Product Information

Specifications:

  • Detector Type: Multi Gas Clip Pump
  • Detectable Gases: Hydrogen sulfide (H2S), carbon monoxide (CO), oxygen (O2), combustible gases (LEL)
  • Alarm Thresholds:
    • Sensor: H2S
    • Low Alarm: 10 ppm
    • High Alarm: 15 ppm
    • TWA: 10 ppm
    • STEL: 15 ppm
    • Sensor: CO
    • Low Alarm: 35 ppm
    • High Alarm: 200 ppm
    • TWA: 35 ppm
    • STEL: 50 ppm
    • Sensor: O2
    • Low Alarm: 19.5%
    • High Alarm: 23.5%
    • Sensor: LEL
    • Low Alarm: 10%
    • High Alarm: 20%
  • Battery Information:
    • Low Battery Warning: With 20 minutes of run time left, the detector will beep, flash, and display “LOW BAT”. Notifications will continue periodically until the battery expires and the detector performs an automatic shutdown.
    • Charging: While charging, the battery icon will cycle and the detector will display an approximate charge percentage. Full charge occurs in 4-6 hours. For fastest charging, charge while detector is turned off.
    • Warning: DO NOT charge the detector in a combustible atmosphere.

Product Usage Instructions

When a Gas Reading (5) exceeds its alarm threshold, the detector will go into visual, vibrating, and audible alarm. On the display, the icon will appear beside the Alarm Condition (1), and the associated Gas Identifier (6) will flash off and on continuously.

Warning: Users must familiarize themselves with the icons in both alarm and non-alarm states.

Warning: If the display is missing icons or cannot be clearly read, discontinue use and contact Gas Clip Technologies.

Turning On the Detector:

  1. To activate the detector, press and briefly hold down the power/menu button. An activation tone will sound.
  2. Each Gas Identifier will show a chasing 0 for the Gas Reading while it is stabilizing and self-testing.
  3. Once all Gas Identifiers have completed the warm-up and stabilization sequence (less than 65 seconds), the detector is ready to detect all applicable gases.
  4. The display will show: “Stabilizing” and then “Fully operational”.

Before each use, check that the gas inlet and alarm port are clear of any debris or obstructions. To confirm that there are no leaks in the sampling system, perform a blocked flow test daily and any time the sampling system is changed.

**READ FIRST BEFORE OPERATION

**

The Multi Gas Clip Pump is a personal safety device designed to detect the presence of hydrogen sulfide (H2S), carbon monoxide (CO), oxygen (O2) and combustible gases (LEL). Before operation, ensure that you have been properly trained on the use of the detector and the appropriate actions to take in the event of an alarm condition.

Display Components

Icon Description
1 Alarm Condition
2 Pump Status
3 Calibration/Test Mode
4 Battery Indicator
5 Gas Readings
6 Gas Identifiers

When a Gas Reading (5) exceeds its alarm threshold, the detector will go into visual, vibrating and audible alarm. On the display, the icon will appear beside the Alarm Condition (1) and the associated Gas Identifier (6) will flash off and on continuously.

Warning Users must familiarize themselves with the icons in both alarm and non-alarm states.

Warning If the display is missing icons, or cannot be clearly read, discontinue use and contact Gas Clip Technologies.

Warning   The IR LEL sensor will not detect hydrogen (H2) or acetylene (C2H2) gases however, due to their cross-sensitive nature, the CO sensor will go into alarm well below 10% LEL.

Factory Default Alarms

Turning On the Detector

To activate the detector, press and briefly hold down the power/menu button. Activation tone will sound. Each Gas Identifier will show a chasing “0” for the Gas Reading while it is stabilizing and self-testing. Once all Gas Identifiers have completed the warm up and stabilization sequence, < 65 seconds, the detector is ready to detect all applicable gases.

Display will show:

gas-clip-technologies-MGC-PUMP-v1-06-Multi-Gas-Clip-Pump-fig-
\(4\)

****Before each use, check that the gas inlet and alarm port are clear of any debris or obstructions.

****To confirm that there are no leaks in the sampling system, perform a blocked flow test 1.) daily and 2.) any time the sampling system is changed.

When the gas flow is blocked, the detector will go into alarm and disable the pump. Press the power/menu button to reactivate the pump.

Turning Off the Detector

Press and hold down the power/menu button until the “OFF” countdown is complete. During the countdown, the detector will beep and flash 4 times then turn off.

Display will show:

gas-clip-technologies-MGC-PUMP-v1-06-Multi-Gas-Clip-Pump-fig-
\(5\)

Manually Calibrating the Detector
Continuously hold down the power/menu button through the “OFF” countdown and the “CAL” countdown that follows. During these countdowns, the detector will beep and flash 8 times with a pause between countdowns.

Display will show:

gas-clip-technologies-MGC-PUMP-v1-06-Multi-Gas-Clip-Pump-fig-
\(6\)

Release the power/menu button when the display shows “AUTO ZERO” followed by “APPLY GAS”. Connect a hose from a demand flow regulator on a gas cylinder to the detector’s gas inlet and apply gas. Calibration is complete once the detector goes into alarm and the next calibration due date is displayed. Turn gas off and remove hose from regulator.

Display will show:

gas-clip-technologies-MGC-PUMP-v1-06-Multi-Gas-Clip-Pump-fig-
\(7\)

Note: Calibration can also be done using the MGC Pump Clip Dock.

Battery Information

Low Battery Warning   With 20 minutes of run time left, the detector will beep, flash and display “LOW BAT”. Notifications will continue periodically until the battery expires and the detector performs an automatic shutdown.
Charging While charging, the battery icon will cycle and the detector will display an approximate charge percentage. Full charge occurs in 4-6 hours. For fastest charging, charge while detector is turned off.

Warning – DO NOT charge the detector in a combustible atmosphere.

Warning – DO NOT charge the detector in temperatures above 104°F (40°C) or below 32°F (0°C).

FAQ’s

Additional Questions?
For additional information on the Multi Gas Clip Pump gas detector regarding programming, bump testing, calibrating, etc., please refer to the Multi Gas Clip Pump User’s Manual. The manual, videos, and other helpful information are available for viewing or download at www.gascliptech.com.