Cardinal Scale Detecto 6495 Platform Wheelchair Scale USER MANUAL
- June 4, 2024
- Cardinal Scale
Table of Contents
**
Cardinal Scale-Detecto 6495 Platform Wheelchair Scale**
SPECIFICATIONS
- CAPACITY: 400 lb. x 0.2 lb. / 180 kg x .1 kg.
- PLATFORM: 30 x 26 / 762mm x 660 mm with nonskid removable mat.
- DISPLAY: .7 high–contrast LCD.
- POWER SOURCE: 6 C size alkaline batteries or optional AC adapter.
- PACKAGE QUANTITY: 1
INTRODUCTION
With the model 6495 digital wheelchair scale, you may weigh pounds or kilograms with the push of a button quickly and with increased accuracy which only electronic weighing can offer. The weight display is fixed to a tilt mount, allowing it to be adjusted so that it is easily viewed. Additionally, a keypad is available for entering known empty chair weights, providing an exact and “constant” estimate of the patient’s weight. 13 mm / 5-inch digits. 400 lb x.2 lb or 180 kg x.1 kilogram of capacity. The platform is 30″ by 26″ (762mm by 660mm) and has a removable nonskid mat. .7″ high-contrast LCD display. 6 “C” size alkaline batteries or an optional AC converter are the power sources.
**SCALE ASSEMBLY INSTRUCTIONS
**
SCALE ASSEMBLY INSTRUCTIONS, CONT.
Refer to Figures No. 1, 2, 3, and 4, and then follow the instructions below for scale assembly.
- Remove the scale from its packing and inspect it for signs of shipping damage. Keep and use the original carton and packing material for return shipment if it should become necessary. Note that it is the responsibility of the purchaser to file all claims for any damage or losses incurred during transit.
- Set the scale base on a flat, level surface. Slide ramp pins in T-Slot on the base, and drop the pins into the groove.
- Remove the four (4) 1/4-20 screws from the column bracket (A1) and set aside.
- Set column (B) onto the column bracket (A1). See Figure No.1. Be sure the back (wide side) of the column is facing the scale deck. Secure the column with the four (4) 1/4-20 screws removed in step 3. Do not tighten the screws at this time.
- Hook end of shock absorber and draft rod assembly to one end of the load cell assembly (C). See Figure No. 3. NOTE: Handle load cell assembly with care! Do not drop, bend, or twist.
- Slide the grommet away from the load cell assembly and slide into the slot on the column. Note: Slot should be on the back of the column (wide side).
- Lower the load cell assembly, shock absorber, and draft rod assembly into the column. The hanger must pass through the opening in the column bracket. Note: Position of hanger (B1).
- Fasten the top hook bolt of the load cell assembly (C) to the column (B). (Note: Load cell assembly should not touch sides of the column.)
- If available, place the scale on a table or workbench, otherwise place the scale on the floor. In either case, DO NOT tilt or place the scale on its side. Next, push the bottom of the hanger (B1) back towards scale weighbridge (A), lift the long lever and nose iron (A2) up, then slide the hanger (B1) under the nose iron pivot (A3). See Figure No. 2. After the hanger is attached, tighten the four (4) 1/4-20 screws. See Figure No. 1 for column (B) mounting.
- Pull excess load cell cable in the column leaving enough slack in the cable so as not to have any side pull on the load cell.
- Attach digital indicator and mounting bracket to the column with two (2) #10-32 button head screws and lock washers. See Figure No. 4.
- To attach handrail (D), remove two (2) 3/8-16 screws from the top of the weighbridge side. Set handrail on weighbridge side, align holes, reinstall the two (2) 3/8-16 screws and tighten.
- Check the scale platform for ease of movement. If the platform does not move freely, check all mechanical connections.
- Refer to the included indicator manual, for indicator setup and operating instructions.
- The scale is now ready for operation.
FREQUENTLY ASKED QUESTIONS
Are Cardinal Detecto scales accurate?
Digital precision balance scales from DETECTO are designed for extremely accurate weighing applications and have an accuracy of 10 milligrams.
How do you calibrate a Cardinal Detecto scale?
Press and hold the 0 key while the software version is displayed on the display. When the calibration enunciator is turned on, the display will change to indicate CAP in the larger lower seven segment display, the current scale capacity setting in the smaller upper seven segment display, and the current scale capacity setting. You are now prepared to calibrate.
What is a Detecto scale used for?
In order to comfortably measure, weigh, and determine Body Mass Index in hospitals, clinics, and emergency care facilities, DETECTO’s digital physician scales have low, flat platforms, models with or without columns, and mechanical or digital stadiometers.
How accurate are wheelchair scales?
Patients can be correctly weighed to 0.2 lb. while in their wheelchairs, which have a maximum weight capacity of 800 pounds. There are hundreds of weight values for different wheelchair manufacturers and models stored in the scale’s huge internal tare memory.
How do you weigh someone in a wheelchair at home?
Press the Tare button after pushing the empty wheelchair onto the scale. A minus value will then appear on the indicator when you remove the wheelchair from the scale. Roll the patient and wheelchair onto the scale at this point, and only the patient’s weight will be shown.
What is a wheelchair scale used for?
This wheelchair/platform scale, which is large and powered by an A/C UL adaptor, has several uses, including weighing people in wheelchairs or patients who are particularly heavy. It is particularly helpful for patients who are seated, such as during dialysis.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>