Omron Relion 7400REL Automatic Blood Pressure Monitor User Manual
INTRODUCTION
Thank you for purchasing the ReliOn® 7400REL Automatic Blood Pressure Monitor. Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.
The 7400REL comes with the following components:
- Monitor
- Wide Range Cuff
- AC Adapter
- Instruction Manual printed in English and Spanish
- Quick Start Guide
Please read this instruction manual thoroughly before using the unit. Please keep it for future reference. For specific information about your blood pressure, CONSULT YOUR DOCTOR.
INTENDED USE
This monitor is designed to measure the blood pressure and pulse rate of
adults within the range of the designated arm cuff, following the directions
provided in this instruction manual. The monitor is designed for home use.
Read the Safety Information in this instruction manual before using this
monitor. To ensure the correct use of the product, basic safety measures
should always be followed including the warnings and cautions listed in this
instruction manual.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could
result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result
in minor or moderate injury to the user or patient or damage to the equipment
or other property.
OPERATING THE DEVICE
Contact your physician for specific information about your blood pressure. Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider. DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure. The monitor is not intended to be a diagnostic device.
If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Contact a physician immediately. Read all of the information in the instruction manual and any other literature in the box before operating the unit. Do not use this device on infants or persons who cannot express their intentions. Read the Special Conditions section of the instruction manual if your systolic pressure is known to be more than 220 mmHg. Inflating to a higher pressure than necessary may result in bruising where the cuff is applied.
Operate the device only as intended. Do not use the device for any other
purpose. Dispose of the device, components and optional accessories according
to applicable local regulations. Unlawful disposal may cause environmental
pollution.
Do not use a cellular phone near the device. It may result in an operational
failure. Use only authorized parts and accessories. Parts and accessories not
approved for use with the device may damage the unit. Use only 1.5V Alkaline
batteries with this device. Other types of batteries may damage the unit. Do
not use it in a location with moisture or a location where water may splash on
the unit.
RISK OF ELECTRICAL SHOCK
Do not plug or unplug the power cord into the electrical outlet with wet
hands. Do not overload power outlets. Plug the device into the appropriate
voltage outlet.
CARE AND MAINTENANCE
Do not subject the monitor to strong shocks, such as dropping the unit on the
floor. Do not submerge the device or any of the components in water. Store the
device and the components in a clean, safe location. Changes or modifications
not approved by the manufacturer will void the user warranty. Do not
disassemble or attempt to repair the unit or components.
BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
- Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes before taking a measurement. Rest for at least 15 minutes before taking the measurement.
- Stress raises blood pressure. Avoid taking measurements during stressful times.
- Measurements should be taken in a quiet place.
- Remove tight-fitting clothing from your left arm.
- Sit on a chair with your feet flat on the floor. Rest your left arm on the table so that the cuff is at the same level as your heart.
- Remain still and do not talk during the measurement.
- Wait 2-3 minutes between measurements. The wait time allows the arteries to return to the condition before taking the blood pressure measurement.
- You may need to increase the wait time depending on your physiological characteristics.
- Keep a record of your blood pressure and pulse readings for your physician. A single measurement does not provide an accurate indication of your true blood pressure.
- You need to take and record several readings over some time. Try to measure your blood pressure at the same time each day for consistency.
UNIT DISPLAY KNOW YOUR UNIT
Main Unit
Components
UNIT DISPLAY
DISPLAY SYMBOLS IRREGULAR HEARTBEAT SYMBOL ( )
When the monitor detects an irregular rhythm two or more times during the
measurement, the Irregular Heartbeat Symbol ( ) will appear on the display
with the measurement values. An irregular heartbeat rhythm is defined as a
rhythm that varies by less than 25% from the average rhythm or 25% more than
the average rhythm detected while the monitor is measuring the systolic and
diastolic blood pressure.
If the irregular heartbeat symbol ( )
displays with your measurement results, we recommend you consult your
physician. Follow the directions of your doctor.
MOVEMENT ERROR SYMBOL ( )
The Movement Error Symbol ( ) is displayed if you move your body during the
measurement. Please remove the arm cuff, and wait 2-3 minutes. Reapply the arm
cuff and take another measurement.
AVERAGE VALUE SYMBOL ( )
The Average Value Symbol ( ) is displayed when you press the memory button.
The most recent average reading appears on the display screen.
HEARTBEAT SYMBOL ( )
The Heartbeat symbol flashes on the display at every heartbeat during the
measurement.
DANGER
TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK, CAREFULLY FOLLOW THESE
INSTRUCTIONS. For connection to a power supply not in the U.S.A., use an
attachment plug adapter of the proper configuration for the power outlet. The
power unit (AC Adapter) is intended to be correctly oriented in a vertical or
floor-mount position.
NOTES
- Use only the authorized AC Adapter included with this monitor.
- We recommend you install the batteries even if the AC Adapter is used. If no batteries are installed, you may need to reset the date and time if the AC Adapter is disconnected. The measurement results will not be deleted.
CAUTION
Use only authorized parts and accessories. Parts and accessories not approved
for use with the device may damage the unit.
TO CONNECT THE AC ADAPTER
- Insert the AC Adapter Plug into the AC Adapter Jack on the right side of the monitor as illustrated.
- Plug the AC Adapter into a 120V AC outlet (60 cycles).
TO DISCONNECT THE AC ADAPTER
- Hold and pull the Housing to remove the AC Adapter from the electrical outlet. Do not remove by pulling the cord.
- Remove the AC Adapter plug from the monitor.
WARNING
Do not plug or unplug the power cord into the electrical outlet with wet hands
WARNING
Do not overload power outlets. Plug the device into the appropriate voltage
outlet
BATTERY INSTALLATION
- Press the hook on the top of the battery cover and lift the cover off in the direction of the arrow.
- Install 4 “AA” size batteries so the + (positive) and – (negative) polarities match the polarities ofthe battery compartment as indicated.
- Replace the battery cover.
BATTERY REPLACEMENT
Low Battery Indicator When the Low Battery Indicator appears on the display, turn the monitor off and remove all the batteries. Replace with four new batteries at the same time. Long-life alkaline batteries are recommended.
WARNING
If battery fluid should get in your eyes, immediately rinse with plenty of
clean water. Contact a physician immediately.
NOTES: Turn the unit off before replacing the batteries. If the batteries are removed while the unit is still on, the date and time will be reset to that of the previous use. However, the measurement results are not deleted. When the batteries are replaced, you may need to reset the date and time. If appears on the display, refer to “Setting the Date and Time”.
CAUTION
Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
SETTING THE DATE AND TIME
Set the monitor to the current date and time before measuring the first time.
TO START THE SETTING
Press the SET button to adjust the date and time The year flashes on the display.
SETTING THE YEAR
The year can be set between 2013 and 2040. When the display reaches 2040, it
will return to 2013. Press the Up button ( ) to advance by one year. Press the
SET button to set the current year. The month flashes on the display.
NOTES : Press the Down button ( ) to decrease the date and time values.
Press and hold the button to increase or decrease the date and time values
faster.
SETTING THE MONTH
Press the Up button ( ) to advance by one month. Press the SET button to
set the current month. The day flashes on the display.
SETTING THE DAY
Press the Up button ( ) to advance by one day. Press the SET button to set the
current day. The hour flashes on the display.
SETTING THE HOUR
The time is set using AM or PM. Press the Up button ( ) to advance by one
hour. Press the SET button to set the current hour. The minute flashes on the
display.
SETTING THE MINUTE
Press the Up button ( ) to advance by one minute. Press the SET button to set
the minute setting.
TO TURN THE MONITOR OFF
Press the ON/OFF button.
NOTE: The monitor will automatically turn off 2 minutes after the minute
has been set.
APPLYING THE ARM CUFF
CAUTION
Read all of the information in the instruction manual and any other literature
in the box before operating the unit.
CAUTION
This device is intended for use in measuring blood pressure and pulse rate in
the adult population. Do not use this device on infants or persons who cannot
express their intentions.
APPLYING THE CUFF ON THE LEFT ARM
- Make sure the air plug is securely inserted into the main unit.
- Remove tight-fitting clothing from your upper arm.
- Apply the cuff to your left upper arm so the arrow is centred on the inside of your arm and aligned with your middle finger.
- The air tube runs down the inside of your arm. The bottom of the cuff should be approximately 1/2 ” above your elbow.
- Wrap the cuff firmly in place around your arm using the cloth fastening.
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm, use this instruction for Step
3. Apply the cuff to your right upper arm so the arrow and the air tube run
along the inside of your arm. The bottom of the cuff should be approximately
1/2 ” above your elbow.
NOTE: Be careful not to rest your arm on the air tube. This will restrict
the flow of air to the cuff.
TAKING A MEASUREMENT
The monitor is designed to store the blood pressure and pulse rate in the
memory of two people (USER A or USER B) every time a measurement is completed.
-
Sit on a chair with your feet flat on the floor. Place your arm on a table so the cuff is level with your heart. Keep still and do not talk during measurement.
-
Select your USER ID (A or B).
-
Press the ON/OFF button. All symbols appear on the display. The cuff starts to inflate automatically. As the cuff inflates, the monitor automatically determines your ideal inflation level.
-
Remain still and do not move your arm until the entire measurement process is completed.
NOTE : To stop the inflation or measurement, push the ON/OFF button. The monitor will stop inflating, start deflating, and will turn off. -
Inflation stops automatically and the measurement is started. As the cuff deflates, decreasing numbers appear on the display and the Heartbeat Symbol ( ) flashes.
-
When the measurement is complete, the arm cuff completely deflates. Your blood pressure and pulse rate are displayed.
The 1999 WHO/ISH guidelines for the management of hypertension are as follows:
Category / Blood pressure
- Hypertension
- High-Normal
- Normal
Systolic blood pressure
- above 139 mmHg
- 130 – 139 mmHg
- 129 mmHg or less
Diastolic blood pressure
- above 89 mmHg
- 85 – 89 mmHg
- 84 mmHg or less
This is a general guideline because blood pressure depends on age and morbidity. A lower home blood pressure goal may be recommended for certain patients, including diabetic patients, pregnant women, and patients with renal failure. Press the ON/OFF button to turn the monitor off.
NOTE: The monitor will automatically turn off after 2 minutes.
WARNING
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous. Follow
the instructions of your physician or licensed healthcare provider.
WARNING
DO NOT adjust medication based on measurement results from this blood pressure
monitor. Take medication as prescribed by your physician. Only a physician is
qualified to diagnose and treat High Blood Pressure.
WARNING
The monitor is not intended to be a diagnostic device.
SPECIAL CONDITIONS
CAUTION
Inflating to a higher pressure than necessary may result in bruising where the
cuff is applied. If your systolic pressure is known to be more than 220 mmHg,
press and hold the ON/OFF button until the monitor inflates 30 to 40 mmHg
higher than your expected systolic pressure after the cuff starts to inflate.
NOTES: Do not apply more pressure than necessary. The monitor will not inflate above 299 mmHg. When the measurement is complete, the arm cuff completely deflates. Your blood pressure and pulse rate are displayed.
USING THE MEMORY FUNCTION
The monitor automatically stores up to 100 sets of measurement values (blood
pressure and pulse rate) for each user (A or B). When 100 sets of measurement
values are stored, the oldest record is deleted to save the most recent
values.
AVERAGING FUNCTION
The monitor calculates the average reading based on the three most recent sets
of measurement values taken within 10 minutes of the most recent reading.
NOTES: If two sets of measurement values are stored in the memory for 10
minutes, the average is based on the two sets of measurement values. If one
set of measurement values is stored, this is displayed. When viewing the
reading taken without setting the date and time, “-:–” is displayed instead of
the date and time. The average reading will not be calculated without setting
the date and time.
TO DISPLAY THE MEASUREMENT VALUES
Select your USER ID (A or B). Press the Memory button ( ). The Average Symbol
( ) appears on the display when the average reading is shown. Press the Down
button ( ) to display the most recent set of measurement values on the screen.
The values are displayed from the most recent to the oldest.
NOTES: The Memory number appears for a second before the pulse rate is
displayed. The newest set is number “1”. The date and time are alternately
displayed with the measurement values. Press the Up ( ) or Down ( ) button to
display the next set of values.
NOTE: Press and hold the Up or Down button to display the values faster.
TO DELETE ALL VALUES STORED IN THE MEMORY
The values stored in the memory are deleted by USER ID. You cannot partially
delete values stored in the memory. All values for the user you select will be
deleted.
- Select your USER ID (A or B).
- While holding the Memory button ( ), press the ON/OFF button down for more than 2 seconds until the display appears as it does in the following illustration. Press the ON/OFF button to turn the monitor off.
NOTE : The monitor will automatically turn off after 2 minutes.
CARE AND MAINTENANCE
To keep your blood pressure monitor in the best condition and protect the unit
from damage, follow the directions listed below: Do not forcefully bend the
arm cuff or air tube. Do not fold tightly. Clean the monitor with a soft dry
cloth.
Do not use any abrasive or volatile cleaners. Do not attempt to clean the
cuff.
CAUTION
Do not submerge the device or any of the components in water. Do not subject
the monitor to extreme hot or cold temperatures, humidity or direct sunlight.
CAUTION
Store the device and the components in a clean, safe location.
CAUTION
Do not subject the monitor to strong shocks, such as dropping the unit on the
floor. Remove the batteries if the unit will not be used for three months or
longer. Always replace all the batteries with new ones at the same time. Use
the unit in a manner consistent with the instructions provided in this manual.
CAUTION
Use only authorized parts and accessories. Parts and accessories not approved
for use with the device may damage the unit.
CAUTION
Changes or modifications not approved by the manufacturer will void the user
warranty. Do not disassemble or attempt to repair the unit or components.
ERROR INDICATORS
TROUBLESHOOTING TIPS
PROBLEM | CAUSES AND SOLUTIONS |
---|
No power.
No display appears on the unit.
| Replace all four batteries with new ones.
Check the battery installation for proper placement of the battery polarities.
Measurement values appear too high or too low.
| Blood pressure varies constantly. Many factors including stress, time of day, how you wrap the cuff, may affect your blood pressure.
Review the sections “Before Taking a
Measurement” and “Taking a Measurement”.
FCC STATEMENT
Note:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has
been tested and found to comply with the limits for a Class B digital device,
under part 15 of the FCC rules. These limits are designed to provide
reasonable protection against harmful interference in a residential
installation. The product generates, uses, and can radiate radio frequency
energy and, if not installed and used by the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If the product does
cause harmful interference to radio or television reception, which can be
determined by turning the product on and off, the user is encouraged to try to
correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the product and the receiver.
- Connect the product to an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise
emissions from digital apparatus as set out in the interference-causing
equipment standard entitled “Digital Apparatus”, ICES-003 of the Canadian
Department of Communication Changes or modifications not expressly approved by
the party responsible for compliance could void the user’s authority to
operate the equipment.
LIMITED WARRANTY
Your ReliOn® 7400REL Automatic Blood Pressure Monitor, excluding the arm cuff,
is warranted to be free from defects in materials and workmanship appearing
within 5 years from the date of purchase when used by the instructions
provided with the monitor. The arm cuff is warranted to be free from defects
in materials and workmanship appearing within one year from the date of
purchase when the monitor is used by the instructions provided with the
monitor. The above warranties extend only to the original retail purchaser.
We will, at our option, repair or replace without charge any monitor or arm cuff covered by the above warranties. Repair or replacement is our only responsibility and your only remedy under the above warranties. To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the address of the Inspection centre. Enclosed is the Proof of Purchase. Include a letter, with your name, address, phone number, and description of the specific problem. Pack the product carefully to prevent damage in transit. Because of possible loss in transit, we recommend insuring the product with the return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY. OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other rights that vary by jurisdiction. Because of special local requirements, some of the above limitations and exclusions may not apply to you. FOR CUSTOMER SERVICE Visit our website at: www.omronhealthcare.com Call toll-free: 1-800-634-4350
SPECIFICATIONS
NOTE: These specifications are subject to change without notice.
Manufactured for OMRON HEALTHCARE Co., Ltd.
Made in China
Distributed by OMRON HEALTHCARE, INC.
1925 West Field Court
Lake Forest, IL 60045 U.S.A.
www.omronhealthcare.com
© 2012 OMRON HEALTHCARE, INC.
ReliOn® is a registered trademark used under license by OMRON HEALTHCARE
Distributed by
Wal-Mart® Stores, Inc.
Bentonville, AR 72716
Download PDF:Omron Relion 7400REL Automatic Blood Pressure Monitor User Manual
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>