HIOKI P9000-01 Differential Probe Instruction Manual

June 1, 2024
HIOKI

P9000-01 Differential Probe

“`html

Product Specifications:

  • Model: P9000-01, P9000-02

  • Type: Differential Probe

  • Input Voltage Range: High voltages divided by 100 or 1000

  • Output: Signal relative to ground

  • Additional Feature: P9000-02 includes AC RMS value output
    functionality

  • Power Supply: Optional AC adapter or USB cable

Product Usage Instructions:

1. Verifying Package Contents:

Upon receiving the product, carefully inspect the package
contents for any damage during shipping. Verify that all
accessories and connectors are included.

2. Cleaning:

If the product gets dirty, use a soft cloth moistened with water
or a neutral detergent to wipe it clean. Avoid using solvents like
benzene or alcohol as they can damage the product.

3. Repair:

If the product malfunctions, ensure proper connections of the AC
adapter and USB cable before contacting your authorized Hioki
distributor or reseller for assistance.

4. Calibration:

Periodic calibration is essential for accurate measurement
results. Determine a calibration period based on usage conditions
and environment, and have the product regularly calibrated by
Hioki. Contact your Hioki distributor for calibration services.

5. Safety Notes:

Read and understand the safety notes provided in the manual to
prevent bodily injury, product damage, or other hazards. Do not
attempt to modify or repair the product yourself.

Frequently Asked Questions (FAQ):

Q: What is the warranty period for the Hioki P9000-01, P9000-02

Differential Probe?

A: The warranty is valid for three (3) years from the date of
purchase under normal use conditions as per the Instruction Manual
and Product Precautionary Markings.

Q: How can I supply power to the probe?

A: Power can be supplied using the optional AC adapter or a
commercially available USB cable.

Q: What should I do if the product seems damaged upon

arrival?

A: If you suspect any damage or if the product is not
functioning correctly, contact your authorized Hioki distributor or
reseller for further assistance.

“`

P9000-01 P9000-02
DIFFERENTIAL PROBE
Instruction Manual
EN
Apr. 2021 Revised edition 2 P9000A981-02 21-04H
www.hioki.com/
[email protected]
Warranty
Warranty malfunctions occurring under conditions of normal use in conformity with the Instruction Manual and Product Precautionary Markings will be repaired free of charge. This warranty is valid for a period of three (3) years from the date of purchase. Please contact the distributor from which you purchased the product for further information on warranty provisions.
Introduction
Thank you for choosing the Hioki P9000-01, P9000-02 Differential Probe. Preserve this manual carefully and keep it handy to make full use of this product for along time.
Overview
The product is a differential probe that is connected to a Memory HiCorder input unit. The probe divides two high voltages input to its high and low terminals by 100 or 1000 and then outputs the difference as a signal relative to ground. The P9000-02 includes AC RMS value output functionality. Power can be supplied via the optional AC adapter or a commercially available USB cable.
Verifying Package Contents
When you open the package, carefully inspect the product to ensure that everything is in good condition, and that no damage occurred during shipping. Carefully check the accessories and connectors. If the product seems to have been damaged or does not work as specified, contact your authorized Hioki distributor or reseller.
Cleaning
If the product becomes dirty, slightly moisten a soft cloth moistened with water or a neutral detergent and wipe the product clean.
IMPORTANT Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case.

Repair
If the product seems to be malfunctioning, confirm that the AC adapter and the USB cable are connected properly before contacting your authorized Hioki distributor or reseller.
WARNING
Touching any of the high-voltage points inside the product is very dangerous. Do not attempt to modify, disassemble, or try to repair the product. Doing so may cause a fire, electric shock, or injury.
Calibration
IMPORTANT Periodic calibration is necessary in order to ensure that the instrument provides correct measurement results of the specified accuracy.
The calibration period varies with the conditions and environment of use. It is recommended to determine a calibration period based on those factors and to have the product regularly calibrated by Hioki. Please contact your Hioki distributor to have your product periodically calibrated.
Precautions during shipment
Be sure to follow these precautions when transporting the product: · Remove accessories and optional equipment from the product
in order to avoid damage. Additionally, be sure to double box the product. Accidental damage suffered in transit is not covered by the warranty. · Attach a description of the issue when sending out your product for repair.
Disposal
Dispose of the product in accordance with local regulations.
Safety Notes
This product is designed to conform to IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, using the product in a way not described in this manual may negate the provided safety features. Carefully read the following safety notes before using the product.
WARNING
Mishandling product could result in bodily injury or even death, as well as damage to the product. Familiarize yourself with the instructions and precautions in this manual before using the product.
WARNING
· Protective Gear Performing measurement using this product involves live-line work. To prevent an electric shock, use appropriate protective insulation and adhere to applicable laws and regulations.
· Electricity can potentially cause serious events such as an electric shock, heat generation, fire, and an arc flash due to a short-circuit. If you have not used electrical measuring instrument before, you should be supervised by a technician who has experience in electrical measurement.

Notations
In this document, the severity levels of risk and hazard are classified as follows.

DANGER

Indicates an imminently hazardous situation that will result in death of or serious injury to the
operator.

WARNING

Indicates a potentially hazardous situation that may result in death of or serious injury to the
operator.

CAUTION

Indicates a potentially hazardous situation that may result in minor or moderate injury to the operator or damage to the product or malfunction.

IMPORTANT

Indicates information or content that is particularly important from the standpoint of operating or maintaining the product.

Indicates cautions and hazards. When the symbol is printed on the product, refer to a corresponding topic in the Instruction Manual.

Indicates the ON side of the power switch.

Indicates the OFF side of the power switch.

Indicates a grounding terminal. Indicates DC (Direct Current). Indicates an action that must not be performed.

Indicates action that must be performed.
Indicates that the product conforms to regulations required by the EU Directive.
Indicates the Waste Electrical and Electronic Equipment Directive (WEEE Directive) in EU member states.

Measurement categories
To ensure safe operation of measuring instruments, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT II to CAT IV, and called measurement categories.
WARNING
· Using a measuring product in an environment designated with a higher- numbered category than that for which the product is rated could result in a severe accident, and must be carefully avoided.
· Never use a measuring product that lacks category labeling in a CAT II to CAT IV measurement environment. Doing so could result in a serious accident.

This product conforms to the safety requirements for CAT III 1000 V, CAT IV 600 V measuring instruments.

CAT II: CAT III: CAT IV:

When directly measuring the electrical outlet receptacles of the primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.) When measuring the primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets When measuring the circuit from the service drop to the service entrance, and to the power meter and primary overcurrent protection device (distribution panel)

Distribution panel Service entrance Service drop
CAT IV Power meter

Internal wiring

CAT II

CAT III

T

Outlet

Fixed installation

Usage Notes
Observe the following precautionary information to ensure that the product can be used safely and in a manner that allows it to perform as described in its specifications.
WARNING
To prevent an electric shock, confirm that the white or red portion (insulation layer) inside the cable is not exposed. If a color inside the cable is exposed, do not use the cable.
WARNING
Do not allow the product to get wet, and do not take measurements with wet hands. This may cause an electric shock.
CAUTION
· To avoid damage to the product, avoid subjecting it to vibration or mechanical shock during transportation and handling. Exercise particular care to avoid subjecting the product to mechanical shock, for example by dropping.
· Do not store or use the product where it could be exposed to direct sunlight, high temperatures or humidity, or condensation. Under such conditions, the product may be damaged and insulation may deteriorate so that it no longer meets specifications.
· This product is not designed to be entirely waterproof or dustproof. Do not use it in an especially dusty environment, nor where it might be splashed with liquid. This may cause damage.
This product complies with EN 61326 Class A. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.

HIOKI P9000A981-02

Parts Names

3 Step

Alligator clip
Maximum rated voltage to earth: 1000 V (CAT lll) 600 V (CAT IV)

4 Step

Red lead (HIGH)

POWER switch Selects the voltage division ratio (power on). Turns off the product.
5 Step
Output selector switch (Model P9000-02 only)

1 Step

Power jack Connect a model Z1008 AC adapter.

7 Step
To measurement target
(pictured: P9000-02)
Black lead (LOW)
OVER LED Lights up when excessively high input is received. POWER LED Lights up when the product is receiving power.

BNC connector

USB-microB jack Connect a USB cable (commercially available).

6 Step

Range Scaling setting

2 Step

For more information about devices that can be connected, contact your authorized Hioki distributor or reseller.

Connected instrument (pictured: Model MR8875)
Inspection Before Use
To ensure that the product is properly operating, conduct an inspection and check product operation before using product to ensure that no damage has occurred during storage or transport. If there is any damage to the product, contact your authorized Hioki distributor or reseller for repair.
WARNING
Confirm that the insulation on the using the probes are undamaged and that no bare conductors are improperly exposed before using the product. If there is any damage to the product that leads to an electric shock, contact your authorized Hioki distributor or reseller for repair.

Measurement Method
WARNING
· For your safe operation, do not connect the product to the primary of the distribution panel.
· Do not connect the product to a voltage or current that exceeds the rating indicated on the product labeling or the measurement range indicated in its specifications. Doing so may damage the product or cause it to become hot, resulting in electric shock or bodily injury.
· To avoid short circuits and electric shock, do not short-circuit two wires to be measured by bringing the metal part of the test lead into contact with them. Never touch the metal end.
WARNING
When supplying power through the USB-micro B jack, make sure that the USB ground pin of the power source is connected to the ground. Using an ungrounded USB terminal could cause an electric shock.
CAUTION
To prevent damage to the connector, be sure to release the locking mechanism, grip the head of the connector (not the cord), and pull it out.
1 Select which of the following two sources you will
use to power the product:
Method 1 Connect a model Z1008 AC adapter

Model Z1008

To commercial power supply

Method 2 Connect a USB cable (commercially available, USB (type A male)USB (type B female))

To USB terminal on connected instrument

2 Connect the product’s BNC connector to the input
terminal on the connected instrument. (1) Line up (2) Lock
Line up

3 Connect the alligator clips to the input leads.

Purchase the tip option that is best suited to the measurement target. These options can be purchased via authorized Hioki distributor or reseller.
CAUTION
Precautions when using the L4932 Test Pin Set To prevent a short circuit accident, be sure to use the test leads with the sleeves attached when performing measurements in the CAT III measurement category. (For the measurement categories, see “Measurement categories”.) If the sleeves are inadvertently removed during measurement, stop the measurement.
600V 1000V
600V 1000V

IMPORTANT
Use only the specified tip options. Using a non-specified tip option may result in incorrect measurements due to poor connection or other reasons.

4 Select the voltage division ratio with the POWER

switch.

Voltage division ratio

Description

1/1000

Divides the high signal and low signal by 1,000 and then outputs the difference between the two signals as a signal relative to ground.

1/100

Divides the high signal and low signal by 100

and then outputs the difference between the two signals as a signal relative to ground.

5 Select the output type (P9000-02 only).

Output Wave

Description
Outputs the signal using the selected voltage division ratio. Choose this setting when you wish to view the power supply waveform.

AC-RMS

Outputs the RMS value using the selected voltage division ratio. Input is AC coupled.

6 Set the input range scaling for the connected in-

strument.

7 Connect the alligator clips to the measurement target.

8 After measurement is complete, turn off the POW-

ER switch.

Specifications

Product warranty duration
Accuracy guarantee duration
DC output accuracy
Temperature characteristic
Frequency characteristic
Voltage division ratio

3 years (except for breaks in cables caused by use)
1 year
±0.5% f.s. f.s. = 1.0 V (ratio 1000:1), 3.5 V (ratio 100:1) ±0.05% f.s./C DC to 100 kHz, -3 dB
1000:1, 100:1 (user-switchable)

Common-mode 80 dB or more (50 Hz/60 Hz shorted input) rejection ratio

Input resistance 10.5 M ±1% (between HIGH and LOW terminals)

Input capacitance 5 pF or less (at 100 kHz)

Input jacks

Banana plug

Output connector Insulated BNC

Measuring range 1000 V AC/DC (ratio 1000:1) 250 V AC, 350 V DC (ratio 100:1)

Maximum input 1000 V AC/DC voltage

Maximum rated line- 1000 V AC/DC, measurement category III to-ground voltage anticipated transient overvoltage: 8000 V

Operating temper- Temperature: -40C to 80C/ -40F to 176F

ature and humidity Humidity: -40C to 40C [non-inclusive]

range

/ -40F to 104F, 80% RH or less

40C to 45C [non-inclusive]

/ 104F to 113F, 60% RH or less

45C to 80C/ 113F to 176F,

50% RH or less (no condensation)

Storage tempera- Temperature: -40C to 80C/ -40F to 176F ture and humidity Humidity: 80% RH or less (no condensation) range

Operating environment

Indoors, Pollution Degree II, altitude up to 2000 m (6562 ft.)

Dimensions

Approx. 128.0W×36.0H×22.0D mm

/ Approx. 5.04W×1.42H×0.87D

Weight

Approx. 170 g/ 6.0 oz.

Power supply

(1) Model Z1008 AC adapter

(12 V DC, DC jack 5.5×2.1 mm)

Rated supply voltage: 100 V AC to 240 V AC

(Voltage fluctuations of ±10% from the rated supply

voltage are taken into account.)

Rated supply frequency: 50 Hz/60 Hz

Anticipated transient overvoltage: 2500 V

(2) USB bus power (5 V DC USB-microB connector)

(3) External power supply, 2.7 V DC to 15 V DC

Maximum rated power

(1) When using a model Z1008 AC Adapter: 6 VA (model P9000-01/-02 with a AC adapter)

0.9 VA (model P9000-01/-02 only)

(2) When using a USB bus power: 0.8 VA

(3) When using a 15 V DC external power supply: 1 VA

Effects of radiated radio-frequency electromagnetic field

±5% f.s. max. at 3 V/m

Effects of conducted radio-frequency electromagnetic field

±5% f.s. max. at 3 V

Standards

EMC EN61326 Class A Safety EN61010

Accessories

Instruction manual ×1 Alligator clips×1 set (red and black/ one each) Carrying case×1

Options

The following options are available for the product. To purchase an option, please contact your authorized Hioki distributor or reseller. Options are subject to change. Please check Hioki’s website for the latest
information.

Z1008 AC Adapter
L9243 Grabber Clip (1000 V CAT II, 1 A) L4936 Bus Bar Clip Set (600 V CAT III, 5 A) L4937 Magnetic Adapter Set (1000 V CAT III, 2 A) L4931 Extension Cable Set (For extending the length of the test leads, 1.5 m)
(600 V CAT IV, 1000 V CAT III, 10 A) L4932 Test Pin Set (600 V CAT IV, 1000 V CAT III, 1000 V CAT II, 10 A) L4934 Small Alligator Clip Set (300 V CAT III, 600 V CAT II, 3 A)

(L4932 is required when using L4934.)

Model P9000-02 only

Measurement modes Convertible voltage RMS value measurement accuracy
RMS value measurement response time

Waveform, AC-RMS (selected with a slide switch)
1000 V AC (ratio 1000:1), 250 V AC (ratio 100:1)
±1% f.s. (30 Hz to 1 kHz [non-inclusive], sine wave) ±3% f.s. (1 kHz to 10 kHz, sine wave) f.s. = 1.0 V (ratio 1000:1), 2.5 V (ratio 100:1)
300 ms Rising 0% f.s. 100% f.s. Response time for input from 0% f.s. to 90% f.s. 600 ms Falling 100% f.s. 0% f.s. Response time for input from 100% f.s. to 10% f.s.

f.s.: maximum display value or scale length (This is usually the maximum value of the currently selected range.)

HIOKI P9000A981-02

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

HIOKI User Manuals

Related Manuals