HIOKI DT4250 Series Digital Multimeter User Manual
- June 9, 2024
- HIOKI
Table of Contents
HIOKI E. E. CORPORATION
Aug. 2021
Revised edition 3
DT4250 series DIGITAL MULTIMETER Remote
Operation Manual
Remote Interface Overview
Step1. Connect the USB cable to the USB port on the computer.
Step2. Once the USB cable is connected and the USB driver is installed, a
virtual COM port is created on the computer.
Interface Parameters
In order to operate the meter via a host computer or terminal, the parameters
in interface within the DT4251,DT4252,DT4253,DT4254,DT4255,DT4256 have to
match the parameters the serial interface provided by the host or terminal.
The following procedures will guide to set up interface parameters within
these models.
Table.1
Item| Parameter|
Setting
---|---|---
1| Baud Rate| 9600 bps
2| Data Length| 8 bit
3| Parity Check| None
4| Stop Bit| 1 bit
About Command
3.1. Terminator
A terminator is a character sent by a host, which identified the end of a
command string.
A valid terminator consists of two-byte data.
Data Type| Explanation|
Example
---|---|---
Summary of Commands
Table.3
Command| Parameter|
Description
---|---|---
QPID| | Query the Meter model.
*IDN| | Query the Meter identification.
:SYST:RST| | Put the meter to reset.
:SYST:LLO| | Put the meter into the local lockout state.
:SYST:GTL| | Put the meter into the local state.
:CONF?| | Query the main function of the display.
:CONF2?| | Query the sub function of the display.
:CONF| <function,range>| Configure the range.
:FETCCNT?| | Query main measured count value.
:FETCCNT2?| | Query sub measured count value.
:CALC:STAT:MAX?| | Query measured maximum count value.
:CALC:STAT:MIN?| | Query measured minimum count value.
:CALC:STAT: AVER?| | Query measured average count value.
:CALC:REL:OFFS?| | Query the offset value of the relative.
:SYST:APS| <0|1>| Set the APS.
:SYST:BEEP| <0|1>| Set the action of the beeper.
:SYST:BLIT| <0|1>| Set the backlight.
:SYST:BLA| <0|1>| Set the auto backlight.
:SYST:BATT?| | Query the battery level.
:SYST:REL| <0|1>| Set the relative.
:SYST:FILTER| <0|1>,<100|500>| Set the filter.
:MEAS:AUTOV?| | Query the status DC or AC of AutoV function.
:SYST:INIT| | Put the meter to power-on-reset state.
:STAT?| | Query the status of the meter.
Detail of Commands
5.1. Description of Commands
The DT4251,DT4252,DT4253,DT4254,DT4255,DT4256 only accepts the UPPERCASE
command. (Except Unit etc.)
Table.4
Command|
Explanation
---|---
QPID| Query the Meter model. Syntax QPID Response “DT4251” or “DT4252” or
“DT4253” or ”DT4254” or “DT4255” or “DT4256”
IDN| Query the Meter identification. Syntax IDN?
Response <data×3>
(1) Since the internal count value is returned, the value beyond the display range is not guaranteed to be accurate.
:SYST:BLIT| Set the backlight.
Syntax :SYST:BLIT <data×1>
Response “
| A—Recording| “0”(OFF)
or “1”(MAX) or “2”(MIN) or “3”(AVG)
---|---
B—Relative value (REL)| “0”(OFF) or “1”(ON)
C—Filter| “0”(OFF) or “1”(ON)
D—Beep| “0”(OFF) or “1”(ON)
E—APS| “0”(OFF) or “1”(ON)
F—Battery level| “0”~“3”
G—Input warning| “0”(normal) or “1”(warn)
HI—Rotary position| “00”~“99”(from OFF)
J—HOLD| “0”(OFF) or “1”(ON)
K—Auto Hold| “0”(OFF) or “1”(ON)
L—Auto Range| “0”(OFF) or “1”(ON)
M—Backlight| “0”(OFF) or “1”(ON)
N—Backlight Auto OFF| “0”(OFF) or “1”(ON)
O—Filter Cut-off frequency| “0”(100Hz) or “1”(500Hz)
P—reserved| “0”
Q— reserved| “0”
R— reserved| “0”
S— reserved| “0”
T— reserved| “0”
U— reserved| “0”
V— reserved| “0”
W—reserved| “0” or “1”
X—reserved| “0”
5.2. Combination response string function and range
Table.5
Function
|
Range
---|---
ACV| 6, 60, 600, 1000
DCV| 600m(※1), 6, 60, 600, 1000
DCmV| 600m
AutoV| 600
CONT| 600
RES| 600, 6k, 60k, 600k, 6M, 60M
CAP| 1u, 10u, 100u, 1m, 10m
DIODE| 1500
TEMP| 400
CLAMP| 10, 20, 50, 100, 200, 500, 1000
ACA| 600m(※2), 6, 10
DCA| 60m(※3), 600m(※4), 6, 10
DCmA| 6m, 60m
DCuA| 60u, 600u
VDET| 0 (Lo, Hi) (※5)
FREQ| 100, 1k, 10k, 100k
(※1)600mV range is for DT4251 or DT4253 or DT4254 or DT4255 or DT4256.
(※2)600mA range is only DT4256
(※3)60mA range is only DT4256
(※4)600mA range is only DT4256
(※5)1 is for DT4254 or DT4255 or DT4256
5.3. Count value of abnormal data
Response of count value query return the values shown in Table.6 in the
case of abnormal data.
Table.6
Type of abnormal value | Count value |
---|---|
Over Range value | 1000000 |
Invalid data | 2000000 |
Open value (DT4253, TEMP function) | 3000000 |
Internal error value (DT4253, TEMP function) | 4000000 |
6. Compatible command with earlier products
Table.7 shows the command compatibility with earlier products 3800
series.
Table.7
Command | Description |
---|---|
*CLS | Clear the system Error Queue. |
*RST | Put the meter to power-on-reset state. |
LLO | Put the meter into the local lockout state when in remote control. This |
means no local key operation at the front panel is allowed during remote
control.
GTL| Put the meter into the local state, clearing the remote state and front
panel lockout.
FETC? [@2]| Return the primary or secondary function value of output butter.
Response:
Please note
-1. Contents subject to change without notice.
-2. This document contains copyrighted content.
-3. It is prohibited to copy, reproduce, or modify the content of this document without permission.