Format] TCC? [Response] d1: ASCII format d2: NR2 format or
d1: SKIP (ASCII format, when a skip setting in memory source)
ROV?| Query the overload status
d1 (Query of overload status: 0 to 1) 0: Overload not detected
1: Overload detected| [Format] ROV? [Response] d1: NR1 format

Output monitor

Command Description Formats
MON Set the monitor ON or OFF d1 (Setting of monitor: 0 to 1)

0: Monitor OFF
1: Monitor ON| [Format] MON d1 d1: NR1 format [Response] OK or CMD ERR
MON?| Query the setting of monitor
The contents of responses are the same as the settings.| [Format] MON? [Response] d1: NR1 format
RMV?| Query the current monitor value during output of voltage This command is valid only in CV:2.5V or CV:25V. d1 (Current monitor value: unit; mA) -28.00 to 28.00| [Format] RMV? [Response] d1: NR2 format or
CMD ERR: When the input is out of the measurement range.
RMC?| Query the voltage monitor value during output of current This command is valid only in CC:25mA. d1 (Voltage monitor value: unit; V) -28.00 to 28.00| [Format] RMC? [Response] d1: NR2 format or CMD ERR: When the input is out of the measurement range.
RMT?| Query the temperature monitor value
This command is valid only in TC:0°C or TC:RJ.
d1 (Temperature monitor value (in TC:RJ): unit; °C) -25.0 to 80.0
d1 (Temperature monitor value (in TC:0°C): unit; °C) 0.0| [Format] RMT? [Response] d1: NR2 format or
CMD ERR: When the input is out of the measurement range.

Memory source

Command Description Formats
MMD Set the source mode.

d1 (Setting of source mode: 0 to 2) 0: Normal source mode
1: Recall source mode 2: Scan source mode| [Format] MMD d1 d1: NR1 format [Response] OK or CMD ERR
MMD?| Query the setting of source mode.
The contents of responses are the same as the settings.| [Format] MMD [Response] d1: NR1 format
MEM| Saving setting value to memory (Refer to the caution on p.11)
Set the source function to store the setting value ahead of time.
This command is valid only in normal source mode. d1 (Address) 01 to 20 d2 (Source voltage value (in CV:2.5V): unit; V) -2.5000 to 2.5000
d2 (Source voltage value (in CV:25V): unit; V) -25.000 to 25.000
d2 (Source current value (in CC:25mA): unit; mA) -25.000 to 25.000
d2 (Source temperature value (in TC:0°C or TC:RJ): unit; °C) The temperature values for the following thermocouples

  • A SKIP string can be set for d2
    d3 (Thermocouple type: K, E, J, T, R, S, B, N) K (K thermocouple) temperature range: -174.0 to 1372.0 E (E thermocouple) temperature range: -220.0 to 839.0 J (J thermocouple) temperature range: -208.0 to 1108.0 T (T thermocouple) temperature range: -169.0 to 400.0 R (R thermocouple) temperature range: -50 to 1768 S (S thermocouple) temperature range: -50 to 1768 B (B thermocouple) temperature range: 300 to 1820 N (N thermocouple) temperature range: -113.0 to 1300.0| [Format 1] MEM d1, d2, d3 d1: NR1 format d2: NR2 format d3: ASCII format [Format 2] MEM d1, d2
    d1: NR1 format d2: NR2 format [Format 3] MEM d1, SKIP [Response] OK or CMD ERR
    MEM?| Query setting values saved to memory d1 (Source voltage value (in CV:2.5V): unit; V) -2.5000 to 2.5000
    d1 (Source voltage value (in CV:25V): unit; V) -25.000 to 25.000
    d1 (Source current value (in CC:25mA): unit; mA) -25.000 to 25.000
    d1 (Source temperature value (in TC:0°C or TC:RJ): unit; °C)
    Temperature value corresponding to each thermocouple d2 (Thermocouple type (in TC:0°C or TC:RJ)) K, E, J, T, R, S, B, N| [Format] MEM? d1 d1: Address (NR1 format)[Response] d1: Source value (NR2 format) d2: Thermocouple type (ASCII format, in TC:0°C or TC:RJ)
    MRM| Initialize setting values saved to memory (Refer to the caution on p.11)
    This command is valid only in normal source mode. d1 (Source function: 0 to 4)

    • 0: CV:2.5V
    • 1: CV:25V
    • 2: CC:25mA
    • 3: TC:0°C, TC:RJ
    • 4: ALL

| [Format] MRM d1 d1: NR1 format [Response] OK or CMD ERR
Command| Description| Formats
---|---|---
RCL| Set the address for recall sourcing
This command is valid only in recall source mode. d1 (Address: 01 to 20)| [Format] RCL d1 d1: NR1 format [Response] OK or CMD ERR
SCN| Set the start address for sourcing and sourcing time interval for scan sourcing (Refer to the caution on p.11)
This command is valid only in scan source mode. d1 (Start address for sourcing) 01 to 20 d2 (Sourcing time interval: unit; sec.) 01 to 99| [Format] SCN d1, d2
d1: NR1 format
d2: NR1 format [Response] OK or CMD ERR
SCN?| Query the scan sourcing The contents of responses are the same as the settings.| [Format] SCN?
[Response] d1: Address (NR1 format) d2: Sourcing time interval
(NR1 format)

Measurement

Command Description Formats
FCM Set the measure function. d1 (Setting of measure function: 0 to 4)

0: OFF

  • 1: V:2.5V
  • 2: V:25V
  • 3: A:25mA
  • 4: TEMP

| [Format] FCM d1 d1: NR1 format [Response] OK or CMD ERR
FCM?| Query the setting of measure function.
The contents of responses are the same as the settings.| [Format] FCM? [Response] d1: NR1 format
RDV?| Query the measured voltage value
This command is valid only in V:2.5V or V:25V. d1 (Measured voltage value (in V:2.5V): unit; V) -2.8000 to 2.8000 d1 (Measured voltage value (in V:25V): unit; V) -28.000 to 28.000| [Format] RDV? [Response] d1: NR2 format or CMD ERR: When the input is out of the measurement range.
RDC?| Query the measured current value This command is valid only in A:25mA. d1 (Measured current value: unit; mA) -28.000 to 28.000| [Format] RDC? [Response] d1: NR2 format or
CMD ERR: When the input is out of the measurement range.
RDT?| Query the measured temperature value This command is valid only in TEMP.
d1 (Measured temperature value: unit; °C) -25.0 to 80.0| [Format] RDT? [Response] d1: NR2 format or
CMD ERR: When the input is out of the measurement range.
ADJ| Zero adjustment
Zero adjust the current measurement value.
This command is valid when the measurement value is within ±100 counts in V:2.5V, V:25V or A:25mA.| [Format] ADJ [Response] OK or CMD ERR

Others

Command Description Formats
RBT? Query the battery low status

d1 (Query of battery low status: 0 to 1) 0: Battery low status not detected 1: Battery low status detected| [Format] RBT? [Response] d1: NR1 format
RRJ?| Query the temperature probe connection status This command is valid only in TC:RJ sourcing or temperature measurement.
d1 (Query of temperature probe status: 0 to 1)
0: Temperature probe not connected, the input temperature is outside the measurement range (of -25°C to 80°C).
1: Temperature probe connected| [Format] RRJ? [Response] d1: NR1 format
ERR?| Query the error register.
d1 (error register query: 0 to 255) bit7: unused bit6: MLE    Message length error bit5: HDE Message header error bit4: DFE     Data notation error bit3: DRE Data range error bit2: CNE Command not enforceable bit1: ISE Internal signal error bit0: BDE     Environment data corrupted| [Format] ERR? [Response] d1: NR1 format
IDN?| Query the Device ID.
d1: Query the Device ID.
Manufacturer’s name, Model name, Software version (Response example) HIOKI,SS7012, Ver 1.01| [Format]
IDN? [Response] d1: ASCII format

Caurion

Do not turn off the power switch of SS7012 while executing MEM, MRM or SCN command. Do not execute these commands while Battery low warning indicator flashing. These may damage SS7012 and cause the internal error.

Examples of sending commands

Sourcing
Normal sourcing

When sourcing 4 mA

Command Description
FCC 2 Sets [CC:25mA].
CCA 4 Sets the setting value to 4 mA.
OUT 1 Sets output to ON.
OUT 0 Sets output to OFF.

Output monitor
When monitoring the output current while sourcing 24 V.

Command Description
FCC 1 Sets [CV:25V].
CVV 24 Sets the setting value to 24 V.
OUT 1 Sets output to ON.
MON 1 Switches monitor on.
RMV? Query the output current.
MON 0 Switches monitor off.
OUT 0 Sets output to OFF.

Memory sourcing
When sourcing the setting value in address 01 with recall sourcing in the current source function.

Command Description
FCC 2 Sets [CC:25mA].
MMD 1 Sets recall source mode.
RCL 1 Sets the address to 1.
OUT 1 Sets output to ON.
OUT 0 Sets output to OFF.

Measurement

When performing voltage measurement at [V:25V]

Command Description
FCM 2 Sets [V:25V].
ADJ Execute the zero adjustment. Short out the ends of the test

leads.
RDV? | Query the measured voltage value.

All regional contact informationwww.hioki.com/

HIOKI E.E. CORPORATION
81 Koizumi, Ueda, Nagano 386-1192 Japan

Edited and published by HIOKI E.E. CORPORATION Printed in Japan

  • Contents subject to change without notice.
  • This document contains copyrighted content.
  • It is prohibited to copy, reproduce, or modify the content of this document without permission.
  • Company names, product names, etc. mentioned in this document are trademarks or registered trademarks of their respective companies. Europe only
  • EU declaration of conformity can be downloaded from our website.
  • Contact in Europe: HIOKI EURPOPE GmbH Helfmann-Park 2, 65760 Eschborn, Germany hioki@hioki.eu

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals