Winsen ZP211 Refrigerant Gas Detection Module User Manual

June 14, 2024
Winsen

Winsen ZP211 Refrigerant Gas Detection Module

Product Description

ZP211 refrigerant gas detection module adopts advanced multi-layer thick film manufacturing process semiconductor gas sensor element. The gas sensor element contains a filter adsorption layer to reduce the interference of alcohol and other gases, and has extremely high sensitivity to refrigerant gas R290. The module has been aged , calibration, calibration, with good consistency and high sensitivity.

Winsen ZP211 Refrigerant Gas Detection Module 1

Features

High sensitivity, fast response, strong anti-interference ability, long life, stable operation, factory calibrated, sensor fault self-diagnosis

Application
Used for air conditioning, refrigeration system refrigerant leak detection.

Parameters

Model ZP211
Target Gas Refrigerant gas R290
Physical interface S05B-PASK-2
Output Data PWM
Working Voltage 5.0±0.2V DC (No voltage

reverse connect protection)

Working Current| ≤80mA
Warm Up Time| ≤3 min
Recovery time| ≤120s
Set alarm

concentration

| 2000ppm
Initial alarm accuracy| 1340~2660ppm
Operating Temperature| -20~55℃
Operating Humidity| ≤95%RH
Storage Temperature| -20~60℃
Storage Humidity| ≤60%RH
Size| 25×34×14.1mm(L×W×H)
Weight| 5g

Winsen ZP211 Refrigerant Gas Detection Module 2

Terminal definition

Pin Name Function
G GND Input power –
R / /
T / /

O

|

Output

| PWM signal output (need to connect external pull-up resistor)
V| 5V| Input power +

Winsen ZP211 Refrigerant Gas Detection Module 3

Definition of output signal

Status PWM output
Preheating H:75ms
Work H:75ms
Alarm H:225ms
Fault H:300ms

Note:

  1. After the module enters the alarm state, it will keep the alarm state and will not switch to other states. It needs to be powered on again to exit the alarm state.
  2. If the sensor is exposed to high concentrations (above 10000ppm) of R290 gas, the sensor may be damaged and its performance may be affected.
  3. The module fault signal takes priority over the alarm signal.

Sensor basic testing circuit
The module uses an MP511D sensor, and Figure 3 shows the basic testing circuit of the MP511D sensor.
The sensor requires two voltages to be applied: heating voltage (VH) and testing voltage (VC).

Winsen ZP211 Refrigerant Gas Detection Module 4

Among them, VH is used to provide a specific operating temperature for the sensor, and the voltage is applied to both ends of the heating electrode using a DC power supply. VC is used to measure the circuit’s loop voltage. VRL is the voltage on the load resistor (RL) connected in series with the sensor, which is the output voltage VOUT. The module determines the concentration of refrigerant R290 in the current environment by detecting the voltage of VOUT. On the premise of meeting the electrical characteristics of the sensor, VH and VC can share a common power supply circuit.

Cautions

  • Please do not put the module in organic solvent (include silica gel and other cementing compounds), painting, medicament, oils and fuels, high concentration gas etc.
  • Please do not impact or vibrate the module seriously.
  • Please warm up for 5 min before first use.
  • Please do not use the module related with personal safety.
  • Please do not install the module in the severe convection environment.
  • Please do not put in the module in high concentration organic gas for long time.
  • Please supply the module with requested voltage as manual strictly, if voltage is higher than 5.5V, the module will be destroyed irreversibly.

Zhengzhou Winsen Electronics Technology Co., Ltd
Add: No.299, Jinsuo Road, National Hi-Tech Zone, Zhengzhou 450001 China
Tel: +86-371-67169097/67169670
Fax: +86-371-60932988
E-mail: sales@winsensor.com
Website: www.winsen-sensor.com

Winsen ZP211 Refrigerant Gas Detection Module 5

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals