PCE-PFG-20 Force Gage User Manual
- June 15, 2024
- PCE
Table of Contents
- PCE-PFG-20 Force Gage
- Product Usage Instructions
- Safety notes
- Specifications
- Delivery scope
- Dimensions
- Key functions
- Battery
- Switching on and off
- Measuring modes
- Communication interface and output interface
- More settings
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
PCE-PFG-20 Force Gage
Specifications
-
Model: PCE-PFG Series Force Gauge
-
Measurement Range:
- PCE-PFG 20: 0 … 20 N
- PCE-PFG 50: 0 … 50 N
- PCE-PFG 100: 0 … 100 N
- PCE-PFG 200: 0 … 200 N
- PCE-PFG 500: 0 … 500 N
- PCE-PFG 2K: 0 … 2000 N
-
Resolution:
- PCE-PFG 20: 0.005 N
- PCE-PFG 50: 0.01 N
- PCE-PFG 100: 0.02 N
- PCE-PFG 200: 0.05 N
- PCE-PFG 500: 0.1 N
- PCE-PFG 2K: 0.5 N
-
Accuracy: 0.3% of measurement range
-
Units: N, kgF, lbF
-
Sampling Rate: 500 Hz
-
Display: 1.8″ graphic display
-
Alarm Modes: Below, Inside, Outside
-
Memory: 100 measurements
-
Power Supply: Lithium battery 3.7 V / 1500 mAh
-
Battery Life: Up to 36 hours
-
Mains Adaptor / USB Charging Adaptor: 5V / 1A
-
Outputs:
- Interface: USB B
- Switching Output / Alarm Modes: MD6 with 2.85V when active
-
Protection Class: IP54
-
Operating and Storage Conditions: 35 … 65% RH,
non-condensing -
Dimensions: 189 x 707 x 34 mm
-
Weight: 450 g
-
Models with External Load Cell:
- Dimensions / Weight Load Cell: L 52 mm / H 72 mm / W 19 mm /M12 thread / 490 g
- Cable Length External Load Cell: Approx. 1.8 m
-
Handheld Dimensions: 189 x 707 x 34 mm
-
Handheld Weight: 240 g
Product Information
The PCE-PFG Series Force Gauge is a handheld device used for measuring force. It is available in different models with varying measurement ranges and resolutions. The force gauge features a 1.8″ graphic display and offers measurement units in N, kgF, and lbF.
Product Usage Instructions
Safety Notes
Please read the manual carefully before using the device for the first
time. It is important to note that only qualified personnel should use and
repair the device. Non-observance of the manual may result in damage or
injuries, which are not covered by the warranty.
Dimensions
The dimensions of the force gauge vary depending on the model. For models with
an internal measuring cell, the dimensions are as follows:
- Handheld Dimensions: 189 x 707 x 34 mm
For models with an external load cell, the dimensions are as follows:
- Load Cell Dimensions: L 52 mm / H 72 mm / W 19 mm / M12 thread
Key Functions
The force gauge is equipped with various key functions for easy operation. The key functions include:
- Power Key: Used to switch the meter on and off
- Mode Key: Used to select between different modes
- Del Key: Used to delete individual values
- Unit Key: Used to set the unit in measuring mode, increase displayed parameter, and select the next item in the menu
- Enter Key: Used to confirm entries or open menu items
- Dir Key: Used to switch between compressive and tensile alarm display
FAQ
Q: Where can I find more information about the product?
A: You can find more information about the product by using our product
search on our website: www.pce-instruments.com.
Q: How can I contact PCE Instruments?
A: You can contact PCE Instruments by referring to the contact details
provided at the end of the manual.
Q: What are the delivery scope items included with the force gauge?
A: The delivery scope for the PCE-PFG 2K force gauge includes the
following items:
- 1 x 2000 N S measuring cell
- 2 x M12 joint eyes
- 2 x pressure fittings
- 1 x USB cable
- 1 x USB charger
- 1 x user manual
- 1 x plastic carrying case
User manuals in various languages (français, italiano, español, português,
nederlands, türk, polski, русский, 中文) can be found by using our
product search on:
Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non- observance of the manual are excluded from our liability and not covered by our warranty.
- The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
- The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
- Do not expose the device to shocks or strong vibrations.
- The case should only be opened by qualified PCE Instruments personnel.
- Never use the instrument when your hands are wet.
- You must not make any technical changes to the device.
- The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
- The device must only be used with accessories from PCE Instruments or equivalent.
- Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
- Do not use the instrument in explosive atmospheres.
- The measurement range as stated in the specifications must not be exceeded under any circumstances.
- Non-observance of the safety notes can cause damage to the device and injuries to the user.
We do not assume liability for printing errors or any other mistakes in this
manual.
We expressly point to our general guarantee terms which can be found in our
general terms of business.
If you have any questions please contact PCE Instruments. The contact details
can be found at the end of this manual.
Specifications
Model | Measurement range | Resolution |
---|---|---|
PCE-PFG 20 | 0 … 20 N | 0.005 N |
PCE-PFG 50 | 0 … 50 N | 0.01 N |
PCE-PFG 100 | 0 … 100 N | 0.02 N |
PCE-PFG 200 | 0 … 200 N | 0.05 N |
PCE-PFG 500 | 0 … 500 N | 0.1 N |
PCE-PFG 2K | 0 … 2000 N | 0.5 N |
Accuracy | 0.3 % of measurement range | |
Units | N, kgF, lbF | |
Sampling rate | 500 Hz | |
Display | 1.8″ graphic display | |
Alarm modes | Below, Inside, Outside | |
Memory | 100 measurements | |
Power supply | Lithium battery 3.7 V / 1500 mAh | |
Battery life | up to 36 h | |
Mains adaptor / USB charging adaptor | 5 V / 1 A | |
Outputs | Interface: USB B, |
switching output / alarm modes: MD6 with
2.85 V when active
Protection class| IP 54
Operating and storage conditions| 5 … 45 °C,
35 … 65 % RH, non-condensing
Dimensions| 189 x 707 x 34 mm
Weight| 450 g
Models with external load cell
Dimensions / weight load cell| L 52 mm / H 72 mm / W 19 mm / M12 thread
/ 490 g
Cable length external load cell| approx. 1.8 m
Handheld dimensions| 189 x 707 x 34 mm
Handheld weight| 240 g
Delivery scope
For force gauges PCE-PFG 20, 50, 100, 200 and 500:
- 1 x flat head adaptor Ø13 mm
- 1 x pointed head adaptor
- 1 x hook adaptor
- 1 x extension rod (75 mm)
- 1 x USB cable
- 1 x USB charger
- 1 x user manual
- 1 x plastic carrying case
For the force gauge PCE-PFG 2K:
- 1 x 2000 N S measuring cell
- 2 x M12 joint eyes
- 2 x pressure fittings
- 1 x USB cable
- 1 x USB charger
- 1 x user manual
- 1 x plastic carrying case
Dimensions
Dimensions for models with internal measuring cell
Key functions
Key | Description |
---|---|
“Power” key for switching the meter on and off | |
“MODE” key to select between the individual modes “DEL” key to delete |
individual values
| “UNIT” key
– Setting the unit in measuring mode
– Increasing displayed parameter
– Selecting the next item in the menu
| “Enter” key to confirm the entry or to open a menu item “DIR” key for
switching between compressive and tensile alarm display
****| “ESC” key to exit the menu “Zero” key to reset the zero point
| “DATA” key:
– Select the next item in the menu
– Call up data memory
| “SET” key:
– to enter the menu
– to select a menu item
– to apply settings
Battery
The force gauge has a built-in 3.7 V rechargeable battery. If the meter is stored for a longer period of time, it should be recharged because the battery discharges itself. The battery is recharged after approx. 6 hours.
When the battery is charging, the following icon is displayed:
When the battery is fully charged, the icon appears on the display.
When switched on, the battery indicator shows the current battery capacity
when no charger is connected.
As soon as the battery is discharged, the force gauge switches itself off.
Switching on and off
To switch the force gauge on / off, press and release the key once. When the force gauge is switched on, the model, version number and serial number are displayed.
Measuring modes
There are four different measuring modes in this force gauge. If the tensile or compressive force is outside the measurement range, “OVER” is shown on the display. An acoustic signal is also generated. Only when the measured value is back within the measurement range, a normal measurement can be resumed.
To select between the modes, press the key in the current measuring mode. The current measuring mode is displayed below the measured value.
Note: Zeroing can only be performed withing the range of 10 % of the total capacity.
Real Time
In Real Time (RT) measuring mode, the current measured value is continuously
displayed.
Peak
In peak mode (PK), the highest measured value is displayed and held. This
measuring mode can be used for tensile and compressive force. The peak value
is reset with the key .
Average mode
In Average (AVG) mode, the average value of a measurement is displayed. This
mode can be used, for example, for peel force, friction force and other tests
where an average value is required. There are two different functions in this
measuring mode.
MOD1: With this function, the average value of the force curve is displayed from the set smallest force and over the set period of time.
MOD2: This function calculates the average above the set smallest measured value. When the measured value falls below the set smallest measured value again, the measurement is finished. This measuring procedure is possible over a period of 10 minutes.
As long as the measurement time of 10 minutes is not exceeded, this measurement can be resumed at any time.
To make settings in this measuring mode, press the key twice.
To select a parameter, use the arrow keys. Press the key to select a parameter. Use the arrow keys again to change the parameter properties. Press the key to apply the settings you have made.
Setting | Meaning |
---|
A.M.C.F.
Average Minimum Capture Force
| Here you set the force at which the average measurement is to start.
Neg. Time Neglectful Time| Here you enter the time span at the beginning of
the measurement that is not yet to be taken into account in the average
measurement. Available settings: 0.0…. 300.0 s.
Resolution: 0.1 s. This parameter only affects the MOD1 function.
Cap. Time Capture Time| Here you set the measuring time for the average measurement. Available settings: 0.0…. 300.0 s. Resolution:
0.1 s. This parameter only affects the MOD1 function.
Ave. Mode| Here you select between the MOD1 and MOD2 function.
To reset and save the average value, press the key. Up to 20 average values can be saved. To view, delete or export these data, press the key. Now select “Average” with the arrow keys. Confirm the selection with the key.
Now you get a choice of four options.
Selection | Description |
---|
View Ave. Data
(Read out the memory of the average measurement)
| Here you can read out the memory of the average measurement and delete
individual saved measured values.
View Ave. Stat.| An average evaluation of all average measurements is carried
out here.
Send Ave. Date| Here, the saved measurement data are transferred directly to a
PC.
Clear All Ave.| Deletes all average measurement data.
Select the desired function with the arrow keys. Press the key to open the function. Press the key to return.
When viewing all saved measured values, you can select the desired measured value with the arrow keys. Use the key to switch between the individual pages. Use the key to delete the selected value.
In the summary of the average readings, you will see the highest, the lowest, the number of readings and the overall average.
Under “Send Ave. Data”, all data are sent to the PC. After all data have been sent, “SENT FINISH” appears on the display.
With the function “Clear All Ave.”, you can clear the memory. As soon as clearing is completed, “CLEAR OVER” appears on the display.
Measuring procedure
If “WAIT” is displayed on the screen, the meter waits until the set minimum
load is applied.
If “I.DLY” is shown on the display, the force gauge will wait until the set minimum time has elapsed.
If the minimum load is present and the minimum time has elapsed, the actual measurement begins. “GATHER” appears on the display. The measurement is made. During this measurement, it is not possible to see the current measured value.
When the measurement is completed, the display shows “DONE”.
Save mode (Automatic memory)
In “SAVE Mode”, the highest measured values can be saved in a single
measurement run. Sufficient memory space for 100 measured values is available
(memory items no. 00 … 99). The number of memory items used is displayed to
the left of “SAVE”. As soon as a single measurement run is completed, the
highest measurement value is automatically saved. The minimum load for this
function is specified in the settings under P.M.C.F. To evaluate the
data,press the key. Then press to select “PEAK”.
The following options are now available:
Selection | Description |
---|
View Peak Data
(Read out the memory of the peak values)
| Here you can read out the memory of the peak values and delete individual
saved measured values.
View Peak Stat.| The highest, lowest and average of all peak values are
displayed here.
Send Peak Date| Here, the saved measurement data are transferred directly to a
PC.
Clear All Peak| Deletes all peak readings.
Use the arrow keys to make your selection here. Use the to select the desired function.
When viewing all saved measured values, you can select the desired measured value with the arrow keys. Use the key to switch between the individual pages. Use the key to delete the selected value.
The peak summary shows you the highest, the loweest, the number of readings and the total peak.
Under “Send Peak Data”, all data are sent to the PC. After all data has been sent, “SENT FINISH” appears on the display.
Use the “Clear All Peak” function to clear the memory. Once clearing is complete, the display shows “CLEAR OVER”.
Alarm limits
The alarm limits function is useful, for example, to check during quality
control whether the sample is working within the specified tolerances. Two
limits can be set here. If the measured value is lower than the set “Lower
Limit”, this is signalled by the red and green LEDs lighting up. If the
measured value is between the set “Higher Limit” and “Lower Limit”, only the
green LED lights up. If the “Higher Limit” value is also exceeded, only the
red LED lights up.
Note:
This function is only available in the measuring modes RT, PK and Save. If the
“Stop Force” function is activated, no LED or only the green LED will light
up, depending on the setting.
To set the alarm limits, press the key in the current measuring mode.
Display | Meaning |
---|---|
This setting refers to tensile force | |
This setting refers to compressive force | |
L. Limit | Lower Limit. Here you set the first limit value. This value cannot |
be higher than the H. Limit value.
H. Limit| Higher Limit. Here you set the second limit value. This value cannot
be smaller than the L. Limit value.
Number| Here you set the desired limit.
Use the arrow keys to select the desired parameter. Now press the key to make changes to this value. You can now use the arrow keys to change the value as desired. Confirm the entry with the key. Press the key to return to the measuring mode.
Note:
The second limit value must always be higher than the first set limit value.
The set values are shown above the display in measuring mode. Pressing the key
selects between the display of the set alarm tension or the set alarm
compression.
Rotate display
To rotate the display, press and hold the key in the current measuring mode.
Then release the key. The display will rotate by 180 °.
Communication interface and output interface
With this force gauge, the saved measurement data can be transferred to a PC. To do this, connect the USB cable to the meter and to the PC. To activate this function, first press the key four times. Now select the “Online” function with the arrow keys. Now press the key. With the arrow keys, you can now change the setting from „OFF“ to “ON”. Now confirm your entry with the key. The USB interface is now activated with this setting.
After installing the drivers and the software, you can establish a serial
connection to the force gauge via the software. To do so, set the parameters
of the force gauge under “Serial Port Setting”. Now use the “OPEN” button to
establish a connection. The force gauge has a memory for the peak values and a
memory for the average values.
To read out the memory for the peak values, type a “P” into the text field of
the software and then click on “Send”. The memory content is then displayed
below the software. If you want to read out the memory for the average values,
type “A” intp the text field and click on “Send”.
To remove the read data, click on “Clear”.
You can also save the data to your PC permanently. To do this, click on the
“SAVE” button. The data can be saved in “.xls” and “.txt” file format.
More settings
To make further settings to the meter, press the key three times in measuring mode. Use the arrow keys to select your desired parameter. Once you have found your desired parameter, press the key to select the function. Use the arrow keys again to make the changes. Confirm the entry with the key.
The displayed functions have the following meanings:
Function | Meaning |
---|---|
St. Force | Here you can set the limit weight at which a test stand is to stop. |
Stop Mode | Here you can activate and deactivate the control of the test stand |
via the PCE-PFG.
S. Point| This is the starting point where data recording is to begin.
F. Point| This is the end point at which data recording is to be stopped.
The key must be pressed again for the following functions.
Function | Meaning |
---|---|
P.M.C.F | Here you set the minimum load for the |
measuring mode “SAVE MODE” to save a peak value.
Shutoff| Here you can select when the force gauge switches itself off. Selectable range: 0 … 30 min. Selecting 0 minutes switches off this
function. The force gauge is then permanently switched on.
Backlight| Here you can switch the backlight on and off.
Online| This activates the data interface.
To exit the menu again, press the key.
Product information
To inform you about the serial number, product name and version number, this
is displayed each time the meter is started.
Contact
If you have any questions, suggestions or technical problems, please do not
hesitate to contact us. You will find the relevant contact information at the
end of this user manual.
Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the
European Parliament applies. Due to the contained pollutants, batteries must
not be disposed of as household waste. They must be given to collection points
designed for that purpose. In order to comply with the EU directive 2012/19/EU
we take our devices back. We either re-use them or give them to a recycling
company which disposes of the devices in line with law. For countries outside
the EU, batteries and devices should be disposed of in accordance with your
local waste regulations. If you have any questions, please contact PCE
Instruments.
PCE Instruments contact information
Germany
PCE Deutschland GmbH
Im Langel 4
D-59872 Meschede
Deutschland
Tel.: +49 (0) 2903 976 99 0
Fax: +49 (0) 2903 976 99 29 info@pce-instruments.com
www.pce-instruments.com/deutsch
References
- France.fr : Explore France and its wonders - Explore France
- iberica.es
- instruments.cn - The domain is available for purchase
- Make an offer on the domain instruments.co.uk - Domains.co.uk
- Computer Instruments | Home
- Discover Italy: Official Tourism Website - Italia.it
- PCE(北京)科技有é™å…¬å¸
- PCE Teknik Cihazlar Paz. Tic. Ltd.Şti. | PCE Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>