Winsen ME2-CO-14×50-C Electrochemical Carbon Monoxide Sensor Instruction Manual

June 15, 2024
Winsen

Electrochemical
Carbon Monoxide Sensor
(Model:ME2-CO-Ф14×50-C)
Manual
Version: 1.1

Issue Date: 2023-3-20

ME2-CO-Ф14×50-C Carbon Monoxide sensor

ME2-CO-Ф14×15-C Carbon monoxide sensor is a fuel cell type sensor. Carbon monoxide and oxygen undergo corresponding REDOX reaction on the working electrode and the opposite electrode and release charge to form current. The current generated is proportional to the concentration of carbon monoxide and follows Faraday’s law.

Features

  • Low consumption
  • High precision
  • High sensitivity
  • Wide linear range
  • Good anti-interference ability
  • Excellent repeatability and stability

Application

Widely used for home safety, underground garage, generator and other fields of carbon monoxide concentration detection.

Technical Parameter

Table1.

Detection gas CO
Measurement Range 0-1000ppm

Max detecting
concentration| 2000ppm
Sensitivity| ( 0.8-4) nA/ppm
Resolution| 1ppm
Response time ( T90 )| 530S
Load resistance
( recommended )| 10000
Repeatability| <3 % output value
Output Linearity| linear
Zero drift (-20°C–40°C)| 510ppm
Temperature range| -20°C-50°C
Humidity range| 15%-90%RH
Pressure range| standard atmosphere± 10%
Lifespan| 7 years

Winsen ME2-CO-14x50-C Electrochemical Carbon Monoxide Sensor -
ApplicationFig1. Sensor Structure
Unit is mm, tolerance is ±0.15mm

Basic circuit.Winsen ME2-CO-14x50-C Electrochemical Carbon Monoxide
Sensor - Basic circuit.

Characterization

Winsen ME2-CO-14x50-C Electrochemical Carbon Monoxide Sensor -
Characterization

Cross sensitivity:
ME2-CO-Ф14×50-C sensor also responds to other gases besides CO. Below are the response characteristics of interferential gases

Gas Concentration ME2-CO-Ф14×50-C
H2S 100ppm 0ppm
C2H4 100ppm 80ppm
NO 35ppm 6ppm
NO2 5ppm 0ppm
C2H5OH 1000ppm 0ppm
CL2 10ppm 1ppm
SO2 20ppm 0.6ppm
H2 500ppm 43ppm
NH3 50ppm 1ppm
CH3CL 5ppm 0ppm
ETO 10ppm 0ppm
C6H6 100ppm 1.5ppm
C3H6O 100ppm 3.5ppm
CH3OH 200ppm 0ppm

Application Notes:

  • During installation, the lead can be welded, and the solder cannot contact the sensor.
  • Before using, power on to aging for more than 48 hours is necessary.
  • Don’t disassemble the sensor.
  • Avoid contacting organic solvent (including Silicone rubber and other adhesive), coatings, medicine, oil and high concentration gases.
  • All the electrochemical sensors shall not be encapsulated completely by resin materials, and shall not immerse in non-oxygen environment, otherwise, it will damage the function of sensor.
  • All electrochemical sensors shall not be applied in corrosive gas environment, or the sensor will be damaged.
  • Zero calibration should be done in clean air.
  • During test and usage, sensors should avoid the gas inflow vertically.
  • The inlet hole can’t be choked and polluted.
  • The sensor shall not be subjected to excessive impact or vibration.
  • Do not use the sensor if its shell is damaged or deformed.
  • In high concentration gas environment, the recovery to the initial state is slow after a long time of use.
  • When the sensor is stored, the working electrode and the counter electrode should be in short circuit.
  • Do not use hot melt adhesive or sealant whose curing temperature is higher than 80℃ to encapsulate sensors.
  • Do not store or use the sensor in high concentration alkaline gas for a long time.

Note: To keep continual product development, we reserve right to change design features without prior notice!

Zhengzhou Winsen Electronics Technology Co., Ltd
Add: No.299, Jinsuo Road, National Hi-Tech Zone, Zhengzhou 450001 China
Tel: +86-371-67169097/67169670
Fax: +86-371-60932988
E-mail: sales@winsensor.com
Website: www.winsen-sensor.com

Winsen ME2-CO-14x50-C Electrochemical Carbon Monoxide Sensor - QR
CODEhttps://www.winsen- sensor.com/

Leading gas sensing solutions supplier in China
Tel: 86-371-67169097
Fax: 86-371-60932988
Email: sales@winsensor.com