RICE LAKE OIML Class III Healthweigh Indicators Instructions

June 13, 2024
RICE LAKE

RICE LAKE OIML Class III Healthweigh Indicators Instructions
Healthweigh Indicators

Disabling Calibration Message

Product: Healthweigh Indicator OIML Class III
Issue: Unwanted calibration prompt
Solution: Set next verification date (MONT) to 0

Manuals are available from Rice Lake Weighing Systems at www.ricelake.com/manuals
Warranty information is available at www.ricelake.com/warranties

Figure 1. Healthweigh Indicator Display
Healthweigh Indicator Display
A CNTL message may appear on an OIML Class III indicator to alert the user that it is time for a calibration. Use the following procedure to disable the CNTL calibration message

  1. Power on the indicator.
  2. Immediately hold the ENTER button for three seconds until firmware ID number flashes.
  3. Press the BMI button. Verification menu displays.
  4. Press ENTER to access Verification menu.
  5. Press ENTER to access MONT parameter.
  6. Use the BMI and arrow buttons to set the MONT parameter to 0.
  7. Press Enter two times to save.
  8. Press Enter to select DONE. Scale will reboot after three seconds.

NOTE: The CNTL reminder is disabled after the procedure is complete. Menu may still continue to display 11 in the MONT parameter after successful procedure.

© Rice Lake Weighing Systems Content subject to change without notice
230 W. Coleman St. • Rice Lake, WI 54868 • USA USA: 800-472-6703International: +1-715-234-9171

www.ricelake.com
RICE LAKE Logo

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

RICE LAKE User Manuals

Related Manuals