FALCON W202333 Vermeiren Patient Hoist User Manual
- June 13, 2024
- Falcon
Table of Contents
- FALCON W202333 Vermeiren Patient Hoist User Manual
- Preface
- 1. Your product
- 2. Before use
- 3. Using the patient hoist
- 4. Installation and adjustment
- 5 Maintenance
- 6. Troubleshooting
- 7. Technical specifications
- Service registration form
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
FALCON W202333 Vermeiren Patient Hoist User Manual
Instructions for specialist dealer
This instruction manual is part and parcel of the product and must accompany every product sold. Version: A, 2022-11
All rights reserved, including translation.
Preface
Congratulations! You are now owner of a Vermeiren patient hoist!
This product is made by qualified and committed personnel. It is designed and
produced according to high quality standards, guarded by Vermeiren.
Thank you for your trust in the products of Vermeiren. To support you on the
use of this patient hoist and its operating options, this manual is offered.
Please read it carefully; it will help you to get familiar with the operation,
capabilities and limitations of your product.
If you still have questions after reading this manual, do not hesitate to
contact your specialist dealer. He/she will be glad to help you.
Important note
To ensure your safety and to prolong the lifetime of your product, please take
good care of it and have it checked and serviced on a regular basis.
This manual reflects the latest product developments. Vermeiren has the right
to implement changes to this type of product without any obligation to adapt
or replace similar products previously delivered.
Pictures of the product are used to clarify the instructions in this manual.
Details of the depicted product may deviate from your product.
Information available
On our website http://www.vermeiren.com/ you will always find the most recent
version of the information in this manual. Please consult this website
regularly for possible updates.
Visually impaired people can download the electronic version of this manual
and have it read out by means of a text-to-speech software application.
This user manual
For user and specialist dealer
Service manual
For specialist dealer
EC declaration of conformity
1. Your product
- Cantilever arm
- Spreader bar
- Motor
- Chassis
- Roller
- Control box
- Handle for pushing
- Manual operation
- Emergency stop
- Pole
- Chassis expansion pedal
- Caster with brake
2. Before use
2.1 Intended use
In this paragraph a brief description of the intended use of your patient
hoist is given. Additionally, relevant warnings are added to the instructions
in the other paragraphs. In this way we would like to make you aware of the
possible misuse that may appear.
- Indications and contraindications: This patient hoist is meant to be operated by an attendant to transfer sitting patients between e.g. wheelchairs, nursing beds and bathrooms. The patient hoist is designed and produced to be a transport/transfer aid for elderly persons or persons who suffer from paralysis, limb loss or defects, stiff or damaged joints, cardiovascular problems, cachexia, … .
- This patient hoist is suitable for indoor use.
- This patient hoist is designed and produced solely to transport/transfer one (1) person with a maximum weight according to the specifications in §7. It is not designed for transportation of goods or objects, nor for any use other than previously described.
- Only use accessories and spare parts approved by Vermeiren.
- Please read all technical details and limits of your patient hoist in chapter 7.
- The warranty on this product is based on normal use and maintenance as described in this manual. Damage to your product caused by improper use or lack of maintenance will cause the warranty to lapse.
2.2 General safety instructions
Risk of injuries and/or damage
Please read and follow the instructions in this manual. Otherwise you may get
injured or your patient hoist may get damaged.
Keep the following general warnings in mind during use:
- The patient hoist must only be operated by qualified staff who have been instructed or trained in its specific application.
- Do not exceed the maximum load for the patient hoist. This will shut down the control box.
- In case of different user weights for patient hoist and slings, always observe the lowest user weight.
- An attendant needs to be present during the lifting operation.
- Be aware that some parts of the patient hoist may get very hot or cold due to ambient temperature, solar radiation or heating devices. Be careful when touching.
- The patient hoist has been tested for electromagnetic compatibility and complies with the standard. Nevertheless, sources of electromagnetic fields may influence the performance of the patient hoist, such as the fields of mobile phones, power generators or high-power energy sources. On the other hand, the electronics of the patient hoist can affect other electronic appliances too.
- Only use the patient hoist on flat surfaces where all castors touch the ground and where there is sufficient contact to operate the patient hoist safely. Do not take obstacles with the patient hoist while transferring a patient.
- Do not use the patient hoist in wet or humid environments.
- Alterations or substitutions should not be made to the patient hoist securement points or to structural and frame parts or components without consulting the manufacturer.
- Make sure that your hands, clothes, belts, buckles or jewellery don’t get caught by wheels or other moving parts during use.
- Be careful with sources of ignition such as lit cigarettes as they may set the sling alight.
In case a serious incident has occurred involving your product, notify Vermeiren or your specialist dealer as well as the competent authority in your country.
3. Using the patient hoist
3.1 Operating the lift
CAUTION Risk of injuries and/or damage
- When using the lift, make sure that there is sufficient room around and above it, as the adjustment movements could otherwise lead to damage or injury.
- Always be aware of the lifting arm to avoid injury.
- Only slings that have been designed and approved for the patients must be used (see usage instructions for the various slings). Use of any other slings is at your own risk.
- Only use the handles to push/pull the patient hoist, no other components.
Please observe the technical information according to which the patient hoists may be operated. The patient hoist must only be operated by authorised staff who have been trained in its use and operation.
3.1.1 Altering the chassis width
To get the lift around a wheelchair or other seating furniture, or to increase
the stability of the lift in the standing position, the separation of the
chassis legs can be increased.
Stand behind the assembled lift and take hold of the handles for pushing the
patient hoist (on the left and right, next to the control box). Press the
lever at the bottom of the chassis (left or right) down with your foot, and
the separation of the legs of the chassis can be increased or decreased.
3.1.2 Parking brakes
Secure the two casters at the back of the chassis by pressing the caster brake plates down gently to their end stops with the tip of your foot. To release the brake, push the brake plate gently back up again with the front of your foot until the casters are free.
3.1.3 Raising and lowering the cantilever arm
The manual control allows the cantilever arm to be adjusted smoothly to any
point.
When using the lift function until the end of the hoist motor, the limit switch is activated by electronics (security function). Use the lifting function only in the middle of the adjustment range.
3.1.4 Lowering button
If the manual operation mechanism should be damaged, it is possible to lower
the cantilever arm from the control box.
The arm moves down for as long as the button (3) on the control box panel is
pressed.
3.1.5 Moving the patient hoist
Risk of injuries
- Be careful passing through restricted passages (e.g. doors).
- Stand behind the patient hoist and grasp the handles (left and right, next to the control box) with both hands.
- Make sure that both back caster brakes have been released.
- Push the patient hoist slowly into the desired position.
CAUTION Risk of injuries
• Be careful passing through restricted passages (e.g. doors).
- Stand behind the patient hoist and grasp the handles (left and right, next to the control box) with both hands.
- Make sure that both back caster brakes have been released.
- Push the patient hoist slowly into the desired position.
When it is not loaded, we advise you to pull the patient hoist backwards, which makes it easier to get round any obstacles (door frames, corners of rooms or furniture).
3.1.6 Turn the patient hoist around his axle
CAUTION Risk of injuries
- Be careful there are no people or objects in the swing range of the patient hoist.
- Turn the patient hoist carefully and slowly around his axle, so that the patient cannot swing out.
- Open the legs of the patient hoist to give more stability.
- Activate the brakes.
- Place the patient in the patient hoist with the appropriate sling. (refer to sling manual for instructions of placing the patient in the hoist).
- Lift the patient from the ground, chair, bed, … .
- Loosen both brakes of the rear wheels.
- Stand behind the patient hoist and grasp the handgrips (left and right next to the control box) with both hands.
- Now turn the patient hoist carefully, slowly around on the same place. Be careful you do not touch any objects with the legs of the patient hoist.
- When you are turned in the good position, you can lift the patient down.
3.2 Slings
CAUTION Risk of injuries
- Only use appropriate slings for the patients.
- Do not use damaged slings.
- Read the full slings manual first.
- Only use compatible slings approved by Vermeiren.
The patient hoist must only be used with slings designed for the Vermeiren
Falcon patient hoists (for 4-point speader bar).
The applicability and instructions for use for the specific slings should be
followed, see the separate manual.
3.2.1 Attaching to the spreader bar of the patient hoists
The slings shall be attached by the mounting straps. These mounting straps
shall be suspended on the spreader bar of the patient hoist.
The mounting straps of the slings have 4 loops in different colors.
With these loops it is possible to lift the patient in different positions. These positions depends on the length and weight of the person.
The legrests can be attached on the patient hoist according different methods.
-
Crossed legrests
This is the most used method to attach the sling. The attachment of the sling with crossed legrests shall be according to the instructions below:
1. Put the mounting strap of the legrest through the loop of the other legrest.
2. Attach the mounting straps of the legrests on the spreader bar of the patient hoist. -
Crossed under both thighs
CAUTION: Risk of falling – By a patient with amputated legs increases the
risk of falling because the patient loose easy his balance.
This method can be used for lifting a patient with amputated legs. The
attachment of the sling crossed under both thighs shall be according to the
instructions below:
1. Cross the mounting strap of the legrest under the thighs.
2. Attach the mounting straps of the legrests on the spreader bar of the
patient hoist.
• Each legrest separately on the spreader bar
CAUTION: Risk of falling- This method increases the chance of sliding out the
sling.
Attaching each legrest separately on the spreader bar can be used for patient
with sensitive genitals. The attachment of the sling with this method shall be
according instructions below:
1. Put the legrest around each leg separately, but not crossed.
2. Attach the mounting straps of the legrests on the spreader bar of the
patient hoist.
3.2.2 Position of spreader bar
The spreader bar must always be perpendicular to the body of the patient. The
left figure is the good method. The right figure is the wrong way.
3.3
Battery and charging
3.3.1 Battery charger
3.3.2 Charging the batteries
Use only the control box with included battery charger.
We recommend that you should regularly recharge the batteries in order to
ensure that the hoist remains usable and to extend the lifespan of the
batteries. The control box will give a warning sound to tell you when the
battery charge is low.
First use
First, insert the supply cable with the connector side into the appropriate
connector on the control box. The plug side of the supply cable must be insert
in the socket outlet. The loading time is about 24 hours.
Recharging
WARNING: Risk of injury – Do not use the lift when the supply cable is connected in the socket outlet.
When the battery is running low, only the left LED will light green. Once the battery level gets very low, the LED will light red and a beeping sound can be heard every 2 seconds.
The battery is fully charged once all LEDs are fully green.
After the recharging process is completed, you should always remove the mains
plug from the socket first and then remove the connector plug from the control
box.
Do not use the lift while it is being recharged!
If the batteries are not used for a significant period, they discharge slowly by themselves (deep discharge). It then becomes impossible to recharge them. You should therefore recharge the batteries at least once a month even if they are not being used.
3.4 Emergency
In cases for emergency the emergency stop and the function for lowering in an
emergency must be used.
3.4.1 Emergency stop
In emergencies, and during assembly and disassembly, the emergency stop is
activated by pressing the red button on the control box.
Push the button and turn in the direction of the arrows. If the button is
fully pressed to the back turn back in the opposite direction of the arrows.
The emergency stop is deactivated by turning the red button in the direction of the arrows.
3.4.2 Emergency lowering
WARNING: Risk of injury – Adjust the manual emergency lowering for the
weight of the patient.
Emergency lowering if the power fails or the batteries are empty can be done
by turning the red button at the upper end of the motor.
Please note that the manual emergency lowering is only possible when the
patient is sitting in the patient hoist.
4. Installation and adjustment
The patient hoist is delivered fully assembled by your specialist dealer.
The instructions in this chapter are for the specialist dealer.
To find a service facility or specialist dealer near you, contact the nearest
Vermeiren facility.
WARNING: Risk of unsafe limitations – Use only the limitations described in this manual.
4.1 Delivery
The Vermeiren Falcon patient hoist shall be delivered with:
- chassis including 4 casters (2 with brakes)
- pole, including handles for pushing
- cantilever arm and spreader bar
- control box (incl. 2 accumulator batteries and manual operation)
- battery charger
- motor
- instruction manual
Before use check if everything is included and that no products are damaged
(example by transport, …).
Please note that the basic configuration may vary from one European country to
another. Please contact the specialist dealer in your country for more
information.
4.2 Assembly or dismantling
4.2.1 Unpacking
WARNING: Risk of injury – After unpacking and before further assembly, always first confirm that the emergency stop (the red pushbutton on the control box) is pressed in.
The patient hoist packaging has been selected to provide the best possible protection during transport.
4.2.2 Assembly
WARNING: Risk of clamping – Take care that no body parts and wires get
trapped, crushed or cut during assembly.
1. Insert the vertical pole into the lower frame and tighten the lever
securely.
2. Place the motor in its attachments on the vertical pole and the hoist arm
(4) and secure into place with the pins.
3. For first use: Mount the control box to the vertical pole by tightening
the two bolts.
4.2.3 Dismantling
WARNING: Risk of clamping – Take care that no body parts and wires get
trapped, crushed or cut during dismantling.
WARNING: Risk of injury – Check the emergency stop (red pushbutton on the
control box) is activated, to avoid unintentional adjustment movements.
1. Secure the two casters at the back of the chassis by pressing the caster
brake plates down gently to their end stops with the tip of your foot. To
release the brake, push the brake plate gently back up again with the front of
your foot until the casters are free.
2. Release the motor from the hoist arm and vertical pole by removing the
pins (4).
3. Remove the lever and take the vertical pole out of the lower frame.
5 Maintenance
5.1 Time of maintenance
CAUTION Risk of injuries and damage
Repairs and replacements may only be undertaken by trained persons and only
genuine replacement parts of Vermeiren should be used (see the maintenance
manual).
The last page of this manual contains a registration form for the specialist
dealer to record each service.
The service frequency depends on the frequency and intensity of use. Contact
your dealer to agree to a common timetable for inspection/maintenance/repair.
For the maintenance manual, refer to the Vermeiren website: www.vermeiren.com.
Before each use
Inspect the following points:
- All parts: present and undamaged or unworn.
- All parts: clean, see §5.3.
- Condition of frame parts: no deformation, instability, weakness or loose connections.
- Brakes: no visible damage and/or dirt.
- Control box, manual control, battery charger, relevant cables: no damage such as wires that have become frayed, broken or exposed. Check that the electrical adjustment of the cantilever arm is working properly, using the manual control.
- Battery status: charge the battery when necessary, see §3.3.
Approx. every 8 weeks
Depending on the frequency of use, check the following:
- Lubrication of the joints of the cantilever arm
- Condition of the casters
- Whether there is any visible damage to the casing of the control box, battery, battery charger and the manual controls
- Whether the cables are fit for use
- Recharge the battery
Approx. every 6 months or for every new user
Depending on the frequency of use check the following:
- Cleanliness
- General condition
- That the battery charger is working
- Casters are working
If the resistance to rotation is too great, clean the casters. If this is insufficient, please consult your dealer.
Yearly or more often
Have the patient hoist inspected and serviced by your specialist dealer, at
least once a year, or more often. The minimum maintenance frequency depends on
use and should therefore be commonly agreed upon with your specialist dealer.
5.2 Shipping and storage
- Shipping and storage should happen according to the technical specifications in §7. Make sure that the patient hoist is stored dry.
- Prevent cables from being crushed or bent with kinks.
- Disconnect the patient hoist from the socket.
- Provide sufficient covering or packaging to protect the patient hoist against rust and foreign bodies. (e.g. salt water, sea air, sand, dust).
- Store all removed parts together in one place (or mark them if necessary) to avoid mixing up with other products when re-assembling (e.g. the charger).
- Make sure no objects are stored on or against the patient hoist and its components to prevent damage.
5.3 Cleaning
CAUTION Risk of damage by moisture
- Never use a hose or high-pressure cleaner.
- Use a mild commercial detergent to remove stubborn dirt.
- Stains can be removed by using a sponge or a soft brush.
- Do not use strong cleaning liquids like solvents, nor use hard brushes.
5.3.1 Plastic parts
Clean plastic parts of your patient hoist with commercial plastic cleaners.
Read the specific product information and only use a soft brush or soft
sponge.
5.3.2 Coating
The high quality of the surface layer guarantees optimal protection against
corrosion. If the outer coat is damaged by scratching or in some other way,
get your specialist dealer to repair the affected surface.
When cleaning, only use warm water and normal household detergents and soft
brushes and cloths. Ensure that no wetness gets into the tubes.
5.3.3 Electronics casing
WARNING: Risk of injury – The emergency stop must be activated before
maintenance work, so that no unintentional adjustment movements can be made.
Wipe the control box, motor casing and manual controls down with a cloth
moistened by a few drops of a commercial domestic cleaner. Do not use
abrasives or sharp-edged polishing equipment like a metal scrubber or brush,
as these can scratch the surface of the manual control and destroy its water
repelling property.
Regularly check the plug connectors for corrosion or damage, as these could
affect the functional integrity of the electronics.
The manufacturer will not accept liability for damages caused by insufficient
maintenance.
5.4 Disinfection
WARNING: Dangerous Products – The use of disinfectants is restricted to
authorized personnel.
All parts of the patient hoist can be treated by scrubbing with a commercial
disinfectant. Follow the instructions on the disinfectant.
Washing instructions for slings can be found in the slings manual and on the
label of each sling.
5.5 Inspection
In principle we recommend one inspection every year, and a minimum of one
before usage is resumed. All of the following checks must be performed and
documented by authorized persons:
- Check the cabling (in particular for crushing, wear, cuts, visible parts of the insulation of the inner wires, visible metal wire cores, kinks, bulges, colour changes of the outer sleeve, brittle spots, and safe positioning so that mechanical effects such as being cut or crushed are not probable).
- Visual inspection of the frame parts to check for plastic deformation and/or wear and tear (frame, motor suspension, cantilever arm, spreader bar).
- Visually check for damage to the paintwork (danger of corrosion).
- Visual inspection of all casings for damage; screws must be tight.
- Verify the amount of grease on the metal joints of movable parts
- Visually check all plastic parts for cracks and brittle spots
- Review the control box + included charger residual discharge current (A) based on VDE 0702
- Review the control box + included charger insulation resistance (MO) based on VDE 0702
- Functional test of the cantilever arm (lubrication of the joint parts, range of adjustability, deformation, wear and tear)
- Check the functioning of the hoist drives (during a test drive → noises, speed, free running, etc.), if necessary: Measure the performance, first with no load and then with the nominal load (“SWL”), to investigate the wear and tear of the motors by comparing the values for the electric current against the values when it was delivered.
- Functional test of the emergency stop
- Functional test of the emergency lowering
- Completeness of the delivery condition, instruction manual available?
Checking measurements may only be carried out by skilled persons trained on
the patient hoist at least and at least under the supervision of an
electrician who knows the checking instruments and processes. Only an
electrician can release the patient hoist for use after making the
measurements and carrying out the servicing.
The service must only be signed off in the maintenance plan if a minimum of
all the above-mentioned aspects have been checked.
5.6 Expected lifespan
The average lifespan of the patient hoist is 8 years. Depending on the
frequency of use, storage, maintenance, servicing and cleaning, the lifespan
of the patient hoist will increase or decrease.
5.7 Reuse
Before each reuse, have the patient hoist disinfected, inspected and serviced
according to the instructions in this chapter.
5.8 End of use
At end of life, you need to dispose of the patient hoist according to the
local environmental legislation. The best way to do so, is to disassemble the
patient hoist to facilitate the transport of recyclable parts.
6. Troubleshooting
Table 1: Troubleshooting
7. Technical specifications
Table 2: Technical specifications
We reserve the right to introduce technical changes. Measurement tolerance ±
15 mm / 1,5 kg / 1,5°
1 = Highest position, 2 = Maximum reach, 3 = Lowest position
- CSP = Central suspension point
Service registration form
This product (name): ……………………………………..
was inspected (I), serviced (S), repaired (R) or disinfected (D):
Vermeiren GROUP
Vermeirenplein 1 / 15
2920 Kalmthout
BE
website: www.vermeiren.com
User manual – Falcon_MULTI_W202333_A5_2022-11_A
Read More About This Manual & Download PDF:
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>