Medtronic MiniMed 670G Insulin Pump System User Manual

June 13, 2024
Medtronic

Medtronic MiniMed 670G Insulin Pump System

Medtronic-MiniMed-670G-Insulin-Pump-System-PRODUCT

Product Information

The MiniMedTM 670G Pump and MiniMedTM 780G Pump are insulin pumps used for managing insulin delivery in individuals with diabetes. The settings mentioned in the user manual may vary from the actual settings on your device.

Transfer Settings
To transfer settings from the MiniMed 670G pump to the MiniMed 780G pump, it is recommended to record all settings on the provided form. This form can serve as a reference when entering the settings into the new pump.

Delivery Settings

  • Preset Bolus Setup: Menu > Options > Delivery Settings > Preset Bolus Setup. This allows you to set the preset bolus name, bolus amount, U type, bolus timing, and square duration.
  • Bolus Increment: Menu > Options > Delivery Settings > Bolus Increment. You can select the bolus increment value from 0.025 U, 0.05 U, or 0.1 U.
  • Max Basal/Bolus: Menu > Options > Delivery Settings > Max Basal/Bolus. This setting allows you to set the maximum basal rate and maximum bolus amount.
  • Easy Bolus: Menu > Options > Delivery Settings > Easy Bolus. You can enable or disable the easy bolus feature and set the step size for easy blousing.
  • Auto Suspend: Menu > Options > Delivery Settings > Auto Suspend. This setting enables or disables the auto-suspend feature and allows you to set the auto-suspend alarm and time.
  • Bolus Speed: Menu > Options > Delivery Settings > Bolus Speed. You can select either standard or quick bolus speed.

Utilities

  • Turn Sensor On: Menu > Options > Utilities > Sensor Settings. This setting allows you to turn the sensor on or off.
  • Display Options: Menu > Options > Utilities. You can adjust the brightness and auto backlight duration.

SmartGuard

  • Auto Mode: Menu > Options > SmartGuard. This setting enables or disables the SmartGuard auto mode and allows you to set the auto mode BG alert.
  • High Setup: Menu > Options > SmartGuard > High Setup. This setting allows you to configure high-limit alerts and snooze duration.
  • Low Setup: Menu > Options > SmartGuard > Low Setup. This setting allows you to configure low-limit alerts and snooze duration.

Product Usage Instructions

Programming the MiniMed 780G insulin pump Startup Wizard

  1. Select Menu > Delivery Settings.
  2. Select Max Basal/Bolus and press OK.
  3. Select Max Basal and confirm by selecting Continue.
  4. Enter the desired Max Basal amount in U/hr using the up and down buttons.
  5. Select Save to save the Max Basal setting.

Setting Max Bolus

  1. Select Menu > Delivery Settings.
  2. Select Max Basal/Bolus and press OK.
  3. Select Max Bolus and confirm by selecting Continue.
  4. Enter the desired Max Bolus amount in units (U) using the up and down buttons.
  5. Select Save to save the Max Bolus setting.

Transfer Settings

Transfer Settings: MiniMed 670G pump to MiniMed 780G pump

  • STEP 1 – Copy your current settings onto the insulin pump settings form. If for any reason you cannot copy the settings, please contact your healthcare professional.
  • STEP 2 – Insert an AA battery into your new pump. You can use either a lithium, alkaline, or rechargeable battery.
  • STEP 3 – Follow the instructions found in this guide to enter the settings that you have recorded on the insulin pump settings form into your new pump. For additional information on your pump, please consult the MiniMed pump system user guide.

Note: The settings shown in this guide are for illustration purposes only — your settings may be different.

Settings from the MiniMed 670G insulin pump
We recommend that you record all settings on this form and use this form as a reference when you enter your settings into the new pump

Delivery Settings

Delivery Settings, Utilities, SmartGuard

DELIVERY SETTINGS

  • BOLUS WIZARD SETUP:
    • Menu > Bolus > Delivery Settings > Bolus Estimate Setup
    • Bolus Wizard: On Off
  • Carb Ratio:
    • 12:00 A ____ ____ g/U
    • ____ ____ ____ g/U
    • ____ ____ ____ g/U
  • Insulin Sensitivity Factor:
    • 12:00 A ____ ____ mg/dL/U
    • ____ ____ ____ mg/dL/U
    • ____ ____ ____ mg/dL/U
  • BG Target:
    • 12:00 A ____ ____-____ mg/dL
    • ____ ____ ____-____ mg/dL
    • ____ ____ ____-____ mg/dL
  • Active Insulin Time:
    • ____ hr
  • BASAL PATTERN SETUP:
    • Menu > Basal > Delivery Settings > Basal Pattern Setup
  • Basal 1:
    • 12:00 A ____ ____ U/hr
    • ____ ____ ____ U/hr
    • ____ ____ ____ U/hr
  • Basal Pattern Name: ____
    • 12:00 A ____ ____ U/hr
    • ____ ____ ____ U/hr
    • ____ ____ ____ U/hr
  • PRESET TEMP SETUP:
    • Menu > Options > Delivery Settings > Preset Temp Setup
  • Preset Temp Name: ____
    • Type: Rate __ U/hr Percent __ %
    • Duration ____ hr
  • DUAL/SQUARE WAVE:
    • Menu > Options > Delivery Settings > Dual/Square Wave
    • Dual: On Off
    • Square: On Off
  • PRESET BOLUS SETUP:
    • Menu > Options > Delivery Settings > Preset Bolus Setup
    • Preset Bolus Name: ____
    • Bolus: __ U Type: __
    • __ Now __ Square
    • Duration ____ hr
  • BOLUS INCREMENT:
    • Menu > Options > Delivery Settings > Bolus Increment 0.025 U 0.05 U 0.1 U
  • MAX BASAL/BOLUS:
    • Menu > Options > Delivery Settings > Max Basal/Bolus
    • Max Basal: __ U/hr Max Bolus: __ U
    • EASY BOLUS:
    • Menu > Options > Delivery Settings > Easy Bolus
    • On Off Step Size ____ U
  • AUTO SUSPEND:
    • Menu > Options > Delivery Settings > Auto Suspend
    • Alarm: On Off
    • Time: ____ hr
  • BOLUS SPEED:
    • Menu > Options > Delivery Settings > Bolus Speed
  • Standard Quic utilities **TURN SENSOR ON:**
    • Menu > Options > Utilities > Sensor Settings
    • Sensor: On Off
  • DISPLAY OPTIONS:
    • Menu > Options > Utilities
    • Brightness ____ Auto
    • Backlight ____ sec/min
  • SMARTGUARD AUTO MODE:
    • Menu > Options > SmartGuard
    • Auto Mode: On Off
    • Auto Mode BG alert: On Off

Sensor Settings

HIGH SETTINGS Menu > Options > SmartGuard > High Setup

High Limit

| Alert before high| Time before high| Alert on high| Rise Alert| Rise Limita
12:00A – mg/dL| On| min| On| On| Custom mg/dL/min or | ↑↑ | ↑↑↑
mg/dL| On| min| On| On| Custom mg/dL/min or | ↑↑ | ↑↑↑
mg/dL| On| min| On| On| Custom mg/dL/min or | ↑↑ | ↑↑↑
SNOOZE _Menu > Options > SmartGuard > Snooze > High Snooze: ___
hr

  • a. Option only available when the Rise Alert feature is set to on

LOW SETTINGS Menu > Options > SmartGuard > Low Setup

Low Limit| Alert before low| Alert on low| Suspend before low| Suspend on low| Resume basal alerta
12:00A – mg/dL| On| On| On| On| On
mg/dL| On| On| On| On| On
mg/dL| On| On| On| On| On
SNOOZE _Menu > Options > SmartGuard > Snooze > Low Snooze: ___
hr

  • a. Option only available when the Suspend on a low feature or the Suspend before low feature is set to on

Startup Wizard
Note: The settings shown in this guide are for illustration purposes only — your settings may be different.
Once you’ve inserted the battery, the pump powers up and the Startup Wizard begins; it will guide you through setting language, time format, and current time and date.

  1. Select English.
  2. When the Select Time Format screen appears, select a 12-hour or a 24-hour time format.
  3. When the Enter Time screen appears, the hour will be flashing. Press UP or DOWN to the correct hour and press .
  4. The minutes will be flashing. Press  UP or  DOWN to the correct minutes and press .
  5. The AM / PM will be flashing. Press  UP or DOWN if needed press . Select Next.
  6. When the Enter Date screen appears, select Year. Press UP or  DOWN to the correct year and press. Repeat this step to set up Month and Day. Select Next.
  7. The Rewinding screen will briefly appear, followed by a confirmation message. Select OK.Medtronic-MiniMed-670G-Insulin-Pump-System-FIG-1 \(2\)

Setting Max Basal
Note: The settings shown in this guide are for illustration purposes only — your settings may be different.

Menu > > Delivery Settings

  1. Select Max Basal/Bolus and press .
  2. Select Max Basal. A confirmation message appears. Select Continue.
  3. Select Max Basal to enter the U/hr amount. Use UP and FOEN to set the Max Basal amount. Press .
  4. Select Save.Medtronic-MiniMed-670G-Insulin-Pump-System-FIG-1 \(3\)

Setting Max Bolus
Note: The settings shown in this guide are for illustration purposes only — your settings may be different.
Menu > > Delivery Settings

  1. Select Max Basal/Bolus and press .
  2. Select Max Bolus. A confirmation message appears. Select Continue.
  3. Select Max Bolus to enter the units (U) amount. Use UP and DOWN to set the Max Bolus amount. Press .
  4. Select Save.Medtronic-MiniMed-670G-Insulin-Pump-System-FIG-1 \(4\)

Setting a Basal Pattern
WARNING: The pump is intended to be used with a basal pattern. The basal pattern must be manually entered and saved into the pump. The pump will operate with a basal rate of 0.0 U/hr until a basal pattern is entered and saved. There is no reminder message to program basal rates. Consult a healthcare professional to determine what basal pattern is needed. For more information about basal patterns, consult the MiniMed 780G System User Guide.
Note: The settings shown in this guide are for illustration purposes only — your settings may be different.

Menu > > Delivery Settings > Basal Pattern Setup

  1. Select Basal 1.
  2. Select Options, then select Edit.
  3. Press to enter End time. Press to leave the end time at 12:00A. If you are setting multiple rates UP use DOWN or to change the end time and press .
  4. Press  UP to enter U/hr and press .
  5. Continue adding end times and basal rates as needed. The end time for your last rate must be 12:00.
  6. The Review option appears only when the last end time in your basal pattern is set to 12:00A. Select Review after all rates are entered to go to the next screen.
    • Note: Your basal rates have not yet been saved.
  7. A screen appears that lets you review your basal pattern. Press  DOWN to review all the settings. If you need to make any changes, press ****to return to the edit screen.
    • Note: If you press ****to return to the edit screen without saving, your changes will not be saved. After making any changes on the edit screen, select Review to return to the review screen.
  8. Select Save. If you do not select Save, your changes are not saved.Medtronic-MiniMed-670G-Insulin-Pump-System-FIG-1 \(5\)

CAUTION: If you have not selected Save after settings are entered and the screen goes dark, the entered settings will not be saved.
Turn On and Set Up the Bolus Wizard Feature
Note: The settings shown in this guide are for illustration purposes only — your settings may be different.

Menu > > Delivery Settings > Bolus Wizard Setup

  1. Select Bolus Wizard to turn the feature on.
  2. Press DOWN to see the rest of the message then select Next.
  3. Review the description of the carb ratio and select Next.
  4. Press to enter an End time. Press again to leave the end time at 12:00A. If you are setting multiple carb ratios use  UP or DOWN to change the end time, then press .
  5. Press UP or  DOWN to enter g/U then press. Continue adding end times and carb ratios as needed.
  6. Select Next when done.
  7. Review the description of the insulin sensitivity factor and select Next.
  8. Press to enter an End time. Press again to leave the end time at 12:00A. If you are setting multiple insulin sensitivity factors, use UP or DOWN to change the end time, then press .
  9. Press UP or DOWN enter mg/dL per U, then press . Continue adding end times and insulin sensitivity factors as needed.
  10. Select Next when done.
  11. Review the description of the BG target and select Next.
  12. Press to enter an End time. Press again to leave the end time at 12:00A. If you are setting multiple BG targets, use UP or DOWN to change the end time, then press .
  13. Press UP or DOWN to enter the low limit (Lo) and then press .
  14. Press UP or DOWN to enter the high limit (Hi) and then press . Continue adding end times and BG targets as needed.
  15. Select Next when done.
  16. Review the description of active insulin time and select Next.
  17. Press to enter the duration for the active insulin time.
  18. Press UP or DOWN to enter the hours of duration then press .
  19. Select Save.

The Bolus Wizard feature setup is now complete.Medtronic-MiniMed-670G-
Insulin-Pump-System-FIG-1 \(6\)

Turn Sensor On
Note: The settings shown in this guide are for illustration purposes only — your settings may be different.
Menu > > Device Settings

  •  Select Sensor to turn the feature on. You can now access the SmartGuard menu and enter the SmartGuard settings.Medtronic-MiniMed-670G-Insulin-Pump-System-FIG-1 \(7\)

Pair the Transmitter
Note: The settings shown in this guide are for illustration purposes only — your settings may be different.
Menu >
1. Attach the Guardian™ 4 transmitter to the charger and place the transmitter (attached to the charger) next to the pump.
2. Select Pair New Device on the pump and immediately remove the transmitter from the charger.
3. The Select Device screen appears with a list of available devices. Select the CGM device that matches the serial number indicated on the back of the transmitter.
Your pump displays a message if the pump and transmitter are paired successfully. When the transmitter is communicating with the pump, the Sensor feature is turned on and appears on the Home screen.

Programming High SG Settings
Note: The settings shown in this guide are for illustration purposes only — your settings may be different.
Menu > > Alert Settings

  1. Select High Alert. The High Setup screen appears.
  2. Press on the time segment. Press again to leave the End time at 12:00A. If setting multiple time segments, press UP to enter the end time of the first segment, and press .
  3. Press UP or DOWN to set the high limit (Hi) and press . Press to continue.
  4. Select each setting you wish to turn on. If a setting is on, select it again to turn it off.
  5. Once settings are selected, select Next. If needed, continue adding time segments and settings to complete the 24-hour period. When done, select Review.
  6. Verify that settings are correct and select Save.
  7. If the snooze time needs to be changed, press DOWN Snooze High & Low and press
    • Select High Snooze. Press UP or DOWN to set up the desired duration and press
    • Select Save.

Your High Setup is now complete.Medtronic-MiniMed-670G-Insulin-Pump-System-
FIG-1 \(9\)

Programming Low SG Settings
Note: The settings shown in this guide are for illustration purposes only — your settings may be different.

Menu > > Alert Settings

  1. Select Low Alert. The Low Setup screen appears.
  2. Press on the time segment. Press again to leave the End time at 12:00A. If setting multiple time segments, press UP to enter the end time of the first segment, and press .
  3. Press UP or DOWN set the low limit (Lo) and press . Press to continue.
  4. Select each setting you wish to turn on. If a setting is on, select it again to turn it off.
  5. Once settings are selected, select Next. If needed, continue adding time segments and settings to complete the 24-hour period. When done, select Review.
  6. Verify that settings are correct and select Save.
  7. If snooze time needs to be changed, press DOWN Snooze High & Low and press
    • Select Low Snooze. Press UP or DOWN to set up the desired duration and press
    • Select Save.

Your Low Setup is now complete.Medtronic-MiniMed-670G-Insulin-Pump-System-
FIG-1 \(10\)

Turn SmartGuard On
Note: The settings shown in this guide are for illustration purposes only — your settings may be different.
Menu >

  1. Select SmartGuard to turn the SmartGuard feature on.
  2. Review the message on the screen and select OK.
  3. Review the SmartGuard Checklist to see if any additional requirements must be met before the SmartGuard feature can be activated. Press DOWN to view the checklist.Medtronic-MiniMed-670G-Insulin-Pump-System-FIG-1 \(11\)

Note: The SmartGuard warm-up period begins at the first midnight after the pump starts delivering insulin. If you turn SmartGuard on before the warm up is complete,
you may receive an alert at midnight to enter a BG for SmartGuard. You may want to wait until day 3 of your pump use to avoid this alert at midnight.

Programming SmartGuard Settings

Note: The settings shown in this guide are for illustration purposes only — your settings may be different.
Note: Make sure you consult with your healthcare provider on your personalized settings before programming the SmartGuard settings.

Menu > > SmartGuard Settings

  1. Review the Target.
  2. Press to change the target.
  3. Press UP or DOWN to the desired target and then press.
  4. Check to make sure that Auto-Correction is set to On. Select Save.Medtronic-MiniMed-670G-Insulin-Pump-System-FIG-1 \(12\)

Note: Please take some time to review ALL your settings and ensure they have been correctly saved.

Technical Support

  • In Need of Technical Assistance?
  • 24-Hour Technical Support is here to help
  • Call 1.800.646.4633
  • For medical advice, contact your healthcare team.

CONTACT

  • Medtronic MiniMed
  • 18000 Devonshire Street
  • Northridge, CA 91325
  • USA
  • 1 800 646 4633
  • +1 818 576 5555
  • www.medtronicdiabetes.com.
  • RF: M994838A001
  • © 2023 Medtronic M037934C001_1 2023-05-10