AGM E122 Forehead Infrared Thermometer User Manual

June 14, 2024
AGM

AGM E122 Forehead Infrared Thermometer

AGM-E122-Forehead-Infrared-Thermometer-PRODUCT

INFRARED THERMOMETER

  • Congratulations on your purchase of our Infrared thermometer. This thermometer is intended for scanning groups of individuals or monitoring an individual for elevated temperatures.
  • Always use a clinical thermometer when high-accuracy body temperature measurements are required.

considerations and Warnings:

  1. Read this manual before using this Infrared thermometer.
  2. For the most accurate results, make reading with an ambient condition of 15’C to 40’C, RHS85%(non­condense), 70kPa-106kPa.
  3. To keep the performance, please store with an ambient condition of -20’C to 55’C, RHS93%(non-condense), 50kPa-106kPa.
  4. Make sure the skin is dry and that no hair interferes with the measurements.
  5. Do not immerse the thermometer in water.
  6. Avoid touching and/or scratching the infrared sensor lens;
  7. Always use a clinical thermometer to verify any abnormal temperature measurements.
  8. Clean the lens area by gently blowing with compressed air and using a damp swab to wipe the lens.
  9. Do not use any solvents to clean the lens.
  10. If you have any problems with this device, such as setting up, maintaining, or using it, please contact with manufacturer or agency.
  11. Don’t open or repair the device by yourself.
  12. Don’t use the device if it is damaged in any way, the continuous use of a damaged unit may cause injury, improper results, or serious danger.
  13. When not in use, store the device in a dry room and protect it against extreme moisture, heat, lint, dust, and direct sunlight. Never place any heavy objects on the storage case.
  14. The internal electrical power source can’t be changed by yourself, if you need it, please contact the manufacturer. This infrared thermometer is an adjusted mode thermometer.
  15. The patient also is an intended operator.
  16. Prohibit patient from servicing and maintaining while the IR thermometer is in use.
  17. The patient will contract the button and outer shell of IR while the IR thermometer is in use.
  18. If the thermometer has been stored in a cold or hot environment, allow it at least 15-20 minutes to acclimate to room temperature before making measurements
  19. After heavy exercise always wait at least 10 minutes before taking forehead temperature readings.
  20. The measuring site and reference body site of the clinical thermometer are the forehead and oxter respectively.
  21. Please dispose of the device/ battery/ packing in accordance with the legal obligation in your area.
  22. Object temperature mode equals direct mode, human mode equals correction mode;
  23. If the environmental condition (eg, temperature and humidity) changes suddenly, it could lead to inaccuracy in measured results.
  24. Prohibit patients or operators from sucking and swallowing the small parts left from this infrared thermometer.
  25. If you take temperature measurements during a particular diagnosis or treatment, there may be mutual interference that may cause the infrared thermometer to be inaccurate.
  26. Repeated cleaning and disinfection of the equipment will not affect the accuracy of the infrared thermometer.
  27. This device is not intended to use in oxygen enrichment conditions.
  28. Degraded sensors and electrodes, or loosened electrodes can degrade performance or cause other problems.
  29. The alkaline batteries are consumable and may be have battery leakage if used for a long time.

Product Description

Operating Principle

  • All objects above absolute zero temperature emit a certain percentage of infrared radiation energy based on their temperature. The amount of the radiation energy and the distribution of the wavelength have a very close relationship.
  • When the human forehead’s temperature is 36-37’C, ii emits wavelength 9-13um of infrared radiation. Based on this principle, according to the relationship between surface forehead temperature, we are able to measure the human forehead’s actual temperature by measuring surface forehead temperature.

Intended use

  • The infrared thermometer is an electronic thermometer using an infrared sensor to measure human body temperature for people of all ages. Its operation is based on measuring the natural thermal radiation emanating from the forehead.

Structure and Composition

  • It mainly contains a shell, temperature probe, button, display screen, and battery.

Product Interface

AGM-E122-Forehead-Infrared-Thermometer-FIG-1

  • In the Off state, press the setting button for 3 seconds, temperature symbol “‘ C” or “‘F” will flash, press again setting button to switch ‘C and oF mode again, and press the test button to confirm.

Display

AGM-E122-Forehead-Infrared-Thermometer-FIG-2

Performance Parameter

  • Note: If the device is used other than in accordance with the specifications, perfect functioning cannot be guaranteed!
  • We reserve the right to make technical changes to improve and develop the product.
  • The accuracy of this thermometer has been carefully checked and developed with regard to a long useful life.
  • Precise instructions for checking accuracy may be requested from the service address.
Product name Infrared thermometer
Product model E122
Rated output range 34.0°C-43.0°C

Rated extended output

range

| 32.0°C-33.9°C
Laboratory measurement

precision

| ±0.3°C (34.0°C-43.°C)

±0.4°C (32.0°C-33.9°C)

Display Precision| 0.1°C/°F
Measurement units| oc/oF
Measure time| 1s
Auto power off time| 15s
Ambient operation| 15-40°C, RHS85%(non-condense),

?0kPa -106kPa

Storage environment| -20-55°C, RHS93%(non-condense),

50kPa -106kPa

Dimensions| 140x40x40(mm)
weight| About 95g (including battery)
battery| AAAx2, DC 3V
Accessory| manual, AAA battery x2
Protection against

electric shock

| Internally powered equipment
Mode of operation| continuous operation
Expected service life| 6 years

Operating Instructions

  1. Before measuring the body temperature, be sure choose the correct mode.
  2. Press the test button to turn on the machine, the machine will self-test, and all symbols will be displayed normally (Figure 1).
  3. The device must keep a 1-5 cm distance from the forehead of the patient(Figure 2), and press the test button.
  4. In one second, the results will displayed{Figure 3). At this time, the data can be read according to the display result to determine whether the temperature is abnormal.
  5. If the machine is not operated within ten minutes after the measurement, it will automatically shut down.AGM-E122-Forehead-Infrared-Thermometer-FIG-3

Caution:

  1. To keep the sensor lens clean, do not block the front with anything.
  2. In order to maintain accurate measurement, please wait for  15- 20 minutes to measure after you enter a new environment with a large ambient temperature change.
  3. When measuring a fairly high or relatively low object temperature, wait for ten minutes before measuring body temperature.
  4. The sweat and stains on the forehead will affect the measurement accuracy. Do not measure the body temperature in 30 minutes after a meal, shower, or exercise.

Unit switch

  • This product is available in two temperature measurement units, Celsius and Fahrenheit. In the off state, long press the setting button for about 3 seconds, and the temperature symbol ‘C or ‘F flashes, continue to press the setting button, and you can switch ‘C and ‘F modes, and press the test button to confirm.

Battery replacement

  • When the battery symbol shows on the screen, please replac the battery. Otherwise, the measurement result will be affected Open the battery compartment, take out the exhausted battery put in 2 AAA batteries, please pay attention to the polarity of the battery, do not reverse.AGM-E122-Forehead-Infrared-Thermometer-FIG-4

Caution:

  1. Remove the battery when not in use for a long time.
  2. Please do not throw the used battery into the fire.
  3. Please dispose of used batteries in accordance with local laws and regulations.
Display Message

AGM-E122-Forehead-Infrared-Thermometer-FIG-15

Cleaning and Maintenance

Cleaning Method

  • Check the probe if damaged or dirty before cleaning the device. Use a clean cotton swab be moisten with alcohol or warm water.
  • To clean the entire device please use a soft cloth to clean the display screen, do not use disinfectants, and do not soak in water or other liquids.
Transport and storage
  1. The device transport and storage should be under temperature control: Humidity -20°c-ss·c, relative humidityS93%, no condensation environment.
  2. Must do package transport simulation test after device package, will allow to use ordinary transport but must avoid rain, moisture, extrusion, and collision.
  3. The device must not be stored or used at an excessively high or low temperature or humidity (see technical data), in sunlight, in association with an electrical current, or in dusty locations. Otherwise, inaccuracies can occur.

Clinical Information

  • Infrared thermometers have been tested with effective clinical accuracy and reproducibility, enabling temperature measurements for adults, children, and newborns.
  • The clinical bias, limits of agreement, and clinical repeatability of the infrared thermometer are shown in the table below.AGM-E122-Forehead-Infrared-Thermometer-FIG-5
  • The reference body site of the infrared thermometer is oxter.

Suggestion

  • a. When you tell the doctor about the measured temperature, please indicate that you are measuring with an infrared thermometer.
  • b. Do not force fall or step on the thermometer.
  • c. Do not disassemble or repair the thermometer.
  • d. The product is not waterproof, please be careful not to let liquid (alcohol, water droplets, hot water, etc.) enter inside.
  • e. The product must be kept clean and dry.
  • If you find any problems, you should contact the seller and you cannot repair the product by yourself.
  • g. Do not use in an electromagnetic interference environment.
  • h. Dispose of waste and residue at the end of its useful life in accordance with local laws and regulations.

Normal body temperature range for different measurement body partAGM-E122
-Forehead-Infrared-Thermometer-FIG-6

People’s body temperature varies with different times of the day and is also affected by other external conditions, such as age, gender, and skin color.
Normal body temperature range for different ageAGM-E122-Forehead-
Infrared-Thermometer-FIG-7

Symbols information

AGM-E122-Forehead-Infrared-Thermometer-FIG-8

EMC statement

  • This device has been tested and homologated in accordance with EN 60601-1-2. This does not guarantee in any way that the device will not be affected by electromagnetic interference. Avoid using the device in a high electromagnetic environment.
  • Recommended separation distances between portable and mobile RF communications equipment and the E122 Infrared Thermometer.
  • The E122 Infrared Thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the E122 Infrared Thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the E122 Infrared Thermometer as recommended below, according to the maximum output power of the communications equipment.AGM-E122-Forehead-Infrared-Thermometer-FIG-9
  • For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where Pis is the maximum output power rating of the transmitter in watts (Y’,/) according to the transmitter manufacturer.
  • NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
  • NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Guidance and manufacturer’s declaration-electromagnetic emission

  • The E122 Infrared Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the device should ensure that it is used in such an environment.AGM-E122-Forehead-Infrared-Thermometer-FIG-10

Guidance and manufacture’s declaration-electromagnetic immunity

  • The E122 Infrared Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the device should ensure that it is used in such an environment.AGM-E122-Forehead-Infrared-Thermometer-FIG-11 AGM-E122-Forehead-Infrared-Thermometer-FIG-12

Guidance and manufacturer’s declaration-electromagnetic immunity

  • The E122 Infrared Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the device should ensure that it is used in such an environmentAGM-E122-Forehead-Infrared-Thermometer-FIG-13 AGM-E122-Forehead-Infrared-Thermometer-FIG-14

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals