CASTON-III Crane Scale Instruction Manual

June 13, 2024
CAS

CASTON-III Crane Scale

CASTON-III Crane Scale

PRECAUTIONS

Warning
Turn power switch off on rear panel when you don’t use the scale for a long time.

  • Do not disassemble the scale.
    When any damage or defect occurs, contact your CAS authorized dealer immediate for proper repair.
    Precautions

  • Do not overload beyond the maximum weight limit
    Precautions

  • Scale must be grounded to minimize electricity static.
    This will minimize defect or electric shock.
    Precautions

  • Do not pull the plug by its cord when unplugging.
    Damaged cord could cause electric shock or fire.
    Precautions

  • To prevent from fire occurring, Do not place or use the scale near flammable or corrosive gas.
    Precautions

  • To reduce electric shock or incorrect reading, Do not spill water on the scale or place it in humid condition.
    Precautions

  • Avoid placing the scale near heater or in direct sunlight.
    Precautions

  • Insert plug firmly to wall outlet to prevent electric shock.
    Precautions

  • Use proper Adapter.
    Incorrect adapter could damage the scale.
    Precautions

Attention

Do not stand under the scale while weighing the load, and you should be careful to provide against a safety accident. To install, safety pin in hook should be fixed to prevent dropping out from lift eye bolt.

  • For consistent and accurate reading, maintain periodical check by your CAS authorized dealer.
    Precautions

  • Avoid sudden shock to the scale.
    Precautions

  • Take the battery out when scale is not in use for long time. Leakage from the batteries is hazardous.
    Precautions

  • Place the scale on firm and temperature consistent environment.
    Precautions

  • Keep the scale away from other electromagnetic enerating devices.
    Precautions

WARNING: The load which hang on the hook should bea vertical load. (Refer to the next features.)
Precautions

OVERALL VIEW (CASTON-III)

Overall View \(Caston-iii\)

SPECIFICATIONS

MAX. TARE WEIGHT FULL TARE
DISPLAY L.E.D(1.5 inch)
OPERATION TEMPERATURE – 10 ~ +40 ℃ ℃
POWER SOURCE DC 12 V Rechargeable Battery / AC Adaptor
POWER CONSUMPTION 1.2 ~ 2.4W
DISPLAY LAMP ZERO, TARE, HOLD, LOW BATTERY LAMP
CONTINUOUS USING TIME 40 HOURS

Overall View \(Caston-iii\)

MAX. TARE WEIGHT FULL TARE
DISPLAY L.E.D(1.2 inch)
OPERATION TEMPERATURE – 10 ~ +40 ℃ ℃
POWER SOURCE DC 6 V Rechargable Battery / AC Adaptor
POWER CONSUMPTION 0.6 ~ 1W
DISPLAY LAMP ZERO, TARE, HOLD, LOW BATTERY LAMP
CONTINUOUS USING TIME 60 HOURS

DISPLAY & KEY FUNCTIONS

Display & Key Functions

DISPLAY

WEIGHT DISPLAY:Displays weight or messages (5 Digit).
ZERO LAMP:Indicates that scale is stable and at zero.
NET LAMP:Indicates that scale is currently using a tare.
HOLD LAMP:Indicates that the HOLD function is activated.
LOW BATTERY LAMP:Indicates that voltage of battery low and should be changed soon.

KEY FUNCTIONS
KEYS FUNCTIONS
Used to turn the scale ON / OFF or used to released from AP/HL mode
Used to reset the scale zero or enter the AP mode
Used to activate tare function and to clear tare entry
Used to weigh unstable things (livestock, liquid, etc.) or used to store the

set value in ‘AP/HL’ mode
| Used to increase the set value in ‘AP/HL’ mode

WARNING: Do not press the keys with excessive force, the keys are to be pressed with a soft touch.

BATTERY USAGE

CASTON-III

  • Open the battery lid located in back of the scale and turn main switch off.

  • Separate the battery connector and take out the discharged battery.

  • Put a charged battery in rear and connect the battery connector.
    Turn main power switch on and cover the battery lid.

  • The drop of battery voltage can affect the brightness of display.
    To prevent this, the regular battery recharging is recommended.

  • In case the battery will be left without recharging for a long time, it is desirable the main power switch is set off. Otherwise, it also can affect the battery life.

Battery Usage

CASTON-II

  • Couple connector in main body with connector in battery portion.
  • Insert battery portion to the main body and turn fixing bolt to the right to fix.
    Battery Usage

USE OF BATTERY CHARGER

  • Check the voltage of battery charger before use.
    (It is fixed to 220V before shipping)

  • Insert charger A.C cord into the power source and connect the discharged battery to the connector.

  • Turn ON the switch on the battery charger, charging will start with Yellow lamp (CHARGE) turn on.

  • When the charging is completed, GREEN lamp(FULL) turns on.

  • Estimated time for charging is 10 hours(CASTON-III).
    Estimated time for charging is 6 hours(CASTON-II).
    (Charging time can be varying according to the battery condition.)

  • Using period when the battery is charged fully : About 15~30 hours

FUNCTION & DESCRIPTIONS

  1. POWER ON

    • Press the ON key located in real panel of the scale.
      (While the power is off, the ON/OFF key on the remote control wouldn’t operate.)

    • Press ON/ OFF key.

    • LED display will be on and then, it will show from 00000 to 99999 continuously

  2. POWER OFF

    • Press ON/OFF key
    • LED display will show off message and then power turns off
  3. ZERO FUNCTION

    • Used to correct drifted zero nalue when the scale is unloaded motion is not detected You can adjust the zero up to ±2% of the maximum capacity
    • The function does not work when weight is fluctuated ort unstable.
  4. TARE FUNCTION

    • Use tare function after removing the weighing material from container being used.
    • The function does not work when weight is fluctuated ort unstable.
    • The weight including TARE weight can’t exceed the maximum capacity.
  5. HOLD FUNCTION
    There are three HOLD functions : Average / Manual / Peak Hold function To use this function , must be check Hold-type of ‘HL’ mode.

    • Auto Hold (HL -> 0)
    • Press the HOLD key before weighing an unstable item
    • “AHold” will display and then, HOLD lamp is on.
    • The averaged value of the weight fluctuation will be displayed after few seconds.
    • Peak Hold (HL -> 1)
    • Press the HOLD key while checking peak value.
    • “PHold” will display and then, peak value of the weight will be updated continuously until HOLD function is released
    •  Average Hold (HL ->2)
    • Press the HOLD key while weighing an unstable item
    • “Hold” will display and then the averaged value of the weight fluctuation will be displayed after few seconds.
    • Release of Hold Functions
    • Press the HOLD key while HOLD functions are activated
    • Remove item on the hook.
    • hdoFF will display and then, the scale return to normal weighing mode.

AP MODE

Press ON key while ZERO key is pressed, then you can enter into ‘AP’ mode. (‘AP’ means ‘Auto Power off’.)

Key Usage

STAR KEY **** Used to increase the input value
HOLD KEY Used to save the input value and Return to Normal Mode.
ZERO KEY Go to Next Mode.

Setting Menu

DISPLAY FUNCTION
AP-00 Not use
AP-10 Power off after keeping ‘0’kg 10 minute
AP-20 Power off after keeping ‘0’kg 20 minute
AP-30 Power off after keeping ‘0’kg 30 minute

SLEEP MODE

Press ON key while ZERO key is pressed and Press ZERO Again, then you can enter into ‘SLEEP’ mode.
(‘Sb’ means ‘SLEEP’.)

Key Usage

STAR KEY Used to increase the input value
HOLD KEY Used to save the input value and Return to Normal Mode
ZERO KEY Go to Next Mode.

Setting Menu.

DISPLAY FUNCTION
Sb-00 Sleep mode is deactivated.
Sb-20 Sleep mode is activated after 20 seconds.
Sb-40 Sleep mode is activated after 40 seconds.
Sb-60 Sleep mode is activated after 1 minute.

AUTO ZERO TRACKING MODE

Press ON key while ZERO key is pressed and Press ZERO Again, then you can enter into ‘AZ’ mode.
(‘AZ’ means ‘Auto Zero Tracking’.)

Key Usage

STAR KEY Used to increase the input value
HOLD KEY Used to save the input value and Return to Normal Mode
ZERO KEY Go to Next Mode.

Setting Menu.

DISPLAY FUNCTION
AZ-1 Zero Tracking mode is activated within 0.5 div.
AZ-2 Zero Tracking mode is activated within 1 div.
AZ-5 Zero Tracking mode is activated within 2.5 div.
AZ-9 Zero Tracking mode is activated within 4.5 div.

Ex>When you set the value 9, Zerotracking will change the weight value slowly within 1 div.

BA MODE

Press ON key while ZERO key is pressed and Press ZERO Again, then you can enter into ‘bA’ mode. (‘BA’ means ‘Weight BAck up’.)

Key Usage

STAR KEY Used to change the input value
HOLD KEY Used to save the input value and Return to Normal Mode

Setting Menu

DISPLAY FUNCTION
bA-oF Weight Back Up OFF
bA-oN Weight Back Up ON

※ The CO MODE and ID MODE are an engineer mode.
You have to set CO=tu and ID=0 as the default, when you **
use Bluetooth (TW- 100B).**

HL MODE

Press ON key while HOLD key is pressed, then you can enter into ‘HL’ mode. (‘HL’ means ‘Hold type’.)

Key Usage

STAR KEY Used to increase the input value
HOLD KEY Used to save the input value

Setting Menu

DISPLAY FUNCTION
HL-00 Auto Hold / Manual Hold
HL-01 Peak Hold use
HL-02 Average Hold use

While using the HOLD function, the weight displayed may be not exact.

oL MODE
This mode is only for Engineer mode.

REMOTE CONTROL

  • ON/OFF: Button to operate power on/off.
  • ZERO: Button to correct zero value.
  • TARE: Used to tare value memorizing.
  • HOLD: Used to weigh moving thing.
  • CLEAR : Used to clear summing weight.
  • SUM: Used to sum item

Remote Control

REMOTE CONTROL

AVAILABLE DISTANCE| 6m   ~ 9m
AVAILABLE ANGLE| 60°
POWER| 3V(1.5V AMM x 2)

OPTIONS

Please refer to handy terminal TW-100 B manual of CASTON-III B (Crane Scale : Bluetooth Type)

Options

TW-100 Specification

Frequency 2400~2483.5 MHz
Channel interval 1 Mhz
Channel 78 channel
Communication Type GFSK (Gaussina Frequency Shift Keying)
Communication Speed 9600
Actual Communication Distance Maximum. Approximately 50~100M
Display LCD (5 digit)
Size of letter 12 mm (Height)
Display below zero “-“ displayed
Battery life time Approx. 6 hours
Battery charging time Approx. 3~5 hours
Power TW -100 : Rechargeable DC Battery.2V750mA Charger : AC220V/60Hz
Power consumption 0.43~0.72 W
Temperature range -10℃ ~ +40℃
Function Printer interface, zero, tare, hold, print, ID setting, weight sum

PRODUCT SIZE (CASTON-III (B))

0.5 TON 5 TON

Product Size \(Caston-iii\(B\)\)

MODEL MAX.CAPA DIVISION A WEIGHT
0.5THD 0.5TON 200g 510 26kg
1THD 1TON 500g 510 26kg
2THD 2TON 1kg 515 28kg
3THD 3TON 1kg 515 28kg
5THD 5TON 2kg 538 31kg

10 TON 50 TON(CASTON-III(B))

10 Ton 50 Ton \(Caston-iii\(B\)\)

MODEL MAX.CAPA DIVISION A B C D E F G WEIGHT
10TH D 10TON 5kg 667 360 C 83 845 57.2 42 57.2 44kg
15TH D 15TON 10kg 682 360 C 99 886 63.5 50 63.5 53kg
20THD 20TON 10kg 725 360 C 126 990 86 60 86 76kg
30TH D 30TON 20kg 895 440 C 138 1200 108 92 108 220kg
50TH D 50TON 20kg 1400 440 C 180 1520 146 112.7 146 390kg

PRODUCT SIZE (CASTON-II)

Product Size \(Caston-ii

MODEL| CAPACITY| GRADUATION| A| B| C| D| E| WEIGHT| F
---|---|---|---|---|---|---|---|---|---
0.5THB| 500kg| 200g| 350| 57.1| 82| 30| 26| 15kg| 26
1000LB| 0.5LB| 6.8LB
1THB| 1TON| 500g| 350| 57.1| 82| 30| 26| 15kg| 26
2000LB| 1LB
2THB| 2TON| 1kg| 430| 57.1| 91| 47| 45| 20kg| 30
5000LB| 2LB| 9.1LB
3THB| 3TON| 1kg| 430| 57.1| 91| 47| 45| 20kg(9.1LB)| 30
5THB| 5TON| 2kg| 474| 57.1| 1| 47| 45| 20kg| 30
10000 LB| 5LB| 9.1LB

ERROR MESSAGE

Error Message Description Solution
“Err 0” The “Err 0” occurs when scale is not stable. Remove unstable

facts.
“Err 1”| The “Err 1” occurs when a current zero point has shifted from the last span calibration.| Please call your CAS dealer.
“Err 3”| The “Err 3” is an overload error.| Please remove the weight.

▶ Notice : Specifications are subject to change for improvement without notice.

CAS BLDG., # 440-1, SUNGNAE-DONG, GANGDONG-GU, SEOUL, KOREA
TEL 82 2 2225 3500
FAX
82 2 475 4668
www.globaIcas.com

CASTON Logo

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

CAS User Manuals

Related Manuals