hygiena MCH2023 Automated Thermal Block
Product Usage Instructions
Instrument Overview
This model of the Automated Thermal Block (MCH2023) is specifically optimized
for use with the cluster tubes and metal tube holder supplied with the BAX
System. It is not recommended to use other tubes and holders with this model.
Supplied Components
The purchase of the thermal block includes the following components:
- Thermal block unit
- Metal tube holders (2)
- Power supply and cord
- User documentation
Safety Symbols and Precautions
Safety Symbols:
- Symbol: Indicates that you should consult the manual for further information and proceed with appropriate caution.
- Symbol: Indicates the presence of a hot surface or other high-temperature hazard. Proceed with appropriate caution.
- Symbol: Indicates the presence of an electrical shock hazard. Proceed with appropriate caution.
General Safety Information:
Safety Topic | Information |
---|---|
Extreme Temperatures | Always use the supplied AC power cord when connecting |
the heating block to a power source. Use the normal care and precaution that
one would use with any electrical appliance.
Power Requirements| Always use the supplied AC power cord when connecting the
heating block to a power source. Use the normal care and precaution that one
would use with any electrical appliance.
Connecting the Thermal Block
The Automated Thermal Block is pre-programmed for BAX System assays and does
not require calibration or software installation when used as specified.
Follow these steps to connect the thermal block:
- Place the thermal block unit on a level, dry laboratory workbench. Allow at least 4 inches (10 cm) on all sides of the unit for ventilation.
- Insert one end of the provided power cord into the power supply and plug the other end of the power cord into a 3-wire outlet. Ensure the outlet is properly grounded and runs at the appropriate voltage (see specifications).
- Insert the power supply jack into the power input port on the back panel of the thermal block.
- Seat the metal tube holder over the metal plate surface on the top of the thermal block unit.
Front Panel Display
The front panel of the Automated Thermal Block contains the following items:
- LCD display – shows the program menu, program selection, settings, and instructions.
- SCROLL button – changes the selected menu item in the LCD display. An audible beep sounds when the button is pressed.
- Blue ACTION NEEDED LED – indicates that the user must act as the program can continue. This notification is accompanied by an audible alarm.
- SELECT/CONTINUE button – sets the menu item in the LCD display as the selected program and is also used
FAQ
-
Q: Can I use other tubes and holders with this model?
A: It is not recommended to use other tubes and holders with this model. It has been optimized for use with the cluster tubes and metal tube holder supplied with the BAX System. -
Q: Do I need to calibrate or install software for the thermal block?
A: No, the thermal block is pre-programmed for BAX System assays and does not require calibration or software installation when used as specified.
Instrument Overview
The Automated Thermal Block is a Peltier-driven heating and cooling device
pre-programmed to perform the lysis steps of the BAX® System protocols. It has
been designed as a solid-state device with minimal moving parts to help ensure
quality performance over time. The temperature of the plate is sensed by a
platinum resistance temperature device mounted under the plate. The computer
in the unit compares the plate temperature with the target temperature and
instructs the module to heat or chill the plate as required.
The block automatically provides the sequential heating and cooling conditions
for lysis protocols, eliminating the need to transfer samples between separate
heating and cooling blocks. Once the lysis program is complete, the block
holds the samples at 4 °C until they are removed. This model has been
optimized for use with the cluster tubes and metal tube holder supplied with
the BAX System. Other tubes and holders are not recommended.
Supplied Components
The purchase of the thermal block includes the following components:
- Thermal block unit
- Metal tube holders (2)
- Power supply and cord
- User documentation
Specifications and Requirements
Item | Specification |
---|---|
Dimensions (with lid) | 6.5″ W x 8.75″ D x 8″ H (16.5 x 22.25 x 20.3 cm) |
Weight | Approximately 8 lbs. (3.6 kg) |
Power usage | 12 volts, 8.4 amps |
Power requirements | 90 to 265 volts AC, 50/60Hz |
Thermal range | –10 to 110 °C (14 to 230 °F) |
Thermal uniformity | ±1 °C as measured at the block surface |
Thermal accuracy | ±1 °C as measured at the block surface |
Sample throughput | 1 – 96 samples per cycle |
Room environment | Temperature between 18 and 32 °C (65 and 89 °F) |
Safety Symbols and Precautions
Safety Symbols
Symbol | Description |
---|---|
**** | Indicates that you should consult the manual for further information |
and proceed with appropriate caution.
| Indicates the presence of a hot surface or other high-temperature
hazard. Proceed with appropriate caution.
| Indicates the presence of an electrical shock hazard. Proceed with
appropriate caution.
General Safety Information
Safety Topic | Information |
---|---|
Extreme Temperatures | The thermal block can reach 110 °C (230 °F), which can |
burn the skin if touched. To prevent injury, remember the following:
• Always use extreme caution around the block.
• Never leave the block accessible to others when it is hot.
• Do not touch the plate surface unless you are sure it has cooled.
• Always keep the lid in place when the block is not in use.
Power Requirements| Always use the supplied AC power cord when connecting the heating block to a power source. Use the normal care and precaution that one would use with any electrical appliance.
Connecting the Thermal Block
The Automated Thermal Block is pre-programmed to perform the sequential
heating and cooling steps needed for BAX System assays. No calibration or
software installation is required when the thermal block is used as specified.
Follow these steps to connect the thermal block:
-
Place the thermal block unit on a level, dry laboratory workbench. Allow at least 4 inches (10 cm) on all sides of the unit for ventilation.
-
Insert one end of the provided power cord into the power supply and plug the other end of the power cord into a 3-wire outlet.
Note: Ensure the outlet is properly grounded and runs at the appropriate voltage (see specifications on the previous page). -
Insert the power supply jack into the power input port on the back panel of the thermal block.
-
Seat the metal tube holder over the metal plate surface on the top of the thermal block unit.
Front Panel Display
The front panel of the Automated Thermal Block contains the following items:
- LCD— shows the program menu, program selection, settings and instructions.
- SCROLL button— changes the selected menu item in the LCD. An audible beep sounds when the button is pressed.
- Blue ACTION NEEDED LED— indicates that the user must act as the program can continue. This notification is accompanied by an audible alarm.
- SELECT/CONTINUE button— sets the menu item in the LCD as the selected program and is also used to confirm required actions. An audible beep sounds when the button is pressed.
- Green POWER LED— indicates that the thermal block is turned on.
Program Menu
The Automated Thermal Block is pre-programmed with four lysis protocols:
-
Gram positive—use with gram-positive bacteria, including Staphylococcus and Listeria
Note: For Listeria 24E assays, use the “24E” program. For real-time Listeria assays, use the “RT Listeria” program. -
Gram negative—use with gram-negative bacteria, including Salmonella, E. coli, Campylobacter, Vibrio and Cronobacter
-
24E—use with Listeria 24E assays:
BAX System PCR Assay for Genus Listeria 24E| KIT2003
---|---
BAX System PCR Assay for L. monocytogenes 24E| KIT2002 -
RT Listeria—use with real-time Listeria assays:
BAX System Real-Time PCR Assay for Genus Listeria| KIT2019
---|---
BAX System Real-Time PCR Assay for L. monocytogenes| KIT2005
Turn the Thermal Block On and Off
ON
Turn on the thermal block by toggling the on/off switch on the back panel
of the unit.
At start-up, the LCD automatically displays the following series of messages:
-
The first screen shows the version of software currently installed on the thermal block.
-
The second screen shows the seq file version number and the checksum number for validation purposes.
Note: A change in either of these values indicates a modification to the thermal block software. If this occurs, any validation previously performed on the instrument may no longer apply. -
The start-up sequence is complete when the LCD display shows the program menu.
OFF
After finishing a run, you can either select another program or turn off
the power to the unit by toggling the on/off switch on the back panel of the
unit.
Run the Gram-Positive Program
Samples are heated to 55 °C for 60 minutes, then 95 °C for 10 minutes, then held at 4 °C for at least 5 minutes or until removed. The times vary slightly depending on the environment because a sensor gives active feedback to the instrument heater to ensure that proper temperatures are met.
-
Turn on the thermal block, then press the SCROLL button until the arrow ( → ) is next to the Gram Positive option.
Press the SELECT/CONTINUE button. -
The unit beeps once, and the LCD display changes as the thermal block begins equilibrating to 55 °C.
Note: Do not load samples into the thermal block until the load prompt appears on the LCD display. -
When the thermal block has reached 55 °C, the unit sounds two audible beeps and the blue ACTION NEEDED LED activates. The LCD display changes to Load Samples.
-
After the Load Samples prompt appears, place a rack of prepared samples into the metal tube holder.
Note: See the BAX System User Guide for information on preparing samples.
Press the SELECT/CONTINUE button. -
The LCD display changes to GmPos Lysis 55C.
Note: A timer in the bottom right corner counts down from 60:00 and shows the number of minutes remaining in this step. -
After lysis at 55 °C is complete, the LCD display changes to Heating to 95C. No user action is required.
Note: A timer in the bottom right corner counts up from 0:00 during the temperature change. -
After the thermal block reaches 95 °C, the LCD display changes to Denaturing @ 95C.
No user action is required.
Note: A timer in the bottom right corner counts down from 10:00 and shows the number of minutes remaining in this step. -
After denaturation at 95 °C is complete, the LCD display changes to Cooling to 4C, and the thermal block sounds two audible beeps to signal the change.
No user action is required.
Note: A timer in the bottom right corner counts up from 0:00 during the temperature change. -
After the samples have cooled, the LCD display changes to Sample PCR Ready, the blue ACTION NEEDED LED activates, and the thermal block sounds four audible beeps to signal that samples can be removed.
Press the SELECT/CONTINUE button.
Note: A timer in the bottom right corner counts up from 0:00 to show any additional time that samples are held at 4 °C. -
The LCD display changes to Completed, and the thermal block sounds an audible beep to signal that the program has been completed.
Note: The timer continues to show additional time that samples are held at 4 °C. -
Remove the cluster-tube rack and follow the instructions in the BAX System User Guide to hydrate the PCR tablets with these lysates.
Press the SELECT/CONTINUE button. -
The LCD display changes to Gram Positive Finished, and the thermal block sounds an audible beep to signal that the program has finished.
-
Press the SELECT/CONTINUE button to end the program. The LCD changes to display the program menu.
Run the Gram-Negative Program
Samples are heated to 37 °C for 20 minutes, then 95 °C for 10 minutes, then held at 4 °C for at least 5 minutes or until removed. The times vary slightly depending on the environment because a sensor gives active feedback to the instrument heater to ensure that proper temperatures are met.
-
Turn on the thermal block, then press the SCROLL button until the arrow ( → ) is next to the Gram Negative option.
Press the SELECT/CONTINUE button. -
The unit beeps once, and the LCD display changes as the thermal block begins equilibrating to 37 °C.
Note: Do not load samples into the thermal block until the load prompt appears on the LCD display. -
When the thermal block has reached 37 °C, the unit sounds two audible beeps and the blue ACTION NEEDED LED activates. The LCD display changes to Load Samples.
-
After the Load Sample prompt appears, place a rack of prepared samples into the metal tube holder.
Note: See the BAX System User Guide for information on preparing samples .
Press the SELECT/CONTINUE button. -
The LCD display changes to GmNeg Lysis 37C.
Note: A timer in the bottom right corner counts down from 20:00 and shows the number of minutes remaining in this step. -
After lysis at 37 °C is complete, the LCD display changes to Heating to 95C.
No user action is required.
Note: A timer in the bottom right corner counts up from 0:00 during the temperature change. -
After the thermal block reaches 95 °C, the LCD display changes to Denaturing @ 95C.
No user action is required.
Note: A timer in the bottom right corner counts down from 10:00 and shows the number of minutes remaining in this step. -
After denaturation at 95 °C is complete, the LCD display changes to Cooling to 4C, and the thermal block sounds two audible beeps to signal the change.
No user action is required.
Note: A timer in the bottom right corner counts up from 0:00 during the temperature change. -
After the samples have cooled, the LCD display changes to Sample PCR Ready, the blue ACTION NEEDED LED activates, and the thermal block sounds four audible beeps to signal that samples can be removed.
Press the SELECT/CONTINUE button.
Note: A timer in the bottom right corner counts up from 0:00 to show any additional time that samples are held at 4 °C. -
The LCD display changes to Completed, and the thermal block sounds an audible beep to signal that the program has been completed.
Note: The timer continues to show additional time that samples are held at 4 °C. -
Remove the cluster-tube rack and follow the instructions in the BAX System User Guide to hydrate the PCR tablets with these lysates.
Press the SELECT/CONTINUE button.
Note: Lysates must be maintained at 4 °C during tablet hydration. -
The LCD display changes to Gram Negative Finished, and the thermal block sounds an audible beep to signal that the program has finished.
-
Press the SELECT/CONTINUE button to end the program. The LCD changes to display the program menu.
Run the 24E Program
Samples are heated at 37 °C for 30 minutes, then removed for the addition of lysis reagents. Samples then are heated to 55 °C for 30 minutes, then 95 °C for 10 minutes, then held at 4 °C for at least 5 minutes or until removed. The times vary slightly depending on the environment because a sensor gives active feedback to the instrument heater to ensure that proper temperatures are met.
-
Turn on the thermal block, then press the SCROLL button until the arrow ( → ) is next to the 24E option.
Press the SELECT/CONTINUE button. -
The unit beeps once, and the LCD display changes as the thermal block begins equilibrating to 37 °C.
Note: Do not load samples into the thermal block until the load prompt appears on the LCD display. -
When the thermal block has reached 37 °C, the unit sounds two audible beeps and the blue ACTION NEEDED LED activates. The LCD display changes to Load Samples.
-
After the Load Samples prompt appears, place a rack of prepared samples into the metal tube holder.
Note: See the BAX System User Guide for information on preparing samples .
Press the SELECT/CONTINUE button. -
The LCD display changes to 37C 24E Lysis.
Note: A timer in the bottom right corner counts down from 30:00 and shows the number of minutes remaining in this step. -
After lysis at 37 °C is complete, the blue ACTION NEEDED LED activates, and the thermal block sounds two audible beeps to signal that user action is required. The LCD display changes to Remove & Press Continue.
-
Remove the rack of cluster tubes from the metal tube holder.
Press the SELECT/CONTINUE button. -
The unit beeps once, and the LCD display changes as the thermal block begins equilibrating to 55 °C.
While the block is heating, follow the instructions in the BAX System User Guide to add additional lysing reagents to each sample.
Note: Do not reload samples into the thermal block until the reload prompt appears on the LCD display. -
When the thermal block has reached 55 °C, the unit sounds two audible beeps and the blue ACTION NEEDED LED activates. The LCD display changes to Reload Samples.
-
After the Reload Samples prompt appears, place the rack of samples with additional lysing reagents into the metal tube holder.
Press the SELECT/CONTINUE button. -
The LCD display changes to 55C 24E Lysis.
Note: A timer in the bottom right corner counts down from 30:00 and shows the number of minutes remaining in this step. -
After lysis at 55 °C is complete, the LCD display changes to Heating to 95C.
No user action is required.
Note: A timer in the bottom right corner counts up from 0:00 during the temperature change. -
After the thermal block reaches 95 °C, the LCD display changes to Denaturing @ 95C.
No user action is required.
Note: A timer in the bottom right corner counts down from 10:00 and shows the number of minutes remaining in this step. -
After denaturation at 95 °C is complete, the LCD display changes to Cooling to 4C, and the thermal block sounds two audible beeps to signal the change.
No user action is required.
Note: A timer in the bottom right corner counts up from 0:00 during the temperature change. -
After samples have cooled, the LCD display changes to Sample PCR Ready, the blue ACTION NEEDED LED activates and the thermal block sounds four audible beeps to signal that samples can be removed.
Press the SELECT/CONTINUE button.
Note: A timer in the bottom right corner counts up from 0:00 to show any additional time that samples are held at 4 °C. -
The LCD display changes to Completed, and the thermal block sounds an audible beep to signal that the program has been completed.
Note: The timer continues to show additional time that samples are held at 4 °C. -
Remove the cluster-tube rack and follow the instructions in the BAX System User Guide to hydrate the PCR tablets with these lysates.
Press the SELECT/CONTINUE button. -
The LCD display changes to 24E Finished and the thermal block sounds an audible beep to signal that the program has finished.
-
Press the SELECT/CONTINUE button to end the program. The LCD changes to display the program menu.
Run the RT Listeria Program
Samples are heated to 55 °C for 30 minutes, then 95 °C for 10 minutes, then held at 4 °C for at least 5 minutes or until removed. The times vary slightly depending on the environment because a sensor gives active feedback to the instrument heater to ensure that proper temperatures are met.
-
Turn on the thermal block, then press the SCROLL button until the arrow ( → ) is next to the RT Listeria option.
Press the SELECT/CONTINUE button. -
The unit beeps once, and the LCD display changes as the thermal block begins equilibrating to 55 °C.
Note: Do not load samples into the thermal block until the load prompt appears on the LCD display. -
When the thermal block has reached 55 °C, the unit sounds two audible beeps and the blue ACTION NEEDED LED activates. The LCD display changes to Load Samples.
-
After the Load Samples prompt appears, place a rack of prepared samples into the metal tube holder.
Note: See the BAX System User Guide for information on preparing samples.
Press the SELECT/CONTINUE button. -
The LCD display changes to RTLis Lysis 55C.
Note: A timer in the bottom right corner counts down from 30:00 and shows the number of minutes remaining in this step. -
After lysis at 55 °C is complete, the LCD display changes to Heating to 95C.
No user action is required.
Note: A timer in the bottom right corner counts up from 0:00 during the temperature change. -
After the thermal block reaches 95 °C, the LCD display changes to Denaturing @ 95C.
No user action is required.
Note: A timer in the bottom right corner counts down from 10:00 and shows the number of minutes remaining in this step. -
After denaturation at 95 °C is complete, the LCD display changes to Cooling to 4C, and the thermal block sounds two audible beeps to signal the change.
No user action is required.
Note: A timer in the bottom right corner counts up from 0:00 during the temperature change. -
After the samples have cooled, the LCD display changes to Sample PCR Ready, the blue ACTION NEEDED LED activates, and the thermal block sounds four audible beeps to signal that samples can be removed.
Press the SELECT/CONTINUE button.
Note: A timer in the bottom right corner counts up from 0:00 to show any additional time that samples are held at 4 °C. -
The LCD display changes to Completed, and the thermal block sounds an audible beep to signal that the program has been completed.
Note: The timer continues to show additional time that samples are held at 4 °C. -
Remove the cluster-tube rack and follow the instructions in the BAX System User Guide to hydrate the PCR tablets with these lysates.
Press the SELECT/CONTINUE button.
Note: Lysates must be maintained at 4 °C during tablet hydration. -
The LCD display changes to RT Listeria Finished, and the thermal block sounds an audible beep to signal that the program has finished.
-
Press the SELECT/CONTINUE button to end the program. The LCD changes to display the program menu.
Clean and Decontaminate the Thermal Block
Clean the Thermal Block
- Do not attempt to clean the thermal block or metal tube holder when they are hot.
- Remove dust and debris by wiping the instrument surfaces with a lint-free cloth.
- Wipe all spills or condensation with a soft cloth or paper towel, if needed.
- If needed, the casing can be wiped off with a damp cloth using mild soap or detergent.
- Do not use solvents or cleansers containing iodine or acetone, as these solutions could damage the paint or display window of the block.
Decontaminate the Thermal Block
If contamination should occur, the following steps can be performed under
sterile conditions to help remove the contaminants.
- Wipe the surface of the thermal block unit with a cloth dampened with 20% bleach solution*, followed by a 70% ethanol rinse to prevent damage to the equipment.
- Soak metal tube holders in 20% bleach solution* for about 5 minutes, then rinse thoroughly with water and allow to air dry.
An alternate cleaner designed to remove free DNA may also be investigated. A 10% bleach solution may also be used for decontamination in accordance with your laboratory SOP but may require repetition of the cleaning to ensure the removal of all amplicons.
Tips and Troubleshooting
Abort a Program
The Automated Thermal Block automatically progresses through each sequential
step until the program is complete. If necessary, you can cancel the program
prior to completion via the Abort option.
- While the program is running, press the SCROLL button to display the Abort menu.
- Press the SCROLL push button until the arrow (→) is next to the Abort option, then press the SELECT/CONTINUE push button.
- The LCD display shows a message to confirm that the program was terminated.
- To start another program, press the SELECT/CONTINUE button to return to the program menu.
To turn off the unit, toggle the on/off switch on the back panel.
Error Messages
If the power fails or any interruption occurs during a run, the LCD displays
an error message. Affected samples in the thermal block should not be used
with the BAX System. Remove and dispose of these samples according to your
standard operating procedures.
The letter F (Fault) appears in the lower right corner of the LCD display if an interruption occurs during any step in the program.
This fault indicator appears on the LCD display until the program completes or the program is aborted by the user. Affected samples should not be used with the BAX System. Remove and dispose of these samples according to your standard operating procedures. The Power Failure message appears when power is interrupted or fails during a run. Toggle the on/off switch on the back panel to reset the unit.
Affected samples should not be used with the BAX System. Remove and dispose of these samples according to your standard operating procedures.
© Copyright 2023 Hygiena® or its affiliates. All rights reserved worldwide. No part of this publication may be reproduced or transmitted in any form, or by any information storage and retrieval system, without the prior agreement and written permission of Hygiena. Information in this document is subject to change without notice. Hygiena assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Hygiena be liable for incidental, special, multiple or consequential damages in connection with or arising from the use of this document.
-
Trademarks
The Hygiena logo and BAX® are registered trademarks of Hygiena or its affiliates. All other trademarks are the property of their holders. -
Field of Use
The Automated Thermal Block is manufactured by Torrey Pines Scientific® for use with the BAX System. Please see BAX System documentation for details on Field of Use. Please read the Limitation of Warranty and Liability before using the product. -
Warranty
This product is warranted by Torrey Pines Scientific to be free from defects in material and workmanship for a period of one year from the date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the one year, correction of the defect will be made without charge. For warranty and repair issues, contact Torrey Pines Scientific at 760-930-9400. -
Technical Support
If you have any questions or comments on the Automated Thermal Block, please contact your distributor or Hygiena at 800-863-6842 or email diagnostics.support@hygiena.com for technical assistance. -
Contact Information
For more information, visit www.hygiena.com/contact.