godrej GBR 100 DC Blood Refrigerator
Product Information
The product is a multifunctional device that offers a wide range of features and capabilities. It is designed to provide convenience and efficiency in various tasks. Whether you need to perform basic functions or more advanced operations, this product is equipped to meet your needs.
Specifications
- Model: [Model Number]
- Dimensions: [Dimensions]
- Weight: [Weight]
- Power Supply: [Power Supply]
- Display: [Display Specifications]
- Connectivity: [Connectivity Options]
Features
The product comes with a variety of features that enhance its
functionality:
- Feature 1: [Description of Feature 1]
- Feature 2: [Description of Feature 2]
- Feature 3: [Description of Feature 3]
- Feature 4: [Description of Feature 4]
Included Accessories
The product package includes the following accessories:
- Accessory 1: [Description of Accessory 1]
- Accessory 2: [Description of Accessory 2]
- Accessory 3: [Description of Accessory 3]
Product Usage Instructions
-
Step 1: Unboxing
Start by opening the product package carefully. Remove all packaging materials and verify that all the included accessories are present. -
Step 2: Powering On
Connect the product to a power source using the provided power supply. Press the power button located [location of power button] to turn on the device. -
Step 3: Initial Setup
Follow the on-screen instructions to complete the initial setup process. This may include selecting language preferences, connecting to a Wi-Fi network, and signing in with your user account.
Step 4: Basic Functions
The product offers various basic functions that can be accessed through the
intuitive user interface. These functions may include:
- Function 1: [Description of Function 1]
- Function 2: [Description of Function 2]
- Function 3: [Description of Function 3]
Step 5: Advanced Operations
If you wish to explore more advanced operations, refer to the product’s
user manual for detailed instructions on specific features and settings.
Step 6: Maintenance and Care
To ensure optimal performance and longevity of the product, follow these
maintenance guidelines:
- Guideline 1: [Maintenance Guideline 1]
- Guideline 2: [Maintenance Guideline 2]
- Guideline 3: [Maintenance Guideline 3]
Frequently Asked Questions (FAQ)
-
Q: How do I update the software?
A: To update the software, go to the Settings menu and look for the “Software Update” option. Follow the on-screen instructions to download and install the latest software version. -
Q: Can I connect the product to my mobile device?
A: Yes, the product supports connectivity with mobile devices. Install the companion app on your mobile device and follow the instructions for pairing and connecting. -
Q: How do I clean the product?
A: Use a soft, lint-free cloth slightly dampened with water or a mild cleaning solution to clean the exterior surfaces of the product. Avoid using abrasive materials or harsh chemicals that may damage the device.
Installer’s instruction manual
- Keep these instructions close to the refrigerator.
- “For water quantity, refer page 31.
GBR 100 DC – Solar Direct Drive Blood Bank Refrigerator
MODEL NUMBER: GBR 100 DC
Sure Chill9 is a patented technology, licensed by The Sure Chill Company
——————————-
GBR 100 DC – Solar Direct Drive Blood Bank Refrigerator
www.godrejappliances.com/medical
SAFETY INSTRUCTIONS
- Do not climb on appliances
- Be careful not to trap fingers
- Do not keep flammable substances near appliance
- Servicing must be carried out by qualified maintenance technician
- Caution! Appliance contains refrigerant and electrolyte capacitor
- Warning! Appliance is solar powered all parts are live in daylight
TOOLS REQUIRED
- Screwdrivers
- Spanners
- Hammer
- Knife
- Multimeter
HANDLING INSTRUCTIONS
- Keep appliance vertical- do not tilt more than 30°
- Keep door closed when moving appliance
- Comply with manual handling regulations when moving appliance
- Empty the appliance of liquid before moving
MOUNTING FRAME COMPONENTS
- Module end and middle clamps
- Mounting rail
- Foot
- Security bolts, washers and nuts
INSTALLATION ADVICE
- Take precautions when working at height
- Keep solar modules completely covered during installation process
- Use appropriate fixings to attach array to surface
- Not all mounting frame components may be required for installation
SOLAR PANEL INSTALLATION DETAILS
SOLAR ARRAY INSTALLATION MANUAUL
Important Safety Instruction
- Take precautions when working at height
- Make sure that the mounting surface is strong enough to carry your weight
- Use fall restraint system when working on roofs
- PV modules generate high voltages in sunlight – cover all modules until all wires are connected
Important Note
Use mounting system supplied to tilt solar modules to 15° or the locations
latitude, whichever is greater
SOLAR ARRAY INSTALLATION MANUAUL
Attach feet to the surface so that they are correctly spaced for required
angle
Angle | x(mm) | 2x(mm) |
---|---|---|
15· | 1340 | 2680 |
20· | 1285.5 | 2571 |
25· | 1214 | 2428 |
30· | 1124 | 2248 |
35· | 1012 | 2024 |
SOLAR ARRAY INSTALLATION MANUAUL
- Solar modules can be installed on a flat or sloping surface facing the equator
- Install solar modules in an unshaded location
- Only put fixing in to solid structure
- Use proper fixings to attach array to surface
- Remove all the cardboard packing before installation
- Ensure all electrical connections are firm with correct polarity
LIST OF ITEMS
Solar kit item list
INSTALLATION PROCEDURE
- Unpack the solar panel box
- Insert foot in the lower end of the rail
- Fix the foot with M8X70 bolt
- Insert and fix foot in smaller leg
- Insert and fix another foot in longer leg
- Fix all the feet
- Fix and tighten all the feet and erect the frame
- Insert the angle brace as shown in figure
- Similarly fix all the feet on rail on the other side
- After fixing all the feet in the mounting rails, attach feet of the frame to the supporting surface using the appropriate fixing – expansion bolts for concrete, coach screws for timber joists, J-bolt for steel sections, etc.
- Insert and slide the solar panel mounting end clamps on both sides with proper direction
- Rest the solar panel on the lower part of the frame with the help of mounting clamps and tighten them with theft deterrent fasteners using theft deterrent key as shown in above figure
- Take solar panel mounting mid clamp and slide spring nut inside the slotted rail on the both sides
- Fix the middle solar panel mounting clamp on both sides and rest the other solar pane on the frame
- Fix and tighten the bracket with the help of theft deterrent fastener after making sure both the panels are clamped and slide the top clamp on both sides
- Clamp the solar panel with clamps and tighten them with theft deterrent bolts
- Connect the connectors from the solar panel to the branch connector, ensure proper polarity
- Fix/ Clamp earth cable (20Mtrs) in the lay in lug fixed to the vent opening
- Similarly fix lay in lug to the solar array angle and clamp the other (5mtr) earth cable
EARTHING THE ARRAY
- Bond the solar modules and mounting frame together
- On the following pages we indicate how we would advise you to install an earthing rod.
- However we appreciate that existing methods in your locality might serve very well and you might well choose to follow those.
OPTIONS
- If the ground in the locality is solid, then using equipment provided, we suggest using a charcoal and salt pit
- Excavate to a depth of about 1.35m (top of your chest), with a diameter of at least 1m
- Dampen the bottom of the pit. Put a layer of charcoal with a layer of salt on top such that the remaining depth is about 1.2m (the earthing rod should just stick out above ground level); then a further good soaking of water.
- If you so wish, you can make this more effective by adding the following features (although you will have to source these components as we don’t supply them)
- After the last layer of salt has been laid, as the remaining depth is filled with earth, insert a pipe and connecting funnel or bucket. The funnel or bucket should be mounted such that it can be covered and not get damaged. By using this funnel and pipe, water can be delivered to the top layers of charcoal and salt, which will then fall naturally through the layers.
- However if the ground conditions are regularly damp but not overly wet, you could use the following method if you so choose
CONNECTING SOLAR MODULES
- Ensure the solar modules are completely covered before touching solar cables!
- Remove cover from solar modules and test the voltage with a multimeter
POSITIONING INSTRUCTIONS
- Position appliance to allow free air movement all a round
- Position appliance away from direct sunlight
- Allow for access to remove appliance lid
- Ensure unit is level before filling
- Check that support bridge and condensation baffle are correctly positioned
- Note: The baffle is designed to be loose
FILLING INSTRUCTIONS
- Remove ventilated lid and screw cap and fill with clean drinking water 230mm from top of pipe
- Observe safe working practices when filling the appliance
- Do not overfill the unit
- Replace screw cap and securely attach ventilated lid before starting the compressor
Fill the Device with 135L clean drinking water. Good examples of clean drinking water would be municipal water ( mains water supply), Reverse Osmosis (RO) water, distilled water or clean, bottled drinking water. The water should not be dirty nor contaminated. A mild dose of chlorine, consistent with drinking water quality standards, is helpful in disinfecting both the water and the container.
For a full description refer to the latest World Health Organisation publication on the subject:
- Guidelines for drinking-water quality, fourth edition
- Editors: WHO
- Publication date: 2011
- Languages: English
- ISBN: 978 92 4 154815 1
- http://www.who.int/watersanitationhealth/publications/dwqguidelines/en/
CONNECTING APPLIANCE
- Connect the appliance to the power isolator using cable and fixings provided
- Please double check all the system’s visible connections done, in both the factory and your location
- Allow 24 hours after transportation before starting the compressor
- Confirm appliance is operational by checking temperatures are displayed on the solar display* and check door is securely closed
- Appliance cools slowly and will reach safe storage temperature within 6- 8 days after commissioning.
CUSTOMER SUPPORT
- Godrej & Boyce Manufacturing Company Limited,
- Pirojshanagar,
- Vikhroli West,
- Mumbai – 400 079
- http://www.godrejappliances.com/medical
- Board: +91 22 6796 5656 I 5959
- Fax: +91 22 6796 1700 / 1800
- E-mail:
- Sales Support: medicalref@godrej.com / hello@surechill.com
- Service Support: smartcare@godrej.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>