tynetec FM0847 Enuresis Sensor User Guide

June 13, 2024
tynetec

tynetec FM0847 Enuresis Sensor - Icon 1 869MHz
Enuresis Sensor Trusted Technology
Caring for People

FM0847 Enuresis Sensor

tynetec FM0847 Enuresis Sensor

  • Detects instances of incontinence in bed
  • Compatible with Reach IP, Advent XT2, Carer ResponseTouchsafe Pro V4.01 onward and NOVO
  • Battery powered radio interface unit
  • Replaceable battery: 9V Duracell MN1604
  • Low battery warning
  • Operating Temperature: +5 o o C to +40 C
  • Telecare Transceiver: 869.2125MHz Class 1.5
  • Digital Heartbeat
  • Weight: 170 grams
  • Dimensions: 92 x 66 x 28mm (HxWxD)
  • 1.2m x 0.9m cotton bed sensor sheet
  • Weight 325 grams
  • 2m connecting lead with press stud fixings
  • Weight: 35 grams
  • Product Code: ZXT847 (or Neat NE1020006-01)

The absorbent cotton sheet with a breathable waterproof back is the perfect solution for detecting an incontinence episode with comfort and discretion. The sheet is machine washable with a long-life assured by the very fine embedded silver coated sensing wires. The connecting lead should be removed before laundering.

Replacement Cotton Sheets

Cotton sheets are non-returnable.
Replacement sheets are available separately.
Order Part No. W00860 (or Neat NE1020006-04)
Positioning
The cotton sheet should be placed between the bottom bed sheet and the mattress at a point where the pelvic area of the person will be when lying down.
Alarm Condition
An alarm will be raised when moisture is detected thereby alleviating Carers from making regular checks which can be intrusive and unwelcome.
IMPORTANT: It is advisable to disconnect the cotton sheet as soon as possible after an alarm. Prolonged connection while damp can result in some cosmetic staining.
Low Battery Condition
A low battery condition will be reported automatically to the Alarm Receiving Centre (ARC) or the local Manager. replaced within 30 days.
Out of Range Notification (Reach IP/NOVO)
The Reach IP/NOVO can monitor all radio peripheral devices to determine whether they have been damaged or removed from an installation.
Should the Reach IP/NOVO be unable to detect a device’s heartbeat it will automatically generate a “radio out of range” notification. Firstly, investigate if there’s a genuine reason for the alert, if not, arrange for the peripheral device to be replaced immediately.
The “Radio out of range timer” can be set from 30 hours to 99999 hours in CMP. “Radio supervision” can be disabled by unchecking the tick box associated with each peripheral device in CMP.
Cotton Bed Sheet Cleaning
Unplug the connecting lead before laundering. For maximum life (50 washes) wash at 40 o C and hang to dry in a warm room (i.e. do not tumble dry). Alternatively, wash at 85 o C with a medium heat tumble dry but expect a reduced life of 30 washes.
Do not use a fabric softener – this will make the sheet waterproof. If the sheet develops brown staining consult the manufacturer for cleaning instructions.

Maintenance

Test periodically by linking the 2 metal studs on the bed sheet with a metal item e.g. spoon, paper clip, etc.
Failure to alarm may suggest the sensing wires inside the sheet have been damaged in the wash.
The interface unit can be cleaned with a damp cloth or disinfectant wipe, do not use solvents or cleaners.
Disposal
Waste electrical products should not be disposed of with normal household waste. The enuresis sensor is ideally suited for disposal within the waste electronic and electrical equipment (WEEE) recycling scheme. Please recycle where facilities exist.
Check with your local authority or contact your supplier for recycling/disposal advice.

How to install the Enuresis Sensor

  1. Remove the bottom sheet from the bed.
  2. Place the cotton sheet across the width of the bed at a point where the pelvic area of the person will be when lying down.
  3. The side with the visible pattern of electrodes should be placed uppermost with the press stud connections accessible. If necessary a towel can be placed over the sheet to make it less sensitive.
  4. Connect the lead to the cotton sheet using one of the pairs of press studs, it doesn’t matter which.  The other pair of studs are used for testing. tynetec FM0847 Enuresis Sensor - Connect
  5. Ensure the connecting lead exits safely without causing atrip hazard.
  6. Refit the bottom bed sheet.
  7. Remove the battery cover from the interface unit and connect the battery.
  8. Plug the lead into the interface unit and make sure it is placed safely out of the way under the bed.

Telecare Radio ID
The enuresis sensor includes a radio transceiver to communicate with the Reach IP, Advent XT2, Carer
Response or NOVO.
The Radio ID is a 10-digit code printed on the enuresis sensor label…

Test Call
To activate a test call from the enuresis sensor simply link the 2 metal Test Studs on the cotton sheet with a paper clip or spoon or any other metal object. See photo above.
Registering the Enuresis Sensor onto a Reach IP
Login to Pulse CMP ( www.tynetecpulse.com ) and enter the Reach IP Serial No. in the ALARM UNIT view. Press the Accessories button and enter the Radio ID code, Location and Equipment type. Press Save then Synchronize.
IMPORTANT: when the enuresis sensor is installed a test call must be made to check operation and start the digital heartbeat.
Alternatively, with the Reach IP in normal operating mode undo the fixing screw and remove the stand or connector cover to access the ⚫ Power button.

tynetec FM0847 Enuresis Sensor - Power batton

Press the ⚫ Power button to enter Control Mode, then press and hold the button until the unit announces “add radio device” then release.
Press the ⚫ button to confirm then activate the enuresis sensor by linking the 2 metal studs on the cotton sheet.
The Reach IP unit will sound a high beep if it’s a new device and announce “range test mode”.
Press ⚫ to confirm, or press   to step to the next function, or press to exit the mode. Perform a test call to check operation
Registering the Enuresis Sensor onto an XT2
Press the Settings button on the Home page of the Manager’s Panel. Press the Manager (or Engineer) button, enter the password and press the Programming button. Select the Program Radio Devices option from the tabs along the bottom of the screen.

tynetec FM0847 Enuresis Sensor -screen

Press Flat ID and enter the flat number the enuresis sensor is being installed in. Enter 4-digits e.g. 0014.
Press Radio ID and enter the 10-digit code taken from the label on the enuresis sensor.
Press Room Locations and select where the enuresis sensor is located from the on-screen options e.g. Bedroom.
Press the Send button to update the XT2 system.
Perform a test call to check operation.
Registering the Enuresis Sensor onto a Carer Response
Refer to the Carer Response user manual for full details of the button functions and how to navigate the menus.
From the Main menu select
Programming.
From the Programming menu select Add transmitter.
The display will prompt you to Activate the transmitter…
Activate the enuresis sensor by linking the 2 metal studs on the cotton sheet.
The Carer Response will beep and the display will change to the Add transmitter menu with a box showing No ID.
Press then to select Room, then press
Press to select the room number box 0000.
Press to select the first digit, press to move between the digits and press or to change each digit.
Press when the correct room number is displayed.
Press to select the alarm type box None.
Press then to select Assistance then press
Press the button to select Save then press
The display will show Radio ID saved to memory…
Return to the Main menu and select Quit.
Perform a test call to check operation.
Registering the Enuresis Sensor onto a Touchsafe Pro
Login to the Master Touchsafe Pro Nursecall Panel (must be V4.01 onward) using an Engineer or Manager password.
From the Main Menu page select the Telecare option to open thescreen shown below.

tynetec FM0847 Enuresis Sensor - Telecare

Touch the Add Telecare Device box and enter the Telecare device 10-digit Radio ID using the on-screen keypad.
Touch the Filter box and enter the zone you want to add the Telecare device to using the on-screen keypad.
Touch the Room box and select the Callpoint ID from the list that will be used to cancel the Telecare alarm.
Touch the Save button on the keypad.
Touch the Close button to exit.
Perform a test call to check operation.
Registering the Enuresis Sensor onto a Neat NOVO
Start your NOVO with the AC adapter attached.
Pairing using CMPtynetec FM0847 Enuresis Sensor - Telecare
1 Login to CMP ( cmp.neat-cloud.com ) and enter the NOVO serial number in the CAREPHONE view.
Perform a search to locate the NOVO.
Select Accessories and thereafter New.
Enter the Radio ID code and the Equipment type.
The Radio ID can be found on the enuresis sensor label.
Be sure to enter all 10 characters then press Save.
Wait and thereafter perform a Refresh, it may take up to 30 minutes for the accessory to be added.
Activate the enuresis sensor by linking the 2 metal studs on the cotton sheet to check operation.
Pairing using the NOVO Service Mode
Enter Service mode by a short press on the Power switch. (Button B1, B2 & B3 are lit)
Then press and hold button B3 until “Add transmitter” is announced. (Button B1 is lit & B3 is flashing) Press button B1 to confirm.tynetec FM0847 Enuresis
Sensor - Telecare 2 (Button B1, B2 & B3 are flashing)
Activate the enuresis sensor by linking the 2 metal studs on the cotton sheet. The NOVO will sound a high beep and announce “range test mode”.
The pairing is done. Exit Service mode by pressing button B2. (Button B1 is lit). Activate the enuresis sensor by linking the 2 metal studs on the cotton sheet to check operation.
Declaration of Conformity
We declare this product complies with the relevant UK legislation on condition that it is used in the manner intended, and in accordance with the current installation standards and/or the manufacturer’s recommendations. A copy of the full UKCA Declaration of Conformity is available on request.
We declare this product complies with the relevant European legislation on condition that it is used in the manner intended, and in accordance with the current installation standards and/or the manufacturer’s recommendations. A copy of the full CE Declaration of Conformity is available on request.

Chris Dodd
CEO
Legrand Assisted Living & Healthcare

Tynetec, a business unit of Legrand Electric Ltd
Cowley Road, Blyth Riverside Business Park, Blyth, Northumberland, NE24 5TF.
Tel: +44 (0) 1670 352371
www.tynetec.co.uk

Documents / Resources

| tynetec FM0847 Enuresis Sensor [pdf] User Guide
FM0847 Enuresis Sensor, FM0847, Enuresis Sensor, Sensor
---|---

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

tynetec User Manuals

Related Manuals