Dr meter LX1010BS Handheld Split Type Design Digital Light Meter Instruction Manual

June 17, 2024
Dr meter

Digital llluminance Meter
Model: LX1010BS
Operation Manual

Instruction

Your purchase of this DIGITAL LUX METER marks a step forward into the field of precision measurements.
Although this lux meter is a complex and delicate instrument, its ruggedness makes it durable if it is operated properly. Please read the following instructions carefully and keep this manual handy.

Features

  • Precise and easy reading.

  • High accuracy in measurement.

  • LSI-circuit use provides high reliability and durability.
    IVWide range of light measurements.

  • LOW BATTERY indicator.

  • Auto zero adjustment.

  • LCD display consumes low power and can be read clearly even in high ambient light.

  • Detached LIGHT SENSOR enables user to take measurements at an optimum position.

Specifications

Display 31/2 digits LCD display, 1,999 counts
Measuring Range 1-100,000L0x. (3 Ranges)

2,000 Lux range: reading x 1
20,000Lux range: readingx 10
100,000Lux range: readingxl 00

Range Resolution Accuracy (23±5°C)
0- 2,000 Lux 1 Lux ± (4%rdg+2d)
2,000- 19,999 Lux 10 Lux + (4%rdg+2d)
20,000-100,000 Lux 100 Lux +±(5%+2d)

NOTE: Accuracy tested by a standard parallel light tungsten lamp of 2854°K temperature

Repeatability ± 2%

Temperature
Characteristic| ±0.1d/C
Testrate| 0.2 times/sec
Photo detector| One silicon photo diode with filter
Operation
temperature-humidity| -10C to 40C (32 F -104 “F) 0-70%Rh
Storage
temperature-humidity| -10C to 50C (14 “F -140 F) 0-80%Rh D159
Over-input| Indication of “1” (2,000Lux/20,000Lux)
Indication of “OVER” (100,000Lux)
Dimension| 1 06x57x26mm (Photo detector)
130x72x30mm (Meter body)
1 50cm (Photo detector lead)
Weight| 170g
Power Supply| One 9V battery, Consumption current approx.2mA
Accessories| Instruction-manual , battery, carrying case

Name of Parts & Positions

Dr meter LX1010BS Handheld Split Type Design Digital Light Meter - Name of
Parts & Positions

  1. LCD display
  2. Turn off, turn on, value hold
  3. Range select button
  4. Photo detector

Operating Instructions

  1.  Install the battery, and then switch the button 2 to ON”
  2. Push down the range selection switch to desired range
  3. Remove the photo detector cap and face it to the light source in a horizontal position
  4. Read the test value from the LCD display.
  5. Out of range: If the instrument only displays one 1 in the LCD, it means the light source is too bright, and a higher range should be selected.
  6. Data-Hold mode: Switch the button 2 to “HOLD” to select Hold mode. When “HOLD” mode is selected, the LUX meter stops all further measurements and the test value will be hold on the display. Switch the button HOLD” to the ON’, the hold value will be cancelled.
  7. When the measurement is completed, snap on the cap on the photo detector and switch the power selector (button 2) t0 “OFF”

Battery Check-up & Replacement

  1. It is necessary to replace with a new 9V battery, when left corner of LCD display show “ “
  2. After turning off the meter, remove the battery cover
  3. Disconnect the battery from the meter and replace it with a standard 9-volt battery and press the cover back into the place

Spectral Sensitivity Characteristic

Dr meter LX1010BS Handheld Split Type Design Digital Light Meter - pectral
Sensitivity Characteristic

Maintenance

  1. Clean the white plastic dome on the top of the detector with a damp cloth when necessary.
  2. Don’t store the instrument where temperature of humidity is excessively high.

Recommended Illuninaton

OFFICE Conference, reception room 200 ∼ 750Lux
Clerical work 700 ∼ 1,500Lux
Typing drafting 1,000 ∼ 2,000Lux
FACTORY Packing work, entrance passage 150 ∼ 300Lux
Visual work at production line 300 ∼ 750Lux
Inspection work 750 ∼ 1,500Lux
Electronic parts assembly line 1,500 ∼ 3,000Lux
HOTEL Public room, cloakroom 100 ∼ 200Lux
Reception, cashier 220 ∼ 1,000Lux
STORE Indoors stairs corridor 150 ∼ ‘ 200Lux
Show window, packing table 750 ∼ 1,500Lux
Forefront of show window 1,500 ∼ 3,000Lux
HOSPITAL Sickroom, warehouse 100 ∼ 200Lux
Medical examination room 300 ∼ 750Lux
Operation room, emergency treatment 750 ∼ 1,500Lux
SCHOOL Auditorium, indoor gymnasium 100 ∼ 300Lux
Class room 200 ∼ 750Lux
Laboratory, library, drafting room 500 ∼ 1,500Lux

FCC Statement

This device complies with Part 1 5 of the FCC Rules.
Operation is subject to the following two conditions:

  1. this device may not cause harmful interference, and
  2. this device must accept any interference received, including interference that may cause undesired operation.

Learn more about the UK Declaration of Conformity, Please click below link to download.
http://files.drmeter.com/lxl010bs-uk-doc.pdf
Learn more about the EU Declaration of Conformity, Please click below link to download.
http://files.drmeter.com/161-61010-07-DOC-UK.pdf

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals