FLUKE B01LW4NFMV FoodPro Plus Probe Thermometer User Manual
- June 17, 2024
- FLUKE
Table of Contents
- FLUKE B01LW4NFMV FoodPro Plus Probe Thermometer
- LIMITED WARRANTY AND LIMITATION OF LIABILITY
- Product Statement
- FoodPro Plus Specifications and Features
- Proper Application and Operation
- Noncontact Temperature Measurement Applications
- Field Verification of Accuracy
- Setting/Changing ºC and ºF Temperature Scales
- Battery Insertion and Replacement
- Troubleshooting
- Certifications
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
FLUKE B01LW4NFMV FoodPro Plus Probe Thermometer
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for two years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090 U.S.A.
11/99
Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands
Product Statement
Thank you for purchasing this food safety thermometer. We are confident that you will be pleased with the quality and performance of this product. Please take a few moments to familiarize yourself with the operation, features, and benefits of this fine product. This food safety thermometer combines two precision thermometers in one unit: a specially calibrated noncontact infrared thermometer (IRT) and a calibrated resistance thermal device (RTD) penetration probe. The noncontact mode can be used for quick scans of a surface temperature, while the probe mode is used to accurately measure internal temperatures.
FoodPro Plus Specifications and Features
See Table 1 for specifications and features. Specifications are subject to
change without notice.
Table 1. Specifications and Features
Infrared
Accuracy
Assumes ambient operating temperature of 23 °c ±2 °c (73 °F
±4 °F)
| Between O °Cand 65 °C(32 – 150
°F): ±1 °C(± 2 °F)Below O °C(32
°F): ±1 oc (± 2 OF)
± 0.1 degree/degree Above 65 °C (150 °F): ±1.5 % of reading
---|---
Response Time| <500 ms after initial reading
Spectral response| 8-14 microns
Emissivity 111| Pre-set for Food Service Applications
Distance to Spot Size Optical Resolution (D:S)| 2.5:1 @ 90 % energy, typical
Typical Working Range (target illumination)| a=25 mm to 250 mm (“‘1″ to 1O”)
Minimum Target Size| 12 mm (0.5 “)
Illumination to IA Channel Offset| 13 mm (0.52 “)
Probe
Accuracy
Assumes ambient operating temperature of 23 °c ± 2 °c (73 °F
± 4 °F)
| Between -5 °Cand 65 °c (23 °F-
150 °F):
±0.5 oc (±1 °F)
Below -5 °C(23 °F): ±1 °C(± 2 °F)
Above 65 °C(150 °F): ±1 % of reading
---|---
Temperature Range| -40 °c to+150 °c (-40 °Fto
+302 °F)
Probe
Response Time| 7-8 seconds (3 time constants)
Probe Dimensions| Ø: 3.0 mm (0.118 “) Length:
80 mm (3.0 “)
Sensor| Thin Film, Platinum Class A RTD
Operational
Repeatability| Within accuracy specifications of the unit
Ambient Operating Range| 0 °C to 50 °C (32 °F to 122 °F)
Relative Humidity| 10 – 90 % (± 5 %)RH
noncondensing @ 30 °C (86 °F)
Storage Temperature| -20 °C to +60 °C (-4 °F to +140 °F)
Weight/Dimensions (with Battery)| 165 mm x 32 mm x 50 mm
(6.5 ” x 1.25 ” x 2 “) 150 g
(0.33 lbs)
Power| 9V Alkaline
Battery Life (Alkaline)| 10 hours minimum @ 23 °C
(73 °F)
Target Illumination| High Brightness LED
Display Resolution| 4 digits, 0.1 °C (0.2 °F)
Display Hold (7 sec)| •
LCD Backlit Display| •
Standards| Conforms to:EN 61326-1 Electromagnetic Emissions and
Susceptibility, Criteria B, EN 61010-1 General Safety, IP54 Sealing (hand
washable, non-submersible)
Other
Certifications| CE, NSF, CMC
Warranty| 2 years
Accessories| Nylon Carrying Pouch, QuickStart Guide, and Battery
[1] Calibration Geometry: >22 °C 144 mm source @ 200 mm (1.45:1)
E=0.97DD <25 °C 140 mm source @ 100 mm (0.7:1) E=1.00 BB.
Proper Application and Operation
Before You Start
First time users should reference “Battery Insertion Replacement” before using
the unit.
Noncontact (Infrared) Mode
Pressing and holding the ON button starts the food safety thermometer in
noncontact (infrared) measurement mode. The measurement will continue as long
as the ON button is depressed, indicated by the blinking “SCAN”.
When the button is released, “HOLD” appears on the display and the last
measurement remains visible for 7 seconds before the display goes blank.
The unit will display the temperature of an area highlighted by the target
illuminator. The maximum reading is indicated in the lower portion of the
display (Figure 1).
Note The infrared temperature measurements are used for screening and measuring surface temperatures only. Critical internal temperatures must be verified by using the probe.
Mode Selection
The SELECT button has two functions. Pressing the SELECT button after the
display has gone blank will recall the last display. Pressing the SELECT
button with an active display cycles through the three different operating
modes: Noncontact, Probe, and Countdown Timer (Figure 2).
Probe Mode
To measure the internal temperature of an object, extend the probe (see Figure
3) and press the SELECT button until the Probe Icon is displayed (see Figure
4). Insert the probe at least 12 mm (1/2 “) into the target and press the ON
button to measure the core temperature.
The Probe Icon on the display will blink for approximately 15 seconds as the probe comes to equilibrium with the object being measured. Three short beeps will sound to indicate the reading is complete and the temperature will be shown on the display. When the highest accuracy is required, it is advisable to repeat the measurement cycle to insure the probe has fully stabilized and reached equilibrium with the object.
Remember if the display blanks the select button will recall the last
measurement.
The probe tip must be sterilized before and between measurements of food
samples to avoid cross contamination.
Countdown (Timer) Mode
The food safety thermometer has a built in countdown timer to conveniently
monitor cooking, cooling, and critical exposure times, as well as Hazard
Analysis Critical Control Points (HACCP) inspection intervals.
HACCP principles and good food safety practices require monitoring the length
of time perishable foods are exposed to temperatures that may support rapid
bacterial growth.
To set the countdown timer, press the SELECT button until the timer icon is
displayed (Figure 5). Press the SET button once to enter the set mode (“SET”
on display blinks). Press the SET button a second time to clear and increment
the timer value. The timer value increases initially by 10 second intervals
then by minutes and hours. The timer can be set for a maximum of 7 hours and
59 minutes.
Once the desired timer value is set, pressing the ON button will start and
stop the countdown. An alarm will sound for approximately 30 seconds when the
timer reaches zero.
Note
The timer can be activated and running in the background (indicated by the
flashing timer icon) during either IR or Probe measurements or when the
display is off. Pressing any button will silence the sounding alarm.
To clear the displayed timer value, press the SET button once to enter set
mode and a second time to zero the timer value.
HACCP Check
This thermometer incorporates a “HACCP Check” feature to graphically display
critical temperature zones. The icons and LED indicators located above the
display indicate a food product is safely held at a sufficient holding
temperature, or if it has fallen within the unsafe HACCP “Danger Zone”
temperatures (Figure 5). The HACCP Check indicators operate in both the
Noncontact and Contact modes. The indicator light will flash during an active
measurement and will be lit steadily during display Hold or Recall (Figure 4).
- A Green LED indicator shows a safe chilled or frozen condition below 4 °C (40 °F) or a safe holding temperature above 60 °C (140 °F).
- The Red LED indicator light is displayed when temperatures are within the HACCP “Danger Zone” of 4 °C to 60 °C (39 °F to 140 ° F) where microbe growth occurs most rapidly (Figure 5).
Notes
When the temperatures within ±1 °C (±2 °F) of the HACCP “Danger Zone”
boundary, both the relevant Green and Red LED indicator lights will be
triggered to alert the user of the border temperature.
Whenever the Red LED indicator light is lit it is necessary to evaluate the
safe storage and handling rules dictated by both exposure time and
temperature.
Noncontact Temperature Measurement Applications
Measuring Liquid Temperatures
To accurately measure the temperature of liquids and semi-liquids, such as
soup, chili, salad dressing, etc., stir the liquid to bring the internal
temperature to the surface, while taking a measurement with the unit in
noncontact mode. Steam, dust, smoke, etc., can prevent accurate measurement by
interfering with the energy emitted from the target. To increase the
reliability of the measurement, do not hold the unit directly over a steaming
or smoking product. Instead, hold the unit back and at an angle to ensure the
most accurate measurement (Figure 6).
Measuring Packaged Products in Cold Storage Cases
Ideally, the temperature of a product should be measured outside of a
refrigerated environment whenever possible. If it is necessary to measure the
product in a refrigerated environment such as a walk-in cooler, either make
rapid measurements (within 1 minute or less) or allow 30 minutes for the unit
to stabilize in the refrigerated environment (above 0 °C / 32 °F) before
measuring. To measure an item within a storage case, open the door or curtain,
and directly scan the product for uniform temperatures. The presence of warmer
areas may indicate improper stocking resulting in blocked airflow in the
cabinet (Figure 7).
Note The unit will not measure through glass or plastic doors.
Measuring Food at the Receiving Dock
Use the food safety thermometer to accurately measure perishable products at
the receiving dock. When a delivery of fresh or frozen food arrives, check
that the products, shipping crates, and internal temperature of the delivery
truck are all at the proper storage temperatures. Check for warm spots in
products that can result from improper stacking and blocked airflow.
Measuring Food in Holding and Serving Areas
Use the noncontact thermometer to easily scan and accurately measure surface
temperature of products held in hot or cold holding areas, such as open-top
refrigeration units, steam tables, salad bars, fresh meat or fish displays and
warming ovens.
Use the HACCP Check feature to quickly identify unsafe temperatures within the
HACCP “Danger Zone” of 4 °C to 60 °C (39 °F to 140 °F) by slowly scanning
across the surface of food, storage containers, the contents of delicases,
chilled salads and desserts, holding ovens, rotisseries, etc.
Note If any questionable temperatures are indicated or if temperature readings are within ±1 °C (±2 °F) of the HACCP “Danger Zone” boundary, use the probe to check internal temperatures.
Field Verification of Accuracy
Infrared (IRT) and Probe channels (RTD)
Follow the steps below to verify the accuracy of your food safety thermometer.
The 0 °C (32 °F) “stirred ice bath” is the recommended verification reference
point. Since it is more difficult to measure the surface of hot water, the hot
verification should be used only as a general accuracy check of the IRT
channel.
Cold Temperature Verification Check
- Fill a large styrofoam cup halfway to the surface with ice cubes. Add cold water to just below the rim of the cup.
- Immerse the tip of a known calibrated probe thermometer (reference probe) into the water and stir the mixture with the probe for one minute, or until the probe temperature stabilizes.
- Continue stirring the water with a straw or swizzle stick while taking simultaneous temperature measurements with the reference probe and the IR thermometer. The unit should be held within 3 inches of the surface of the water (Figure 8). To insure measurement accuracy the probe tip must be immersed to a minimum depth of 12 mm (1/2 “).
The noncontact (IRT) measurement should be within ±1 °C (±2 °F) of the
reference probe reading (nominally 0 °C (32 °F)).
The probe temperature reading should be within ±0.5 °C (±1 °F) of the
reference probe reading.
Hot Temperature Verification Check
- Follow the same procedure as above, substituting hot water (>140 °F/60 °C). Hot tap water is adequate for the procedure.
- Repeat steps two and three from above.
Note Due to evaporative cooling on the surface of the hot water, it is particularly important that the surface of the water be continually stirred while making the IR measurement (Figure 9).
Using this method, the noncontact (IRT) channel should be considered accurately calibrated if the reading is within ±2 °C (±3.5 °F) of the reference probe’s reading; the unit’s probe reading should be within ±0.5 °C (±1 °F) of the reference probe reading.
Cautions
- Hold the noncontact thermometer outside the rim of the cup, approximately 3 inches from the surface of the water.
- Avoid steam condensation on the unit’s lens. If condensation forms, carefully wipe the lens or let it dry at room temperature and resume measurement.
Field of View
The ideal working range of the noncontact thermometer (IRT) is between 25 mm
and 250 mm (1 and 10 inches) or the field of view from the target to the unit,
divided by 2. The built in target illumination beam helps to indicate the
measurement area. To ensure accurate measurements, the measurement target must
fill or exceed the field of view. When conditions permit move in closer to the
subject (Figure 10).
Operating and Ambient Temperature Range
The food safety thermometer is designed to operate in environments between 0
°C – 50 °C (32 °F – 122 °F). Avoid subjecting the unit to extreme or abrupt
changes in ambient temperatures. Allow the unit to stabilize for a minimum of
30 minutes if exposed to rapid temperature changes. Failure to precondition
the instrument to the ambient temperature may result in measurement errors
(Figure 11).
Response Time
The response time of the instrument from initial start up is less than 1
second. The temperature display is updated at approximately 2 times per second
during sustained operation.
Humidity
Noncontact thermometers (IRTs) are not intended for use in extremely humid or
condensing environments. Condensation on the lens window will impede the
optical performance and prevent accurate temperature measurements. If this
occurs, allow the window to dry by evaporation or wipe with a soft cloth and
resume measurements.
Emissivity
A noncontact thermometer (IRT) determines temperature by measuring the emitted
energy of an object. Emissivity (E value) is the measure of an object’s
ability to emit infrared energy.
This noncontact thermometer is specially calibrated for high emissivity
targets and is preset to an E value of ~0.97. This setting is the most
advantageous for capturing the emitted energy from water, oils, shortening,
fat, vegetables, as well as frozen, partially frozen and refrigerated products
in boxes and plastic containers.
Note: Shiny metal surfaces (such as polished or stainless steel) have low emissivities and reflect the energy of their surrounding, resulting in inaccurate temperature readings. The emissivity of shiny metal surfaces can be enhanced to provide more accurate readings by covering the measurement area with masking tape, flat black paint, or a coating of shorting or oil. Blackened cooking surfaces such as griddles and cast iron pans make good targets.
Setting/Changing ºC and ºF Temperature Scales
The food safety thermometer can display temperature in either the °C (Celsius) or °F (Fahrenheit) scales. The desired scale is selected at the time the battery is installed:
Selecting the °C Scale:
When the battery is connected to the battery leads the °C scale indicator
shows on the display for approximately 15 seconds. If the SET button is not
pressed the unit will time out and default to the °C or Celsius scale.
Selecting the °F Scale:
If the SET button is pressed within 15 seconds of connecting the battery to
the battery leads, the temperature scale display will toggle to the °F scale
indicator and the unit will now default to the °F or Fahrenheit scale (Figure
12).
Note The °C/°F temperature scale initialization process must be performed whenever battery power is interrupted or the battery is replaced.
Cleaning Instructions
This food safety thermometer is sealed to IP54 standards. The unit may be
wiped down with a wet sponge or cloth using a mild water based detergent or
anti-bacterial soap and rinsed under a gentle stream of cold water (Figure
13).
Note: This unit is not designed for complete submersion or washing in automated dishwashers.
Probe Replacement
The modular probe of the food safety thermometer is replaceable. To replace
probe, extend the probe partially to access the rubber caps. Carefully pry up
the rubber caps using a pin, and remove the screws using a #2 cross-tip
screwdriver as shown. Grasp the probe and carefully separate the old probe
from the probe base. Install the new probe in the reverse order, firmly
tightening the screws and replacing the rubber plugs. The calibration accuracy
of the unit is not affected by probe replacement (Figure 14).
Battery Insertion and Replacement
To install a new 9V battery, remove the rubberized battery compartment “plug” at the base of the unit by grasping the sides of the cover and pulling straight out, exposing the battery. Gently shake or tap the base of the unit on your palm to gain access to the battery. The 9V battery of the unit is connected to the polarized snap connector (Figure 15).
Note Each time the battery is installed or replaced, the temperature scale will default to the Celsius scale. To select the Fahrenheit scale see ‘Setting/Changing °C and °F Temperature Scales’.
Troubleshooting
- Code: **** — (on display)
- Problem: Target temperature is over or under range
- Action: Select target within range specifications
- Code: **** Battery Symbol
- Problem: **** Possible low battery
- Action: **** Check and/or replace battery
- Code: **** Blank display
- Problem: **** Dead battery or unit failure
- Action: Check and/or replace battery. If the battery is OK, then contact a Fluke service center for repair.
Other Operational Considerations
All models are protected from the following:
- EMI (Electro Magnetic Interference) from induction heaters and microwave ovens
- Electrostatic discharge
- Should the unit become damaged, check the accuracy of the unit by performing the verification process recommended in this manual. If the unit is out of calibration, do not rely on it for critical temperature measurements. Contact a Fluke service center for repair,
- Heat from stovetops, pans, or other hot surfaces (don’t set it on the stove)
Certifications
CE, NSF, CMC
This instrument conforms to the following standards:
- EN 61326-1 Electromagnetic Emissions and Susceptibility
- EN 61010-1 General Safety
- IP54
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>