gas clip technologies MGC-S-PLUS Multi Gas Clip Simple Plus User Guide

June 13, 2024
gas clip technologies

gas clip technologies MGC-S-PLUS Multi Gas Clip Simple Plus

gas-clip-technologies-MGC-S-PLUS-Multi-Gas-Clip-Simple-Plus-
product

Product Information

The Multi Gas Clip Simple Plus is a portable gas detector designed to detect the presence of hydrogen sulfide (H2S), carbon monoxide (CO), oxygen (O2), and combustible gases (LEL). It has a long battery life and does not require charging or battery replacement during its lifetime. The detector displays the remaining life in months, starting from 36 months and counting down until 30 days remain. It then displays the remaining days until the end of life. The detector also features factory default alarm thresholds for each gas.

Product Usage Instructions

  1. Before operating the Multi Gas Clip Simple Plus, ensure that you have received proper training on its use and know the appropriate actions to take in the event of an alarm condition.
  2. Make sure the display components are functioning properly. The display should show icons for alarm conditions, life remaining, calibration/test mode, battery indicator, gas readings, and gas identifiers. If any icons are missing or unreadable, discontinue use and contact Gas Clip Technologies.
  3. To turn on the detector, press and briefly hold down the power/menu button. The detector will start displaying gas readings immediately.
  4. During the warm-up and stabilization sequence, each gas identifier will show a chasing 0 for the gas reading. Wait until all gas identifiers have completed this sequence.
  5. If a gas reading exceeds its alarm threshold, the detector will go into visual, vibrating, and audible alarm. The alarm condition icon will appear on the display, and the associated gas identifier will flash on and off continuously.
  6. Familiarize yourself with the icons in both alarm and non-alarm states to understand the detector’s status.

For additional information on programming, bump testing, calibrating, and other features of the Multi Gas Clip Simple Plus gas detector, refer to the Multi Gas Clip Simple Plus User’s Manual available at www.gascliptech.com. Gas Clip Technologies can also be contacted for further assistance.

READ FIRST BEFORE OPERATION
The Multi Gas Clip Simple Plus is a personal safety device designed to detect the presence of hydrogen sulfide (H2S), carbon monoxide (CO), oxygen (O2) and combustible gases (LEL). Before operation, ensure that you have been properly trained on the use of the detector and the appropriate actions to take in the event of an alarm condition.

Display Components

Icon Description
1 Alarm Condition
2 Life Remaining
3 Calibration/Test Mode
4 Battery Indicator
5 Gas Readings
6 Gas Identifiers

When a Gas Reading (5) exceeds its alarm threshold, the detector will go into visual, vibrating and audible alarm. On the display, the icon will appear beside the Alarm Condition (1) and the associated Gas Identifier (6) will flash off and on continuously.

  • Warning – Users must familiarize themselves with the icons in both alarm and non-alarm states.
  • Warning – If the display is missing icons, or cannot be clearly read, discontinue use and contact Gas Clip Technologies.
  • Warning – The IR LEL sensor will not detect hydrogen (H2) or acetylene (C2H2) gases however, due to their cross-sensitive nature, the CO sensor will go into alarm well below 10% LEL.

Factory Default Alarms

Sensor LOW HIGH TWA STEL
H 2 S 10 ppm 15 ppm 10 ppm 15 ppm
CO 35 ppm 200 ppm 35 ppm 50 ppm
O 2 19.5% 23.5%  
LEL 10% 20%  

Turning On the Detector

Once activated, the detector will run continuously for 36 months. To activate, press and briefly hold down the power/menu button. The detector will begin to display Gas Readings (5) immediately. Each Gas Identifier (6) will show a chasing “0” for the Gas Reading (5) while it is stabilizing and self-testing. Once all Gas Identifiers (6) have completed the warm up and stabilization sequence, <65 seconds, the detector is ready to detect all applicable gases.

Display will show:

Warning
Before each use, check that all sensor and alarm ports are clear of any debris or obstructions.

Bump Testing and Calibrating

These operations can be performed manually or automatically using the MGC Simple Clip Dock. Refer to the Multi Gas Clip Simple Plus User’s Manual for instructions.

Auto Zeroing

Continuously hold down the power/menu button to auto zero the detector. Display will show “ZERO” and count down from 3 to 1, followed by “ZERO OK”. Release button when display shows “ZERO OK”. The display will change to “AUTO ZERO”. Wait a few seconds, then press the button briefly when display shows “APPLY GAS”. Detector has been successfully auto zeroed.

  • Warning – DO NOT auto zero the detector in a combustible atmosphere.
  • Warning – DO NOT auto zero in temperatures above 104°F (40°C) or below 32°F (0°C).

Battery Information

This detector is designed so that the battery won’t need charging or replacing during the life of the detector.

Detector Life Countdown

Once activated, the detector will display 36 months of life remaining. Detector will continue to countdown and display months remaining until 30 days remain. Display will then show days remaining until end of life.

Additional Questions

For additional information on the Multi Gas Clip Simple Plus gas detector regarding programming, bump testing, calibrating, etc., please refer to the Multi Gas Clip Simple Plus User’s Manual. The manual and other helpful videos & information are available for viewing or download at www.gascliptech.com.

Contact Information

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

gas clip technologies User Manuals

Related Manuals