hygiena KIT230042 Strip Kit User Guide

June 17, 2024
Hygiena

hygiena KIT230042 Strip Kit User Guide

hygiena KIT230042 Strip Kit User Guide

PCR kit for the qualitative detection of Escherichia coli O157 DNA using real- time PCR instruments.
Before starting, it is strongly recommended to read the entire product manual available on our website.

PROGRAM SETUP

Program your real-time PCR instrument before setting up the PCR reactions. Select the following channels:
⇒ FAM (E. coli O157) and VIC (Internal Control).
As an alternative to VIC, HEX can be used. For the PikoReal® 24, Yakima Yellow must be selected.

hygiena KIT230042 Strip Kit - PROGRAM SETUP

For some real-time PCR instruments the probe quencher as well as the use of a passive reference dye must be specified. This kit contains probes with TAMRA as quencher and no passive reference dye.

DATA INTERPRETATION

Verify results of positive (Control Template) and negative (H2O) controls, before interpreting the sample results. Always compare samples to positive and negative controls. Review data from each channel and interpret results
as described in the table.

hygiena KIT230042 Strip Kit - DATA INTERPRETATION

foodproof® is a registered trademark of Hygiena. The above mentioned real-time PCR instruments are registered trademarks of their respective holders.

PREPARATION OF THE PCR MIX

Take appropriate precautions to prevent contamination, e.g., by using fi lter tips and wearing gloves.
Thaw reagents, mix (do not vortex!) and briefl y spin vials before opening.

1. PREPARE PCR MIX

Add 18 µL of Master Mix (yellow cap), 1 µL of Enzyme Solution (red cap) and 1 µL of Internal Control (white cap) for each reaction to a suitable tube (n samples + 2 controls + at least one additional reaction to cover pipetting loss).
Mix carefully but thoroughly by pipetting up and down.

2. ADD PCR MIX

Pipet 20 µL of prepared PCR mix into each strip or plate well.

3. ADD SAMPLES AND CONTROLS

Pipet 5 µL of samples, Negative Control (colorless cap) or Control Template (purple cap) into respective wells.

4. SEAL

Seal strips/plate accurately.

5. CENTRIFUGE

Briefly spin strips/plate in a suitable centrifuge.

6. START REAL-TIME PCR RUN

Cycle samples as described above.

foodproof®
E. coli O157 Detection Kit

KIT230042
Kit for 96 reactions
Store kit at -15 to -25 °C

For food testing purposes
FOR IN VITRO USE ONLY
Made in Germany

Hygiena® | Camarillo, CA 93012 USA | diagnostics.support@hygiena.com | www.hygiena.com

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals