tynetec FM0885 Companion Mini Epilepsy Monitor Instruction Manual

June 13, 2024
tynetec

tynetec FM0885 Companion Mini Epilepsy Monitor

Product Information

The 869MHz Companion Mini Trusted Technology Epilepsy Monitor is a device designed to monitor the movements of a user while they are in bed. It is specifically developed for individuals with epilepsy to provide safety and peace of mind. The Companion Mini is equipped with a bed sensor that detects excessive movements and triggers an alarm call if necessary. It has a battery back-up to ensure continuous operation, and it communicates wirelessly with the Connect-iT unit.

The Connect-iT unit serves as the central hub for the Companion Mini system. It maintains charge for the battery back-up and facilitates radio communications between the Companion Mini and other devices. The Connect-iT unit has LED indicators to display
the status of the system.

Positioning

To install the Companion Mini bed sensor, place it under the mattress beneath the user. Since bed types and mattresses vary, it is recommended to perform multiple tests during installation to determine the optimal position. For increased sensitivity, you can place the foam block provided in the carton under the Companion Mini. Further adjustments can be made using the Profile Editor software, which can be downloaded for free from Alert-it.

Normal Operating Mode:

To ensure the battery back-up remains charged, keep the Connect-iT unit plugged into the mains supply. The POWER LED on the Connect-iT unit should be steadily lit green. The MONITOR LED should flash red every 8 seconds to indicate that radio communications between the Companion Mini and Connect-iT are functioning correctly.

Alarm Condition:

If the bed sensor detects excessive movements in the bed for more than 15 seconds, an alarm call will be transmitted. The epilepsy monitor will automatically reset itself and return to normal operation after 1 minute. The delay before the alarm period can be adjusted using the Alert-it Profile Editor software.

Companion Mini Low Battery Condition:

When the bed sensor battery is low, the MONITOR LED on the Connect-iT unit will flash red every second. It is important to replace the 2 x AAA Alkaline batteries inside the Companion Mini as soon as possible. If the bed sensor battery becomes very low, the MONITOR LED will flash red every second, and the ALARM LED will start flashing red every 2 seconds.

Connect-iT Low Battery Condition:

If the Connect-iT unit’s battery becomes low, it will automatically report this to the Alarm Receiving Centre (ARC) or the local Manager. In such cases, the complete Connect-iT unit must be returned to Tynetec for battery replacement.

Maintenance:

To clean the device, use a damp cloth and mild detergent. Do not use solvents or abrasive cleaners. Take care to avoid getting any liquid into the micro-USB connector.

Disposal:

Do not dispose of the product with normal household waste. The epilepsy monitor is suitable for disposal within the waste electronic and electrical equipment (WEEE) recycling scheme. Please recycle it where facilities exist. For recycling/disposal advice, consult your local authority or contact your supplier.

Warning:

The Connect-iT unit contains a rechargeable Nickel Metal Hydride and a Lithium battery. Tampering with these batteries may cause damage and create a fire hazard. Do not incinerate the batteries.

Tynetec, a business unit of Legrand Electric Ltd Cowley Road, Blyth Riverside Business Park, Blyth, Northumberland, NE24 5TF. Tel: +44 (0) 1670 352371 www.tynetec.co.uk

Doc No. FM0885 V1.03 Page 1

  • Detects tonic-clonic, partial or generalised seizures

  • Compatible with Reach IP, Advent XT2, Carer Response and Touchsafe pro V4.01 onward

  • Product Code: ZXT846
    Companion Mini Unit

  • Wireless battery powered bed sensor

  • Robust and tamper-proof casing

  • Frequency: 434.075MHz

  • Radio range: up to 200 metres

  • 2 x AAA Alkaline user replaceable batteries

  • 12 months battery life (typical)

  • Low battery warning

  • B-Micro USB 2.0 (for advanced config setup)

  • Weight: 54 grams

  • Dimensions: 100mm x 45mm x 15mm (HxWxD)

  • Operating Temperature: +5oC to +35oC, 90%RH Connect-iT Unit

  • Mains powered/battery backed alarm unit

  • Plug-top mains adapter: 12V DC (Type: P171A)

  • 2 x AA 1.2V 2500mAh NiMH rechargeable batteries

  • 24 hours battery back-up (typical)

  • Weight: 175 grams

  • Dimensions: 90mm x 90mm x 31mm (HxWxD)

  • Operating Temperature: +5oC to +35oC, 90%RH Telecare Transceiver

  • Factory fitted inside the Connect-iT unit

  • Frequency: 869.2125MHz Class 1.5

  • Digital Heartbeat

  • 5 year Lithium battery (typical)

  • Low battery warning

IMPORTANT: refer to the Alert-it Companion Mini Connect handbook supplied in the carton for more detailed information on installation, operation, and maintenance of the product.

Positioning

The Companion Mini bed sensor should be placed under the mattress beneath the user. Due to the wide variety of bed types and mattresses it is advised to perform several tests during installation to establish the best position. The foam block provided in the carton can be placed under the Companion Mini to increase sensitivity. Further adjustments can be made using the Profile Editor software available as a free download from Alert-it.

Normal Operating Mode

To maintain charge of the battery back-up the Connect-iT unit should be left plugged-in to the mains supply.
The POWER LED should be lit steady green.
The MONITOR LED on the Connect-IT unit should flash red every 8 seconds to indicate radio communications are OK between the Companion Mini and the Connect-iT.

Alarm Condition

An alarm call is transmitted if the bed sensor detects excessive movements in the bed for a period greater than 15 seconds. The epilepsy monitor will reset itself and return to normal operation 1 minute after an alarm.
Note: this delay before alarm period can be changed using the Alert-it Profile Editor software.

Companion Mini Low Battery Condition

When the bed sensor battery gets low the MONITOR LED on the Connect-iT unit will flash red every second. Replace the 2 x AAA Alkaline batteries inside the Companion Mini as soon as possible.
If bed sensor battery gets very low the MONITOR LED will flash red every second and the ALARM LED will start to flash red every 2 seconds.

Connect-iT Low Battery Condition

If the mains supply fails the battery backup will maintain operation for about 24 hours.
When running on the battery backup, the POWER LED will turn from solid green to a green flash every 4 seconds. When the battery backup begins to get low the POWER LED will flash green every second.
When the battery backup gets very low the MONITOR LED will begin to flash red every second and the ALARM LED will flash red every 2 seconds.
It will take about 12 hours to fully recharge the battery backup when the mains supply is restored.

Telecare Transceiver Low Battery Condition

If the Telecare transceiver battery becomes low this will be reported automatically to the Alarm Receiving Centre
(ARC) or the local Manager. The complete Connect-iT unit must be returned to Tynetec for battery replacement.

Out of Range Notification (Reach IP)

The Reach IP can monitor all radio peripheral devices to determine whether they have been damaged or removed from an installation. Should the Reach IP be unable to detect a device’s heartbeat it will automatically generate a “radio out of range” notification. Firstly, investigate if there’s a genuine reason for the alert, if not, arrange for the peripheral device to be replaced immediately.
Note: the “Radio out of range timer” can be set from 30 hours to 99999 hours in Pulse CMP. “Radio supervision” can be disabled by unchecking the tick box associated with each peripheral device in Pulse CMP.

Maintenance

Wipe with a damp cloth and mild detergent, do not use solvents or abrasive cleaners. Avoid getting any liquid into the micro-USB connector.

Disposal

Waste electrical products should not be disposed of with normal household waste. The epilepsy monitor is ideally suited for disposal within the waste electronic and electrical equipment (WEEE)  recycling scheme. Please recycle where facilities exist. Check with your local authority or contact your supplier for recycling/disposal advice.

Warning

The Connect-iT unit contains a rechargeable Nickel Metal Hydride and a Lithium battery.
If tampered with the batteries may be damaged and create a fire hazard. Do not incinerate.

Telecare Radio ID
The epilepsy monitor includes a radio transceiver to communicate with the Reach IP, XT2 or Carer Response.

The Radio ID is a 10-digit code printed on the Connect-iT label…

Alert-it Profile Editor Software

Download free software from www.alert-it.co.uk/support Connect the Companion Mini to your PC using the micro USB lead provided in the carton to adjust the sensitivity and delay before alarm.

Installation

  1. Turn on the Companion Mini by pressing the TEST button with a pen tip. The ALARM light will flash green for 60 seconds during the power-up test mode.
    Note: an alarm call will be transmitted when any movement is detected during the test mode.

  2. Turn on the Connect-iT unit by pressing the PAIR button with a pen tip. The POWER light will flash green when its running on its battery back-up.
    Connect the P171 adapter and plug into the mains supply. The POWER light should now be lit steady green.
    During normal operation the MONITOR light should flash green every 8 seconds when a transmission is received from the Companion Mini.

  3. Put the Companion Mini into test mode and place it under the mattress in a central location. Pat the bed to simulate a seizure and check the MONITOR and ALARM lights flash red on the Connect-iT unit.

  4. If movement is not detected due to the mattress thickness or density, then the foam pad provided in the box can be used to increase sensitivity. Use the sticky pads to secure the foam to the bottom of the Companion Mini. Put the Companion Mini back into test mode and place it back under the mattress and re-test.

  5. If movement is still not detected, then the sensitivity may need to be increased by connecting the Companion Mini to a PC using the Profile Editor software.

  • How to test the Epilepsy Monitor
    Pat the bed with constant rhythm for at least 15 seconds until the MONITOR light comes on steady red and the ALARM light flashes red on the Connect-iT unit.
    An epilepsy alarm signal will be transmitted.
    After approx. 1 minute the epilepsy alarm will reset itself i.e. the red lights will go out and the MONITOR light will flash green every 8 seconds.

  • How to adjust the Sensitivity
    The default sensitivity setting should be suitable for detecting spasms in a domestic bed.
    The sensitivity can be changed using the Alert-it Profile Editor software.

  • How to adjust the Time Delay
    This is the time for which the distress condition must occur before an alarm is transmitted. The delay can be changed using the Alert-it Profile Editor software.
    The delay should be set to minimize false alarms during normal movement. The default time delay of 15 seconds will normally be a good compromise between speed of detection and avoiding false alarms during nocturnal restlessness or coughing, turning, etc.

  • How to turn the Epilepsy Monitor off
    If the epilepsy alarm is being taken out of service and stored for a prolonged period, then it should be powered-down as described below.
    Switch off and disconnect the mains adapter from the Connect-iT unit.
    Press and hold the small white button on the bottom face of the Connect-iT unit, the POWER light will go steady green and the MONITOR light will go steady red, after about 3 seconds a series of tones will sound and the MONITOR light will go out. Keep the button pressed until the ALARM light goes red and the POWER light flashes green. Release the button and the Connect-iT unit is now switched off.
    Remove the batteries from the Companion Mini during storage.

  • Registering the Epilepsy Monitor onto a Reach IP
    Login to Pulse CMP (www.tynetecpulse.com) and enter the Reach IP Serial No. in the ALARM UNIT view. Press the Accessories button and enter the Radio ID code, Location and Equipment type. Press Save then Synchronize. IMPORTANT: when the epilepsy monitor is installed a test call must be made to check operation and start the digital heartbeat.
    Alternatively, with the Reach IP in normal operating mode undo the fixing screw and remove the stand or connector cover to access the ⚫ Power button.tynetec FM0885 Companion Mini Epilepsy Monitor-
fig6

  • Registering the Epilepsy Monitor onto an XT2

    • Press the Settings button on the Home page of the Manager’s Panel. Press the Manager (or Engineer) button, enter the password and press the Programming button. Select the Program Radio Devices option from the tabs along the bottom of the screen.
    • Press Flat ID and enter the flat number the epilepsy monitor is being installed in. Enter 4-digits e.g. 0014. Press Radio ID and enter the 10-digit code taken from the label on the back of the epilepsy monitor.
    • Press Room Locations and select where the epilepsy monitor is located from the on-screen options.
    • Press the Send button to update the XT2 system.
    • Activate the epilepsy monitor to check operation by patting the bed with constant rhythm until the MONITOR light comes on steady red and the ALARM light flashes red on the Connect-iT unit.
  • Registering the Monitor onto a Carer Response
    Refer to the Carer Response user manual for full details of the button functions and how to navigate the menus.
    From the Main menu select

    • Programming.
      From the Programming menu select

    • Add transmitter.
      The display will prompt you to

    • Activate the transmitter…
      Activate the epilepsy monitor by patting the bed with constant rhythm until the MONITOR light comes on steady red and the ALARM light flashes red on the Connect-iT unit. The Carer Response will beep, and the display will change to the Add transmitter menu with a box showing No ID. Press to select Room, then press

    • Press to select the room number box 0000.

    • Press to select the first digit, press to move between the digits and press to change each digit.

    • Press when the correct room number is displayed.

    • Press to select the alarm type box None.

    • Press to select Epilepsy alarm then press Press the button to select Save then press

    • The display will show Radio ID saved to memory…

    • Return to the Main menu and select Quit.
      Activate the epilepsy monitor again to check operation.

  • Registering the Epilepsy Monitor onto a Touchsafe Pro
    Login to the Master Touchsafe Pro Nursecall Panel (must be V4.01 onward) using Engineer or Manager password. From the Main Menu page select the Telecare option to open the screen shown below.

    • Touch the Add Telecare Device box and enter the Telecare device 10-digit Radio ID using the on-screen keypad.
    • Touch the Filter box and enter the zone you want to add the Telecare device to using the on-screen keypad. Touch the Room box  and select the Callpoint ID from the list that will be used to cancel the Telecare alarm. Touch the Save button on the keypad.
    • Touch the Close button to exit.
    • Activate the epilepsy monitor to check operation by patting the bed with constant rhythm until the MONITOR light comes on steady red and the ALARM light flashes red on the Connect-iT unit.

Declaration of Conformity

We declare this product complies with the relevant UK legislation on condition that it is used in the manner intended, and in accordance with the current installation standards and/or the manufacturer’s recommendations. A copy of the full UKCA Declaration of Conformity is available on request.

We declare this product complies with the relevant European legislation on condition that it is used in the manner intended, and in accordance with the current installation standards and/or the manufacturer’s recommendations. A copy of the full CE Declaration of Conformity is available on request.

Chris Dodd
CEO
Legrand Assisted Living & Healthcare

Tynetec, a business unit of Legrand Electric Ltd
Cowley Road, Blyth Riverside Business Park, Blyth, Northumberland, NE24 5TF. Tel: +44 (0) 1670 352371 www.tynetec.co.uk

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

tynetec User Manuals

Related Manuals