TOTO 22163 Lateral Turning System Instruction Manual
- January 4, 2024
- TOTO
Table of Contents
22163 Lateral Turning System
Product Information
The Toto System is a revolutionary bed platform that provides
customizable inflation and rest durations for enhanced comfort and
pressure relief. It consists of a platform, control unit, and air
cells that can be adjusted to meet individual needs.
Specifications
- Inflation Time: 30 minutes to 240 minutes
- Rest Time: 30 minutes to 240 minutes
- Control Unit: Toto Touch Control Unit
- Cleaning: Can be cleaned with chlorine-based solution
Product Usage Instructions
Toto System Installation
To install the Toto System, follow these steps:
-
Place and centralize the platform directly onto the bed
frame. -
Ensure that each side of the platform can freely inflate
without any restrictions from the mattress or bed frame. -
Place the mattress or mattress system directly on top of the
platform.
Connecting the Platform to the Control Unit
To connect the platform to the control unit, follow these
steps:
1. Using the bed hooks located on the rear of the control unit,
securely hang it on the footboard of the bed. Alternatively,
position it on any stable horizontal surface.
2. Connect the control unit to mains power using the supplied
power lead.
Operating Instructions
1. To change the default settings, press the button on the Toto
Touch Control Unit to scroll through the LED lit options until you
have arrived at your preferred choice.
2. Air cell inflation begins, and each side remains inflated for
the duration that is selected.
3. If the control unit is turned off, it reverts to the default
settings when switched back on. Reselect the preferred options
using the above instructions.
Cleaning and Care
All components of the Toto System can be cleaned and
decontaminated with either a 0.1% or 1% chlorine-based solution.
After cleaning, ensure that the control unit is switched off and
unplugged before cleaning begins. Rinse all components with clean
water after using the chlorine-based solution.
FAQ
Problem: No LEDs lit; no audible alarm.
Cause: Initial Checks / Trouble Shooting
Quick Reference Guide
pioneering simplicity
Toto® Lateral Turning System, with Toto® Touch Control Unit
Toto System Installation
Connecting the Platform to the Control Unit
Using the quick connectors, connect the grey inflation hoses to the platform
air cell hoses red to red. Connect the grey inflation hoses to the control
unit. Check the CPR deflation valves are securely in place.
Operating Instructions
1. To turn the unit on, press and hold the power button (1) for three
seconds. The default settings are; Air cell side selection (2) both sides
Air cell inflation time (9) 120 minutes Air cell rest time (8) 120 minutes
2. To change the default settings, press the button to scroll through the LED
lit options until you have arrived at your preferred choice.
3. Air cell inflation begins, and each side remains inflated for the duration
that is selected.
Note – If the control unit is `turned off’ it reverts to the default settings
when switched back on. Reselect the preferred options using the above
instructions.
Cleaning and Care
All components can be cleaned and decontaminated with either a 0.1% or 1%
chlorine-based solution when followed by a thorough rinse with clean water.
Please ensure that the control unit is switched off and unplugged before
cleaning begins.
Toto System Installation
This Quick Reference Guide has been produced to assist Healthcare
Professionals and other users with the functional aspects of the system. This
guide does not replace the general safety warnings, cautions and
recommendations provided in the Toto® Lateral Turning System, with Toto® Touch
Control Unit User Guide. Healthcare providers and carers are often required to
reposition their patients every 4 hours; morning noon and night, so to reduce
the risk of the development of pressure ulcers. This can be very time
consuming, ineffective, be a drain on resources, can introduce personal risk
to staff and patients, and might not be well tolerated. Toto is an automatic
lateral turning system designed to assist with patient turning. Consisting of
a digital control unit and turning platform which is fitted beneath the
patient’s mattress, Toto automatically turns patients relieving pressure over
their most vulnerable areas. Unlike other patient turning systems, Toto
provides regular and consistent complete body turning in a lightweight, easy
to use, set up and access format.
Installation
Place and centralise the platform directly onto the bed frame and then place
the mattress or mattress system directly on top.
! IMPORTANT The platform should be free to inflate. Please ensure that each
side of the platform can freely inflate and not be restricted by the mattress
or the bed frame.
Toto Touch Control Unit
Using the bed hooks located on the rear of the control unit, securely hang on
the footboard of the bed. Alternatively, position on any stable horizontal
surface. Connect to mains power via the power lead supplied.
! IMPORTANT A Healthcare Professional should determine inflation and rest
durations.
ATTENTION!
FOR CLINICIANS AND CARERS USE ONLY
Please tick the following options once installed to indicate the settings
selected on the Toto Touch® Control Unit.
AIR CELL SIDE SELECTION
L&R R L
INFLATE TIME 240mm 180mm 120mm 60mm 30mm
REST TIME 30mm 60mm 120mm 180mm 240mm
Toto Touch Control Unit Interface
1. Start, Stop and Therapy Pause Therapy pause mode can be engaged by
pressing this button once. This temporarily deactivates all control unit
functions, returning the platform to a flat position.
9. Air Cell Inflation Time Length of time the air cells remain inflated;
minimum 30 minutes, maximum 240 minutes. Default is set to 120 minutes (2
hours)
2. Air Cell Side Selection
Left only, right only or both. Defaults to both (left and right inflation)
each time the unit is switched on.
1
touch
2
L & R R
INFLATE TIME
4
240 min
5
180 min
30 min
120 min
60 min
60 min
120 min
30 min
180 min 6
3
L
240 min 7 REST TIME
3. Alarm Mute & Interface Lock
The alarm can be paused for up to 10 minutes while troubleshooting, after
which time, the alarm recommences if the problem remains. This button also
activates the interface lock. (See note below)
4. Low System Pressure Alarm Indicator
Indicates a potential air leak in the system.
8
5. High System Pressure Alarm Indicator
9
Indicates a potential blockage in the system.
8. Air Cell Rest Time
Length of time the air cells remain deflated and the platform is flat; minimum
30 minutes, maximum 240 minutes. Default is set to 120 minutes (2 hours)
7. Interface Lock Indicator
For further information on the system alarms, please refer to the Trouble Shooting Guide overleaf.
6. Service Indicator
An amber LED indicates when routine servicing is due.
System Error Trouble Shooting Guide
Problem
No LEDs lit; no audible alarm.
Cause
Initial Checks / Trouble Shooting
· Control unit may not be attached to a power source.
· Fuse may need replacing.
· Check the control unit is connected to an active mains power outlet which is
switched on.
· Check the control unit is switched on.
· If following points 1 and 2 does not resolve the problem, switch off and
unplug the control unit: – Check the mains plug fuse (3A). – Check the control
unit fuse (1A).
· Restart the control unit.
Low pressure alarm LED (4) permanently lit yellow; an audible alarm can be heard.
· Pressure has fallen below the minimum operational requirements.
· Press the alarm mute button.
· Check the CPR valve is properly closed, ensuring all sealing connectors are
fully pressed in.
· Check each air cell is securely attached to the connecting air cell hose.
· Check all air cells and hoses for any air leakage.
High pressure alarm LED (5) permanently lit yellow; an audible alarm can be heard.
· Air flow between the control unit and the platform is obstructed.
· Press the alarm mute button.
· Check for twists or kinks in the air hoses between the platform and control
unit.
· Resolve and wait for the alarm to reset.
Service LED (6) permanently lit amber.
· Service required.
Please contact Frontier Therapeutics Customer Services for support Tel. +44 (0) 1495 235800 The Control Unit will continue to function even though the service LED remains lit.
Interface panel is lit but unresponsive.
· Interface panel is locked.
Audible alarm sounding; no LEDs lit.
· Accidental removal of power lead.
· Accidental mains wall power is turned off.
· Mains power failure.
· Check that screen lock LED is lit amber (9). · Press and hold the alarm mute
button for four
seconds.
· Reconnect mains power lead to the control unit. · Check the mains power wall
switch is turned on. · Attempt to restart the control unit.
Trouble Shooting Not Working?
If you require any further assistance with the trouble shooting or repair of
your Toto System please call the number below. Additional information can be
found on our website www.frontier-group.co.uk
© January 2021 Frontier Therapeutics Limited. All rights reserved. GP0133_A57_07
Frontier Therapeutics Limited Newbridge Road Industrial Estate,
Blackwood, Caerphilly, NP12 2YN, United Kingdom.
Tel: Email:
+44 (0)330 460 6030 info@frontier-group.co.uk
www.frontier-group.co.uk