tynetec FM0825 Pullcord Alarm Instruction Manual

June 13, 2024
tynetec

tynetec FM0825 Pullcord Alarm Instruction Manual

Product Instruction

  • Manually activated fixed trigger device
  • Compatible with Reach IP, Advent XT2, Career Response and Touch safe Pro V4.01 onward
  • Complies with BS EN 50134-2:2017
  • Ceiling mounted momentary action switch
  • 2.7 metre long orange cord
  • 2 triangular orange grips
  • Anti-ligature safety break
  • 5 year Lithium battery (typical)
  • Low battery warning
  • Operating Temperature: +5oC to +40oC
  • Telecare Transceiver: 869.2125MHz Class 1.5
  • Digital Heartbeat
  • Weight: 85 grams
  • Dimensions: 87mm diameter x 35mm depth
  • Product Code: ZXT820

Positioning
Pullcords should be strategically placed around the home to provide an easy means for the user to activate an emergency alarm call. They will normally be fitted in rooms where personal pendants may not be worn, e.g. in the bathroom or alongside the bed. Pullcords should be ceiling mounted – they are not suitable for wall mounting.
Being wireless, they can be used to speed-up the installation in a home where it would otherwise be difficult
or time consuming to hard-wire a standard pullcord.

Emergency Alarm Call
An emergency alarm call can be made by simply pulling the cord once – no further action is required.

Low Battery Condition
A low battery condition will be reported automatically tothe Alarm Receiving Centre (ARC) or the local Manager. The complete pullcord should be replaced as soon as possible within 30 days. Pullcords can be returned to Tynetec for battery replacement.

Out of Range Notification (Reach IP)
The Reach IP can monitor all radio peripheral devices to determine whether they have been damaged or removed from an installation. Should the Reach IP be unable to detect a device’s heartbeat it will automatically generate a “radio out of range” notification. Firstly, investigate if there’s a genuine reason for the alert, if not, arrange for the peripheral device to be replaced immediately.
The “Radio out of range timer” can be set from 30 hours to 99999 hours in Pulse CMP. “Radio supervision” can be disabled by unchecking the tick box associated with each peripheral device in Pulse CMP.

Maintenance
Test the pullcord once every month.
Clean with a damp cloth, do not use solvents or cleaners on the pullcord. Replace the cord if worn or damaged.

Disposal
Waste electrical products should not be disposedof with normal household waste. The pullcord alarm is ideally suited for disposal within the waste electronic and electrical equipment (WEEE) recycling scheme. Please recycle where facilities exist. Check with your local authority or contact your supplier for recycling/disposal advice.

Warning
The pullcord contains a Lithium battery. If tampered with the battery may be damaged and create a fire hazard. No attempt should be made to remove or replace the battery. Do not incinerate.

How to install the Pullcord
The pullcord is designed for internal ceiling mounting only, it is not suitable for wall mounting or external applications. Remove the cover by twisting it off the base. Select the best location and use the fixing holes in the base as a guide when marking the position of the drill holes. Securely fix the base to the ceiling with two 1½“ screws (not supplied). Where possible fix to a wooden joist, if not cavity fixings will be required.

Refit the cover then knot the cord so one orange triangle is within normal reach and the other just above floor level. Cut-off any excess cord.

Anti-ligature Safety Break
The pullcord includes an anti-ligature safety break in the orange cord. Should excessive force be applied the cord will part from the main body – it can be simply clipped back into place.

  1. Slide the cover down the cord to expose the buckle.
  2. Hold the buckle horizontal and pass the link rod through the hole in the buckle.
  3. Turn the buckle vertical and pull down into place over the link rod.
  4. Slide the cover back up the cord and into place over the buckle.

Telecare Radio ID
The pullcord includes a radio transceiver to communicate with the Reach IP, Advent XT2 or Carer Response.
ZXT820
S:081923638
Iss-1
039308512A
19W8

Registering the Pullcord onto a Reach IP Login to Pulse CMP (www.tynetecpulse.com) and enter the Reach IP Serial No. in the ALARM UNIT view. Press the Accessories button and enter the Radio ID code, Location and Equipment type. Press Save then  Synchronize.
IMPORTANT: when the pullcord is installed a test call must be made to check operation and start the digital heartbeat. Alternatively, with the Reach IP in normal operating mode undo the fixing screw and remove the stand or connector cover to access the ⚫ Power button.

Press the ⚫ Power button to enter Control Mode, then press and hold the button until the unit announces “add radio device” then release.
Press the  button to confirm then activate the alarm by pulling the orange cord.
The Reach IP unit will sound a high beep if it’s a new device and announce “range test mode”.
Press to confirm, or press to step to the next function, or press to exit the mode.

Pull the cord to check operation.

Registering the Pullcord onto an Advent XT2
Press the Settings button on the Home page of the Manager’s Panel. Press the Manager (or Engineer) button, enter the password and press the Programming button.
Select the Program Radio Devices option from the tabs along the bottom of the screen.

Press Flat ID and enter the flat number the pullcord is being installed in. Enter 4-digits e.g. 0014.
Press Radio ID and enter the 10-digit code taken from the label inside the pullcord cover.
Press Room Locations and select where the pullcord is located from the on- screen options e.g. Bedroom.
Press the Send button to update the XT2 system.
Pull the cord to check operation.

Registering the Pullcord onto a Carer Response

Refer to the Carer Response
user manual for full details
of the button functions and
how to navigate the menus.
From the Main menu select
Programming.
From the Programming menu select
Add transmitter.
The display will prompt you to
Activate the transmitter…
Pull the alarm cord to activate.
The Carer Response will beep and the display will change to the
Add transmitter menu with a box showing No ID.
Press  then  to select Room, then press 
Press  to select the room number box 0000.
Press  to select the first digit, press  to move between the
digits and press  or  to change each digit.
Press  when the correct room number is displayed.
Press  to select the alarm type box None.
Press  then to select Trigger then press 
Press the  button to select Save then press 
The display will show Radio ID saved to memory…Return to the Main menu and select Quit.
Pull the alarm cord to check operation.

Registering the Pullcord onto a Touch safe Pro
Login to the Master Touch safe Pro Nursecall Panel (must be V4.01 onward) using an Engineer or Manager password. From the Main Menu page select the Telecare option to open the screen shown below.

Touch the Add Telecare Device box and enter the Telecare device 10-digit Radio ID using the on-screen keypad.
Touch the Filter box and enter the zone you want to add the Telecare device to using the on-screen keypad.
Touch the Room box  and select the Callpoint ID from the list that will be used to cancel the Telecare alarm.
Touch the Save button on the keypad.
Touch the Close button to exit.
Pull the alarm cord to check operation.

Declaration of Conformity

We declare this product complies with the relevant UK legislation on condition that it is used in the manner intended, and in accordance with the current installation standards and/or the manufacturer’s recommendations. A copy of the full UKCA Declaration of Conformity is available on request.

We declare this product complies with the relevant European legislation on condition that it is used in the manner intended, and in accordance with the current installation standards and/or the manufacturer’s recommendations. A copy of the full CE Declaration of Conformity is available on request.

Chris Dodd
CEO
Legrand Assisted Living & Healthcare

Tynetec, a business unit of Legrand Electric Ltd
Cowley Road, Blyth Riverside Business Park, Blyth, Northumberland, NE24 5TF.
Tel: +44 (0) 1670 352371 www.tynetec.co.uk